3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD: Use the 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD in conjunction with the 3M™ Attest™ Auto-reader 490 having software version 4.0.0 or greater or a 3M™ Attest™ Mini Auto-reader 490M or a 3M™ Attest™ Auto-reader 490H as a standard method of routine monitoring and performance qualification of vaporized hydrogen peroxide sterilization processes in the follow systems: STERRAD 100S® Sterilization System, STERRAD NX® Sterilization System (Standard and Advanced cycles), STERRAD 100NX® Sterilization System (Standard, Flex, Express, and Duo cvcles), STERRAD NX® with ALLClear® Technology Sterilization System (Standard and Advanced cycles), STERRAD 100NX® with ALLClear® Technology Sterilization System (Standard, Flex, Express, and Duo cycles). V-PRO® 1 Low Temperature Sterilization System (Lumen cycle), V-PRO® 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles), V-PRO® maX Low Temperature Sterilization System (Lumen, and Flexible cycles), V-PRO® 60 Low Temperature Sterilization System (Lumen, and Flexible cycles), V-PRO® maX 2 Low Temperature Sterilization System (Lumen, Flexible, and Fast Non Lumen cycles). V-PRO® s2 Low Temperature Sterilization System (Lumen, Flexible, and Fast cycles) 3M™ Attest™ Auto-reader (490): The 3M™ Attest™ Auto reader 490 is designed to incubate and automatically read 3M™ Attest™ Rapid Readout Biological Indicators 1295 and 3M™ Attest™ Super Rapid Readout Biological Indicators, catalog numbers 1491, 1492V, and 1592 at 60°C for a final fluorescent result at 24 minutes. 3M™ Attest™ Auto-reader (490H): The 3M™ Attest™ Auto-reader 490H is designed to incubate and automatically read 3M™ Attest™ Rapid Readout Biological Indicators 1295, 3M™ Attest™ Super Rapid Readout Biological Indicators, catalog numbers 1491 and 1492V, and 3M™ Attest Super Rapid Steam Biological Indicators 1592 at 60°C for a final fluorescent result at 24 minutes. 3M™ Attest™ Mini Auto-reader (490M): The 3M™ Attest™ Mini Auto-reader 490M is designed to incubate and automatically read 3M™ Attest™ Rapid Readout Biological Indicators 1295 and 3M™ Attest™ Super Rapid Readout Biological Indicators, catalog numbers 1491, 1492V and 1592, at 60°C for a final fluorescent result at 24 minutes.
Device Story
Device monitors vaporized hydrogen peroxide sterilization processes; consists of 1295PCD challenge pack (clear plastic shell, two channels, foil lid) containing 1295 Biological Indicator (BI) and 1348 Tri-Metric Chemical Indicator (CI). 1295 BI uses Geobacillus stearothermophilus spores; enzyme system produces fluorescence detected by 490/490H/490M Auto-readers at 60°C. Auto-reader provides final fluorescent result at 24 minutes; indicates sterilization failure if fluorescence detected. 1348 CI verifies exposure time, temperature, and concentration. Used in healthcare facilities by staff to monitor sterilizer performance. Output affects clinical decision-making regarding sterilization adequacy; ensures patient safety by confirming effective sterilization of instruments.
Clinical Evidence
No clinical data; bench testing only. Performance verified via PCD resistance characterization, functionality in pass/fail cycles, peel force testing (ASTM F88/F2824), and cycle performance testing in claimed sterilizers.
Technological Characteristics
Clear plastic tray with channels and foil lid; contains Geobacillus stearothermophilus spores (>10^6). Sensing principle: enzymatic fluorescence detection. Energy source: Auto-reader (60°C incubation). Connectivity: Standalone. Software: Version 4.0.0 or greater. Standards: ANSI/AAMI/ISO 11138-1:2017, ASTM F88/F88M-15, ASTM F2824-10.
