The REDEMPTION™ Charcot Plating System bone plates and screws are indicated for use in bone fractures, osteotomies, and fixation of bones and bone fragments in the upper and lower extremities, primarily of the hand, wrist, foot, and digits. Specific examples include: medial column fusion (talus, navicular, cuboid, first metatarsal) for patients with osteopenic bone in neuropathic osteoarthropathy (Charcot).
Device Story
System comprises bone plates, screws, and surgical instruments for orthopedic reconstruction; provides polyaxial locking technology and low-profile design. Optimized for osteopenic bone in Charcot patients. Inputs: surgical site anatomy; surgeon selects appropriate plate/screw configuration. Operation: polyaxial locking/non-locking screws (4.0mm/5.5mm) provide fixation; instrumentation (wires, drills, drivers, guides) facilitates implantation. Output: rigid internal fixation of bone fragments. Used in clinical/surgical settings by orthopedic surgeons. Benefits: stabilization of fractures/fusions in compromised bone; enables bone healing.
Clinical Evidence
Bench testing only. Performance data includes static four-point bend testing (ASTM F382-17), screw torsional properties (ASTM F543-17 A1), screw insertion/removal torque (ASTM F543-17 A2), and screw pullout testing (ASTM F543-17 A3). Biocompatibility, cleaning, and sterilization validations were leveraged from predicate devices (K212348, K221558, K230204).
Technological Characteristics
Implants: Titanium alloy (Ti-6Al-4V, ASTM F136). Instruments: Stainless steel, aluminum, silicone. Features: Polyaxial locking and non-locking screws (4.0mm, 5.5mm diameter). Mechanical testing per ASTM F382-17 and ASTM F543-17. Sterilization: Validated per ANSI/AAMI/ISO 17665-1.
Indications for Use
Indicated for bone fractures, osteotomies, and fixation of bone fragments in upper/lower extremities (hand, wrist, foot, digits). Specifically for medial column fusion (talus, navicular, cuboid, first metatarsal) in patients with osteopenic bone due to neuropathic osteoarthropathy (Charcot).
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
August 28, 2024
Vilex, LLC Brock Johnson President 111 Moffitt Street McMinnville, Tennessee 37110
Re: K233816
Trade/Device Name: REDEMPTION Charcot Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS. HWC Dated: August 2, 2024 Received: August 5, 2024
Dear Brock Johnson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements. the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combinationproducts/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-reportmedical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuingeducation/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumereducation-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Christopher Ferreira -S
Christopher Ferreira, MS, RAC Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Submission Number (if known)
K233816
Device Name
REDEMPTION™ Charcot Plating System
Indications for Use (Describe)
The REDEMPTION™ Charcot Plating System bone plates and screws are indicated for use in bone fractures, osteotomies, and fixation of bones and bone fragments in the upper and lower extremities, primarily of the hand, wrist, foot, and digits. Specific examples include: medial column fusion (talus, navicular, cuboid, first metatarsal) for patients with osteopenic bone in neuropathic osteoarthropathy (Charcot).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### I. Submitter
Vilex LLC
111 Moffitt Street
McMinnville, TN 37110
Contact Person: Brock Johnson, President of Vilex
Phone: (801) 916-4157
brock.johnson@vilex.com
Date Prepared: May 27, 2022
#### II. Device
| Device Proprietary Name: | REDEMPTION TM Charcot Plating System |
|--------------------------|------------------------------------------------------------------------------------------|
| Common or Usual Name: | Plate, Fixation, Bone; Screw, Fixation, Bone |
| Classification Name: | Single/multiple component metallic bone fixation appliances and<br>accessories (Primary) |
| | Smooth or threaded metallic bone fixation fastener |
| Regulation Number: | 21 CFR 888.3030 (Primary) |
| | 21 CFR 888.3040 |
| Product Code: | HRS (Primary), HWC |
| Device Classification | II |
{4}------------------------------------------------
Traditional 510(k)
#### III. Predicate Device
Substantial equivalence is claimed to the following devices:
- Vilex ALPHALOKTM Plating System (K221558) Primary Predicate ●
- . Vilex ALPHALOK™ Plating System (K212348, K230204)-Reference Devices
- Stryker [Wright Medical] SALVATION® 3Di Plating System (K140792 and K140408) -Additional Predicate Devices
#### IV. Device Description
The REDEMPTION™ Charcot Plating System contains bone plates, screws, and accompanying instruments to aid in the surgical procedures. The system is a multi-indication reconstruction solution providing polyaxial locking technology and low-profile design. The system is optimized for use in osteopenic bone, such as those of neuropathic osteoarthropathy (Charcot) patients.
The system contains poly-axial locking and non-locking 4.0mm and 5.5mm diameter screws with threads optimized for osteopenic bone purchase.
All implant components are manufactured from titanium (Ti-6AI-4V, ASTM F136).
Specific instrumentation including wires, drills, torx drivers, and drill guides are required for use with the system. The REDEMPTION™ instruments are manufactured from stainless steel, with a few Aluminum and Silicone components.
#### V. Indications for Use
The REDEMPTION™ Charcot Plating System bone plates and screws are indicated for use in bone fractures, osteotomies, and fixation of bones and bone fragments in the upper and lower extremities, primarily of the hand, wrist, foot, ankle, and digits. Specific examples include: medial column fusion (talus, navicular, cuboid, first metatarsal) for patients with osteopenic bone in neuropathic osteoarthropathy (Charcot).
#### VI. Comparison of Technological Characteristics
The subject and predicate devices have similar intended uses and include explicit Charcot indications. Both systems also share equivalent core features and mechanical characteristics.
The systems are intended to be used in the hand, foot, and ankle along with other bones of the upper and lower extremities. Specific examples include: medial column fusion (talus, navicular, cuboid, first metatarsal) for patients with osteopenic bone in neuropathic osteoarthropathy (Charcot). The subject and
{5}------------------------------------------------
| K233816 | |
|---------|--|
|---------|--|
predicate systems include bone plates along with locking and non-locking screws with similar implant designs made from titanium alloy material. Similar instrumentation is included in all the system.
Any minor technological differences between the subject device and predicate devices do not raise different questions of safety or effectiveness and substantial equivalence is demonstrated through the testing described below.
#### VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination:
- Static four point bend testing per ASTM F382-17
- Screw Torsional Properties per ASTM F543-17 A1
- . Screw Insertion and Removal Torque per ASTM F543-17 A2
- . Screw Pullout Testing per ASTM F543-17 A3
In addition, cleaning and sterilization validations, performed in accordance with ANSI/AAMI/ISO 17665-1, from the applicant's own predicate device were leveraged. Biocompatibility, cleaning and sterilization are identical to K212348, K221558, and K230204. No changes have been made with respect to material, manufacturing, cleaning, or sterilization.
#### VIII. Conclusion
The information provided above supports that the REDEMPTION™ Charcot Plating System is as safe and effective as the predicate device. Any minor differences in design between the subject and predicate devices do not raise any new questions of safety and effectiveness as demonstrated by mechanical testing. Therefore, it is concluded that the REDEMPTION™ Charcot Plating System is substantially equivalent to the predicate device.
-------------------------This space intentionally left blank------------------------------------------------------------------------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.