K233761 · Longwood Valley Medical , Ltd. · LLZ · Aug 5, 2024 · Radiology
Device Facts
Record ID
K233761
Device Name
Orthopaedic Surgery Planning Software (AIJOINT)
Applicant
Longwood Valley Medical , Ltd.
Product Code
LLZ · Radiology
Decision Date
Aug 5, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Key point identification
—
—
Actual error distance below 1mm; 95% CI for all points.
—
—
—
>1 (professional physicians)
Anatomical segmentation
—
—
Mean Dice coefficient 0.9483; mean Hausdorff Distance 1.2917.
—
—
200 patient cases (100 cases per region).
—
Surgical planning parameters
—
—
Length measurement accuracy within 1mm; angle measurement accuracy within 1°.
—
—
—
—
Acetabular cup coverage
—
—
Accuracy within ±3%.
—
—
—
—
Indications for Use
This product is used in medical institutions, it can carry out 3D reconstruction, segmentation and semi-automatic identification of key points on CT images conforming to DICOM3.0 standard format, and assist physicians to overlay digital templates to assist surgeons with preoperative planning of adult hip and total knee replacement surgery. This product is intended for use by trained and qualified physicians and cannot be used as a basis for clinical diagnosis and treatment decisions alone.
Device Story
Standalone software for preoperative planning of adult hip and total knee replacement surgery; inputs CT images in DICOM 3.0 format; performs 3D reconstruction, image segmentation, and semi-automated identification of anatomical key points; allows surgeons to overlay digital templates of prosthetic materials; used in medical institutions by trained physicians; output assists surgeons in surgical simulation and planning; does not provide clinical diagnosis or treatment decisions; benefits include improved surgical planning efficiency and accuracy.
Clinical Evidence
No clinical data. Bench testing only. Validation included accuracy testing of key point recognition (error <1mm; 95% CI), image segmentation (mean Dice coefficient 0.9483; mean Hausdorff Distance 1.2917 across 200 cases), output parameter accuracy (length <1mm, angle <1°), prosthesis model consistency (deviation <0.5mm), and acetabular cup coverage accuracy (±3%).
Technological Characteristics
Standalone software; runs on Windows PC; processes DICOM 3.0 CT images; utilizes 3D reconstruction, segmentation, and semi-automated anatomical identification algorithms; supports digital prosthetic templates; software verification and validation performed per FDA guidance.
Indications for Use
Indicated for adult patients undergoing hip and total knee replacement surgery. Intended for use by trained and qualified physicians in medical institutions for preoperative planning. Not for use as a sole basis for clinical diagnosis or treatment decisions.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
Longwood Valley Medical Ltd Wei Yuanyuan RA Manager 2201 Floor 22, Tower 1, No.2 Ronghua South Rd., BDA 100176 Beijing, China
August 5, 2024
## Re: K233761
Trade/Device Name: Orthopaedic Surgery Planning Software (AIJOINT) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: July 1, 2024 Received: July 1, 2024
Dear Wei Yuanyuan :
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming
{1}------------------------------------------------
product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K233761
Device Name
Orthopaedic Surgery Planning Software (AIJOINT)
#### Indications for Use (Describe)
This product is used in medical institutions, it can carry out 3D reconstruction, segmentation and semi-automatic identification of key points on CT images conforming to DICOM3.0 standard format, and assist physicians to overlay digital templates to assist surgeons with preoperative planning of adult hip and total knee replacement surgery. This product is intended for use by trained and qualified physicians and cannot be used as a basis for clinical diagnosis and treatment decisions alone.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K233761-510(k) Summary
# Date of Preparation: 7/31/2024
- 1. Sponsor Identification
#### Longwood Valley Medical LTD.
Address: 2201 Floor 22, Tower 1, No.2 Ronghua South Rd., BDA 100176 Beijing, China Contact Person: Peng Chen Position: RA manager Tel: +86 13511033835 Email: pchen@changmugu.com
- 2. Designated Submission Correspondent
Mr. Peng Chen (Primary Contact Person) Ms. Yuanyuan Wei (Alternative Contact Person)
#### Longwood Valley Medical LTD.
2201 Floor 22, Tower 1, No.2 Ronghua South Rd., BDA 100176 Beijing, China
Tel: +86 4001004670 Fax: +86 (010)62692997 Email: pchen@changmugu.com
{4}------------------------------------------------
#### Identification of Proposed Device 3.
Trade Name: Orthopaedic Surgery Planning Software (AIJOINT) Common Name: Pre-operative planning software
#### Regulatory Information
Regulation Name: Medical image management and processing system Classification: Class II Product Code: LLZ Regulation Number: 21 CFR 892.2050 Review Panel: Radiology
#### Indication for Use:
This product is used in medical institutions, it can carry out 3D reconstruction, segmentation and semi-automatic identification of key points on CT images conforming to DICOM3.0 standard format, and assist physicians to overlay digital templates to assist surgeons with preoperative planning of adult hip and total knee replacement surgery. This product is intended for use by trained and qualified physicians and cannot be used as a basis for clinical diagnosis and treatment decisions alone.
- Identification of Predicate Devices 4.
510(k) Number: K182464 Product Name: PeekMed
Indications for Use:
PeekMed is a software system designed to help surgeons' specialists carry out the pre-operative planning in a prompt and efficient manner for several surgical procedures, based on their patients' imaging studies.
