K233678 · Philips Medical Systems Dmc GmbH · IZL · Feb 14, 2024 · Radiology
Device Facts
Record ID
K233678
Device Name
Radiography 7000 M
Applicant
Philips Medical Systems Dmc GmbH
Product Code
IZL · Radiology
Decision Date
Feb 14, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1720
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Radiography 7000 M is intended for diagnostic procedures on both adult and pediatric patients who are unable to be transferred to the stationary X-ray system. It enables radiographic imaging of various body parts, including the skull, chest, spine, shoulders, pelvis, extremities, abdomen, and other body regions. The system allows for applications in different positions, such as sitting, standing, and lying in either a prone or supine position. However, it should be noted that the system is not suitable for mammography applications.
Device Story
Motorized mobile X-ray system; utilizes wireless flat panel detectors (SkyPlate family) for digital image acquisition; supports CR or traditional film cassettes as backup. System components include base unit (Sedecal SM-IV), wireless detectors, and Eleva Workspot 43.0. Used in clinical environments (e.g., ICU, trauma, OR) by physicians or technologists. Operates via battery power (Li-ion) with smart management; features motorized filters, manual shutter tracking, and electromagnetic brakes for head assembly. Images transmitted to PACS/RIS/HIS via wireless technology. Output viewed on 21.5" DICOM multi-touch main screen; assists clinicians in diagnostic decision-making for non-transferable patients. Includes RFID-based login and status light indicators.
Clinical Evidence
No clinical data submitted. Substantial equivalence supported by non-clinical performance testing, including system verification, software verification, risk control, and cybersecurity assessments, following FDA-recognized consensus standards (e.g., IEC 60601 series, ISO 14971, ISO 10993-1).
Technological Characteristics
Mobile X-ray system; 20kW or 40kW generator; Li-ion battery; wireless flat panel detectors (Csl scintillator, TFT readout); 21.5" DICOM multi-touch display; electromagnetic head assembly brakes; motorized collimator filters. Connectivity: 1Gbit/s Ethernet, WLAN (Aruba 303 WAP). Software: Eleva Workspot 43.0 with UNIQUE 2 image processing. Complies with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-54, 62304, 62366-1, and ISO 14971.
Indications for Use
Indicated for adult and pediatric patients unable to be transferred to stationary X-ray systems. Enables radiographic imaging of skull, chest, spine, shoulders, pelvis, extremities, abdomen, and other body regions in sitting, standing, prone, or supine positions. Not for mammography.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 14, 2024
Philips Medical Systems DMC GmbH % Shalin Dave Head of Regulatory Affairs DXR Röntgenstraße 24 Hamburg, 22335 GERMANY
## Re: K233678
Trade/Device Name: Radiography 7000 M Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile X-Ray System Regulatory Class: Class II Product Code: IZL Dated: November 8, 2023 Received: November 16, 2023
## Dear Shalin Dave:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233678
Device Name
Radiography 7000 M
## Indications for Use (Describe)
The Radiography 7000 M is intended for diagnostic procedures on both adult and pediatric patients who are unable to be transferred to the stationary X-ray system. It enables radiographic imaging of various body parts, including the skull, chest, spine, shoulders, pelvis, extremities, abdomen, and other body regions. The system allows for applications in different positions, such as sitting, standing, and lying in either a prone or supine position. However, it should be noted that the system is not suitable for mammography applications.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# PHILIPS
Page 1 of 18
# 510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness is prepared in accordance with 21 CFR §807.92.
| Preparation Date: | November 14th , 2023 |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner: | Philips Medical Systems DMC GmbH<br>Röntgenstrasse 24<br>22335 Hamburg GERMANY<br>Establishment registration number: 3003768251 |
| Primary Contact: | Shalin Dave<br>Head of Regulatory Affairs DXR<br>Phone: +31 (6) 4253260<br>E-mail: shalin.dave@philips.com |
| Secondary Contact: | Ming Xiao<br>Regulatory Affairs Manager<br>Phone: +49(40) 34971-2306<br>E-mail: ming.xiao@philips.com |
| Proposed Device | |
| Device Name | Radiography 7000 M |
| Legal Manufacturer | Philips Medical Systems DMC GmbH<br>Röntgenstrasse 24<br>22335 Hamburg<br>GERMANY |
| Classification Name: | Mobile x-ray system |
| Classification Regulation: | 892.1720 |
| Classification Panel: | 90 - Radiology |
| Device Class: | Class II |
| Product Code: | IZL |
| Predicate Device | |
| Device Name | MobileDiagnost wDR 2.2 |
| Legal Manufacturer | Philips Medical Systems DMC GmbH<br>Röntgenstrasse 24<br>22335 Hamburg<br>GERMANY |
| Classification Name: | Mobile x-ray system |
| Classification Regulation: | 892.1720 |
| Classification Panel: | 90 - Radiology |
| Device Class: | Class II |
| Product Code: | IZL, MQB |
| Reference Device # 1 | |
| Device Name | SM-IV |
| Legal Manufacturer | SEDECAL SA |
## Philips Medical Systems DMC GmbH
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in blue, with a horizontal line running through the middle of the letters. The font is bold and sans-serif. The letters are evenly spaced and the word is centered.