Indications for Use
Indicated for routine monitoring and performance qualification of vaporized hydrogen peroxide sterilization processes in healthcare facilities using specified STERRAD and V-PRO sterilization systems.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
Sterilucent Self-Contained Biological Indicator, Sterilucent Lumen Cycle Process Challenge Device, Sterilucent Flexible Cycle Process Challenge Device (K192001)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
March 15, 2024
3M Company Michelle Larsen Advanced Regulatory Affairs Specialist 3M Center 2510 Conway Ave., Building 275-5W-06 St. Paul, Minnesota 55144
#### Re: K233814
Trade/Device Name: 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack (1295PCD); 3MTM Attest™ Auto-reader (490); 3MTM Attest™ Auto-reader (490H); 3MTM Attest™ Mini Auto-reader (490M) Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: February 14, 2024 Received: February 14, 2024
Dear Michelle Larsen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen A. Anisko -S
Digitally signed by Stephen A. Anisko -S Date: 2024.03.15 11:42:02 -04'00'
for Christopher Dugard Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices
{2}------------------------------------------------
OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K233814
Device Name
3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack (1295PCD);
3M™ Attest™ Auto-reader (490);
3M™ Attest™ Auto-reader (490H);
3MTM Attest™ Mini Auto-reader (490M)
ndications for Use (Describe)
3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack (1295PCD): Use the 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD in conjunction with the 3M™ Attest™ Auto-reader 490 having software version 4.0.0 or greater or a 3M™ Attest™ Mini Auto-reader 490M or a 3M™ Attest™ Auto-reader 490H as a standard method of routine monitoring and performance qualification of vaporized hydrogen peroxide sterilization processes in the follow systems:
STERRAD 100S® Sterilization System,
STERRAD NX® Sterilization System (Standard and Advanced cycles),
STERRAD 100NX® Sterilization System (Standard, Flex, Express, and Duo cvcles),
STERRAD NX® with ALLClear® Technology Sterilization System (Standard and Advanced cycles), STERRAD 100NX® with ALLClear® Technology Sterilization System (Standard, Flex, Express, and Duo cycles).
V-PRO® 1 Low Temperature Sterilization System (Lumen cycle),
V-PRO® 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles),
V-PRO® maX Low Temperature Sterilization System (Lumen, and Flexible cycles),
V-PRO® 60 Low Temperature Sterilization System (Lumen, and Flexible cycles),
V-PRO® maX 2 Low Temperature Sterilization System (Lumen, Flexible, and Fast Non Lumen cycles).
V-PRO® s2 Low Temperature Sterilization System (Lumen, Flexible, and Fast cycles)
3M™ Attest™ Auto-reader (490):
The 3M™ Attest™ Auto reader 490 is designed to incubate and automatically read 3M™ Attest™ Rapid Readout Biological Indicators 1295 and 3M™ Attest™ Super Rapid Readout Biological Indicators, catalog numbers 1491, 1492V, and 1592 at 60°C for a final fluorescent result at 24 minutes.
3M™ Attest™ Auto-reader (490H):
The 3M™ Attest™ Auto-reader 490H is designed to incubate and automatically read 3M™ Attest™ Rapid Readout Biological Indicators 1295, 3M™ Attest™ Super Rapid Readout Biological Indicators, catalog numbers 1491 and 1492V, and 3M™ Attest Super Rapid Steam Biological Indicators 1592 at 60°C for a final fluorescent result at 24 minutes.
3M™ Attest™ Mini Auto-reader (490M):
The 3M™ Attest™ Mini Auto-reader 490M is designed to incubate and automatically read 3M™ Attest™ Rapid Readout Biological Indicators 1295 and 3M™ Attest™ Super Rapid Readout Biological Indicators, catalog numbers 1491, 1492V and 1592, at 60°C for a final fluorescent result at 24 minutes.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
{4}------------------------------------------------
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the 3M logo. The logo is red and features the number "3" followed by the letter "M". The logo is simple and recognizable. The color red is bold and eye-catching.
