K233784 · Hinge Health, Inc. · NUH · Feb 23, 2024 · Neurology
Device Facts
Record ID
K233784
Device Name
ENSO (Model 2)
Applicant
Hinge Health, Inc.
Product Code
NUH · Neurology
Decision Date
Feb 23, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
A transcutaneous electrical nerve stimulation (TENS) Mode which is indicated for the symptomatic relief and management of chronic intractable pain, and for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities., and A powered muscle stimulation (PMS) mode which is indicated to improve and facilitate muscle performance in healthy muscles. The ENSO Model 2 should be applied to normal, healthy, dry and clean skin of adult patients.
Device Story
ENSO (Model 2) is a wearable, battery-powered transcutaneous electrical nerve stimulation (TENS) and powered muscle stimulation (PMS) device. It consists of a main device unit, disposable gel pads, and a charging cable. The device is controlled wirelessly via the Hinge Health mobile app. It delivers electrical pulses to the skin to manage pain or facilitate muscle performance. The device features automatic overload/no-load trips and automatic shut-off. It is intended for over-the-counter use by adult patients. The system supports over-the-air (OTA) firmware updates. The device provides symptomatic relief of chronic intractable pain and temporary relief of muscle pain, or improves muscle performance in healthy muscles.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included electrical safety (IEC 60601-1, IEC 60601-2-10), EMC (IEC 60601-1-2), biocompatibility (ISO 10993-5, -10), and shelf-life validation (ASTM D6862, D3330, F1980).
Technological Characteristics
Wearable TENS/PMS device; polycarbonate/PET housing; lithium-ion polymer battery; wireless connectivity via mobile app; OTA firmware update capability. Output: Biphasic symmetrical/asymmetrical waveforms; 40-160 Hz (TENS), 2-15 Hz (EMS). Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ISO 10993-1/5/10. Non-sterile.
Indications for Use
Indicated for adult patients with chronic intractable pain or pain associated with sore/aching muscles in shoulder, waist, back, neck, arms, or legs due to exercise/household strain. Also indicated for muscle performance improvement in healthy muscles. Contraindicated for patients with cardiac pacemakers, implanted defibrillators, or other implanted electronic devices.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
February 23, 2024
Hinge Health, Inc. % Dawn Norman Regulatory Consultant EBG Advisors, Inc. 1227 25th Street, NW Suite 700 Washington, District of Columbia 20037
Re: K233784
Trade/Device Name: ENSO (Model 2) Regulation Number: 21 CFR 882.5890, 21 CFR 890.5850 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, NGX Dated: November 10, 2023 Received: November 27, 2023
Dear Dawn Norman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Robert Kang -S
for Pamela Scott
Assistant Director DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233784
Device Name ENSO (Model 2)
#### Indications for Use (Describe)
A transcutaneous electrical nerve stimulation (TENS) Mode which is indicated for the symptomatic relief and management of chronic intractable pain, and for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities., and
A powered muscle stimulation (PMS) mode which is indicated to improve and facilitate muscle performance in healthy muscles.
The ENSO Model 2 should be applied to normal, healthy, dry and clean skin of adult patients.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | |
|----------------------------------------------|-------------------------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span> <span style="text-decoration: overline;">X</span> </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Traditional 510(k) Summary
| Manufacturer: | Hinge Health ™ Incorporated<br>455 Market Street, Suite 700<br>San Francisco, CA 94105<br>USA<br>Phone (855) 902-2777 |
|-----------------|-----------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Dawn N. Norman, MS |
- Partner MRC Global, LLC Phone (618) 604-3064 Email: Dawn.Norman@askmrcglobal.com
February 21, 2024 Date Prepared:
#### DEVICE INFORMATION
| Trade/Proprietary Name: | ENSO® Device System 510(k) |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Common Name:<br>Classification Names: | Transcutaneous Nerve Stimulator<br>Transcutaneous Electrical Nerve Stimulator for Pain Relief, Powered Muscle Stimulator |
| | 21 CFR 882.5890 and 21 CFR 890.5850<br>Class II<br>Device Product Codes: NUH, NGX |
#### PREDICATE DEVICE INFORMATION
| Predicate Device: | K160052 (CUR Model 1) |
|---------------------------------------|--------------------------------------------------------------------------------------------|
| Trade/Proprietary Name: | CUR Model 1 |
| Common Name:<br>Classification Names: | Transcutaneous Nerve Stimulator<br>Stimulator, Nerve, Transcutaneous, Over-The-<br>Counter |
21 CFR 882.5890 and 21 CFR 890.5850 Class II Device Product Codes: NUH, NGX
ENSO® Device System 510(k)
{5}------------------------------------------------
Product Description:
The Hinge Health ENSO® Device System consists of the main ENSO Device, a disposable Gel Pad, and a charging cable with AC adapter for recharging. The ENSO App is used to access and control the device.
