LoopView® Single-Use Digital Flexible Bronchoscope (B38, B38-C, B50, B50-C, B58, B58-C), ViewHub® Video Processor (S13)
Applicant
MacroLux Medical Technology Co., Ltd.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Apr 5, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Indications for Use
The LoopView® is a sterile, single-use, flexible endoscope intended for endoscopic procedures and examination within the airways and tracheobronchial tree. The endoscope is intended to provide visualization via MACROLUX™ video processor. The endoscope is intended for use in a hospital environment. It is designed for use in adults. The ViewHub® is designed to display live imaging data from compatible MACROLUX™ endoscopes.
Device Story
LoopView® is a sterile, single-use, flexible bronchoscope; ViewHub® is a reusable video processor. Bronchoscope features distal LED light source, CMOS camera, and manual control lever for tip articulation. Processor receives video signals from endoscope; provides power; outputs live imaging to external monitor via HDMI. Used in hospital settings by clinicians for airway/tracheobronchial examination. Instrument channel allows suctioning of mucus/saliva and passage of accessories. System provides real-time visualization to assist clinical decision-making during bronchoscopy. Benefits include single-use sterility to reduce cross-contamination risk.
Clinical Evidence
Bench testing only. Includes optical performance (ISO 8600), color performance, geometric distortion, resolution, depth of field, image intensity, SNR, dynamic range, mechanical testing (suction, bending), biocompatibility (cytotoxicity, sensitization, intradermal reactivity), sterilization validation (ISO 11135), and electrical safety/EMC (IEC 60601-1, 60601-2-18, 60601-1-2).
Technological Characteristics
Single-use flexible bronchoscope with CMOS sensor and LED illumination. ViewHub processor supports HDMI output, touch panel, snapshot, white balance, and zoom. Sterilization via Ethylene Oxide (ISO 11135). Biocompatibility per ISO 10993-1. Shelf life 3 years (ASTM F1980). Packaging per ISO 11607. Electrical safety per IEC 60601-1/2-18; EMC per IEC 60601-1-2.
Indications for Use
Indicated for endoscopic procedures and examination within the airways and tracheobronchial tree in adults in a hospital environment.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 5, 2024
MacroLux Medical Technology Co., Ltd. % Jie Yang Consultant Chonconn Medical Device Consulting Co., Ltd. Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong 518067 China
Re: K233779
Trade/Device Name: LoopView@ Single-Use Digital Flexible Bronchoscope (B38, B30-C, B58, B58-C), ViewHub® Video Processor (S13) Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) and accessories Regulatory Class: Class II Product Code: EOQ Dated: March 7, 2024 Received: March 7, 2024
Dear Jie Yang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shuchen Peng -S
Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
Submission Number (if known)
K233779
Device Name
LoopView® Single-Use Digital Flexible Bronchoscope (B38, B38-C, B50, B50-C, B58-C); ViewHub® Video Processor (S13)
Indications for Use (Describe)
The LoopView® is a sterile, single-use, flexible endoscope intended for endoscopic procedures and examination within the airways and tracheobronchial tree.
The endoscope is intended to provide visualization via MACROLUX™ video processor.
The endoscope is intended for use in a hospital environment. It is designed for use in adults. The ViewHub® is designed to display live imaging data from compatible MACROLUX™
endoscopes.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) #: | K233779 | 510(k) Summary | Prepared on: | 2024-04-04 |
|-----------|---------|----------------|--------------|------------|
|-----------|---------|----------------|--------------|------------|
| Contact Details | 21 CFR 807.92(a)(1) |
|-----------------|---------------------|
|-----------------|---------------------|
| Applicant Name | MacroLux Medical Technology Co., Ltd. |
|-----------------------------|---------------------------------------------------------------------------------------------------|
| Applicant Address | 301, Building 3, NamTai Inno Park In Guang Ming Avenue, Guangming Shenzhen Guangdong 518107 China |
| Applicant Contact Telephone | 86-755-88656680 |
| Applicant Contact | Mr. Linbin Ye |
| Applicant Contact Email | yelinbin@microlite.cn |
| Device Name | 21 CFR 807.92(a)(2) |
|-------------|---------------------|
|-------------|---------------------|
| Device Trade Name | LoopView® Single-Use Digital Flexible Bronchoscope (B38, B38-C, B50, B50-C, B58, B58-C); ViewHub® Video Processor (S13) |
|---------------------|-------------------------------------------------------------------------------------------------------------------------|
| Common Name | Bronchoscope (flexible or rigid) and accessories |
| Classification Name | Bronchoscope (Flexible Or Rigid) |
| Regulation Number | 874.4680 |
| Product Code(s) | EOQ |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) |
|------------------------------------|---------------------|
|------------------------------------|---------------------|
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K173727 | Ambu® aScope™ 4 Broncho, Ambu® aView™ Monitor | EOQ |
| Device Description Summary | 21 CFR 807.92(a)(4) |
|----------------------------|---------------------|
|----------------------------|---------------------|
## Device Description Summary
The Single-Use Diaital Flexible Bronchoscope System consists of LoopView® Single Bronchoscope (consists of a handle with control lever, endoscope cable and access port for accessories, and a flexible body portion) and ViewHub® Video Processor with its accessories including power cables. The LoopView® is provided sterile (sterilized by EO) and intended to be single-use. The ViewHub® is a reusable multi-patient use device.
