← Product Code [GEX](/productcode/GEX) · K233803

# UltraClear Laser System (N/A) (K233803)

_Acclaro Corporation · GEX · May 8, 2024 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K233803

## Device Facts

- **Applicant:** Acclaro Corporation
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** May 8, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The UltraClear Fractional Laser System with its accessories is intended for use in dermatological procedures requiring coagulation, resurfacing, and ablation of soft tissue. Procedures include skin resurfacing and treatment of wrinkles, rhytids, furrows, fine lines, textural irregularites, epider veins, actinic cheilitis, keloids, verrucae, skin tags, anal tags, keratoses, scar revision (including acne scars), benign pigmented lesions, and vascular dyschromia.

## Device Story

UltraClear Fractional Laser System is a transportable 2910 nm laser device for dermatological soft tissue ablation, coagulation, and resurfacing. System inputs include user-defined settings via a 15-inch touch screen; device utilizes a laser module, scanner drivers, and software controls to deliver laser energy through a handpiece. A 635 nm red diode aiming beam assists in visualizing treatment location. Operated by clinicians in a clinical setting, the device provides fractional effects during skin resurfacing. Output consists of controlled laser energy delivery to target tissue, enabling scar revision, lesion removal, and skin rejuvenation. System includes internal TEC cooling, water pump, and fans for thermal management. Clinical benefit is achieved through precise tissue ablation and coagulation, allowing for targeted dermatological treatments.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including biocompatibility, electrical safety, and electromagnetic compatibility per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1. Software verification and validation confirmed system performance against design specifications.

## Technological Characteristics

2910 nm Er:YAG laser system; 10W power; 170 µm beam diameter; 0.1-3.0 ms pulse duration. Features include 15" touch screen, 635 nm aiming beam, and water-to-air heat exchanger cooling. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1. Patient-contacting materials tested per ISO 10993-1 for transient use.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Joule Profractional System ([K173285](/device/K173285.md))

## Reference Devices

- UltraClear Fractional Laser System ([K210847](/device/K210847.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Acclaro Corporation Shlomo Assa President 333 George Washington Highway Smithfield. Rhode Island 02865

Re: K233803

Trade/Device Name: UltraClear Fractional Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: November 29, 2023 Received: November 29, 2023

Dear Shlomo Assa:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).

{2}------------------------------------------------

Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tanisha L. Digitally signed by
Tanisha L. Hithe -S
Hithe -S Date: 2024.05.08
 15:28:53 -04'00'

Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K233803

Device Name UltraClear Fractional Laser System

#### Indications for Use (Describe)

The UltraClear Fractional Laser System with its accessories is intended for use in dermatological procedures requiring coagulation, resurfacing, and ablation of soft tissue. Procedures include skin resurfacing and treatment of wrinkles, rhytids, furrows, fine lines, textural irregularites, epider veins, actinic cheilitis, keloids, verrucae, skin tags, anal tags, keratoses, scar revision (including acne scars), benign pigmented lesions, and vascular dyschromia.

Fractional Effects:

The UltraClear Fractional Laser System with its accessories when used in intended to produce fractional effects during skin resurfacing

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary

## UltraClear Fractional Laser System

# K233803

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92.

| Owner/Applicant:       | Acclaro Corporation<br>333 George Washington Highway<br>Smithfield, RI 02917                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Correspondent | Shlomo Assa, President<br>333 George Washington Highway<br>Smithfield, RI 02917<br>619-988-4796<br>sassa@acclaromd.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Date of Summary:       | May 7, 2024                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device Trade Name:     | UltraClear Fractional Laser System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Common or Usual Name:  | 2910 nm laser system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Regulation Number:     | 21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Device Class:          | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Product Code:          | GEX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Panel:                 | General and Plastic Surgery                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Predicate Device:      | Joule Profractional System K173285                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Reference Device:      | UltraClear Fractional Laser System K210847                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device Description:    | UltraClear Fractional Laser System is a transportable device. The system includes the 15" touch screen, on/off switch, foot peddle, an emergency stop button, remote interlock, and calibration port.<br>The device console houses most of the power consuming components, including the laser module, medical grade power supply, the scanner drivers, software, TEC cooling module, water pump, fans, software controls, and all other electrical control components.<br>The laser energy is delivered during treatment via the handpiece. A 635 nm visible red laser diode aiming beam is used to visualize the location of the beam during laser treatment. |

{5}------------------------------------------------

| Indications for Use: | The UltraClear Fractional Laser System with its accessories is<br>intended for use in dermatological procedures requiring<br>coagulation, resurfacing, and ablation of soft tissue. Procedures<br>include skin resurfacing and treatment of wrinkles, rhytids, furrows,<br>fine lines, textural irregularities, epidermal nevi, telangiectasia,<br>spider veins, actinic cheilitis, keloids, verrucae, skin tags, anal tags,<br>keratoses, scar revision (including acne scars), benign pigmented<br>lesions, and vascular dyschromia. |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Fractional Effects:  | The UltraClear Fractional Laser System with its accessories when<br>used in Ultra mode is intended to produce fractional effects during<br>skin resurfacing.                                                                                                                                                                                                                                                                                                                                                                           |

