The FDR Visionary Suite is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. The FDR Visionary Suite enables radiographic or tomographic exposures of the whole body of all ages including pediatric patients. Exposures may be taken with the patient sitting, standing, or lying in the prone or supine position. The FDR Visionary Suite uses portable or integrated flat panel detectors to generate diagnostic images by converting x-rays into electronic signals. The device is also designed to be used with conventional film/screen or computed radiography (CR) cassettes. The Tomosynthesis option is intended to generate tomographic images of human anatomies. Tomosynthesis technique is used to produce a specific cross-sectional plane of the body by reconstruction of tomographic acquisition. The device is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/trained professionals. The device is not intended for mammographic applications.
Device Story
FDR Visionary Suite is a stationary X-ray system for general radiography and tomosynthesis. Input: X-ray radiation converted to electronic signals via flat panel detectors (FPDs) or CR cassettes. Operation: System includes X-ray generator, tube support, collimator, and optional radiographic table/stand. Tomosynthesis uses Filtered Back-Projection (FBP) or Shift Addition (SA) modes for cross-sectional imaging. Used in hospitals/clinics by trained professionals. Modifications include larger LCD displays on collimator/generator, addition of HD video camera for patient positioning (with superimposed guidelines), and wireless remote controls for collimator and exposure. Output: Diagnostic radiographic/tomographic images. Clinical use: Images support diagnostic decision-making. Benefits: Improved positioning assistance and workflow via wireless controls and updated UI.
Clinical Evidence
Bench testing only. No clinical data presented. Safety and effectiveness supported by hardware verification, software validation, and EMC/electrical safety testing (IEC 60601-1-2, IEC 60601-2-54, IEC 60601-1-3, IEC 60601-1-6).
Technological Characteristics
Stationary X-ray system; 50/65/80 kW generators; 40-150 kV range. Uses FUJIFILM flat panel detectors. Tomosynthesis reconstruction via FBP or SA modes. Connectivity: DICOM. Software: Multi-objective frequency processing, grid removal, shuttering, gradation, dynamic range control, tomographic artifact suppression, flexible noise control, retake analysis. Infrastructure: Linux OS, C/Golang/JavaScript. Compliance: IEC 60601-2-54, IEC 60601-1-2.
Indications for Use
Indicated for digital or conventional radiographic and tomographic imaging of the skull, spine, chest, abdomen, extremities, and other body parts. Suitable for all ages, including pediatric patients. Used in sitting, standing, or recumbent positions. Not for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 28, 2024
Shimadzu Corporation % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct. NAPLES, FL 34114
Re: K233719
Trade/Device Name: FDR Visionary Suite Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: KPR, MQB Dated: January 26, 2024 Received: January 26, 2024
## Dear Daniel Kamm:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K233719
Device Name
FDR Visionary Suite
#### Indications for Use (Describe)
The FDR Visionary Suite is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. The FDR Visionary Suite enables radiographic or tomographic exposures of the whole body of all ages including pediatric patients. Exposures may be taken with the patient sitting, standing, or lying in the prone or supine position. The FDR Visionary Suite uses portable or integrated flat panel detectors to generate diagnostic images by converting x-rays into electronic signals. The device is also designed to be used with conventional film/screen or computed radiography (CR) cassettes. The Tomosynthesis option is intended to generate tomographic images of human anatomies. Tomosynthesis technique is used to produce a specific cross-sectional plane of the body by reconstruction of tomographic acquisition. The device is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/trained professionals. The device is not intended for mammographic applications.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Shimadzu logo and the text "510(k) Summary K233719". The text is located above the logo. The Shimadzu logo is a black square with a white cross inside, followed by the company name in bold, black letters.
SHIMADZU CORP. 1, Nishinokyo-Kuwabaracho, Nakagyo-ku, Kyoto City, JAPAN 604-8511 Date submitted: February 12, 2024 Prepared by: Koichi Kataoka, General Manager, Quality Assurance Department, Medical Systems Division k_koichi@shimadzu.co.jp; Tel +81-75-8231305
- 1. Identification of the Device: Trade/Device Names: FDR Visionary Suite Regulation Number: 21CFR892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: II Product Code: KPR, MQB
- 2. Equivalent legally marketed device: K152294 Trade/Device Name: FDR Visionary Suite Manufacturer: Shimadzu. Regulation Number: 21CFR892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: II Product Code: KPR, MQB
- 3. Purpose of Submission: Modified Device. The fundamental technological features are the same for the subject and the predicate systems.
