K233669 · Ge Oec Medical Systems, Inc. · OXO · Mar 28, 2024 · Radiology
Device Facts
Record ID
K233669
Device Name
OEC 3D
Applicant
Ge Oec Medical Systems, Inc.
Product Code
OXO · Radiology
Decision Date
Mar 28, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The OEC 3D mobile fluoroscopy system is designed to provide fluoroscopic and digital spot images of adult and pediatric populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include: orthopedic, gastrointestinal.endoscopic, urologic, neurologic, vascular, cardiac, critical care, and emergency procedures.
Device Story
OEC 3D is a mobile C-Arm fluoroscopy system for intraoperative imaging. It captures X-ray inputs via a 31cm x 31cm CsI(Tl) flat panel detector and high-voltage generator. The system performs volumetric reconstructions and provides real-time 2D fluoroscopic and 3D images to assist surgeons in visualizing anatomy and interventional tools. Operated by physicians/surgeons in OR/clinical settings, the system uses a tethered workstation for image processing and display. New features include a 3D Spine Centerline Tool for vertebrae labeling, a 3D Screw Evaluation Tool, and Augmented Fluoroscopy, which overlays 3D points of interest onto live 2D fluoroscopy. These tools aid in localizing pathology and guiding surgical interventions, potentially improving procedural accuracy and outcomes.
Clinical Evidence
Bench testing only. No clinical data required. Performance of new 3D tools evaluated using cadaveric spine volume datasets; Augmented Fluoroscopy accuracy quantified using rigid phantom testing.
Technological Characteristics
Mobile C-Arm with 31cm x 31cm CMOS flat panel detector (CsI(Tl) scintillator). Rotating anode X-ray tube (0.3/0.6mm focal spots). 15kW high-frequency generator. Connectivity via DICOM. Software architecture compliant with IEC 62304. System meets IEC 60601-1 series, NEMA XR-27, and 21 CFR Subchapter J.
Indications for Use
Indicated for adult and pediatric populations undergoing diagnostic, interventional, and surgical procedures, including orthopedic, gastrointestinal, endoscopic, urologic, neurologic, vascular, cardiac, critical care, and emergency applications.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 28, 2024
GE OEC Medical Systems, Inc. % Shawn Quigley Regulatory Affairs Manager 384 Wright Brothers Drive SALT LAKE CITY, UT 84116
Re: K233669
Trade/Device Name: OEC 3D Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OXO, JAA, OWB Dated: February 27, 2024 Received: February 28, 2024
Dear Shawn Quigley:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
| Submission Number (if known) | |
|------------------------------|--|
|------------------------------|--|
K233669
Device Name
OEC 3D
Indications for Use (Describe)
The OEC 3D mobile fluoroscopy system is designed to provide fluoroscopic and digital spot images of adult and pediatric populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include: orthopedic, gastrointestinal.endoscopic, urologic, neurologic, vascular, cardiac, critical care, and emergency procedures.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the GE Healthcare logo, which consists of a circular emblem with the letters 'GE' inside, followed by the text 'GE HealthCare'. Below the logo and text, there is an alphanumeric code 'K233669'. The logo and text are in a purple color.
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) Summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h).
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | March 27, 2024 |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE OEC Medical Systems, Inc (GE Healthcare)<br>384 Wright Brothers Drive<br>Salt Lake City, Utah 84116 |
| Primary Contact: | Shawn Quigley<br>Regulatory Affairs Manager<br>Tel: 385-475-8853<br>e-mail: shawn.quigley@gehealthcare.com |
| Secondary Contacts: | Michelle Huettner<br>Director, Regulatory Affairs - IGT and Oncology<br>GE HealthCare<br>Tel: 901-558-8035<br>e-mail: michelle.huettner@gehealthcare.com |
# PRODUCT IDENTIFICATION
| Device Trade Name: | OEC 3D |
|--------------------------|---------------------------------------------------------------------------------------------------------|
| Regulation Name: | Image-intensified Fluoroscopic x-ray system |
| Classification Panel: | Radiology |
| Regulation: | 21CFR 892.1650 |
| Classification: | Class II |
| Product Code: | OXO |
| Subsequent Product Codes | JAA, OWB |
| Manufacturer: | GE OEC Medical Systems, Inc (GE HealthCare)<br>384 Wright Brothers Drive.<br>Salt Lake City, Utah 84116 |
| Manufacturing Location: | GE OEC Medical Systems, Inc (GE HealthCare)<br>384 Wright Brothers Drive.<br>Salt Lake City, Utah 84116 |
| Predicate Device: | |
| Device Name: | OEC 3D |
| 510(k) number: | K203346 |
| Manufacturer: | GE OEC Medical Systems, Inc |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the GE HealthCare logo. The logo consists of a purple circular emblem with a stylized "GE" inside, followed by the text "GE HealthCare" in a matching purple color. The text is in a clean, sans-serif font, and the overall design is modern and corporate.