## Technological Comparison to the predicate and reference devices

| Device & Predicate<br>Device(s):                                                                                          | K233803<br>(Subject Device)                                                                                                 | K173285<br>(Predicate) | K210847<br>(Reference)                                                                                                      |
|---------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Light Source                                                                                                              | Diode, Er:YAG                                                                                                               | Er:YAG                 | Diode, Er:YAG                                                                                                               |
| Wavelength (nm)                                                                                                           | 2910                                                                                                                        | 2940                   | 2910                                                                                                                        |
| Modes                                                                                                                     | CW, pulsed                                                                                                                  | Pulsed                 | CW, pulsed                                                                                                                  |
| Laser power (W)                                                                                                           | 10W                                                                                                                         | 30W                    | 10W                                                                                                                         |
| Energy/pulse/µbeam<br>(mJ/pulse/µbeam)                                                                                    | 0.6-1.5 (Clear<br>Mode)<br>1.5-3.0 (Silk<br>Mode)<br>3.6-30.0 (Ultra<br>Mode)<br>0.6-1.5, 3.6-30.0<br>(UltraClear-<br>Mode) | ≤ 70 mJ                | 0.6-1.5 (Clear<br>Mode)<br>1.5-3.0 (Silk<br>Mode)<br>3.6-30.0 (Ultra<br>Mode)<br>0.6-1.5, 3.6-30.0<br>(UltraClear-<br>Mode) |
| µbeam dia. (µm)                                                                                                           | 170                                                                                                                         | 450                    | 170                                                                                                                         |
| Fluence per µ-beam (J/cm²)                                                                                                | 15.9-132.2                                                                                                                  | ~3.0                   | 15.9-132.2                                                                                                                  |
| No. of circles in a spot of<br>10x10mm                                                                                    | 49-81 rings (clear)<br>36-64 rings (silk)                                                                                   | -                      | 49-81 rings (clear)<br>36-64 rings (silk)                                                                                   |
| Inter µbeam spacing (mm)<br>(In Ultra mode for subject<br>device)<br>Inter ring spacing (mm)<br>(in clear and ultra mode) | 1.25-0.7 (1.5%-<br>5%, Ultra)<br>1.25-1.66 (30-<br>50%, Clear)<br>1.43-2.0 (20-40%, Clear)                                  |                        | 1.25-0.7 (1.5%-<br>5%, Ultra)<br>1.25-1.66 (30-<br>50%, Clear)<br>1.43-2.0 (20-40%, Clear)                                  |

Confidential

{6}------------------------------------------------

| Device & Predicate                                 | K233803                        | K173285                        | K210847                        |
|----------------------------------------------------|--------------------------------|--------------------------------|--------------------------------|
| Device(s):                                         | (Subject Device)               | (Predicate)                    | (Reference)                    |
|                                                    | Silk)                          |                                | Silk)                          |
|                                                    | 30-50% (Clear<br>mode)         |                                | 30-50% (Clear<br>mode)         |
| µbeam density or coverage<br>(in a circle or spot) | 20-40% (Silk<br>mode)          | -                              | 20-40% (Silk<br>mode)          |
|                                                    | 1.5-5% (Ultra<br>mode)         |                                | 1.5-5% (Ultra<br>mode)         |
|                                                    | 30-50%<br>(UltraClear mode)    |                                | 30-50%<br>(UltraClear mode)    |
| Spot size (mm X mm)                                | 2X2 – 15X15                    | 1.3X1.3 – 20X20                | 2X2 – 15X15                    |
| Spot size patterns                                 | Same as above                  | Same as above                  | Same as above                  |
| Repetition Rate (Hz)                               | 3                              | 3                              | 3                              |
| Pulse duration (ms)                                | 0.1-3.0                        | 0.5-1.5                        | 0.1-3.0                        |
| Aiming beam                                        | Yes, 650 nm                    | Yes, red/green                 | Yes, 650 nm                    |
| Diode Cooling                                      | Water to air heat<br>exchanger | Water to air heat<br>exchanger | Water to air heat<br>exchanger |
| Skin Cooling                                       | No                             | No                             | No                             |

{7}------------------------------------------------

## Technological Characteristics / Substantial Equivalence

The UltraClear Fractional Laser System is substantially equivalent in design, function, operating principles, and intended use to the Joule ProFractional System (K173285) predicate device, based on the information presented. The devices share the same design and technical features, which includes calibration port, handpiece, similar wavelength, laser medium, fiber delivery, power supply, internal cooling system, hardware, electronics, firmware, software and user display screen.

Both the new device and predicate device share the same laser operating principles, such as energy, pulse width, repetition rate, fractional delivery and offers a range of spot sizes. Any minor design differences do not raise any new types of safety or effectiveness questions, thus rendering substantial equivalence.

#### Performance Data

Acclaro Corporation conducted non-clinical performance studies to confirm the overall functional specification testing of the UltraClear Fractional Laser System against its design specifications and intended use.

The following testing was conducted in support substantial equivalence.

#### Biocompatibility Testing

Testing was conducted in accordance with ISO-10993-1 to demonstrate the patient contacting material of the handpiece is safe for transient use (≤24 hours of contacting).

#### Electromagnetic Compatibility & Electrical Safety

UltraClear Fractional Laser System underwent electrical safety and electromagnetic compatibility with passing results according to the following recognized standards.

- IEC 60825-1, 2007: Safety of Laser Products – Part 1: Equipment classification and requirements.
- IEC 60601-1, 2005+A1:2012: Medical electrical equipment-- Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2 Ed. 3, 2007-03: Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
- . IEC 60601-2-22 Third Edition, 2007-05: Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment.

{8}------------------------------------------------

UltraClear Fractional Laser System underwent software verification and validation testing, which demonstrated the software is appropriate for release and that the system performed as intended. Furthermore, the testing verified the energy outputs of the system meet its design specifications.

### Conclusion

The UltraClear Fractional Laser System has the same Intended Use/Indications for Use as the predicate device. The minor differences between the two devices do not raise new or different questions about safety and effectiveness. The performance data presented in this 510(k) Premarket Notification support the safety of the new device. Additionally, the data further supports the new device should perform as intended and to its specifications. In sum, the UltraClear Fractional Laser System is as safe and effective as its predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K233803](https://fda-staging.innolitics.com/device/K233803)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).