- 4. Device Description: The FDR Visionary Suite is an X-ray radiography system that is mainly used for the radiography of various regions of the patient's body in a standing or recumbent position. The FDR Visionary Suite can be used in a wide range of applications from general radiography using X-ray film or Computed Radiography (CR) cassettes, to digital radiography. The FDR Visionary Suite consists of an Xray high voltage generator, X-ray tube support and collimator. The system can be configured with radiographic table, radiographic stand and digital radiography system as well. Optionally, the device is also used to perform tomosynthesis radiography by two different reconstruction modes. Filtered Back-Projection (FBP) mode is used to obtain a tomosynthesis image by performing back-projection after correcting the projection data. Shift Addition (SA) mode is used to obtain a tomosynthesis image at an arbitrary slice plane height by shifting each image according to projection angle of the tube based on the reconstruction height, and by applying image addition processing to them. FBP mode is generally recommended for all body parts. In case an artifact is observed at joints and other similar places, SA mode may remedy this artifact. In case metal artifact is obviously displayed. This description is the same as our predicate. Key technological characteristics of the main components: X-Ray Generators: Available in 80 kW, 65 kW, and 50 kW versions. Image receptor panels: More recently cleared versions of Fujifilm panels are now employed. Tomography remains completely unchanged, including the digital receptor panel DR-ID 911SE. Details are in the comparison table below.
{4}------------------------------------------------
Software Description: These functions are unchanged:
- DICOM Connection Software,
- Basic Grayscale Print Management o
- Media Storage о
- Image Processing Software
- o Multi-objective Frequency Processing
- Grid Removal Processing o
- o Shuttering Processing
- o Gradation Processing
- o Dynamic Range Control Processing
- Tomographic Artifact Suppression Processing o
- Flexible Noise Control o
- Function Software
- o Retake Analysis Software
But the fundamental technological features are the same for the subject and the predicate systems. The changes are described in Section 6, below.
- 5. Indications for Use: The FDR Visionary Suite is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. The FDR Visionary Suite enables radiographic or tomographic exposures of the whole body of all ages including pediatric patients. Exposures may be taken with the patient sitting, standing, or lying in the prone or supine position. The FDR Visionary Suite uses portable or integrated flat panel detectors to generate diagnostic images by converting x-rays into electronic signals. The device is also designed to be used with conventional film/screen or computed radiography (CR) cassettes. The Tomosynthesis option is intended to generate tomographic images of human anatomies. Tomosynthesis technique is used to produce a specific cross-sectional plane of the body by reconstruction of tomographic acquisition. The device is intended to be used in hospitals, clinics, imaging centers, and/or other healthcare facilities by qualified/trained professionals. The device is not intended for mammographic applications.
6. Comparison of Technological Characteristics with the predicate device: At a high level, new device and its predicate device is based on the following same technological elements: Energy emission to the patient - X-ray Power requiremental requirement; Mechanism to generate X-ray; Mechanism to acquire, process and store image data; Use of the hardware components. The imaging components were replaced by more recently cleared digital x-ray receptor panels made by FUJIFILM. A detailed list is provided in the comparison table below.
The following technological differences exist between modified and its predicate device:
A larger LCD display is employed in the collimator.
A larger LCD display is employed in the generator control.
A video camera has been added to the collimator to assist in patient positioning. Functions include: movie display, rotation, zooming, guideline overlay.
A wireless remote control collimator control has been added.
A wireless remote exposure control has been added.
The software for the two touchscreens (Collimator and Generator Control) has been updated.
Infrastructure: Change programming language (C => C, Golang, JavaScript)
{5}------------------------------------------------
## Architecture: Port to new OS (Real-time OS => Linux)
The software functionality remains unchanged.
The generator technical characteristics have not changed (kW, kV, mA, exposure times) Imaging components: The updated versions of the FUJIFILM X-ray receptor panels (since 2015) have been employed except for the tomography digital panel which remains unchanged. All the imaging panels have 510(k) clearance as shown in the table below. The panels all use the same integration software which comes with the panel. The integration software, although updated, has the identical functionality to the predicate.
