K233713 · Teleflex Medical · BTO · Mar 22, 2024 · Anesthesiology
Device Facts
Record ID
K233713
Device Name
Pilling Tracheostomy Tubes
Applicant
Teleflex Medical
Product Code
BTO · Anesthesiology
Decision Date
Mar 22, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Pilling™ Tracheostomy Tubes are intended to provide tracheal access for airway management of tracheostomized patients.
Device Story
Pilling™ Tracheostomy Tubes are stainless-steel, reusable, single-patient devices providing tracheal access for airway management. The device consists of an outer cannula, inner cannula, and pilot/obturator. Available in various diameters, curvatures, and neck plate configurations (Jackson, Jackson Improved, Mayo styles). Some models are fenestrated to facilitate speech; others include 15mm adapters. Used in home and clinical settings by healthcare providers or patients. The device is inserted into the trachea to maintain an open airway. The inner and outer tubes fit closely to allow for complete cleaning. The device is provided non-sterile and must be cleaned prior to initial use and reprocessed per instructions for use.
Clinical Evidence
Bench testing only. Testing included biocompatibility per ISO 10993-1:2018, locking mechanism tests, fitted connector checks, pilot/obturator tests, and neckplate/adapter separation tests per ISO 5366:2016.
Technological Characteristics
Metallic (stainless steel), uncuffed, reusable tracheostomy tube. Available with/without 15mm adapter and fenestration. Sizes 00 to 12. Compliant with ISO 10993-1:2018 for biocompatibility and ISO 5366:2016 for mechanical security of attachments.
Indications for Use
Indicated for adult tracheostomized patients requiring tracheal access for airway management.
Regulatory Classification
Identification
A tracheostomy tube and tube cuff is a device intended to be placed into a surgical opening of the trachea to facilitate ventilation to the lungs. The cuff may be a separate or integral part of the tracheostomy tube and is, when inflated, intended to establish a seal between the tracheal wall and the tracheostomy tube. The cuff is used to prevent the patient's aspiration of substances, such as blood or vomit, or to provide a means for positive-pressure ventilation of the patient. This device is made of either stainless steel or plastic.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a human figure, and the text on the right is in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line.
March 22, 2024
Teleflex Medical Hannah Boning Regulatory Affairs Specialist 3015 Carrington Mill Blvd. Morrisville, North Carolina 27560
# Re: K233713
Trade/Device Name: Pilling Tracheostomy Tubes Regulation Number: 21 CFR 868.5800 Regulation Name: Tracheostomy Tube And Tube Cuff Regulatory Class: Class II Product Code: BTO Dated: February 26, 2024 Received: February 27, 2024
# Dear Hannah Boning:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}------------------------------------------------
See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**Bradley Q. Quinn -S**
Bradley Quinn Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
# Indications for Use
Submission Number (if known)
K233713
Device Name
Pilling Tracheostomy Tubes
Indications for Use (Describe)
Tracheostomy tubes are intended to provide tracheal access for airway management of tracheostomized patients.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Teleflex" in a bold, dark blue font. The word is slightly stylized, with the "T" having a unique shape. A small, circled "R" symbol is located to the right of the "x" in "Teleflex", indicating that the name is a registered trademark.
#### Teleflex Medical, Inc
3015 Carrington Mill Blvd Morrisville NC 27509 USA 919-544-8000 Phone: Fax· 866-804-9881
eleflex.com
### 510(k) SUMMARY
# Pilling™ Tracheostomy Tube
### A. Name, Address, Phone and Fax Number of Applicant
Teleflex Medical, Inc. 3015 Carrington Mill Blvd Morrisville, NC 27560 Phone: 919-544-8000 919-433-3914 Fax:
### B. Contact Person
Hannah Boning, RAC-Devices, Requlatory Affairs Specialist RA Product Management Surgical Business Unit Tel – (919) 228-2437 Fax – (919) 433-4996 Email - hannah.boning@teleflex.com
#### C. Date Prepared
22 March 2024
#### D. Device Name
| Trade Name: | Pilling™ Tracheostomy Tube |
|----------------------|-------------------------------------|
| Common Name: | Tracheostomy Tube |
| Product Code: | BTO |
| Regulation: | 21 CFR 868.5800 |
| Classification Name: | Tube, Tracheostomy (w/wo connector) |
### E. Predicate Device
K123699 – Fahl Tracheostomy Tube, BTO and JOH
# F. Device Description
Pilling™ Tracheostomy Tubes are stainless-steel tracheostomy tubes that are intended to provide tracheal access for airway management of tracheostomized patients. The tracheostomy tubes are reusable, single-patient devices that are provided non-sterile, and intended to be cleaned prior to initial use. Each tube consists of an outer cannula, inner cannula, and pilot/obturator. The inner and outer tubes fit closely to provide maximum airway. The distal edge of the inner tube is flush with, and fits very closely to, that of the outer tube to allow for complete cleaning. Variations in diameter, shape of the
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Teleflex" in a bold, blue font. The word is written in all lowercase letters, except for the first letter, which is capitalized. To the right of the word is a small, circled "R" symbol, indicating that the word is a registered trademark. The logo is simple and modern, and the blue color gives it a sense of trustworthiness and reliability.
inner and outer tubes (radius of curvature), locking mechanism, and neck plate configurations are the basis of the various styles available.
