K233712 · Lima Corporate S.P.A. · MBF · Jan 11, 2024 · Orthopedic
Device Facts
Record ID
K233712
Device Name
PRIMA Humeral System; PRIMA TT Glenoid
Applicant
Lima Corporate S.P.A.
Product Code
MBF · Orthopedic
Decision Date
Jan 11, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3670
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PRIMA Humeral System is intended for partial or total, primary or revision, shoulder joint replacement in skeletally mature patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device. The PRIMA Short Stem and PRIMA Short Stem Plus are intended for use in cementless and cemented applications, at the discretion of the surgeon. The PRIMA Anatomic implant is indicated for partial or total, primary or revision shoulder joint replacement, in patients suffering from pain and disability due to: Non-inflammatory degenerative joint disease (i.e. osteoarthritis), Inflammatory arthritis of the glenohumeral join including rheumatoid arthritis, Avascular necrosis of the humeral head, Traumatic/post-traumatic arthritis, Fractures of the humeral head where adequate fixation can be achieved and adequate bone stock remains, Post-fracture deformity with intact rotator cuff, where adequate fixation can be achieved and adequate bone stock remains, Cuff tear arthropathy (CTA Heads only). The PRIMA Reverse implant is indicated for primary reverse total shoulder replacement or for revision when converting an anatomic PRIMA arthroplasty to a reverse total shoulder arthroplasty (i.e. in case of cuff tear arthropathy or in a grossly rotator cuff deficiency joint with severe arthropathy). Revision surgery with retention of the PRIMA Short Stem and PRIMA Short Stem Plus are intended as conversion surgery from anatomic to reverse, where the stem is stable, well positioned and tissue integrated. Other revisions of the humeral prosthesis part should be treated with traditional shoulder prostheses. The PRIMA reverse implant is indicated for patients suffering from pain and disability due to: Rotator cuff tear arthropathy, Osteoarthritis with rotator cuff tear, Rheumatoid arthritis with rotator cuff tear, Massive irreparable rotator cuff tear, Avascular necrosis of the humeral head, Correction of functional deformity, where adequate fixation can be achieved and adequate bone stock remains, Fractures of the humeral head where adequate fixation can be achieved and adequate bone stock remains. The PRIMA Glenoid System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy (disabled shoulder). The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device. The PRIMA Glenoid System components are intended for uncemented use with the addition of screw fixation.
Device Story
PRIMA Humeral System and PRIMA TT Glenoid are shoulder arthroplasty implants; used for primary or revision total/partial shoulder replacement. System includes humeral stems (Short Stem/Short Stem Plus), humeral head adaptors, reverse trays, reverse inserts, and glenoid baseplates with central compressive screws. Implants provide structural replacement for damaged glenohumeral joints; restore joint function in patients with severe arthritis, fractures, or rotator cuff deficiency. Surgeon selects anatomic or reverse configuration based on patient pathology; implants are surgically implanted into humeral and glenoid bone. Short stems feature Trabecular Titanium for proximal fixation; reverse inserts utilize Vitamin E highly crosslinked UHMWPE. Devices are intended for use with SMR and PRIMA glenoid systems. Clinical benefit includes pain relief and restoration of shoulder function.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including fatigue testing of humeral stems and dynamic evaluation of glenoid loosening/disassociation per ASTM F2028.
Technological Characteristics
Materials: Ti6Al4V (ISO 5832-3, ASTM F1472), LimaVit™ Vitamin E highly crosslinked UHMWPE (ISO 5834-2, ASTM F648, F2695, F2565). Components: 3D-printed Ti6Al4V humeral stems, reverse trays, inserts, and glenoid screws. Fixation: Cementless (Trabecular Titanium) or cemented. Dimensions: Stems 86-96mm length; screws 20mm length. Single-use, non-active orthopedic implants.
Indications for Use
Indicated for skeletally mature patients requiring partial/total primary or revision shoulder joint replacement. Conditions include osteoarthritis, rheumatoid arthritis, avascular necrosis, traumatic/post-traumatic arthritis, humeral head fractures, post-fracture deformity, and cuff tear arthropathy. Requires functional deltoid muscle and adequate bone stock/anatomy. Reverse configuration indicated for rotator cuff deficiency/tear arthropathy.