In conclusion, the main technological characteristics have not changed. The modified system has a more modern, updated "look."
| Name of<br>Modified<br>Item | Predicate Device FDR Visionary Suite<br>K152294 | Modified Device FDR Visionary Suite<br>K233719 |
|------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The FDR Visionary Suite is intended to generate<br>digital or conventional radiographic images of the<br>skull, spinal column, chest, abdomen, extremities,<br>and other body parts of human anatomies in all<br>routine radiography examinations. The FDR<br>Visionary Suite enables radiographic or tomographic<br>exposures of the whole body of all ages including<br>pediatric patients. Exposures may be taken with the<br>patient sitting, standing, or lying in the prone or<br>supine position. The FDR Visionary Suite uses<br>portable or integrated flat panel detectors to<br>generate diagnostic images by converting x-rays into<br>electronic signals. The device is also designed to be<br>used with conventional film/screen or computed<br>radiography (CR) cassettes. The Tomosynthesis<br>option is intended to generate tomographic images<br>of human anatomies. Tomosynthesis technique is<br>used to produce a specific cross-sectional plane of<br>the body by reconstruction of tomographic<br>acquisition. The device is intended to be used in<br>hospitals, clinics, imaging centers, and/or other<br>healthcare facilities by qualified/trained<br>professionals. The device is not intended for<br>mammographic applications | The FDR Visionary Suite is intended to generate digital<br>or conventional radiographic images of the skull,<br>spinal column, chest, abdomen, extremities, and other<br>body parts of human anatomies in all routine<br>radiography examinations. The FDR Visionary Suite<br>enables radiographic or tomographic exposures of the<br>whole body of all ages including pediatric patients.<br>Exposures may be taken with the patient sitting,<br>standing, or lying in the prone or supine position. The<br>FDR Visionary Suite uses portable or integrated flat<br>panel detectors to generate diagnostic images by<br>converting x-rays into electronic signals. The device is<br>also designed to be used with conventional<br>film/screen or computed radiography (CR) cassettes.<br>The Tomosynthesis option is intended to generate<br>tomographic images of human anatomies.<br>Tomosynthesis technique is used to produce a specific<br>cross-sectional plane of the body by reconstruction of<br>tomographic acquisition. The device is intended to be<br>used in hospitals, clinics, imaging centers, and/or<br>other healthcare facilities by qualified/trained<br>professionals. The device is not intended for<br>mammographic applications. UNCHANGED |
| Name of<br>Modified<br>Item | Predicate Device FDR Visionary Suite<br>K152294 | Modified Device FDR Visionary Suite<br>K233719 |
| Over-head<br>Tube crane<br>control | Image: Predicate Device FDR Visionary Suite K152294 Over-head Tube crane control | CH-200(L) Touch screen now rectangular 12"<br>instead of square 7"<br>Image: Modified Device FDR Visionary Suite K233719 Over-head Tube crane control |
| Generator<br>Console | 6.7" x 5" display, 640 x 480 pixels<br>Image: Predicate Device FDR Visionary Suite K152294 Generator Console | 8.5"d x 5.4" display, 1280 x 800 pixels<br>Image: Modified Device FDR Visionary Suite K233719 Generator Console<br>Larger display, but same content |
| Name of<br>Modified<br>Item | Predicate Device FDR Visionary Suite<br>K152294 | Modified Device FDR Visionary Suite<br>K233719 |
| Added<br>Camera<br>Option<br>in Colli-<br>mator | R-300 Collimator Before Modification, no camera port<br>Bottom View Shown<br><br>Image: [R-300 Collimator Before Modification, no camera port Bottom View Shown] | Video Camera (HD resolution, 1280x720) added<br>On front of R-300 Collimator, now called RC-300<br>Bottom View shown<br>Camera Functions Patient position, patient<br>movement, plus: The following guide lines are<br>superimposed on the camera image: Gridline,<br>Exposure field, FPD effective area (Described in<br>Section 10 of the System Users Manual)<br><br>Image: [Camera and Control unit in cover] |