# 510(k) Summary for 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD, 3MTM Attest™ Auto-reader 490, 3MTM Attest™ Auto-reader 490H, 3MTM Attest™ Mini Auto-reader 490M K233814
#### Sponsor Information:
3M Company 3M Center, Bldg. 275-5W-06 St. Paul. MN 55144-1000
Contact: Michelle M. Larsen Advanced Regulatory Affairs Specialist Phone Number: (651) 467-3991 Fax Number: (651) 737-5320 Email: mmlarsen@solventum.com
#### Date of Summary: 13, March 2024
#### 1. Device Name and Classification:
Common or Usual Name: Sterilization Biological Indicator
{6}------------------------------------------------
#### TRADITIONAL PREMARKET NOTIFICATION [510(k) 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD M™ Attest™ Auto-readers 490 and 490H 3MTM Attest™ Mini Auto-reader 490M
| Proprietary Name: | 3MTM Attest™ Super Rapid Vaporized Hydrogen Peroxide<br>Clear Challenge Pack 1295PCD,<br>3MTM Attest™ Auto-reader 490,<br>3MTM Attest™ Auto-reader 490H,<br>3MTM Attest™ Mini Auto-reader 490M |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name: | Indicator, Biological Sterilization Process |
| Device Classification: | Class II, 21 CFR 880.2800(a) |
| Product Code: | FRC |
# 2. Predicate Device:
Sterilucent Self-Contained Biological Indicator, Sterilucent Lumen Cycle Process Challenge Device, Sterilucent Flexible Cycle Process Challenge Device, K192001
# 3. Description of Device:
The 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD is designed as a standard method of rapid and reliable routine monitoring and performance qualification of vaporized hydrogen peroxide sterilization processes.
The 1295PCD Challenge Pack consists of a clear plastic shell, with two individual channels allowing for air removal and sterilant penetration. Those two individual channels connect to a cavity containing the monitoring products and all are covered by a foil lid. The 1295PCD is a single-use device.
Each 1295PCD Challenge Pack contains a 3M™ Attest™ Rapid Readout Biological Indicator (BI) 1295 (pink cap) and a 3MTM Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator (CI) 1348. The 1348 CI verifies that the stated values for the three critical parameters of exposure time, temperature, and concentration of vaporized hydrogen peroxide have been achieved and offers an ACCEPT or REJECT reading. The 1295 BI is specifically designed for rapid and reliable routine monitoring of vaporized hydrogen peroxide sterilization processes when used in conjunction with the 3MTM Attest™ Autoreader 490 having software version 4.0.0 or greater or a 3M™ Attest™ Mini Auto-reader 490M or a 3M™ Attest™ Auto-reader 490H. The presence of fluorescence within the specified incubation time for the 1295 BI in the 490, 490M, or the 490H Auto-reader indicates a sterilization process failure. A printed chemical process indicator is present on the cap of the 1295 BI and is visible through the clear plastic shell of the challenge pack. The
{7}------------------------------------------------
process indicator turns color upon exposure to vaporized hydrogen peroxide and is used by the customer to verify that the challenge pack was exposed to vaporized hydrogen peroxide.
# 4. Indications for Use
#### 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack (1295PCD):
Use the 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD in conjunction with the 3M™ Attest™ Auto-reader 490 having software version 4.0.0 or greater or a 3M™ Attest™ Mini Auto-reader 490M or a 3M™ Attest™ Auto-reader 490H as a standard method of routine monitoring and performance qualification of vaporized hydrogen peroxide sterilization processes in the follow systems:
STERRAD 100S® Sterilization System
STERRAD NX® Sterilization System (Standard and Advanced cycles)
STERRAD 100NX® Sterilization System (Standard, Flex, Express, and Duo cycles)
STERRAD NX® with ALLClear® Technology Sterilization System (Standard and Advanced cvcles)
STERRAD 100NX® with ALLClear® Technology Sterilization System (Standard, Flex, Express, and Duo cycles)
V-PRO® 1 Low Temperature Sterilization System (Lumen cycle)
V-PRO® 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles)
V-PRO® maX Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cvcles)
V-PRO® 60 Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles)
V-PRO® maX 2 Low Temperature Sterilization System (Lumen, Non Lumen, Flexible, and Fast Non Lumen cycles)
V-PRO® s2 Low Temperature Sterilization System (Lumen, Flexible, and Fast cycles)
# 3MTM Attest™ Auto-reader (490):
The 3M™Attest™ Auto reader 490 is designed to incubate and automatically read 3M™ Attest™ Rapid Readout Biological Indicators 1295 and 3M™ Attest™ Super Rapid Readout Biological Indicators, catalog numbers 1491, 1492V, and 1592 at 60°C for a final fluorescent result at 24 minutes.
# 3MTM Attest™ Auto-reader (490H):
The 3M™ Attest™ Auto-reader 490H is designed to incubate and automatically read 3M™ Attest™ Rapid Readout Biological Indicators 1295, 3M™ Attest™ Super Rapid Readout Biological Indicators, catalog numbers 1491 and 1492V, and 3M™ Attest Super Rapid Steam Biological Indicators 1592 at 60°C for a final fluorescent result at 24 minutes.