Regulatory Classification
Identification
A shoulder joint metal/polymer/metal nonconstrained or semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits movement in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys such as cobalt-chromium-molybdenum (Co-Cr-Mo) and titanium-aluminum-vanadium (Ti-6Al-4V) alloys, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, or a combination of an articulating ultra-high molecular weight bearing surface fixed in a metal shell made of alloys such as Co-Cr-Mo and Ti-6Al-4V. The humeral component and glenoid backing have a porous coating made of, in the case of Co-Cr-Mo components, beads of the same alloy or commercially pure titanium powder, and in the case of Ti-6Al-4V components, beads or fibers of commercially pure titanium or Ti-6Al-4V alloy, or commercially pure titanium powder. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance: Shoulder Joint Metal/Polymer/Metal Nonconstrained or Semi-Constrained Porous-Coated Uncemented Prosthesis.”
{0}------------------------------------------------
January 11, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
LimaCorporate S.p.A % Kenneth Newman Senior Regulatory Affairs Specialist Lima U.S.A., Inc. 2001 NE Green Oaks Blvd. Ste. 100 Arlington, Texas 76006
# Re: K233712
Trade/Device Name: PRIMA Humeral System; PRIMA TT Glenoid Regulation Number: 21 CFR 888.3670 Regulation Name: Shoulder joint metal/polymer/metal nonconstrained or semi-constrained porouscoated uncemented prosthesis Regulatory Class: Class II Product Code: MBF, PHX, HSD, PAO, KWS, KWT Dated: November 14, 2023 Received: November 20, 2023
# Dear Kenneth Newman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
# Sincerely,
Farzana
Digitally signed by
Farzana Sharmin -S
Sharmin -S Date: 2024.01.11
17:42:36 -05'00'
Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
510(k) Number (if known) K233712
Device Name
PRIMA Humeral System
Indications for Use (Describe)
The PRIMA Humeral System is intended for partial or total, primary or revision, shoulder joint replacement in skeletally mature patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device. The PRIMA Short Stem and PRIMA Short Stem Plus are intended for use in cementless and cemented applications, at the discretion of the surgeon.
The PRIMA Anatomic implant is indicated for partial or total, primary or revision shoulder joint replacement, in patients suffering from pain and disability due to:
- . Non-inflammatory degenerative joint disease (i.e. osteoarthritis),
- Inflammatory arthritis of the glenohumeral join including rheumatoid arthritis, .
- Avascular necrosis of the humeral head, ●
- . Traumatic/post-traumatic arthritis,
- Fractures of the humeral head where adequate fixation can be achieved and adequate bone . stock remains,
- . Post-fracture deformity with intact rotator cuff, where adequate fixation can be achieved and adequate bone stock remains,
- . Cuff tear arthropathy (CTA Heads only).
The PRIMA Reverse implant is indicated for primary reverse total shoulder replacement or for revision when converting an anatomic PRIMA arthroplasty to a reverse total shoulder arthroplasty (i.e. in case of cuff tear arthropathy or in a grossly rotator cuff deficiency joint with severe arthropathy).
Revision surgery with retention of the PRIMA Short Stem and PRIMA Short Stem Plus are intended as conversion surgery from anatomic to reverse, where the stem is stable, well positioned and tissue integrated. Other revisions of the humeral prosthesis part should be treated with traditional shoulder prostheses.
The PRIMA reverse implant is indicated for patients suffering from pain and disability due to:
- Rotator cuff tear arthropathy,
- Osteoarthritis with rotator cuff tear,
- Rheumatoid arthritis with rotator cuff tear,
- Massive irreparable rotator cuff tear,
- Avascular necrosis of the humeral head,
- Correction of functional deformity, where adequate fixation can be achieved and adequate bone stock remains,
{3}------------------------------------------------
- Fractures of the humeral head where adequate fixation can be achieved and adequate bone ● stock remains.
The PRIMA Humeral System consists of the following single use components:
- Anatomic configuration: ●
- । stem
- adaptor for humeral heads. ।
- Reverse configuration: ●
- stem -
- reverse tray
- reverse insert. -
| System | | Components | Material | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|---------------------------|------------------------------------------------------------|--|--|
| A | R | | | | |
| ● | ● | PRIMA Short Stem | Ti6Al4V 3d printed | | |
| ● | ● | PRIMA Short Stem Plus | Ti6Al4V 3d printed | | |
| ● | | ADAPTOR FOR HUMERAL HEADS | Ti6Al4V | | |
| | ● | REVERSE TRAY | Ti6Al4V | | |
| | ● | REVERSE INSERT | LimaVit™ (Vitamin E highly crosslinked<br>UHMWPE), Ti6Al4V | | |
| Material Standards<br>Ti6Al4V (ISO 5832-3 - ASTM F1472), Ti6Al4V 3d printed (ISO 5832-3), LimaVit™ (Vitamin E highly crosslinked UHMWPE) (ISO<br>5834-2 - ASTM F648 - ASTM F2695 - ASTM F2565) | | | | | |
| A = Anatomic, R = Reverse | | | | | |
Table 1 PRIMA Humeral System components materials
The PRIMA Humeral System is intended to be used with all glenoids implants belonging to the SMR Shoulder System and PRIMA Glenoid System.
