Papacarie Duo

K233768 · Elevate Oral Care · LBH · Jul 18, 2024 · Dental

Device Facts

Record IDK233768
Device NamePapacarie Duo
ApplicantElevate Oral Care
Product CodeLBH · Dental
Decision DateJul 18, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3260
Device ClassClass 2

Indications for Use

Papacarie Duo is a gel cleanser intended to clean the interior of a prepared cavity of a tooth prior to self-etch bonding procedures.

Device Story

Papacarie Duo is a light blue gel cleanser used by dental professionals to clean prepared tooth cavities. The device is supplied in a prefilled, multi-use syringe. The clinician dispenses the gel onto a mixing pad and applies it to the interior of the prepared cavity using a flocked or micro brush. After 60 seconds, the cavity is rinsed with water. The gel's blue color aids in visualization during application. By cleaning the cavity surface prior to self-etch bonding, the device facilitates the dental restoration process. The active ingredients include papain and propylene glycol.

Clinical Evidence

Bench testing only. Performance was evaluated according to ISO 29022:2013 to demonstrate cleaning efficacy on prepared tooth surfaces. Biocompatibility was established in accordance with ISO 10993-1:2018.

Technological Characteristics

Gel cleanser containing papain and propylene glycol. Dispensed via prefilled, multi-use syringe. Applied manually with brush tips. Evaluated for biocompatibility per ISO 10993-1:2018 and cleaning performance per ISO 29022:2013.

Indications for Use

Indicated for cleaning the interior of a prepared tooth cavity prior to self-etch bonding procedures in patients undergoing dental restoration.

Regulatory Classification

Identification

Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.

Special Controls

*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. July 18, 2024 Elevate Oral Care Steven Pardue Managing Member 346 Pike Rd Suite 5 West Palm Beach, Florida 33411 Re: K233768 Trade/Device Name: Papacarie Duo Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity Varnish Regulatory Class: Class II Product Code: LBH, KLE Dated: [NOTE: Use date of most recent supplement] Received: June 18, 2024 Dear Steven Pardue: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Bobak Shirmohammadi -S For Michael E. Adjodha, M.ChE., RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, {2}------------------------------------------------ Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use Submission Number (if known) K233768 Device Name Papacarie Duo Indications for Use (Describe) Papacarie Duo is a gel cleanser intended to clean the interior of a prepared cavity of a tooth prior to self-etch bonding procedures. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image is a logo for Elevate Oral Care. The logo features a teal-colored globe on the left side, with white lines dividing the globe into sections. To the right of the globe is the word "elevate" in a teal, sans-serif font. Below "elevate" is the phrase "ORAL CARE" in a smaller, also teal, sans-serif font, with a registered trademark symbol. # 510(k) Summary: K233768 #### Submitter: | Company: | Elevate Oral Care | |------------------|---------------------------------| | Street: | 346 Pike Road, Suite 5 | | City, State Zip: | West Palm Beach, FL 33411 | | Country: | USA | | Correspondent: | Steve Pardue<br>Managing Member | | Phone: | 877-866-9113 | | Fax: | 561-244-1927 | | Email: | spardue@elevateoralcare.com | | Date: | July 18 2024 | #### Name of Device | Proprietary Name: | Papacarie™ Duo | |----------------------|---------------------------------------------------------------------------------| | Classification Name: | Varnish, Cavity<br>21 CFR 872.3260 as Class II device<br>Product Code: LBH, KLE | | Common Name: | Cavity Varnish | ## Predicate Devices | Predicate Device | 510(k) | Product Code | |---------------------------------------------|---------|--------------| | FiteBac Cavity Cleanser | K190271 | LBH | | Clean and Boost Dentin &<br>Enamel Cleanser | K120620 | KLE | ## Description Papacarie Duo is a light blue colored gel cleanser dispensed from a prefilled, multi-use syringe storage system to a mixing pad, then applied to the interior of a prepared tooth cavity by dental professionals using a flocked tip brush or micro brush (not included). The tooth cavity is rinsed with water 60 seconds after application and prior to application of self-etch bonding material. The formula has a light blue color for easier visualization when it is applied on tooth surfaces. For single application prior to self-etch bonding procedure. ## Indications for Use Papacarie Duo is a gel cleanser intended to clean the interior of a prepared cavity of a tooth prior to self-etch bonding procedures. {5}------------------------------------------------ | | | | | Comparison of the Indication for Use of Papacarie Duo to the Predicate Devices | | | | | | | | |--|--|--|--|--------------------------------------------------------------------------------|--|--|--|--|--|--|--| |--|--|--|--|--------------------------------------------------------------------------------|--|--|--|--|--|--|--| | Papacarie Duo<br>Indication for Use | Clean & Boost Dentin &<br>Enamel Cleanser, Indication for<br>Use, K120620 | FiteBac Cavity Cleanser | |------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------| | Papacarie Duo is a gel<br>cleanser intended to<br>clean the interior of a<br>prepared cavity of a<br>tooth prior to self-etch<br>bonding procedures. | Clean and Boost is an acidic,<br>aqueous cleanser that has been<br>designed to help remove<br>contaminants from the surface of a<br>tooth prior to bonding. These<br>contaminants can be hand piece<br>oil, tooth debris, or imaging<br>powder (used to create digital<br>impressions for CAD/CAM<br>restorations). | The FiteBac® Cavity Cleanser<br>is a 2% K21 QAS aqueous<br>ethanolic solution intended for<br>cleansing and<br>moistening/rewetting of cavity<br>preparations. | # Technological Characteristics The chemical composition of Papacarie Duo, while different than the predicate devices has the same intended function as both predicate devices, to clean the prepared cavity in a tooth. The use for Papacarie Duo is highly similar to the predicates indications for use as both devices are used in a prepared cavity in a tooth prior to bonding as part of tooth restoration procedures and any differences do not raise concerns of safety or effectiveness. The predicate device Clean and Boost uses strong acid to clean prepared cavities while the second predicate device uses ethanol and quaternary ammonium silane to clean surfaces. All products use different technological methods to clean prepared tooth surfaces. Papacarie Duo uses papain and propylene glycol as cleansers to clean these surfaces. These different technological characteristics do not present different questions of safety and effectiveness. Papain has been used in food preparation and as ingredients and is considered GRAS for these uses. # Summary of Non-Clinical Performance Tests Non-clinical performance tests were conducted according to ISO 29022:2013 using Papacarie DUO to demonstrate substantial equivalence to the predicate device, and cleaning of prepared tooth surfaces prior to bonding of restorations. Biocompatibility of Papacarie Duo was demonstrated and the data provided in this submission according to ISO 10993-1:2018. # Conclusions {6}------------------------------------------------ Based on the comparison of indications for use, technological characteristics and non-clinical performance testing, we believe that Papacarie Duo is substantially equivalent to the predicate device Clean and Boost and FiteBac Cavity Cleanser.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...