| Remote<br>Controller | Wired<br><br>Image: [Wired Remote Controller] | Wireless Remote Controller (Page 127 and 139 of<br>the Tube Support User Manual)<br>Shown in holder Closeup View<br><br>Image: [Wireless Remote Controller Shown in holder Closeup View] |
| Name of<br>Modified<br>Item | Predicate Device FDR Visionary Suite<br>K152294 | Modified Device FDR Visionary Suite<br>K233719 |
| Exposure<br>switch, part<br>of Generator<br>assembly | Wired<br>Wired, shown attached to control panel | Wireless Option: Battery operated Exposure<br>Switch with holder, Described in Generator User<br>Manual |
| SOFTWARE | X | The software for the two touchscreens:<br>(Colllimator and Generator Control) has been<br>updated.<br>Infrastructure: Change programming language<br>(C ⇒ C, Golang, JavaScript)<br>Architecture: Port to new OS , (Real-time OS ⇒<br>Linux)<br>The software functionality remains unchanged. |
| Tomography<br>Modes | SA mode : Reconstruction is performed using<br>the standard back projection method.<br>FBP mode : Reconstruction is performed using<br>the Filtered Back Projection method | UNCHANGED |
| X-Ray<br>Generator<br>Power | Three available models:<br>50 kW (100 kV, 500 mA)<br>65 kW (100 kV, 650 mA)<br>80 kW (100 kV, 800 mA) | UNCHANGED |
| Generator<br>kV | 40 to 150 kV | UNCHANGED |
| Exposure<br>Times | 81 positions, 0.001 to 10 sec | UNCHANGED |
| Current | (Depending on model)<br>10-500 mA (12 positions)<br>10-630 mA (12 positions)<br>10-800 mA (12 positions) | UNCHANGED |
| Name of<br>Modified<br>ltem | Predicate Device FDR Visionary Suite<br>K152294 | Modified Device FDR Visionary Suite<br>K233719 |
| Filtration | Minimum total filtration<br>IEC 60601-2-28:2017<br>1.7 mm Al/75 kV (Including added filter) | UNCHANGED |
| DR System | DR-ID 900. | UNCHANGED |
| FPD Flat panel<br>detectors | For Tomography: DR-ID 911SE<br>For radiography:<br>DR-ID 601SE<br>DR-ID 602SE<br>DR-ID 611SE | For Tomography: DR-ID 911SE (SAME)<br>For radiography:<br>DR-ID 1201SE (K142003)<br>DR-ID 1202SE (K142003)<br>DR-ID 1211SE (K142003)<br>DR-ID 1212SE (K142003)<br>DR-ID 1831SE (K192932)<br>DR-ID 1832SE (K192932)<br>DR-ID 1811SE (K192932)<br>DR-ID 1812SE (K192932)<br>(All FPD models have previous clearance) |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
- 7. Performance Testing Information: The radiation safety aspects of the device have not changed. The device remains compliant with the FDA Radiation Safety Standards. The following activities were performed as a result of the modifications to the device:
- The User Manuals were modified to reflect the changes detailed above. a.
- b. Software Validation was performed according to the FDA Guidance Documents: Content of Premarket Submissions for Device Software Functions Guidance for Industry and Food and Drug Administration Staff (Basic Documentation Level) AND Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions Guidance for Industry and Food and Drug Administration Staff.
- c. EMC and Electrical Safety Testing was performed according to these IEC Standards: Test Specification: EN 60601-1-2:2015 +A1:2021, IEC 60601-1-2:2014 +A1:2020; EN 60601-2-54:2009 +A1:2015 +A2:2019 clause 202; IEC 60601-2-54:2009 +A1:2015 +A2:2018 clause 202. Test Specification: IEC 60601-2-54:2009+A1+A2, IEC 60601-1-3:2008+A1, IEC 60601-1-3:2008+A1, IEC 60601-1-6:2010+A1, plus a Test Report for National Differences.
- 8. Conclusion: The non-clinical data supports the safety of the device and the hardware verification and validation demonstrate that new device should performs as intended in the specified use. Based on our risk analysis and bench testing, the differences do not affect its clinical safety or effectiveness. From the result of our risk analysis, software verification and validation, and nonclinical testing discussed above, it is our conclusion that, - New device is substantially equivalent to the legally marketed predicate devices K152294. The fundamental technological features are the same for the subject and the predicate systems. Therefore, new device is as safe, as effective, and performs as well as or better than the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.