The detailed device description and Engineering drawings for the ENSO® Device System can be found in Description Section of the eSTAR form.
The ENSO Device System that is the subject of this submission is a modification to the predicate Cur 1 Device System (K160052) that was cleared on May 20, 2016.
The subject of the present Traditional 510(k) submission is a modification of CUR Model 1. The device that is the subject of this 510(k) submission is identical to the predicate device (K160052) in terms of intended use, indications for use, charger function/materials. AC adaptor, gel pad tape, overall functionality. design, type of mating connections, biocompatibility and packaging. The subject device and the predicate share the same fundamental scientific technology and intended use.
Compared to the predicate device (K160052), the subject device uses several enhancements to improve device functionality and operation (see Performance Testing Section for details). The main changes are:
- ·Contraindications, remove implanted metal devices
- ·Extend gel pad expiration date period from 12mo to 18mo and then 18mo to 24mo
- ·Improve battery shelf life
- •Minor software updates to add functionality and change system requirements.
- ·Add over-the-air (OTA) firmware update ability
There are no changes to the product's intended use, indications for use, functionality, biocompatibility, and packaging.
The ENSO Device System that is the subject of this submission meets the specifications for electrical output for TENS and EMS Modes, and IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ISO 10993-1, ISO 10993-5 and -10.
#### Indications for Use (for this submission):
The Hinge Health ENSO Device System is intended to be used in the following: A transcutaneous electrical nerve stimulation (TENS) Mode which is indicated for the symptomatic relief and management of chronic intractable pain, and for temporary relief of pain associated with sore and aching muscles in the shoulder,
ENSO® Device System 510(k)
{6}------------------------------------------------
waist, back, neck, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities. , and
A powered muscle stimulation (PMS) mode which is indicated to improve and facilitate muscle performance in healthy muscles.
The ENSO Device should be applied to normal, healthy, dry and clean skin of adult patients.
With the exception of the addition of Hinge Health's "ENSO" brand name, there are no changes to the indications for use or intended use from the previously cleared CUR Model 1 predicate (K160052). Form FDA 3881. Indications for Use. can be found in the 510(k) submission. The addition of the "ENSO" brand name to the Indications for Use statement is not considered a change because the product line has been branded "ENSO" since it was launched in the USA in 2017.
## Performance Testing
No performance standards applicable to this device have been adopted under Section 514 of the Food, Drug and Cosmetic Act. Performance testing of the ENSO Device System was conducted in accordance with the various international standards listed below and in the Performance Testing - Bench Section of this 510(k) submission.
Performance testing of the subject device modification (ENSO 2) was conducted in accordance with the following standards and test methods as seen in Figure 5-1.
| Test Method | Standard(s) | |
|-------------------|-----------------------------------------------------------------------------------------------------------|--|
| Safety | IEC 60601-1 | |
| | IEC 60601-1-6 | |
| | IEC 60601-2-10 | |
| | IEC 60601-1-11 | |
| | IEC 60529 | |
| | IEC 62133-2 | |
| | IEC 62368 | |
| EMC | IEC 60601-1-2 | |
| | FCC §15.107 & ICES Issue 5, (Nov. 2014 Update),<br>Conducted Emissions | |
| | FCC §15.109 & CAN/CSA-CISPR 22-10 Table 6 and ICES-<br>003 Issue 5 (Nov. 2014 Update), Radiated Emissions | |
| | FCC § 1.1310, RF Exposure Evaluation | |
| | RSS-102, Issue 5:2015+A1:2021, RF Exposure Evaluation | |
| SoftwareIEC 62304 | | |
## Figure 5-1 Test Method and Standards
ENSO® Device System 510(k)
{7}------------------------------------------------
| Test Method | Standard(s) |
|-----------------------------|-----------------------------------------|
| Usability | IEC 62366-1 |
| Biocompatibility | ISO 10993-1 |
| | ISO 10993-5 |
| | ISO 10993-10 |
| Packaging and<br>Shelf-life | ASTM D6862 |
| | ASTM D3330/<br>ASTM D3331<br>ASTM F1980 |
## Basis of Substantial Equivalence
The ENSO® Device System that is the subject of this submission has the following similarities to the predicate CUR Model 1 cleared under K160052:
- similar material ●
- same intended use ●
- same indications for use ●
- similar design ●
- similar sizes and dimensions .
