This breathing set is an accessory to the F&P 950 Respiratory Humidifier to be used within the limits of its stated technical specifications. It is intended for delivery of heated humidified high flow respiratory gases to adult patients in a hospital. This product can be used on multiple patients when used with a hydrophobic filter between the product and the patient interface for a maximum of 24 hours after setup. This product is for use by appropriately qualified healthcare professionals who perform anesthesia care and airway management.
Device Story
F&P Optiflow Oxygen Kit (AA451J) is a multi-patient use breathing circuit accessory for the F&P 950 Respiratory Humidifier. Kit components include a dryline, water bag, humidification chamber, inspiratory limb, tubing clips, and date-change stickers. Gas flows from a source through the dryline to the humidification chamber for heating and humidification; gas then travels through the inspiratory limb, a hydrophobic filter, and a patient interface into the patient's upper airway. Used in hospitals by qualified healthcare professionals for anesthesia care and airway management. Supports flow rates of 5-70 L/min. When used with a hydrophobic filter, the kit is suitable for multiple patients, requiring reprocessing between uses and a maximum 24-hour duration after setup. Benefits include delivery of conditioned respiratory gases to adult patients.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing confirmed compliance with ISO 10993-1, ISO 80601-2-74, IEC 60601-1, ISO 5367, AAMI TIR12, and AAMI ST98.
Technological Characteristics
Breathing circuit kit comprising dryline, humidification chamber, and inspiratory limb. Materials evaluated per ISO 10993-1 and ISO 18562-1. Flow range: 5-70 L/min. Multi-patient use with hydrophobic filter; max 24-hour use; 3-year shelf life. Not provided sterile. Compatible with F&P 950 Respiratory Humidifier.
Indications for Use
Indicated for delivery of heated humidified high flow respiratory gases to adult patients in a hospital setting. For use by qualified healthcare professionals performing anesthesia care and airway management.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 13, 2024
Fisher & Paykel Healthcare Ltd Reena Daken Regulatory Affairs Manager 15 Maurice Paykel Place, East Tamaki Auckland. 2013 New Zealand
Re: K233821
Trade/Device Name: F&P Optiflow Oxygen Kit (AA451J) Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory gas humidifier Regulatory Class: Class II Product Code: BTT Dated: May 29, 2024 Received: May 29, 2024
Dear Reena Daken:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Ethan L. Nyberg -S
Ethan Nyberg, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and
{2}------------------------------------------------
Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233821
Device Name F&P Optiflow Oxygen Kit (AA451J)
### Indications for Use (Describe)
This breathing set is an accessory to the F&P 950 Respiratory Humidifier to be used within the limits of its stated technical specifications.
It is intended for delivery of heated humidified high flow respiratory gases to adult patients in a hospital.
This product can be used on multiple patients when used with a hydrophobic filter between the product and the patient interface for a maximum of 24 hours after setup.
This product is for use by appropriately qualified healthcare professionals who perform anesthesia care and airway management.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
As Required by 21 CFR 807.92
# I. SUBMITTER
| Company Name and<br>Address | Fisher & Paykel Healthcare Limited<br>15 Maurice Paykel Place<br>East Tamaki<br>Auckland 2013, New Zealand<br>Telephone: +64 9 574 0100 |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Prepared and Submitted by | Danica Tung |
| | Regulatory Affairs Market Manager |
| Contact Person | Reena Daken |
| | Regulatory Affairs Manager |
| | Telephone: +64 9 574 0100 |
| | Email: reena.daken@fphcare.co.nz |
| Date Prepared | 13 June 2024 |
| II. DEVICE | |
| Name of Device | F&P Optiflow™ Oxygen Kit (AA451J) |
| Common/Usual Name | Inspiratory tube |
| Classification Name | Respiratory gas humidifier |
| Regulatory Class | Class II (21 CFR §868.5450) |
| Product Code | BTT |
#### PREDICATE DEVICE lll.
- Primary Predicate Device: Optiflow™ Oxygen Kit, K211096 .
- Secondary Predicate: F&P 950 Respiratory Humidifier, K220703
{5}------------------------------------------------
#### DEVICE DESCRIPTION IV.
The Fisher & Paykel Healthcare (F&P) Optiflow™ Oxygen Kit is a multiple-patient use breathing tube kit. This F&P Optiflow™ Oxygen Kit is presented as the AA451J.