{8}------------------------------------------------
### 3MTM Attest™ Mini Auto-reader (490M):
The 3M™ Attest™ Mini Auto-reader 490M is designed to incubate and automatically read 3MTM Attest™ Rapid Readout Biological Indicators 1295 and 3MTM Attest™ Super Rapid Readout Biological Indicators, catalog numbers 1491, 1492V and 1592, at 60ºC for a final fluorescent result at 24 minutes.
# 5. Comparison of Subiect Device to Predicate Device
# Indications for Use
The 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD and the predicate device Sterilucent Process Challenge Device (PCD) test pack have the same intended use. Both the subject device and the predicate device are Single Use biological indicator process challenge devices that are intended to be used in healthcare facilities to accompany products being sterilized and to monitor adequacy of the sterilization process. The Indications for Use are similar in that both the predicate and subject challenge test packs are specifically designed as a method of routine monitoring and performance qualification of vaporized hydrogen peroxide sterilization processes. The subject device covers different sterilizers and cycles compared to the predicate device. The differences in Indications for Use do not alter the fundamental Intended Use of these products as sterilization process monitors and do not raise new questions of safety and effectiveness.
# Technological Characteristics
The 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD has similar technological characteristics compared to the predicate Sterilucent Process Challenge Device (PCD) test pack. The subject device is designed and constructed with a clear plastic tray with channels to the cavity containing the Biological Indicator and Chemical Indicator sealed with a foil lid whereas the predicate device consists of a clear polymer vial containing the Biological Indicator and a challenge tube on one end. Both the subject and predicate challenge packs are designed to increase resistance beyond that measured with a standalone BI and both represent a challenge to the sterilization process based on worse case cycle with load. The process indicator on the predicate challenge pack is present on the PCD whereas the process indicator on the subject device is present on the 1295 BI cap which is visible through the clear test pack. Both process indicators change color upon exposure to vaporized hydrogen peroxide (VH2O2) and is used by the customer to verify the challenge pack was exposed to VH2O2.
The differences in technological characteristics between the subject and predicate challenge pack are illustrated in the Device Comparison Table (Table 1).
{9}------------------------------------------------
| Feature | Submission Device:<br>3MTM AttestTM<br>Super Rapid Vaporized<br>Hydrogen Peroxide<br>Clear Challenge Pack<br>1295PCD and<br>3MTM AttestTM Auto-<br>reader 490 and 490H<br>and 3MTM AttestTM<br>Mini Auto-reader 490M | Predicate Device<br>(K192001):<br>Sterilucent Self-<br>Contained Biological<br>Indicator, Sterilucent<br>Lumen Cycle Process<br>Challenge Device,<br>Sterilucent Flexible<br>Cycle Process<br>Challenge Device | Comparison |
|----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Single Use sterilization<br>process indicator | Single Use sterilization<br>process indicator | Identical |
| Indications for use | Use the 3MTM AttestTM<br>Super Rapid Vaporized<br>Hydrogen Peroxide Clear<br>Challenge Pack<br>1295PCD in conjunction<br>with the 3MTM AttestTM<br>Auto-reader 490 having<br>software version 4.0.0 or<br>greater or a 3MTM<br>AttestTM Mini Auto-<br>reader 490M or a 3MTM<br>AttestTM Auto-reader<br>490H as a standard<br>method of routine<br>monitoring and<br>performance qualification<br>of vaporized hydrogen<br>peroxide sterilization<br>processes in the follow<br>systems:<br>STERRAD 100S®<br>Sterilization System,<br>STERRAD NX®<br>Sterilization System<br>(Standard and Advanced<br>cycles), STERRAD<br>100NX® Sterilization<br>System (Standard, Flex,<br>Express, and Duo<br>cycles), STERRAD NX®<br>with ALLClear®<br>Technology Sterilization | The Sterilucent Process<br>Challenge Device (PCD)<br>test pack is used for<br>routine monitoring of the<br>Sterilucent HC 80TT<br>Sterilizer Lumen and<br>Flexible Cycles. The<br>Sterilucent PCD may<br>also be used for<br>performance<br>qualification of the<br>Sterilucent HC 80TT<br>sterilizer Lumen and<br>Flexible Cycles during<br>initial installation, after<br>relocation, major repairs<br>or mal functions, or after<br>sterilization process<br>failures.