- same type of mating components ●
- same biocompatibility .
- . same packaging methods
{8}------------------------------------------------
#### Comparison Table - Indications for Use
#### Figure 5-2: Indications for Use Comparison Table
| ENSO® Device System | | | |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|
| Attribute | Subject Device<br>(ENSO 2, K233784) | Predicate<br>(CUR Model 1, K160052) | Comments |
| Indications for<br>Use | A transcutaneous electrical nerve stimulation (TENS) Mode<br>which is indicated for the symptomatic relief and management of<br>chronic intractable pain, and for temporary relief of pain<br>associated with sore and aching muscles in the shoulder, waist,<br>back, neck, upper extremities (arm) and lower extremities (leg)<br>due to strain from exercise or normal household work activities.,<br>and<br>A powered muscle stimulation (PMS) mode which is indicated to<br>improve and facilitate muscle performance in healthy muscles.<br>The ENSO® Model 2 should be applied to normal, healthy, dry<br>and clean skin of adult patients. | A transcutaneous electrical nerve stimulation (TENS) Mode<br>which is indicated for the symptomatic relief and management of<br>chronic intractable pain, and for temporary relief of pain<br>associated with sore and aching muscles in the shoulder, waist,<br>back, neck, upper extremities (arm) and lower extremities (leg)<br>due to strain from exercise or normal household work activities.,<br>and<br>A powered muscle stimulation (PMS) mode which is indicated to<br>improve and facilitate muscle performance in healthy muscles.<br>The CUR Model 1 should be applied to normal, healthy, dry and<br>clean skin of adult patients. | Identical<br>except for<br>device name<br>and model |
| Class | II | II | Identical |
| Product Code | NUH; NGX | NUH; NGX | Identical |
| Regulation | 21 CFR 882.5890 and 21 CFR 890.5850 | 21 CFR 882.5890 and 21 CFR 890.5850 | Identical |
{9}------------------------------------------------
#### Comparison Table – Contraindications
| ENSO® Device System | | | |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Attribute | Subject Device<br>(ENSO 2, K233784) | Predicate<br>(CUR Model 1, K160052) | Comments |
| Contraindications | Contraindication: Do not use this device if you have a cardiac<br>pacemaker, implanted defibrillator, or other implanted<br>electronic device. Such use could cause electric shock,<br>burns, electrical interference, or death.<br>Warning: Do not apply stimulation over or near metallic<br>implants | Do not use this device if you have a cardiac pacemaker,<br>implanted defibrillator, or other implanted metallic or<br>electronic device. Such use could cause electric shock,<br>burns, electrical interference, or death. | Identical except "metallic"<br>was removed and the<br>warning was added |
#### Figure 5-3: Contraindications Comparison Table
#### Comparison Table – Design Features
#### Figure 5-4: Design Feature Comparison Table
| ENSO® Device System | | | | |
|-------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|-------------------------------------|
| Attribute | Subject Device<br>(ENSO 2, K233784) | Predicate<br>(CUR Model 1, K160052) | Comments | |
| Design Comparison | ENSO 2 is a wearable medical device that<br>provides professional-grade transcutaneous<br>electrical nerve stimulation (TENS) therapy<br>along with an easy-to-use interface within a<br>compact design. The device system consists of<br>the main ENSO 2 Device, a disposable Gel<br>Pad, and a charging cable with AC adapter for<br>recharging the Device. | CUR 1 is a wearable medical device that provides<br>professional-grade transcutaneous electrical nerve<br>stimulation (TENS) therapy along with an easy-to-use<br>interface within a compact design. The device system<br>consists of the main Cur 1 Device, a disposable Gel Pad,<br>and a charging cable with AC adapter for recharging the<br>Device.<br>Cur 1 is controlled wirelessly using the Cur Mobile app on a | Identical | |