The software imports diagnostics imaging studies such as x-rays, CT or magnetic resonance image (MRI). The import process can retrieve files from a CD ROM, a local folder or the PACS. In parallel, there is a database of digital representations related to prosthetic materials supplied by their producing companies.
PeekMed allows health professional to digitally perform the surgical planning without adding any additional steps to that process. This software system requires no imaging study acquisition specification (no protocol). Experience in usage and a clinical assessment are necessary for a proper use of the software.
#### Device Description న్.
The proposed device, Orthopaedic Surgery Planning Software, is standalone software which is designed to help surgeons' specialists carry out the pre-operative planning for an adult hip and total
{5}------------------------------------------------
knee replacement surgery for several surgical procedures, based on the patient's CT images.
The Orthopaedic Surgery Planning Software can a carry out 3D reconstruction, segmentation and semi-automated identification of anatomical sites of CT images conforming to DICOM3.0 standard format, and assist physicians to simulate adult hip and total knee replacement surgery.
Substantially Equivalent (SE) Comparison 6.
| ITEM | Proposed Device | Predicate Device K182464 | Remark |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Product name | Orthopaedic<br>Surgery<br>Planning Software | PeekMed | / |
| Regulation No. | 21 CFR 892.2050 | 21 CFR 892.2050 | Same |
| Product Code | LLZ | LLZ | Same |
| Class | II | II | Same |
| Intended for Use | A pre-operative planning<br>software for surgery | A pre-operative planning<br>software for surgery | Same |
| Indication for Use | This product is used in<br>medical institutions, it can<br>carry out 3D<br>reconstruction,<br>segmentation and<br>semi-automatic<br>identification of key points<br>on CT images conforming<br>to DICOM3.0 standard<br>format, and assist<br>physicians to overlay<br>digital templates to assist<br>surgeons with preoperative<br>planning of adult hip and<br>total knee replacement<br>surgery. This product is<br>intended for use by trained<br>and qualified physicians<br>and cannot be used as a<br>basis for clinical diagnosis<br>and treatment decisions | PeekMed is a software system<br>designed to help surgeons'<br>specialists carry out the<br>pre-operative planning in a<br>prompt and efficient manner for<br>several surgical procedures, based<br>on their patients' imaging studies.<br>The software imports diagnostics<br>imaging studies such as x-rays,<br>CT or magnetic resonance image<br>(MRI). The import process can<br>retrieve files from a CD ROM, a<br>local folder or the PACS. In<br>parallel, there is a database of<br>digital representations related to<br>prosthetic materials supplied by<br>their producing companies.<br>PeekMed allows health<br>professional to digitally perform<br>the surgical planning without<br>adding any additional steps to | Similar<br>Analysis<br>1 |
| | | | |
| | alone. | that process. This software | |
| | | system requires no imaging study | |
| | | acquisition specification (no | |
| | | protocol). Experience in usage | |
| | | and a clinical assessment are | |
| | | necessary for a proper use of the | |
| | | software. | |
| | | PeekMed allows the surgeon to | |
| | | perform the pre-surgical planning | |
| | AIJOINT allows the | efficiently in the following | |
| | surgeon to perform the | subspecialties: | |
| | pre-surgical planning | | |
| | efficiently in the following | - Hip | |
| | subspecialties: | - Knee | |
| | - Hip | - Spine | |
| | | - Upper Limb | |
| | - Knee | - Foot and Ankle | |
| | | - Trauma | |
| | | For each subspecialty, there are | |
| | For each subspecialty, there | several procedures: | |
| | | | |
| | are several procedures: | | |
| | | -Hip: | |
| | -Hip: | Hip Dysplasia Correction, | |
| | Hip Dysplasia Correction, | Limb Length Discrepancy, | |
| | Limb Length Discrepancy, | Center of Rotation - Ranawat | Different |
| Subspecialties | Acetabular Angle, | method, | Analysis |
| | Hip Arthroplasty | Acetabular Angle, | 2 |
| | | | |
| | | Total Hip Arthroplasty | |
| | | -Knee: Leg Deformity | |
| | -Knee: Knee Arthroplasty | Correction, AP Knee Resection, | |
| | | AP Full Leg | |
| | | resection, High Tibial Osteotomy, | |
| | | ACL Tunnel Reconstruction, | |
| | | Medial Patellofemoral Ligament | |
| | | | |
| | / | -Spine: Single Cobb Angle,<br>Thoracic Kyphosis Angle,<br>Lumbar Lordosis Angle, Sagittal<br>Vertical Axis, Pelvic Angles,<br>Coronal Balance, Sacro- femoral<br>Angle, Sagittal Balance, | |
| | | Smith-Petersen | |
| | Osteotomy, Pedicle Subtraction Osteotomy | | |
| | -Upper Limb: Total Shoulder Replacement, Clavicular Angle, Shoulder Resurfacing | | |
| | -Foot and Ankle: Talar Tilt, Hallux Valgus, Moreau-Costa-Bertani Internal Angle, Moreau- Costa-Bertani External Angle, Maestro Formula | | |
| | -Trauma: Diaphyseal Shaft Fracture Angle, Metaphyseal Shaft Fracture Angle, Roof Arc | | |
| Patient Population | Adults | Adults and pediatrics…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.