The tubes are available in sizes ranging from 00 (4.5mm outside diameter, 2.4mm inside diameter) to 12 (16.3mm outside diameter. 12.7mm inside diameter) based on individual style. Insertable lengths range from 33mm (size 00) to 85mm (size 9 Extra Long) based on individual style. Shape of the device and specific design of the neck plate and locking feature is based on style (Jackson, Jackson Improved, and Mayo).
Some Pilling™ Tracheostomy Tubes are fenestrated to facilitate speech and all tubes can be fenestrated upon request. Certain inner tubes are compatible with the Shikani-French Speaking Valve (Product Code JOH, 510(k) K982128). Pilling™ Tracheostomy Tubes are available in models with or without permanently attached 15mm adapters.
# G. Intended Use
Pilling™ Tracheostomy Tubes are intended to provide tracheal access for airway management of tracheostomized patients.
# H. Substantial Equivalence
Pilling™ Tracheostomy Tubes are substantially equivalent to Fahl Tracheostomy Tubes (K123699), specifically the Silvervent tracheostomy tubes.
| Predicate Device | Manufacturer | 510(k) | Clearance<br>Date |
|-------------------------|---------------------|---------|-------------------|
| Fahl Tracheostomy Tubes | Fahl Medizintechnik | K123699 | 12 August 2013 |
#### . Comparison To Predicate Devices
Pilling™ Tracheostomy Tubes are substantially equivalent to the predicate device with respect to technology, intended use, indications, and functional characteristics.
| Feature | Predicate Device | Subject Device | Equivalence |
|--------------------------|----------------------------------------|--------------------------------|-------------|
| Device Name | Fahl Tracheostomy<br>Tubes, Silvervent | Pilling™ Tracheostomy<br>Tubes | Identical |
| 510(k) Number | K123699 | K233713 | N/A |
| Device<br>Classification | Class II | Class II | Identical |
| Product Code | BTO and JOH | BTO | Equivalent |
| Regulation | 21 CR 868.5800 | 21 CR 868.5800 | Identical |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Teleflex" in a bold, sans-serif font. The word is a dark blue color. There is a registered trademark symbol to the right of the word "Teleflex."
| Intended Use | The Fahl Tracheostomy<br>Tubes are intended to<br>provide tracheal access<br>for airway management<br>of tracheostomized<br>patients. | Pilling™ Tracheostomy<br>Tubes are intended to<br>provide tracheal access<br>for airway management<br>of tracheostomized<br>patients. | Identical |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|------------|
| Size Range | 0 to 14 | 00 to 12 | Equivalent |
| w/wo cuff | w/o cuff | w/o cuff | Identical |
| Provided Sterile | No | No | Identical |
| Material | Sterling Silver | Stainless Steel | Equivalent |
| Technological<br>Characteristics | Metallic, uncuffed<br>tracheostomy tube | Metallic, uncuffed<br>tracheostomy tube | Equivalent |
| Prescription Use | Yes | Yes | Identical |
| Patient<br>Population | Adult | Adult | Identical |
| Use<br>Environment | Home and Clinical Use | Home and Clinical Use | Identical |
| Single/Multiple<br>Use | Multiple Use, Single<br>Patient | Multiple Use, Single<br>Patient | Identical |
| Reprocessing | Reprocess per IFU<br>cleaning instructions. | Reprocess per IFU<br>cleaning instructions | Equivalent |
| Options | Available fenestrated or<br>non-fenestrated<br>Available with or<br>without 15mm adapter | Available fenestrated or<br>non-fenestrated<br>Available with or<br>without 15mm adapter | Identical |
# J. Materials
All patient contacting materials are compliant with ISO 10993-1:2018 according to their nature and duration of contact. Pilling™ Tracheostomy Tube patient contacting materials are equivalent to the predicate device with respect to safety and biocompatibility.
# K. Technological Characteristics
A comparison of the technological characteristics of the Pilling™ Tracheostomy Tubes and the predicate has been performed. Both subject and predicate device are metallic, uncuffed, reusable tracheostomy tubes for single-patient multiple-use. Both predicate and subject devices are available in variations with or without permanently attached 15mm adapter and with or without fenestration. Pilling™ Tracheostomy Tubes additionally include an obturator/pilot for use in inserting the tracheostomy tube. This is used for tube placement only and does not constitute a technological difference from the predicate device as the obturator does not affect the performance or use of the tracheostomy tube. The results of this comparison demonstrate that the Pilling™ Tracheostomy Tubes are equivalent to the marketed predicate device in performance characteristics.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Teleflex" in a bold, sans-serif font. The word is in blue, and there is a registered trademark symbol to the right of the "x". The logo is simple and modern.
# L. Performance Data
Non-clinical performance testing has been conducted in order to ensure the device performed equivalently to the predicate. Biocompatibility testing to ISO 10993-1:2018 demonstrated that Pilling™ Tracheostomy Tubes are equivalent to the predicate device with respect to material safety and biocompatibility. Bench testing was conducted to demonstrate that functional and performance requirements were met, including a Locking Mechanism Test, Fitted Connector Check, and Pilot/Obturator Test. Additionally, a Separation Test was completed on the neckplate and 15mm Adapter to demonstrate compliance to the requirements of ISO 5366:2016 for security of attachment of neckplate and fitted connector to a tracheostomy tube.
# M. Conclusion
Based upon the biocompatibility study, comparison of characteristics, and functional test results, the proposed Pilling™ Tracheostomy Tubes are substantially equivalent in performance, safety, and effectiveness to the predicate device cleared to market via 510(k) K123699.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.