| | ENSO 2 is controlled wirelessly using the Hinge<br>Health app on a compatible Mobile Device. | compatible Mobile Device. | | |
| Sizes | Disposable gel pad- Large<br>Disposable gel pad- Standard | Disposable gel pad | Identical | |
| ENSO® Device System | | | | |
| Attribute | Subject Device<br>(ENSO 2, K233784) | Predicate<br>(CUR Model 1, K160052) | Comments | |
| Anatomical Sites | •shoulder<br>•waist<br>•back<br>•neck<br>•upper extremities (arm)<br>•lower extremities (leg) | •shoulder<br>•waist<br>•back<br>•neck<br>•upper extremities (arm)<br>•lower extremities (leg) | Identical | |
| Biocompatibility/ Material | Tested per ISO 10993-1<br>ISO 10993-5 and -10 | Tested per ISO 10993-1<br>ISO 10993-5 and -10 | Equivalent<br>biocompatibility results<br>for gel pad and tape | |
| Power Source(s) | Lithium-ion Polymer chemistry, and allow<br>charging at 4.20V | 3.7V Lithium-Polymer battery (rechargeable) | Equivalent or better<br>battery life | |
| Method of Line Current<br>Isolation | Physically isolated; device cannot connect to<br>electrodes and battery recharger concurrently | Physically isolated; device cannot connect to electrodes and<br>battery recharger concurrently | Identical | |
| Patient Leakage Current<br>- Normal Condition (μΑ) | Battery powered<br>(< 10 μΑ) | Battery powered<br>(< 10 μΑ) | Identical | |
| Patient Leakage Current<br>- Single Fault Condition (µA) | Battery powered<br>(< 50 μΑ) | Battery powered<br>(< 50 μΑ) | Identical | |
| Average DC current through<br>electrodes when device is on<br>but no pulses are being applied<br>(μΑ) | < 0.1μA | < 0.1μA | Identical | |
| Number of Output Modes | 3 TENS, 1 EMS | 3 TENS, 1 EMS | Identical | |
| Number of Output<br>Channels | 1 | 1 | Identical | |
| Method of<br>Channel Isolation | N/A | N/A | Identical | |
| Regulated Current or<br>Regulated Voltage? | Yes | Yes | Identical | |
| Software/Firmware/Microproces<br>sor Control? | Yes | Yes | Identical | |
| Automatic Overload Trip? | Yes | Yes | Identical | |
| ENSO® Device System | | | | |
| Attribute | Subject Device<br>(ENSO 2, K233784) | Predicate<br>(CUR Model 1, K160052) | Comments | |
| Automatic No-Load Trip? | Yes | Yes | Identical | |
| Automatic Shut Off? | Yes | Yes | Identical | |
| User Override Control? | Yes | Yes | Identical | |
| Indicator<br>Display | On/Off Status?<br>Low Battery?<br>Voltage/Current<br>Level? | Yes<br>Yes<br>Yes | Yes<br>Yes<br>Yes | Identical<br>Identical<br>Identical |
| Timer Range (minutes) | N/A | 60 minutes | No time range | |
| Compliance with Standards? | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-10<br>ISO 10993-5 and -10 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-10<br>ISO 10993-5 and -10 | Identical | |
| Compliance with 21 CFR 898? | N/A | N/A | N/A | |
| Weight (g) | 13.1g | 11.5 g | 1.6g heavier | |
| Dimensions (mm) (W x H x D) | 38 x 38 x 11 | 38 x 38 x 10 | ~1mm taller (D) | |
| Housing Materials and<br>Construction | Polycarbonate/PET Case | Polycarbonate plastic | Equivalent with more<br>durability | |
{10}------------------------------------------------
{11}------------------------------------------------
#### Output Specifications Table for TENS Modes
Figure 5-5: Output Specification for TENS Mode Comparison Table
{12}------------------------------------------------
| Output Specifications Table for TENS Modes | | | | | | | |