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Indications for Use
Submission Number (if known)
# K233712
Device Name
PRIMA TT Glenoid
# Indications for Use (Describe)
The PRIMA Glenoid System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy (disabled shoulder). The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.
The PRIMA Glenoid System components are intended for uncemented use with the addition of screw fixation.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary
Date: January 02, 2024
Manufacturer: LimaCorporate S.p.A. Via Nazionale, 52 33038 – Villanova di San Daniele Udine - Italy
U.S. Contact Person: Kenneth Newman Kenneth.Newman@limacorporate.com Lima USA Inc. 2001 NE Green Oaks Blvd. Ste.100 Arlington, Texas 76006, USA www.limacorporate.com Cell Phone: 682-597-3381
| Product | Product Code | Regulation and Classification Name |
|-------------------------|--------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| PRIMA Humeral<br>System | HSD | Prosthesis, Shoulder, Hemi-, Humeral, Metallic<br>Uncemented per 21 CFR 888.3690 |
| | PAO | Shoulder joint metal/polymer semi-constrained<br>cemented prosthesis, per 21 CFR 888.3660 |
| | PHX | Shoulder Prosthesis, Reverse Configuration per<br>21 CFR 888.3660 |
| | KWS | Prosthesis, Shoulder, Semi-Constrained,<br>Metal/Polymer Cemented per 21 CFR 888.3660 |
| | KWT | Prosthesis, Shoulder, Non-Constrained,<br>Metal/Polymer Cemented per 21 CFR 888.3650 |
| PRIMA TT Glenoid | MBF | Shoulder joint metal/polymer/metal<br>nonconstrained or semi-constrained porous-<br>coated uncemented prosthesis per 21 CFR<br>888.3670 |
| | PHX | Shoulder joint metal/polymer semi-constrained<br>cemented prosthesis per 21 CFR 888.3660 |
#### Common name:
| Product | Common name |
|--------------------------------------------------------------------------------------------|-----------------------------|
| PRIMA Humeral System including<br>PRIMA Short Stem Plus | Humeral shoulder prosthesis |
| PRIMA TT Glenoid including Central<br>Compressive Cortical and Cancellous<br>Screws L 20mm | Glenoid shoulder prosthesis |
# Description:
This 510(k) submission aims at introducing new components to the previously cleared PRIMA Humeral (K212800) and Glenoid (K222427) systems.
The PRIMA Short Stem Plus is introduced as part of the PRIMA Humeral System. As the already cleared PRIMA Short Stem, the PRIMA Short Stem Plus is a convertible short stem component with proximal (metaphyseal) fixation with Trabecular Titanium to be used in both anatomic and reverse configurations. Depending on the configuration, the
{6}------------------------------------------------
stem component can be coupled with an Adaptor for the humeral heads for the anatomic configuration and with a Reverse Tray and Reverse Insert in reverse configuration.
The PRIMA Short Stem Plus is intended for use in cementless and cemented applications.
The PRIMA Short Stem Plus is made of Ti6A14V (ISO 5832-3) and is available in two different lengths: 86 mm and 96 mm. Each length is available in seven different sizes, varying from 28mm to 40mm in the proximal diameter.
The PRIMA TT Glenoid Central Compressive Cortical and Cancellous Screws in length 20mm are introduced as part of the PRIMA Glenoid System, as an additional option to the already available sizes (length 25 to 50mm), cleared via K222427. The PRIMA TT Glenoid Central Compressive Cortical and Cancellous Screws are intended to be inserted in the central hole of the glenoid baseplate and are used to fix the PRIMA TT Glenoid baseplates to the glenoid bone.
As for the already cleared central compressive screws of PRIMA TT Glenoid, the PRIMA TT Glenoid Central Compressive Cortical and Cancellous Screws in length 20 mm are manufactured from Ti6A14V (ASTM F1472, ISO 5832-3) and have respectively diameter 5mm (cortical) and 6.5mm (cancellous).
#### Indications for Use:
#### PRIMA Humeral System
The PRIMA Humeral System is intended for partial or total, primary or revision, shoulder joint replacement in skeletally mature patients. The patient's joint must be anatomically suited to receive the selected implants and a functional deltoid muscle is necessary to use the device. The PRIMA Short Stem Plus are intended for use in cemented applications, at the discretion of the surgeon.