The kit consists of a dry line, water bag, humidification chamber, inspiratory limb, tubing clips and datechange stickers.
The flow source delivers gas through a dryline to a humidification chamber that sits on a humidifier. As the gas passes through the chamber, it is heated and humidified. The gas then flows through the inspiratory limb, through a hydrophobic filter and then through a patient interface into the patients' upper airway via the nose.
The AA451J kit delivers humidified respiratory gases at flows from 5 to 70 L/min. When used with an FDAcleared hydrophobic filter, the AA451J kit can be used on multiple patients. The kit is reprocessed between each patient. The kit can be exposed to a maximum of 30 reprocessing cycles and used for a maximum of 24 hours after setup.
The scope of this 510(k) submission is limited to the AA451J, F&P Optiflow™ Oxygen Kit. The flow source, humidifier, hydrophobic filter and interface are not in the scope of this submission.
#### INDICATIONS FOR USE V.
This breathing set is an accessory to the F&P 950 Respiratory Humidifier to be used within the limits of its stated technical specifications.
It is intended for delivery of heated humidified high flow respiratory gases to adult patients in a hospital.
This product can be used on multiple patients when used with a hydrophobic filter between the product and the patient interface for a maximum of 24 hours after setup.
This product is for use by appropriately qualified healthcare professionals who perform anesthesia care and airway management.
{6}------------------------------------------------
#### COMPARISON OF TECHNOLOGICAL CHARATCERISTICS WITH THE PREDICATE DEVICE VI.
| Design/Technological<br>Characteristic | Subject Device<br>(AA451J - K233821) | Primary Predicate Device<br>(AA403 - K211096) | Secondary Predicate Device<br>(AA451SU - K220703) | Similarity of subject<br>device to<br>primary/secondary<br>predicate device |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | This breathing set is an accessory to the<br>F&P 950 Respiratory Humidifier to be used<br>within the limits of its stated technical<br>specifications.<br>It is intended for delivery of heated humidified<br>high flow respiratory gases to adult patients in<br>a hospital.<br>This product can be used on multiple patients<br>when used with a hydrophobic filter between<br>the product and the patient interface for a<br>maximum of 24 hours after setup.<br>This product is for use by appropriately<br>qualified healthcare professionals who perform<br>anesthesia care and airway management. | This product delivers respiratory gases to<br>adult patients. It is intended for use with an<br>MR810 humidifier at flows from 5 to 70<br>L/min.<br>This product can be used on multiple<br>patients when used with a hydrophobic filter<br>between the product and the patient interface<br>for a maximum of 24 hours after set-up.<br>This product is indicated for the delivery of<br>Nasal High Flow (NHF) by appropriately<br>qualified healthcare professionals under the<br>direction of a physician anesthesiologist in a<br>medical procedure or surgical room.<br>Qualitative carbon dioxide sampling can be<br>used at nasal cannula flow rates from 5 to 50<br>L/min.<br>This product can be used for pre-<br>oxygenation and short-term supplemental<br>oxygenation (up to 10 minutes) during<br>intubation in operating rooms under the<br>direction of a physician anesthesiologist.<br>This product is not intended for apneic<br>ventilation. | The breathing set is an accessory to the<br>F&P 950 Respiratory Humidifier. It is<br>intended for delivery of heated humidified<br>respiratory gases to adult, adolescent and<br>child patients, within the limits of its stated<br>technical specifications. | Equivalent.<br>The subject device and<br>the predicate devices are<br>intended for use with<br>humidifiers.<br>The primary predicate is<br>intended for use with the<br>MR810 humidifier,<br>whereas the subject<br>device is intended for use<br>with the F&P 950<br>humidifier.<br>The subject device and<br>secondary predicate are<br>both intended for use<br>with the F&P 950<br>humidifier. |
| Design/Technological<br>Characteristic | Subject Device<br>(AA451J - K233821) | Primary Predicate Device<br>(AA403 - K211096) | Secondary Predicate Device<br>(AA451SU - K220703) | Similarity of subject<br>device to<br>primary/secondary<br>predicate device |