|----------------------------------------------------------------------------------------|-------------------------------------|----------------------|-----------------------|-------------------------------------|----------------------|----------------------|-----------|
| | Subject Device | | Predicate | | | Comments | |
| Parameter | (ENSO 2, K233784) | | (CUR Model 1, K160052 | | | | |
| | Mode A<br>(Default) | Mode B | Mode C | Mode A<br>(Default) | Mode B | Mode C | |
| Waveform (e.g., pulsed | Biphasic, | Biphasic, | Biphasic, | Biphasic, | Biphasic, | Biphasic, | Identical |
| monophasic, biphasic) | Asymmetrical | Symmetrical | Symmetrical | Asymmetrical | Symmetrical | Symmetrical | |
| Shape (e.g., rectangular, spike) | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Identical |
| | 42.6V @ 500Ω | 43.2V @ 500Ω | 42.9V @ 500Ω | 42.6V @ 500Ω | 43.2V @ 500Ω | 42.9V @ 500Ω | Identical |
| Maximum Output Voltage (volts)<br>(10+/- %) | 85.2V @ 2kΩ | 85.4V @ 2kΩ | 86.0V @ 2kΩ | 85.2V @ 2kΩ | 85.4V @ 2kΩ | 86.0V @ 2kΩ | Identical |
| | 87.0V @ 10kΩ | 85.4V @ 10kΩ | 87.8V @ 10kΩ | 87.0V @ 10kΩ | 85.4V @ 10kΩ | 87.8V @ 10kΩ | Identical |
| | 85.1 mA @<br>500Ω | 86.3 mA @<br>500Ω | 85.6 mA @<br>500Ω | 85.1 mA @<br>500Ω | 86.3 mA @ 500Ω | 85.6 mA @ 500Ω | Identical |
| Maximum Output Current (mA)<br>(10+/- %) | 43.1 mA @<br>2kΩ | 43.2 mA @ 2kΩ | 43.5 mA @ 2kΩ | 43.1 mA @<br>2kΩ | 43.2 mA @ 2kΩ | 43.5 mA @ 2kΩ | Identical |
| | 8.8 mA @<br>10kΩ | 8.8 mA @ 10kΩ | 8.8 mA @ 10kΩ | 8.8 mA @<br>10kΩ | 8.8 mA @ 10kΩ | 8.8 mA @ 10kΩ | Identical |
| Duration of primary (depolarizing)<br>phase (usec) | 25.85 usec | 90.5 usec | 51.0 usec | 25.85 usec | 90.5 usec | 51.0 usec | Identical |
| Pulse Duration (both phases)<br>(usec) | 91.7 usec | 221 usec | 142 usec | 91.7 usec | 221 usec | 142 usec | Identical |
| Frequency (Hz) | 40-160 Hz | 40-160 Hz | 120-160 Hz | 40-160 Hz | 40-160 Hz | 120-160 Hz | Identical |
| For multiphasic waveforms only: | N/A | N/A | N/A | N/A | N/A | N/A | Identical |
| Net Charge (microcoulombs (µC) | Nominally 0uC | Nominally 0uC | Nominally 0uC | Nominally 0uC | Nominally 0uC @ | Nominally 0uC @ | Identical |
| per pulse) (If zero, state how this | @ 500Ω, zero | @ 500Ω, zero | @ 500Ω, zero | @ 500Ω, zero | 500Ω, zero net | 500Ω, zero net | |
| was achieved) | net current | net current | net current | net current | current | current | |
| Maximum Phase Charge, (µC) | 7.16 uC @<br>500Ω | 7.45 uC @ 500Ω | 4.14 uC @ 500Ω | 7.16 uC @<br>500Ω | 7.45 uC @ 500Ω | 4.14 uC @ 500Ω | Identical |
| Maximum Current Density. | 0.009 mA/cm2 | 0.004 | 0.006 | 0.009 mA/cm2 | 0.004 | 0.006 | Identical |
| (mA/cm2, r.m.s.) | @500Ω | mA/cm2 @500Ω | mA/cm2 @500Ω | @500Ω | mA/cm2 @500Ω | mA/cm2 @500Ω | |
| Maximum Average Current<br>(average absolute value), mA | 0.94 mA<br>@500Ω | 1.85 mA<br>@500Ω | 1.28 mA @500Ω | 0.94 mA<br>@500Ω | 1.85 mA @500Ω | 1.28 mA @500Ω | Identical |
| Output Specifications Table for TENS Modes | | | | | | | |
| Parameter | Subject Device<br>(ENSO 2, K233784) | | | Predicate<br>(CUR Model 1, K160052) | | | |
| | Mode A<br>(Default) | Mode B | Mode C | Mode A<br>(Default) | Mode B | Mode C | Comments |
| Maximum Average Power<br>Density, (W/cm2), (using smallest<br>conductive surface area) | 1.50 mW/cm2<br>@500Ω | 3.05 mW/cm2<br>@500Ω | 2.09 mW/cm2<br>@500Ω | 1.50 mW/cm2<br>@500Ω | 3.05 mW/cm2<br>@500Ω | 2.09 mW/cm2<br>@500Ω | Identical |
| Burst Mode<br>(i.e., pulse trains): | N/A | N/A | N/A | N/A | N/A | N/A | Identical |
{13}------------------------------------------------
#### Output Specifications Table for EMS Modes
#### Figure 5-6: Output Specifications for EMS Mode Comparison Table
| Output Specifications Table for EMS Modes | | | |