C/ Pelava, 9 – 13, Pol. Ind. Río de Janeiro22335 Hamburg 28110 Algete, Madrid España (Spain) Tel.- +34 91 6280544 Fax.- +34 91 6280574 Mobile x-ray system 892.1720 90 – Radiology Class II IZL
## Device Description:
Classification Name:
Classification Panel:
Device Class:
Product Code:
Classification Regulation:
The Radiography 7000 M digital mobile will be a new motorized mobile x-ray system that will employ integrated wireless flat panel detectors of two different sizes (large and small) to obtain digital images in mobile environments; the system will also be able to utilize CR or traditional film cassettes in a back-up scenario. The system will primarily use wireless technology for image acquisition and the transmission of images to archive systems like PACS, as well as for the retrieval of patient and exam information from the radiology information system (RIS) or hospital information system (HIS).
The system is available in the following configurations.
- with 20 kW X-ray generator
- with 40 kW X-ray generator
The Radiography 7000 M system is an integrated system consisting of the following subsystems:
- Base Unit ( Sedecal Mobile X-ray system SM-IV, cleared under K232185 August 21, 2023)
- · SkyPlate Family of detectors (Wireless Portable Detectors) and their accessories
- Eleva WorkSpot 43.0
The "Radiography 7000M" complies with the requirements specified in FDA's device specific guidance document entitled, "Guidance for the submission of 510(k) for solid state X-ray Imaging Device – September 1, 2016." Compliance with Sections I through VI, VIII and IX, which includes Table of contents, Scope, Purpose, Description, Regulatory Requirements, Non-Clinical Considerations, Labeling and Quality Assurance Program, of the FDA guidance document for the detectors is demonstrated in this submission.
## Indications for Use:
The Radiography 7000 M is intended for diagnostic procedures on both adult and pediatric patients who are unable to be transferred to the stationary X-ray system. It enables radiographic imaging of various body parts, including the skull, chest, spine, shoulders, pelvis, extremities, abdomen, and other body regions. The system allows for applications in different positions, such as sitting, standing, and lying in either a prone or supine position. However, it should be noted that the system is not suitable for mammography applications.
## Fundamental Scientific Technology
The proposed Radiography 7000 M is substantially equivalent to the manufacturer's legally marketed devices:
. Predicate Device: MobileDiagnost wDR 2.2 (K191813 - August 2, 2019), Philips Medical
Systems DMC
A detailed comparison of the proposed and predicate device (K191813) is provided in Table 1.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The letters are all capitalized and are a bright blue color. The word is centered and takes up most of the frame. The background is white.
| Table 1: Comparison of the design feature of the currently marketed and predicate MobileDiagnost<br>wDR 2.2 versus the proposed Radiography 7000 M | | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Predicate Device:<br>MobileDiagnost<br>wDR 2.2 (K191813, SE<br>date on August 2, 2019) | Proposed Device:<br>Radiography 7000 M<br>(K233678) | Discussion & Conclusion | |
| Clinical characteristics | | | | |
| Intended use | The MobileDiagnost<br>wDR is suitable for all<br>routine radiographic<br>exams, including<br>specialty areas such as<br>intensive care, trauma,<br>operating room, or<br>pediatric work.<br>Standard radiography<br>procedures are, for<br>example:<br>- X-ray examinations of<br>the skeleton including<br>skull, chest, spine,<br>pelvis, upper<br>extremities, lower<br>extremities, etc.<br>- X-ray examinations of<br>the lung<br>- X-ray examinations of<br>soft tissue such as<br>abdomen | The Radiography 7000<br>M is intended for<br>diagnostic procedures<br>on both adult and<br>pediatric<br>patients who are<br>unable to be<br>transferred to the<br>stationary X-ray<br>system. It enables<br>radiographic imaging<br>of various body parts,<br>including the skull,<br>chest, spine,<br>shoulders, pelvis,<br>extremities, abdomen,<br>and other body<br>regions. The system<br>allows for applications<br>in different<br>positions, such as<br>sitting, standing, and<br>lying in either a prone<br>or supine position.<br>However, it should be<br>noted that the system<br>is not suitable for<br>mammography<br>applications. | No difference in<br>characteristics between<br>products.<br>Conclusion: Same | |