The PRIMA Anatomic implant is indicated for partial or total, primary or revision shoulder joint replacement, in patients suffering from pain and disability due to:
- Non-inflammatory degenerative joint disease (i.e. osteoarthritis), ●
- Inflammatory arthritis of the glenohumeral join including rheumatoid arthritis,
- Avascular necrosis of the humeral head.
- Traumatic/post-traumatic arthritis,
- Fractures of the humeral head where adequate fixation can be achieved and adequate bone stock remains, ●
- Post-fracture deformity with intact rotator cuff, where adequate fixation can be achieved and adequate ● bone stock remains.
- . Cuff tear arthropathy (CTA Heads only).
The PRIMA Reverse implant is indicated for primary reverse total shoulder replacement or for revision when converting an anatomic PRIMA arthroplasty to a reverse total shoulder arthroplasty (i.e. in case of cuff tear arthropathy or in a grossly rotator cuff deficiency joint with severe arthropathy).
Revision surgery with retention of the PRIMA Short Stem and PRIMA Short Stem Plus are intended as conversion surgery from anatomic to reverse, where the stable, well positioned and tissue integrated. Other revisions of the humeral prosthesis part should be treated with traditional shoulder prostheses.
The PRIMA reverse implant is indicated for patients suffering from pain and disability due to:
- Rotator cuff tear arthropathy,
- Osteoarthritis with rotator cuff tear, ●
- Rheumatoid arthritis with rotator cuff tear, ●
- Massive irreparable rotator cuff tear, ●
- . Avascular necrosis of the humeral head,
{7}------------------------------------------------
- Correction of functional deformity, where adequate fixation can be achieved and adequate bone stock . remains.
- Fractures of the humeral head where adequate fixation can be achieved and adequate bone stock remains. ●
The PRIMA Humeral System consists of the following single use components:
- Anatomic configuration: ●
- stem
- adaptor for humeral heads. -
- Reverse configuration:
- stem -
●
- reverse tray
- reverse insert. -
| System | | Components | Material |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|---------------------------|------------------------------------------------------------|
| A | R | | |
| • | • | PRIMA Short Stem | Ti6Al4V 3d printed |
| • | • | PRIMA Short Stem Plus | Ti6Al4V 3d printed |
| • | | Adaptor for humeral heads | Ti6Al4V |
| | • | Reverse tray | Ti6Al4V |
| | • | Reverse insert | LimaVit™ (Vitamin E highly crosslinked UHMWPE),<br>Ti6Al4V |
| Material Standards<br>Ti6Al4V (ISO 5832-3 - ASTM F1472), Ti6Al4V 3d printed (ISO 5832-3), LimaVit™ (Vitamin E highly crosslinked UHMWPE) (ISO<br>5834-2 - ASTM F648 - ASTM F2695 - ASTM F2565) | | | |
Table 1 PRIMA Humeral System components materials
The PRIMA Humeral System is intended to be used with all glenoids implants belonging to the SMR Shoulder System and PRIMA Glenoid System.
#### PRIMA TT Glenoid
The PRIMA Glenoid System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy (disabled shoulder). The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device. The PRIMA Glenoid System components are intended for uncemented use with the addition of screw fixation.
#### Predicate Devices:
| No. | Company | Device name | Cleared via |
|-----------------------|---------------------|-----------------------------------|-------------|
| 1 (Primary Predicate) | LimaCorporate S.p.A | PRIMA TT Glenoid | K222427 |
| 2 | LimaCorporate S.p.A | PRIMA Humeral System | K212800 |
| 3 | DJO | Altivate Shoulder Prostheses Stem | K172351 |
#### Summary of technology comparison:
The intended use, design, and materials of the PRIMA Humeral System and PRIMA TT Glenoid are substantially equivalent to the ones of the predicate devices.
#### Non-clinical testing
Mechanical tests demonstrated that device performance fulfilled the intended use and that
{8}------------------------------------------------
the devices are substantially equivalent to the predicate devices.
Mechanical testing was performed on worst case components or constructs:
- Fatigue test of the humeral stem ●
- . Dynamic Evaluation of the Glenoid Loosening or Disassociation (ASTM F2028)
### Clinical testing
Clinical testing was not necessary to demonstrate substantial equivalence of PRIMA Humeral System and PRIMA TT Glenoid to the predicate devices.
# Conclusion
Based upon a comparison of intended use, materials, summary of technological characteristics, and preclinical testing, PRIMA Humeral System and PRIMA TT Glenoid are substantially equivalent to the predicate devices identified in this premarket notification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.