|-------------------------------------------------|-------------------------------------|-------------------------------------|-----------|
| Parameter | Subject Device<br>(ENSO 2, K233784) | Predicate<br>(CUR Model 1, K160052) | Comments |
| Mode or Program Name | Measured<br>(or Nominal as noted) | Measured<br>(or Nominal as noted) | Identical |
| Waveform (e.g., pulsed<br>monophasic, biphasic) | Biphasic, Symmetrical | Biphasic, Symmetrical | Identical |
| Shape (e.g., rectangular, spike) | Rectangular | Rectangular | Identical |
| Maximum Output Voltage<br>(volts) (10+/- %) | 38.4V @ 500Ω | 38.4V @ 500Ω | Identical |
| | 85.0V @ 2kΩ | 85.0V @ 2kΩ | Identical |
| | 85.0V @ 10kΩ | 85.0V @ 10kΩ | Identical |
| Maximum Output Current (mA)<br>(10+/- %) | 76.7 mA @ 500Ω | 76.7 mA @ 500Ω | Identical |
| | 43.0 mA @ 2kΩ | 43.0 mA @ 2kΩ | Identical |
{14}------------------------------------------------
| Output Specifications Table for EMS Modes | | | |
|-------------------------------------------------------------------------------------------|----------------------------------------|----------------------------------------|-----------|
| Parameter | Subject Device<br>(ENSO 2, K233784) | Predicate<br>(CUR Model 1, K160052) | Comments |
| | 8.1 mA @ 10Ω | 8.1 mA @ 10Ω | Identical |
| Duration of primary<br>(depolarizing) phase (usec) | 221.5 usec | 221.5 usec | Identical |
| Pulse Duration (both phases)<br>(usec) | 483.8 usec | 483.8 usec | Identical |
| Frequency (Hz) | 2-15 Hz | 2-15 Hz | Identical |
| For<br>multiphasic<br>waveforms<br>only: | N/A | N/A | Identical |
| Net Charge (microcoulombs<br>(µC) per pulse) (If zero, state<br>how this was achieved) | Nominally 0uC @ 500Ω, zero net current | Nominally 0uC @ 500Ω, zero net current | Identical |
| Maximum Phase Charge, (µC) | 16.21 uC @ 500Ω | 16.21 uC @ 500Ω | Identical |
| Maximum Current Density,<br>(mA/cm2, r.m.s.) | 0.006 mA/cm2 @500Ω | 0.006 mA/cm2 @500Ω | Identical |
| Maximum Average Current<br>(average absolute value), mA | 0.22 mA @500Ω | 0.22 mA @500Ω | Identical |
| Maximum Average Power<br>Density, (W/cm2), (using<br>smallest conductive surface<br>area) | 0.315 mW/cm2 @500Ω | 0.315 mW/cm2 @500Ω | Identical |
| Burst Mode<br>(i.e., pulse<br>trains): | N/A | N/A | Identical |
{15}------------------------------------------------
#### Sterility and Shelf Life Specifications Table
Figure 5-7: Sterility and Shelf Life Comparison Table
| Attribute | Subject Device | Predicate | Comments |
|----------------------------------|--------------------------------|--------------------------------|---------------------------------------------------------------------------------------------------------|
| Device Usage | Reusable | Reusable | Identical |
| Sterility | Provided non-sterile | Provided non-sterile | Identical |
| Expiration dating/<br>Shelf life | 2-year gel pad expiration date | 1 year gel pad expiration date | Gel pad expiry extended<br>based on validation<br>testing. See Sterilization<br>and Shelf-life Section. |
# Clinical Testing
Hinge Health determined that bench and non-clinical testing were sufficient to demonstrate that the ENSO Device System is as safe and effective as the predicate device.
{16}------------------------------------------------
## Conclusion:
The data and information provided in this submission support the conclusion that the ENSO Device System that is the subject of this 510(k) submission is substantially equivalent to its predicate device (K160052), and does not significantly affect the safety or effectiveness of the device. This modification is being proposed to improve device functionality, operation and is not at result of any clinical or performance factors.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.