| Site of the body | Whole Body | Whole Body | No difference in<br>characteristics between<br>products.<br>Conclusion: Same | |
| Population | Any type of patient | Any type of patient | No difference in<br>characteristics between<br>products.<br>Conclusion: Same | |
| Basic Information | | | | |
| Table 1: Comparison of the design feature of the currently marketed and predicate MobileDiagnost<br>WDR 2.2 versus the proposed Radiography 7000 M | | | | |
| | Predicate Device:<br>MobileDiagnost<br>WDR 2.2 (K191813, SE<br>date on August 2, 2019) | Proposed Device:<br>Radiography 7000 M<br>(K233678) | Discussion & Conclusion | |
| Design | Image: MobileDiagnost WDR 2.2 | Image: Radiography 7000 M | This difference does not<br>impact the safety or<br>effectiveness<br>of the device; Hence<br>demonstrating Substantial<br>Equivalence. | |
| Configuration | Battery Operated<br>Mobile | Battery Operated<br>Mobile | No differences in<br>characteristics between<br>subject and predicate<br>device; Hence<br>demonstrating Substantial<br>Equivalence. | |
| Battery | Lead Acid | lithium-ion (Li-ion)<br>with advanced battery<br>management system | Li-Ion battery-based<br>solution brings smart<br>battery<br>management/monitoring<br>system with fast charging.<br>No impact on clinical<br>needs and safety of the<br>device; Hence<br>demonstrating substantial<br>equivalence. | |
| Weight | 580kg | 450 kg (992 lb) | This difference does not<br>impact the safety or<br>effectiveness of the<br>device; Hence<br>demonstrating Substantial<br>Equivalence. | |
| Dimensions: (I x w x h)<br>In operation max. | 2577 mm x 670 mm x<br>2125 mm | 2670 mm x 540 mm x<br>2220 mm | The differences in the<br>dimensions do not impact<br>the safety or effectiveness<br>of the device. Hence<br>demonstrating Substantial<br>Equivalence. | |
| Dimensions: (I x w x h)<br>In transport | 1382 mm x 670 mm x<br>1960 mm | 1340 mm x 540 mm x<br>1285 mm | The differences in the<br>dimensions do not impact<br>the safety or effectiveness<br>of the device. Hence<br>demonstrating Substantial<br>Equivalence. | |
| Table 1: Comparison of the design feature of the currently marketed and predicate MobileDiagnost<br>wDR 2.2 versus the proposed Radiography 7000 M | | | | |
| | Predicate Device:<br>MobileDiagnost<br>wDR 2.2 (K191813, SE<br>date on August 2, 2019) | Proposed Device:<br>Radiography 7000 M<br>(K233678) | Discussion & Conclusion | |
| Dimensions: Source -<br>floor distance | 530 mm to 2020 mm | 530 mm to 2020 mm | No differences in<br>characteristics between<br>subject and predicate<br>device; Hence<br>demonstrating Substantial<br>Equivalence. | |
| Tube Head Display | Not Available | Available with touch<br>screen - 8.4" | This difference does not<br>impact the safety or<br>effectiveness<br>of the device; Hence<br>demonstrating Substantial<br>Equivalence. | |
| Head-assembly<br>movements | Friction based<br>movements | Electromagnetic<br>brakes for<br>omnidirectional<br>Head assembly<br>movement handle<br>with capacitive touch<br>technology based | This difference does not<br>impact the safety or<br>effectiveness<br>of the device; Hence<br>demonstrating Substantial<br>Equivalence. | |
| Monitor (Main<br>Screen) | Non DICOM Touch<br>(single) screen - 17" | DICOM Multi-touch<br>main screen - 21.5" | This difference does not<br>impact the safety or<br>effectiveness<br>of the device; Hence<br>demonstrating Substantial<br>Equivalence. | |
| Status Light<br>Indicator | NA (Not Available) | Status Light Indicator<br>on column assembly to<br>indicate system ready<br>for exposure | This difference does not<br>impact the safety or<br>effectiveness<br>of the device; Hence<br>demonstrating Substantial<br>Equivalence. | |
| Smart Power ON / Login | keypad based | Smart on/off using<br>RFID | This difference does not<br>impact the safety or<br>effectiveness<br>of the device; Hence<br>demonstrating Substantial<br>Equivalence. | |
| X ray Tube assembly<br>rotation around<br>telescope arm axis | +/-180° | ±180° | No differences in<br>characteristics between<br>subject and predicate<br>device; Hence<br>demonstrating Substantial<br>Equivalence. | |
| Table 1: Comparison of the design feature of the currently marketed and predicate MobileDiagnost<br>wDR 2.2 versus the proposed Radiography 7000 M | | | | |
| | Predicate Device:<br>MobileDiagnost<br>wDR 2.2 (K191813, SE<br>date on August 2, 2019)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.