| Regulation Name: | Image-intensified Fluoroscopic x-ray system |
|--------------------------|---------------------------------------------|
| Regulation: | 21CFR 892.1650 |
| Classification: | Class II |
| Product Code: | OXO |
| Subsequent Product Code: | OWB, JAA |
### Device Description:
The OEC 3D is a mobile fluoroscopic C-Arm imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic X-ray images and volumetric reconstructions during diagnostic, interventional, and surgical procedures. These images help the physician visualize the patient's anatomy and interventional tools. This visualization helps to localize clinical regions of interest and pathology. The images provide real-time visualization and records of pre-procedure anatomy, in vivo-clinical activity and post-procedure outcomes.
The system is composed of two major components, a C-Arm and a tethered Workstation.
The C-Arm is a stable mobile platform capable of performing linear motions (vertical, horizontal) and rotational motions (orbital, lateral) that allow the user to position the X-ray image chain at various angles and distances with respect to the patient anatomy to be imaged. The C-Arm is comprised of the high voltage generator, software, X-ray control, and a "C" shaped image gantry, which supports an X-ray tube and a Flat Panel Detector. Its functionality is controlled by software on the Workstation and on the OEC Touch, a digital flat panel controller mounted on the cross-arm.
The workstation is a stable mobile platform with an articulating arm supporting a color image high resolution LCD display monitor. It also includes image processing equipment/software, recording devices, data input/output devices and power control systems. The Workstation is the primary user interface to the system and can be located at a convenient location in the room independent of where the C-Arm is located.
On the C-Arm, the generator remains unchanged from the predicate device OEC 3D. This is also true for the 31 cm x 31 cm image receptor, consisting of a Thallium-doped Cesium Iodide [Cs] (Tl)] solid state flat panel X-ray detector with Complementary Metal Oxide Semiconductor (CMOS) light imager. The X-ray tube housing and insert remains the same as on the predicate OEC 3D (K203346).
C-Arm functionality is managed by a digital flat tablet control panel mounted on the C-arm base. Motion is controlled by a joystick.
On the workstation, the main hardware includes a computer with integrated wireless capability and a dedicated computer for 3D reconstruction located within the storage bay. The OEC 3D employs the same software architecture and platform design that fully supports the flat panel detector as the predicate OEC 3D and complies with IEC 60601-1.
The purpose of this Premarket Notification is to demonstrate that the subject device, OEC 3D, is a modification of and is substantially equivalent to the predicate device OEC 3D (K203346).
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for GE HealthCare. The logo consists of a purple circular emblem with a stylized "GE" monogram inside. To the right of the emblem is the text "GE HealthCare" in a matching purple color.
# Proposed Device Modification:
The device name for the subject device remains the same as the cleared predicate: OEC 3D. (K203346). The proposed modifications are being made to GE HealthCare's cleared predicate OEC 3D system (K203346).
The primary modifications to the cleared OEC 3D device are as follows:
- 1. 3D Spine Centerline Tool with Manual Labeling of the Vertebrae
- 2. 3D Screw Evaluation Tool
- 3. Augmented Fluoroscopy
# Indications for Use:
The OEC 3D mobile fluoroscopy system is designed to provide fluoroscopic and digital spot images of adult and pediatric populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include: orthopedic, gastrointestinal, endoscopic, urologic, neurologic, vascular, cardiac, critical care and emergency procedures.
The intended use and indications for use are unchanged from the predicate device.
# Technology:
The indications for use are identical and technology is similar to the predicate device. X-ray generation and control used with the subject device OEC 3D is identical to the technology used with the predicate device OEC 3D. The subject OEC 3D device employs the same fundamental technology as that of the predicate device. The image chain including the X-ray tube, high voltage generator, collimator, X-ray filters, and detectors, remains unchanged from the predicate, OEC 3D.
The system continues to meet all applicable IEC 60601-1 series of standards, NEMA XR-27, and applicable parts of 21CFR Subchapter J. The new performance claims did not require clinical data in order to establish safety or efficacy.
The OEC 3D device was built upon the existing modular and extensible software architecture, following the same design control process and software development lifecycle process that is compliant to IEC 62304 used in the predicate OEC 3D.
The changes described above do not change the control mechanism, operating principle, energy type, or the scientific technology of the predicate devices.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the GE HealthCare logo. The logo consists of a purple circular emblem with a stylized "GE" monogram inside. To the right of the emblem, the words "GE HealthCare" are written in a purple sans-serif font.