<br>Both the Sterilucent<br>Lumen Cycle PCD and<br>the Sterilucent Flexible<br>Cycle PCD are intended<br>to have greater resistance<br>than the stand alone<br>SCBI. Both devices are<br>designed to have<br>increased resistance<br>beyond the sterilization<br>half-cycle, but complete<br>inactivation upon | The Indications for<br>Use are similar in<br>that the subject<br>device and the<br>predicate device are<br>used for routine<br>monitoring and<br>performance<br>qualification of<br>vaporized hydrogen<br>peroxide sterilization<br>processes. The<br>subject device covers<br>different sterilizers<br>and cycles compared<br>to the predicate<br>device.<br>The Indications for<br>Use for the subject<br>device are identical<br>to the 3M 1295 BI<br>(K210277) and 3M<br>Tri-Metric CI 1348<br>(K203284) in terms<br>of sterilizers and<br>cycles covered.<br>The subject<br>challenge pack<br>contains the 3M<br>1295 BI which is |
| Feature | Submission Device: | Predicate Device (K192001): | Comparison |
| | 3MTM AttestTM Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD and 3MTM AttestTM Auto-reader 490 and 490H and 3MTM AttestTM Mini Auto-reader 490M | Sterilucent Self-Contained Biological Indicator, Sterilucent Lumen Cycle Process Challenge Device, Sterilucent Flexible Cycle Process Challenge Device | |
| | System (Standard and Advanced cycles), STERRAD 100NX® with ALLClear® Technology Sterilization System (Standard, Flex, Express, and Duo cycles), V-PRO® 1 Low Temperature Sterilization System (Lumen cycle), V-PRO® 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles), V-PRO® maX Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles), V-PRO® 60 Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles), V-PRO® maX 2 Low Temperature Sterilization System (Lumen, Non Lumen, Flexible, and Fast Non Lumen cycles), V-PRO® s2 Low Temperature Sterilization System (Lumen, Non Lumen, Flexible, and Fast cycles) | exposure to the full cycle. | intended to be used in conjunction with the 3MTM AttestTM Auto-reader 490 or the 3MTM AttestTM Auto-reader 490H or the 3MTM AttestTM Mini Auto-reader 490M cleared under K210277. |
| Biological Indicator (BI) | 3MTM AttestTM Rapid Readout Biological Indicator 1295 with ≥ 1 x 106 <i> Geobacillus</i> | Sterilucent HC 80TT SCBI, quartz fiber carrier inoculated with >106 <i> Geobacillus</i> | Both the subject device and the predicate device utilize Biological |
| | Submission Device: | Predicate Device (K192001): | |
| Feature | 3MTM AttestTM Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD and 3MTM AttestTM Auto-reader 490 and 490H and 3MTM AttestTM Mini Auto-reader 490M | Sterilucent Self-Contained Biological Indicator, Sterilucent Lumen Cycle Process Challenge Device, Sterilucent Flexible Cycle Process Challenge Device | Comparison |
| | <i>stearothermophilus</i> spores | <i>stearothermophilus</i> spores per disc | Indicators with >106 <i>Geobacillus stearothermophilus</i> spores. |
| BI - Mechanism of Action | The 1295 BI utilizes an enzyme system, which is generated naturally within growing cells of <i>Geobacillus stearothermophilus</i> . The enzyme in its active state is detected by measuring the fluorescence produced by the enzymatic reaction with a non-fluorescent substrate. The resultant fluorescent by product is detected in an Auto-reader. The presence of fluorescence within the specified incubation time in the Auto-reader indicates a sterilization process failure. The 1295 BI can also indicate the presence of <i>G. stearothermophilus</i> organisms by a visual pH color change reaction. | The Sterilucent HC 80TT SCBI indicates the presence of <i>G. stearothermophilus</i> organisms by a visual pH color change reaction. | Similar, the 1295 BI utilizes a fluorescence result and also a visual pH color change reaction. |
