K233674 · Ascension Orthopedics, Inc. · JWJ · May 30, 2024 · Orthopedic
Device Facts
Record ID
K233674
Device Name
Freedom Wrist Arthroplasty System
Applicant
Ascension Orthopedics, Inc.
Product Code
JWJ · Orthopedic
Decision Date
May 30, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Freedom Wrist Arthroplasty System is intractable pain resulting from traumatic arthritis, osteoarthritis, rheumatoid arthritis, and trauma-induced osteoarthritis of the radial/carpal joint and is intended to replace functionality of the joint due to deformity or elements stated above. The Freedom Wrist Arthroplasty System is intended for cemented use.
Device Story
System replaces articulation of distal radius and proximal carpal row; components used as a system. Radial component: Cobalt Chrome Molybdenum (CrCoMo) alloy with concave articulating surface; stem fixed in radial intramedullary canal. Carpal implant: titanium plate fixed to carpal bones via cemented central peg and locking caps; convex Ultra-High-Molecular-Weight Polyethylene (UHMWPE) bearing locked onto plate. Device provides semi-constrained articulation. Used by orthopedic surgeons in clinical settings to treat joint pain and deformity. Output is a physical joint replacement; clinical decision-making involves surgical intervention to restore wrist function and alleviate pain.
Clinical Evidence
Bench testing only. Safety and compatibility in the MR environment were established via non-clinical testing addressing magnetically induced displacement force (ASTM F2052), torque (ASTM F2213), RF-induced heating (ASTM F2182), and image artifacts (ASTM F2119) per ASTM F2503 and FDA guidance.
Technological Characteristics
Semi-constrained cemented wrist prosthesis. Materials: Cobalt Chrome Molybdenum (CrCoMo) alloy (radial component), titanium (carpal plate), and Ultra-High-Molecular-Weight Polyethylene (UHMWPE) (bearing). Fixed via cement. MR Conditional labeling supported by testing per ASTM F2052, F2213, F2182, and F2119.
Indications for Use
Indicated for patients with intractable pain due to traumatic arthritis, osteoarthritis, rheumatoid arthritis, or trauma-induced osteoarthritis of the radial/carpal joint requiring joint replacement for deformity or functional restoration. Intended for cemented use.
Regulatory Classification
Identification
A wrist joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a wrist joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have either a one-part radial component made of alloys, such as cobalt-chromium-molybdenum, with an ultra-high molecular weight polyethylene bearing surface, or a two-part radial component made of alloys and an ultra-high molecular weight polyethylene ball that is mounted on the radial component with a trunnion bearing. The metallic portion of the two-part radial component is inserted into the radius. These devices have a metacarpal component(s) made of alloys, such as cobalt-chromium-molybdenum. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Integra Freedom Wrist Arthroplasty System (K132250)
Submission Summary (Full Text)
{0}------------------------------------------------
May 30, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it. The seal is a symbol of the Department of Health & Human Services.
Ascension Orthopedics, Inc. Stephanie Rincones Regulatory Affairs Specialist 11101 Metric Blvd. Austin, Texas 78758
Re: K233674
Trade/Device Name: Freedom Wrist Arthroplasty System Regulation Number: 21 CFR 888.3800 Regulation Name: Wrist joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWJ Dated: May 1, 2024 Received: May 1, 2024
Dear Stephanie Rincones:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Farzana Farzana Sharmin -S Date: 2024.05.30 Sharmin -S 16:27:17 -04'00'
Farzana Sharmin, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
#### Indications for Use
Submission Number (if known)
K233674
Device Name
Freedom Wrist Arthroplasty System
Indications for Use (Describe)
The Freedom Wrist Arthroplasty System is indicated for intractable pain resulting from traumatic arthritis, osteoarthritis, rheumatoid arthritis, and trauma-induced osteoarthritis of the radial/carpal joint and is intended to replace functionality of the joint due to deformity or elements stated above. The Freedom Wrist Arthroplasty System is intended for cemented use.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| K233674 - Page 1 of 2 | | |
|------------------------------------|----------------------------------------------------------------------------|-------------------------|
| 510(k) Summary | | Prepared on: 2023-11-15 |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Ascension Orthopedics, Inc. | |
| Applicant Address | 11101 Metric Blvd. Austin TX 78758 United States | |
| Applicant Contact Telephone | 737-270-8239 | |
| Applicant Contact | Ms. Stephanie Rincones | |
| Applicant Contact Email | Stephanie.Rincones@smith-nephew.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Freedom Wrist Arthroplasty System | |
| Common Name | Wrist Prosthesis | |
| Classification Name | Wrist joint metal/polymer semi-constrained cemented prosthesis | |
| Regulation Number | 21 CFR 888.3800 | |
| Product Code | JWJ, Prosthesis, Wrist, 3 Part Metal-Plastic-Metal Articulation, Semi-Cons | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K132250 | Integra Freedom Wrist Arthroplasty System | JWJ |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
## Device Description Summary
The Freedom Wrist Arthroplasty System consists of components to replace the articulation of the distal radius and proximal row of carpal bones of the wrist joint. The components are intented together as a system, not individually as hemi-arthroplasty components. The radial component is made of Cobalt Chrome Molybenum Alloy (CrCoMo) and has a concave articulating surface and is fixed by means of a stem which is inserted into the radial intramedullary canal. The carpal implant is an assembly consisting of a titanium carpal plate, which is fixed into the carpal bones with a cemented central peg and locking caps. A convex Ultra-High-Molecular-Weight Polyethylene (UHMWPE) bearing is locked onto the carpal plate to articulate with the radial component.
### Intended Use/Indications for Use
The Freedom Wrist Arthroplasty System is intractable pain resulting from traumatic arthritis, osteoarthritis, rheumatoid arthritis, and trauma-induced osteoarthritis of the radial/carpal joint and is intended to replace functionality of the joint due to deformity or elements stated above. The Freedom Wrist Arthroplasty System is intended for cemented use.
21 CFR 807.92(a)(5)
21 CFR 807.92(a)(5)
21 CFR 807.92(a)(6)
## Indications for Use Comparison
The subject device has the same intended use and indications for use as the predicate device.
# Technological Comparison
The subject device introduces MR conditional labeling when compared to the wise, the subject device has the same
{4}------------------------------------------------
technological characteristics as the predicate device.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Safety and compatibility in the magnetic resonance (MR) environment was established through non-clinical testing that addressed applicable MR safety hazards per ASTM F2503 (Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment) and FDA Guidance (Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment). Testing addressed the following: Magnetically Induced Displacement Force (ASTM F2052), Magnetically Induced Toque (ASTM F2213), RF-Induced Heating (ASTM F2182), and Image Artifacts (ASTM F2119).
Non-clinical testing supports the addition of MR Conditional labeling to the subject device by demonstrating that it can be safely scanned under defined conditions that are clinically relevant.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.