The built-in LED at the Distal tip of the LoopView® Single Bronchoscope provides a light source, the lens module captures the light signal, then the CMOS module converts the light signal; the endoscope cable connects the LoopView® to the ViewHub®, which provides power and processes video signal from the endoscope; the control lever on the handle connected to the controllable portion by a wire rope controls the bending direction and angle of the controllable portion; the instrument channel delivers water and other instruments. Suctioning of saliva and mucus from airway is possible through the suction system.
The ViewHub® Video Processor is a video imaging system that receives video signals from the connected endoscope, controls the light at the endoscope tip and outputs this signal including a graphical user interface to a connected external video monitor.
The LoopView® Single-Use Digital Flexible Bronchoscope has the following physical and performance characteristics:
- Maneuverable tip controlled by the user
- Flexible insertion cord
- Camera and LED light source at the distal tip
- Sterilized by Ethylene Oxide
- For single-use
{4}------------------------------------------------
The ViewHub® Video Processor has the following physical and performance characteristics:
- Snapshot, white balance, zoom and HD video output
- supports HDMI video output formats
- Touch Panel
- Reusable
## Intended Use/Indications for Use
The LoopView® is a sterile, single-use, flexible for endoscopic procedures and examination within the arways and tracheobronchial tree.
The endoscope is intended to provide visualization via MACROLUX™ video processor.
The endoscope is intended for use in a hospital environment. It is designed for use in adults.
The ViewHub® is designed to display live imaging data from compatible MACROLUX™ endoscopes.
# Indications for Use Comparison
The indications for use of the subject device and the predicate device are the same.
## Technological Comparison
The proposed device has the same technological characteristics including technology, parameters for bronchoscope(including direction of view, working length, maximum insertion width,deflection angle and sterility) and parameters for processor(including illumination source, brightness control, white balance, image/video capture and external monitor) as the predicate device. The technological differences exist between the proposed and predicate device on aspects including FOV(proposed/predicate device:125°/85°), depth of view(proposed/predicate device:3-100mm/6-50mm) and Minimum channel width(proposed/predicate device:B38,B38-C:≥Ø 2.0mm; B50,B50-C:≥Ø 2.2mm; B58, B58-C:≥Ø 2.8mm/Slim: 1.2 mm; Large: 2.6 mm).The differences will not raise new questions on safety and effectiveness of the proposed device.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Biocompatibility testing
Biocompatibility of the Single-use Bronchoscope was evaluated in accordance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The following tests were performed, as recommended:
-Cytotoxicity
- Sensitization
-Intradermal reactivity
The Single-use Bronchoscope is considered surface – mucosal membrane contacting for a duration of less than 24 hours.
Sterilization and shelf life testing
The Single-use Bronchoscope is provided sterile and its shelf-life is 3 years.
Sterilization Process has been validated accordance with ISO 11135:2014.
EO/ECH residual test was performed according to ISO 10993-7:2008.
The shelf life is determined by accelerated aging to ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
Package validation was conducted according to ISO 11607-1:2019 and ISO 11607-2:2019, and ASTM F 1886/F1886M-16, ASTM F88/F88M-15, ASTM F 1929-15.
Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Bronchoscope System. The
system complies with the IEC 60601-1 and IEC60601-2-18 for safety and the IEC 60601-1-2 for EMC.
Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was
provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions."
Bench performance testing
21 CFR 807.92(a)(5)
21 CFR 807.92(a)(6)
21 CFR 807.92(a)(5)
{5}------------------------------------------------
1.Optical performance testing according to ISO 8600 series.
2.Color performance, geometric distortion, optical performance
(resolution, depth of field and image intensity), and SNR and dynamic range tests compared with the predicate device. 3. Mechanical testing including physical features, accessories compatibility, suction and bending testing.
It is concluded form the nonclinical tests that the subject devices are as safe, as effective, and performs as well as the legally marketed predicate device identified above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.