| Feature | Submission Device:<br>3MTM AttestTM<br>Super Rapid Vaporized<br>Hydrogen Peroxide<br>Clear Challenge Pack<br>1295PCD and<br>3MTM AttestTM Auto-<br>reader 490 and 490H<br>and 3MTM AttestTM<br>Mini Auto-reader 490M<br>(Tested at 10 mg/L<br>vaporized hydrogen<br>peroxide) | Predicate Device<br>(K192001):<br>Sterilucent Self-<br>Contained Biological<br>Indicator, Sterilucent<br>Lumen Cycle Process<br>Challenge Device,<br>Sterilucent Flexible<br>Cycle Process<br>Challenge Device | Comparison |
| BI - Resistance<br>Characteristics<br>• D-value | ≥ 1 sec. | 5.7 - 9.1 sec. | Similar, both BIs are<br>characterized with a<br>D-value. |
| BI - Resistance<br>Characteristics<br>• Survival/Kill<br>Window | (Tested at 10 mg/L<br>vaporized hydrogen<br>peroxide)<br>Survival Time ≥ 5 sec.<br>Kill Time = 7 min. | 25.0 sec. (All survive)<br>140 sec. (All kill) | Similar, both BIs are<br>characterized with<br>Survival and Kill<br>testing. |
| BI – Culture<br>Conditions | Crushable glass ampoule<br>containing a growth<br>media and a fluorescence<br>indicator which is<br>correlated to an optional<br>use pH indicator | Crushable glass ampoule<br>containing modified<br>Tryptic Soy Broth with a<br>pH indicator | Similar |
| Chemical Indicator<br>(CI) | 3MTM AttestTM Vaporized<br>Hydrogen Peroxide Tri-<br>Metric Chemical<br>Indicator 1348 | | The Chemical<br>Indicator in the<br>subject challenge<br>pack is able to<br>independently<br>monitor three critical<br>parameters: exposure<br>time, temperature,<br>and concentration of<br>vaporized hydrogen<br>peroxide. The<br>predicate device<br>does not contain a<br>separate chemical<br>indicator. |
| CI – Endpoint<br>Specifications<br>(Minimum Stated<br>Values) | VH2O2 Concentration –<br>5.1 mg/L<br>Exposure Time – 1<br>minute<br>Temperature – 50°C | NA | independently<br>monitor three critical<br>parameters: exposure<br>time, temperature,<br>and concentration of<br>vaporized hydrogen<br>peroxide. The |
| CI - Color change | Blue toward pink along<br>the reactive chemistry<br>strip | | predicate device<br>does not contain a<br>separate chemical<br>indicator. |
| General<br>Design/Construction | Clear plastic tray with<br>channels to the cavity | Clear polymer vial with<br>polymer caps and | Both the subject and<br>predicate challenge |
| Feature | Submission Device: 3MTM AttestTM<br>Super Rapid Vaporized<br>Hydrogen Peroxide<br>Clear Challenge Pack<br>1295PCD and<br>3MTM AttestTM Auto-<br>reader 490 and 490H<br>and 3MTM AttestTM<br>Mini Auto-reader 490M | Predicate Device<br>(K192001):<br>Sterilucent Self-<br>Contained Biological<br>Indicator, Sterilucent<br>Lumen Cycle Process<br>Challenge Device,<br>Sterilucent Flexible<br>Cycle Process<br>Challenge Device | Comparison |
| | containing the indicators,<br>acting as a challenge to<br>limit air removal and<br>VH2O2 penetration,<br>sealed with a foil lid. | challenge tube on one<br>end.<br>Tube of various inner<br>diameter and lengths<br>tailored to individual<br>sterilization cycle. | packs are designed<br>to increase resistance<br>beyond that<br>measured with a<br>standalone BI and<br>both represent a<br>challenge to the<br>sterilization process<br>based on worse case<br>cycle with load. |
| Mechanism to<br>distinguish processed<br>and unprocessed<br>challenge pack | Process indicator present<br>on 1295 BI cap is visible<br>through PCD and turns<br>color from blue towards<br>pink upon vaporized<br>hydrogen peroxide<br>exposure. | SPSmedical VH2O2<br>Chemical Indicator<br>placed on the outside of<br>the vial. | Both the subject and<br>predicate device<br>contain a process<br>indicator that<br>changes color upon<br>vaporized hydrogen<br>peroxide exposure<br>that is used by the<br>customer to verify<br>that the challenge<br>pack was exposed to<br>VH2O2. |
| Shelf Life | 6 months | 12 months from date of<br>manufacture | Shelf life will be<br>extended as real time<br>data is available. |
# Table 1: Device Comparison Table
{10}------------------------------------------------
# 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD 3M™ Attest™ Auto-readers 490 and 490H
{11}------------------------------------------------
# 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD 3M™ Attest™ Auto-readers 490 and 490H
{12}------------------------------------------------