# Table 1: Subject and Predicate Device Comparison
| | Subject Device | Predicate Device | Discussion of |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Modified OEC 3D | OEC 3D (K203346) | Differences |
| X-Ray<br>Tube<br>Assembly | ➤ Type: rotating anode<br>➤ Focal spot: 0.3 mm & 0.6 mm<br>➤ Anode heat capacity: 300,000 HU<br>➤ Anode cooling rate: 85,000 HU/min<br>➤ Housing cooling rate: 68,000 HU/min<br>➤ Housing heat capacity: 1,800,000 HU | ➤ Type: rotating anode<br>➤ Focal spot: 0.3 mm & 0.6 mm<br>➤ Anode heat capacity: 300,000 HU<br>➤ Anode cooling rate: 85,000 HU/min<br>➤ Housing cooling rate: 68,000 HU/min<br>➤ Housing heat capacity: 1,800,000 HU | Identical |
| X-ray<br>Generator | ➤ High frequency, DC<br>➤ 15 kW<br>➤ Peak tube potential: 40-120 kVp<br>➤ Fluoroscopy: 0.2-10.0 mA<br>➤ High level fluoro: 0.2-20.0 mA<br>➤ Pulsed fluoro: 4, 8, 15, 30 pulses/sec | ➤ High frequency, DC<br>➤ 15 kW<br>➤ Peak tube potential: 40-120 kVp<br>➤ Fluoroscopy: 0.2-10.0 mA<br>➤ High level fluoro: 0.2-20.0 mA<br>➤ Pulsed fluoro: 4, 8, 15, 30 pulses/sec | Identical |
| Collimator | ➤ Iris<br>➤ Secondary<br>➤ Tungsten dual leaf | ➤ Iris<br>➤ Secondary<br>➤ Tungsten dual leaf | Identical |
| X-ray<br>Control<br>Modes | ➤ Auto mode<br>➤ Manual mode | ➤ Auto mode<br>➤ Manual mode | Identical |
| | Subject Device | Predicate Device | Discussion of |
| | Modified OEC 3D | OEC 3D (K203346) | Differences |
| | ➤ (Normal fluoro, High<br>level fluoro,<br>Low dose fluoro) | ➤ (Normal fluoro, High<br>level fluoro,<br>Low dose fluoro) | |
| Standards<br>Compliance | ➤ ANSI/AAMI ES60601-1<br>➤ IEC 60601-1-2<br>➤ IEC 60601-1-3<br>➤ IEC 60601-2-43<br>➤ IEC 60601-2-54<br>➤ IEC 60601-1-6<br>➤ IEC 60825-1<br>➤ IEC 62304<br>➤ NEMA DICOM PS 3.1-3.20<br>➤ NEMA XR-27 | ➤ ANSI/AAMI ES60601-1<br>➤ IEC 60601-1-2<br>➤ IEC 60601-1-3<br>➤ IEC 60601-2-43<br>➤ IEC 60601-2-54<br>➤ IEC 60601-1-6<br>➤ IEC 60825-1<br>➤ IEC 62304<br>➤ NEMA DICOM PS 3.1-3.20<br>➤ NEMA XR-27 | Identical |
| Imaging<br>Modes | ➤ Continuous<br>fluoroscopy<br>o Normal dose<br>o High level dose<br>o Low dose<br>➤ Pulsed fluoroscopy<br>o Normal dose<br>o High level dose<br>o Low dose | ➤ Continuous<br>fluoroscopy<br>o Normal dose<br>o High level dose<br>o Low dose<br>➤ Pulsed fluoroscopy<br>o Normal dose<br>o High level dose<br>o Low dose | Identical |
| | ➤ Digital spot | ➤ Digital spot | |
| | ➤ Digital cine pulse<br>o Normal dose<br>o Low dose | ➤ Digital cine pulse<br>o Normal dose<br>o Low dose | |
| | ➤ Subtraction<br>o Normal dose<br>o Low dose | ➤ Subtraction<br>o Normal dose<br>o Low dose | |
| | ➤ Roadmap<br>o Normal dose | ➤ Roadmap<br>o Normal dose | |
| | Subject Device | Predicate Device | Discussion of |
| | Modified OEC 3D | OEC 3D (K203346) | Differences |
| | ○ Low dose | ○ Low dose | |
| 3D Imaging<br>Modes | > HD+ (30s, 400 images) | > HD+ (30s, 400 images) | Identical |
| | > HD+ Low dose (30s, 400 images) | > HD+ Low dose (30s, 400 images) | |
| | > HD (30s, 400 images) | > HD (30s, 400 images) | |
| | > HD Low dose (30s, 400 images) | > HD Low dose (30s, 400 images) | |
| | > SD (30s, 200 images) | > SD (30s, 200 images) | |
| | > SD Low dose (30s, 200 images) | > SD Low dose (30s, 200 images) | |
| | | | |
| 3D Imaging<br>Features | > Coronal views | > Coronal views | Substantially<br>Equivalent |
| | > Sagittal views | > Sagittal views | |
| | > Axial views | > Axial views | |
| | > Oblique views (Multi-Planar Reformat) | > Oblique views (Multi-Planar Reformat) | 3D Spine Centerline<br>Tool with labeling<br>identifies vertebrae<br>levels in a 3D volume<br>with centroids and<br>facilitates oblique<br>viewing along the<br>spine centerline<br>defined by the<br>centroids. The tool<br>gives the user the<br>option to label<br>vertebrae levels<br>manually. |
| | > 3D volume render view (VR) | > 3D volume render view (VR) | |
| | > Maximum Intensity<br>Projection view | > Maximum Intensity<br>Projection view | |
| | > Window/Level<br>(brightness/contrast) | > Window/Level<br>(brightness/contrast) | |
| | > Image rotation | > Image rotation | |
| | > Image zoom | > Image zoom | |
| | > Metal artifact reduction | > Metal artifact reduction | |
| | > 3D Spine Centerline<br>Tool with Labeling | | |
| | > 3D Spine Screw<br>Evaluation Tool | | |
| > Augmented<br>Fluoroscopy | | Augmented<br>fluoroscopy is a | |
| Subject Device | Predicate Device | Discussion of<br>Differences | |
