← Product Code [GEI](/productcode/GEI) · K233615

# Electrosurgical Pads (GBS-Dm 1030ac, GBS-Db 1031ac, GBS-Dm 1030a, GBS-Db 1031a, GBS-Dm 1130ac, GBS-Db 1131ac, GBS-Dm 1130a, GBS-Db 1131a, GBS-Dm 1040ac, GBS-Db 1041ac, GBS-Dm 1040a, GBS-Db 1041a, GBS-Dm 1140ac, GBS-Db 1141ac, GBS-Dm 1140a, GBS-Db 1141a, GBS-Db 4030c, GBS-Db 4030, GBS-Db 4040c, GBS-Db 4040, GBS-Dm 1030bc, GBS-Db 1031bc, GBS-Dm 1030b, GBS-Db 1031b, GBS-Dm 1130bc, GBS-Db 1131bc, GBS-Dm 1130b, GBS-Db 1131b, GBS-Dm 1030nc, GBS-Db 1031nc, GBS-Dm 1030n, GBS-Db 1031n, GBS-Dm 1 (K233615)

_Baisheng Medical Co., Ltd. · GEI · Jan 8, 2024 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K233615

## Device Facts

- **Applicant:** Baisheng Medical Co., Ltd.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Jan 8, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Electrosurgical Pads are Neutral Electrode which is also known as plate, plate electrode, passive, return, or dispersive electrode. They are indicated for use to the patient over the entire pad surface to complete an electrical circuit during electrosurgery between the electrosurgical generator. the active electrode and the patient. Solid Electrosurgical Pads are for use with generators that do not have a Contact Quality Monitoring System (CQMS). Split Electrosurgical Pads are for use with generators that have a CQMS (i.e. REM, ARM, NESSY, etc.).

## Device Story

Electrosurgical pads (neutral/dispersive electrodes) function as return path for high-frequency electrical current during electrosurgical procedures. Device applied to patient skin; provides conductive surface to complete circuit between electrosurgical generator and active electrode. Solid pads used with generators lacking Contact Quality Monitoring System (CQMS); split pads used with generators equipped with CQMS (e.g., REM, ARM, NESSY) to monitor pad-to-patient contact quality. Used in clinical settings by surgeons/operating room staff. Benefits include safe return of electrosurgical current, preventing alternate site burns.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Conductive neutral electrode pads; available in solid and split configurations. Split pads compatible with Contact Quality Monitoring Systems (CQMS). Designed for use with electrosurgical generators. Non-sterile or sterile options (implied by model variations).

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 8, 2024

Baisheng Medical Co., Ltd. Sylvia Dong Regulatory Affairs Specialist No.11, Fusheng Road, Xinhui District Jiangmen, 529100 China

## Re: K233615

Trade/Device Name: Electrosurgical Pads (GBS-Dm 1030ac, GBS-Db 1031ac, GBS-Dm 1030a, GBS-Db 1031a, GBS-Dm 1130ac, GBS-Db 1131ac, GBS-Dm 1130a, GBS-Db 1131a, GBS-Dm 1040ac, GBS-Db 1041ac, GBS-Dm 1040a, GBS-Db 1041a, GBS-Dm 1140ac, GBS-Db 1141ac, GBS-Dm 1140a, GBS-Db 1141a, GBS-Db 4030c, GBS-Db 4030, GBS-Db 4040c, GBS-Db 4040, GBS-Dm 1030bc, GBS-Db 1031bc, GBS-Dm 1030b, GBS-Db 1031b, GBS-Dm 1130bc, GBS-Db 1131bc, GBS-Dm 1130b, GBS-Db 1131b, GBS-Dm 1030nc, GBS-Db 1031nc, GBS-Dm 1030n, GBS-Db 1031n, GBS-Dm 1130nc, GBS-Db 1131nc, GBS-Dm 1130n, GBS-Db 1131n, GBS-Dm 2030c, GBS-Db 2031c, GBS-Dm 2030, GBS-Db 2031, GBS-Dm 2130c, GBS-Dm 2130, GBS-Db 2131, GBS-Dm 3030c, GBS-Db 3031c, GBS-Dm 3030, GBS-Db 3031, GBS-Dm 3130, GBS-Db 3131, GBS-Dm 3130c, GBS-Db 3131c, GBS-Db 2131c) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI, ODR Dated: November 10, 2023

Dear Sylvia Dong:

Received: November 13, 2023

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and

{1}------------------------------------------------

adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Mark Digitally signed by Mark Trumbore -S Trumbore -S -Date: 2024.01.08 09:14:27
-05'00' Mark Trumbore, Ph.D.

Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Submission Number (if known)

#### Device Name

Electrosurgical Pads (GBS-Dm 1030ac, GBS-Db 1031ac, GBS-Dm 1030a, GBS-Db 1031a, GBS-Dm 1130ac, GBS-Db 1131ac, GBS-Dm 1130a, GBS-Db 1131a, GBS-Dm 1040ac, GBS-Db 1041ac, GBS-Dm 1040a, GBS-Db 1041a, GBS-Dm 1140ac, GBS-Db 1141ac, GBS-Dm 1140a, GBS-Db 1141a. GBS-Db 4030c. GBS-Db 4030. GBS-Db 4040c. GBS-Db 4040. GBS-Dm 1030bc. GBS-Db 1031bc. GBS-Dm 1030b. GBS-Db 1031b. GBS-Dm 1130bc. GBS-Db 1131bc. GBS-Dm 1130b. GBS-Db 1131b, GBS-Dm 1030nc, GBS-Db 1031nc, GBS-Dm 1030n, GBS-Db 1031n, GBS-Dm 1130nc, GBS-Db 1131nc, GBS-Dm 1130n, GBS-Db 1131n, GBS-Dm 2030c, GBS-Db 2031c, GBS-Dm 2030, GBS-Db 2031, GBS-Dm 2130c, GBS-Dm 2130, GBS-Db 2131, GBS-Dm 3030c, GBS-Db 3031c, GBS-Dm 3030, GBS-Db 3031, GBS-Dm 3130, GBS-Db 3131, GBS-Dm 3130c, GBS-Db 3131c. GBS-Db 2131c)

Indications for Use (Describe)

Electrosurgical Pads are Neutral Electrode which is also known as plate, plate electrode, passive, return, or dispersive electrode. They are indicated for use to the patient over the entire pad surface to complete an electrical circuit during electrosurgery between the electrosurgical generator. the active electrode and the patient. Solid Electrosurgical Pads are for use with generators that do not have a Contact Quality Monitoring System (CQMS). Split Electrosurgical Pads are for use with generators that have a CQMS (i.e. REM, ARM, NESSY, etc.).

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda-staging.innolitics.com/device/K233615](https://fda-staging.innolitics.com/device/K233615)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