# 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD 3M™ Attest™ Auto-readers 490 and 490H
{13}------------------------------------------------
# 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD 3M™ Attest™ Auto-readers 490 and 490H
{14}------------------------------------------------
# 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD 3M™ Attest™ Auto-readers 490 and 490H
| | | 3M™ Attest™ Mini Auto-reader 490M | |
|--|--|-----------------------------------|--|
|--|--|-----------------------------------|--|
| Feature | Submission Device:<br>3MTM AttestTM<br>Super Rapid Vaporized<br>Hydrogen Peroxide<br>Clear Challenge Pack<br>1295PCD and<br>3MTM AttestTM Auto-<br>reader 490 and 490H<br>and 3MTM AttestTM<br>Mini Auto-reader 490M | Predicate Device<br>(K192001):<br>Sterilucent Self-<br>Contained Biological<br>Indicator, Sterilucent<br>Lumen Cycle Process<br>Challenge Device,<br>Sterilucent Flexible<br>Cycle Process<br>Challenge Device | Comparison |
|-------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Accessories | 3MTM AttestTM Auto-<br>reader 490 or the 3MTM<br>AttestTM Auto-reader<br>490H or the 3MTM<br>AttestTM Mini Auto-<br>reader | NA | The subject<br>challenge pack<br>contains the 3M<br>1295 BI which is<br>intended to be used<br>in conjunction with<br>the 3MTM AttestTM<br>Auto-reader 490 or<br>the 3MTM AttestTM<br>Auto-reader 490H or<br>the 3MTM AttestTM<br>Mini Auto-reader<br>490M cleared under<br>K210277. |
______________________________________________________________________________________________________________________________________________________________________________
{15}------------------------------------------------
### 6. Nonclinical Comparison to the Predicate Device
The differences between the subject and predicate device have been evaluated through performance tests for the 3MTM Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD.
Three lots of 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD, containing three lots of 1295 BIs and three lots of 1348 Cls, were tested side by side with standalone indicators 1295 BIs and 1348Cls. Testing was completed per the Guidance for Industry and FDA Staff, Biological Indicator Premarket Notification [510(k)] Submissions, October 4, 2007, and ANSI/AAMI/ISO 11138-1:2017 Sterilization of Health Care Products-Biological Indicators-Part 1: General Requirements.
Four lots of 3MTM Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD were subjected to simulated use testing in healthcare sterilizers. Testing was completed per the Guidance for Industry and FDA Staff, Biological Indicator Premarket Notification [510(k)] Submissions, October 4, 2007.
Three lots of 3M™ Attest™ Super Rapid Vaporized Hydrogen Peroxide Clear Challenge Pack 1295PCD were tested for peel force of the heat seal foil lid from the plastic shell. The validated method used to perform the testing is based on ASTM F88 / F88M-15: Standard Test Method for Seal Strength of Flexible Barrier and ASTM F2824 – 10: Standard Test Method for Mechanical Seal Strength Testing for Round Cups and Bowl Containers with Flexible Peelable Lids.
| Item Tested | Test | Standard | Purpose | Acceptance Criteria | Results |
|---------------------------------------------------------------------------------------------|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|---------|
| 1295PCD<br>Challenge<br>Packs and<br>Standalone<br>chemical and<br>biological<br>indicators | PCD Resistance<br>Characterization | FDA<br>Guidance1 and,<br>ANSI/AAMI/ISO 11138-<br>1:2017<br><i>Sterilization of Health Care<br/>Products—<br/>Biological<br/>Indicators—<br/>Part 1:<br/>General<br/>Requirements</i> | Demonstrate<br>indicators<br>within the<br>3MTM Attest<br>Vaporized<br>Hydrogen<br>Peroxide Clear<br>Test Pack<br>1295PCD have<br>greater<br>resistance<br>compared to<br>standalone<br>indicators in a<br>fractional<br>exposure cycle. | Indicators<br>perform as<br>intended | Passed |
| Item Tested | Test | Standard | Purpose | Acceptance Criteria | Results |
| 1295PCD<br>Challenge<br>Packs | PCD<br>Functionality |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.