| Modified OEC 3D | OEC 3D (K203346) | method to overlay a<br>3D point of interest<br>in an imaging<br>volume on live<br>fluoroscopy (2D X-ray).<br>The new 3D<br>imaging features do<br>not raise new<br>questions for safety<br>and effectiveness. | |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for GE HealthCare. The logo consists of a purple circular emblem with a stylized "GE" monogram inside. To the right of the emblem is the text "GE HealthCare" in a matching purple color. The text is simple and clean, making the logo easily recognizable.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the GE HealthCare logo. The logo consists of the GE monogram in a purple circle on the left, followed by the text "GE HealthCare" in purple. The text is in a simple, sans-serif font.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the GE Healthcare logo. The logo consists of a circular emblem with a stylized "GE" monogram on the left and the text "GE HealthCare" on the right. The emblem and text are both in a purple color.
# Determination of Substantial Equivalence:
GE HealthCare believes the modified OEC 3D is of comparable type and substantially equivalent to the cleared predicate OEC 3D.
## Non-Clinical Performance Testing
The verification and validation testing have been successfully completed as required by design control procedures under GE HealthCare's quality system. The system has been tested and is compliant with the IEC 60601-1, including IEC 60601-1-2, 60601-1-3, 60601-2-43, and 60601-2-54. All applicable 21CFR Subchapter J performance standards are met., 1020.30 Diagnostic X-Ray Systems and their major components, 1020.32 Fluoroscopic equipment, 1040.10 Laser products.
The OEC 3D system was developed under the GE OEC Medical Systems Quality Management System, including design controls, risk management and software development life cycle processes. The following quality assurance measures were applied to the development of the system:
- Risk Analysis
- · Required Reviews
- · Design Reviews
- · Testing on Unit Level (Module verification)
- · Integration testing (System verification)
- · Performance testing (Verification)
- · Safety testing (Verification)
- · Simulated use testing (Validation)
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the GE HealthCare logo. The logo consists of a purple circular emblem with a stylized "GE" inside, followed by the text "GE HealthCare" in purple. The text is in a simple, sans-serif font.
The substantial equivalence was also based on a basic software documentation level.
# Additional Non-Clinical Performance Testing
The 3D Centerline Tool with Labeling and the 3D Screw Evaluation Tool performance were evaluated on cadaveric volume datasets of the spine representing different imaging conditions.
The Augmented Fluoroscopy feature is a method to overlay a 3D point of interest in an imaging volume on live fluoroscopy (2D X-ray). This method projects a 3D point with some error compared to its actual position in the associated fluoroscopic image. Performance testing was done to quantify this error for accuracy using a rigid phantom.
# Clinical Testing
The modified OEC 3D device is substantially equivalent to the cleared predicate for OEC 3D. The indication for use is identical and has equivalent/identical technological characteristics. This type of notification supports using scientific, established, engineering-based performance testing. The system has been fully tested using engineering bench testing. Clinical data is not required to demonstrate substantial equivalence for the modified OEC 3D device compared to the cleared predicate.
# Substantial Equivalence Conclusion:
The OEC 3D device follows the same design control process and software development lifecycle processes as the predicate OEC 3D (K203346).
The differences discussed in this section do not introduce any adverse effects nor raise new questions of safety and effectiveness. Based on the successful verification and validation testing, additional performance bench testing, conformance to standards, and development under GE OEC Medical System's Quality Management System, we believe that the modified OEC 3D device is substantially equivalent to the cleared predicate device OEC 3D (K203346) and therefore, is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.