| Operation and safety features | | | | |
| Patient Population | Adult patients | Adult patients | Adult, adolescent and child patients | Identical to primary<br>predicate |
| Intended Use<br>Environment | Hospitals | Hospitals | Hospital or long-term care facility | Identical to primary<br>predicate |
| Availability | Prescription use | Prescription Use | Prescription use | Identical |
| Operating Principle | Breathing circuit kit contains a dryline,<br>humidification chamber, and inspiratory limb<br>to create a single limb for the delivery of high<br>flow therapy. | Breathing circuit kit contains a dryline,<br>humidification chamber, and inspiratory<br>limb to create a single limb for the delivery<br>of high flow therapy. | Breathing circuit kit contains a dryline,<br>humidification chamber, and inspiratory<br>limb to create a single limb for the delivery<br>of high flow therapy. | Identical |
| Reusability | Multi-patient use | Multi-patient use | Single Use | Identical to primary<br>predicate |
| Duration of Use | < 24 hours | < 24 hours | 14 Days | Identical to primary<br>predicate |
| Specifications | | | | |
| Specified Flow Range | 5 - 70 L/min | 5 to 70 L/min | 5 - 70 L/min | Identical |
| Sterility | Device not provided sterile | Device not provided sterile | Device not provided sterile | Identical |
| Design/Technological<br>Characteristic | Subject Device<br>(AA451J - K233821) | Primary Predicate Device<br>(AA403 - K211096) | Secondary Predicate Device<br>(AA451SU - K220703) | Similarity of subject<br>device to<br>primary/secondary<br>predicate device |
| Shelf Life | 3 years | 12 months | 3 years | Identical to secondary<br>predicate |
| Storage conditions | -10°C - 50°C | -10°C - 50°C | -10°C – 50°C | Identical |
| Biocompatibility and Materials | | | | |
| Assessment | Testing performed according to ISO 10993-1:<br>Fifth edition 2018-08 and ISO 18562-1: First<br>Edition 2017-03 | Testing performed according to ISO 10993-<br>1: Fifth edition 2018-08 and ISO 18562-1:<br>First Edition 2017-03 | Testing performed according to ISO 10993-<br>1: Fifth edition 2018-08 and ISO 18562-1:<br>First Edition 2017-03 | Identical |
| Components | | | | |
| Humidification<br>Chamber | Included | Included | Included | The humidification<br>chamber of the subject<br>device is identical to the<br>secondary predicate. |
| Inspiratory Limb | Included | Included | Included | The inspiratory limb of<br>the subject device is<br>identical to the<br>secondary predicate. |
| Design/Technological<br>Characteristic | Subject Device<br>(AA451J - K233821) | Primary Predicate Device<br>(AA403 - K211096) | Secondary Predicate Device<br>(AA451SU - K220703) | Similarity of subject<br>device to<br>primary/secondary<br>predicate device |
| Dryline | Included | Included | Included | Identical |
| Circuit Clip | Included | Included | Included | Equivalent to primary<br>predicate<br>The tubing clip of the<br>subject device is<br>identical to the<br>secondary predicate. |
3
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
#### VII. PERFORMANCE DATA
## Summary of Non-Clinical Tests
Performance testing of the F&P Optiflow™ Oxygen Kit was completed and confirms the subject device does not raise new questions of safety and effectiveness. The testing provided demonstrates that the reusability characteristic and environment of use of the subject device is substantially equivalent to the predicates.
The F&P Optiflow™ Oxygen Kit has been tested to the applicable requirements to the following standards:
- ISO 10993-1:2018 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- ISO 80601-2-74:2017 Part 2-74 Particular requirements for basic safety and essential performance of respiratory humidifying equipment
- IEC 60601-1 Edition 3.2 (2020) Medical electrical equipment Part 1: General ● requirements for basic safety and essential performance
- ISO 5367:2014
- AAMI TIR12:2010
- AAMI ST98:2022
#### VIII. CONCLUSION
The F&P Optiflow™ Oxygen Kit is substantially equivalent to the predicates based on patient population, intended uses, comparison of the technological characteristics, and performance. In addition, the conclusions drawn from the non-clinical tests demonstrate that the device is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.