Autofuse is a software package that provides physicians a means for comparison of medical data including imaging data that is DICOM compliant. It allows the display, annotation, volume operation, volume rendering, registration and fusion of medical images as an aid during use by diagnostic radiology, oncology, radiation therapy planning, and other medical specialties. Autofuse is not intended for mammography.
Device Story
Autofuse is a standalone medical imaging software application for radiotherapy professionals. It ingests DICOM-compliant medical images (CT, MR, PET, RT Dose) from PACS or other clinical systems. The software provides tools for image display, annotation, volume rendering, and co-registration (including deformable registration) of multi-modal images. It allows users to draw Regions of Interest (ROI) and export results to radiation treatment planning systems. Used in clinical settings on off-the-shelf computers, the device assists clinicians in visualizing and analyzing anatomical and functional data to support treatment planning workflows. It does not perform automated segmentation or manual registration editing. The output is viewed by clinicians to inform clinical decision-making in oncology and radiation therapy.
Clinical Evidence
No clinical data. Bench testing only. Software verification and validation were performed in a simulated clinical environment using virtual machines to confirm device specifications meet user needs and intended use, following FDA guidance and AAPM standards (TG-53, TG-132).
Technological Characteristics
Software-only device; runs on off-the-shelf (OTS) hardware. OS: Ubuntu 22.04. Connectivity: DICOM-compliant (DIMSE/TLS). Features: Rigid and deformable image registration, volume rendering, ROI contouring, and DICOM-RT import/export. No hardware components. Software development follows ISO 13485 and 21 CFR 820 design controls.
Indications for Use
Indicated for physicians in diagnostic radiology, oncology, and radiation therapy planning to compare, display, annotate, register, and fuse DICOM-compliant medical images. Not intended for mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
March 6, 2024
Image /page/0/Picture/1 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Pymedix % Trey Thorsen RA Consultant Lean RAQA 131 E Loch Lomond Pl ORO VALLEY, AZ 85737
### Re: K233572
Trade/Device Name: Autofuse Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: January 30, 2024 Received: February 2, 2024
Dear Trey Thorsen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Loran Weiner
Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
Submission Number (if known)
#### K233572
Device Name
Autofuse
Indications for Use (Describe)
Autofuse is a software package that provides physicians a means for comparison of medical data including imaging data that is DICOM compliant. It allows the display, annotation, volume operation, volume rendering, registration and fusion of medical images as an aid during use by diagnostic radiology, oncology, radiation therapy planning, and other medical specialties. Autofuse is not intended for mammography.
Type of Use (Select one or both, as applicable)
( Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 1. APPLICANT INFORMATION
| Pymedix Company Contact | |
|-------------------------|------------------------|
| Name: | Kris Huang |
| Title: | Co-CEO, CTO |
| Phone Number: | 513-693-5845 |
| Email: | kris.huang@pymedix.com |
| Official Correspondent | |
| Name: | Trey Thorsen |
| Title: | RA consultant |
| Phone Number: | 850-450-3932 |
| Email: | treyt@leanraqa.com |
### 2. DEVICE INFORMATION
| Device Name: | Picture Archiving and Communication System |
|----------------------|------------------------------------------------|
| Common Name: | Medical Imaging Software |
| Classification Name: | Medical image management and processing system |
| Model(s): | Autofuse |
| Regulation Number: | 21 CFR § 892.2050 |
| Product Code: | LLZ |
| Class: | II |
### 3. PREDICATE DEVICES
| Predicate Type | 510(k)<br>Number | Name of Device | Name of Manufacturer |
|------------------|------------------|---------------------------------------------|---------------------------------|
| Primary Device | K081076 | Velocity AIS | Velocity Medical Solutions, LLC |
| Reference Device | K201607 | Eclipse Treatment<br>Planning System, v16.1 | Varian Medical Systems, Inc |
{4}------------------------------------------------
#### 4. DEVICE DESCRIPTION
Autofuse is a software application providing relevant tools for Radiotherapy professionals to use while creating treatment plans.
The Autofuse device is a Picture Archiving and Communication System (medical imaging software). Autofuse provides medical image processing designed to facilitate the oncology or other clinical specialty workflow by allowing the comparison of medical imaging data from different modalities, points in time, and / or scanning protocols. The product provides users with the means to display, co-register and analyze medical images from multiple modalities including PET, CT, RT Dose and MR; draw Regions of Interest (ROI); and import / export results to/from commercially available radiation treatment planning systems and PACS devices. Co-registration includes deformable registration technology that can be applied to DICOM data including diagnostic and planning image volumes, structures, and dose.
Autofuse is used as a stand-alone application on recommended Off-The-Shelf (OTS) computers supplied by the company or by the end-user.
#### 5. INDICATIONS FOR USE
Autofuse is a software package that provides physicians a means for comparison of medical data including imaging data that is DICOM compliant. It allows the display, annotation, volume operation, volume rendering, registration and fusion of medical images as an aid during use by diagnostic radiology, oncology, radiation therapy planning, and other medical specialties.
Autofuse is not intended for mammography.
| Characteristic | Proposed Device | Predicate Device | Remark |
|------------------------|---------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|---------|
| Device | Autofuse | Velocity AIS | N/A |
| Manufacturer | Pymedix, Inc. | Velocity Medical<br>Solutions, LLC | N/A |
| 510(k) Number | TBD | K081076 | N/A |
| Product Common<br>Name | Picture Archiving and<br>Communication System<br>(Medical Imaging<br>Software) | Picture Archiving and<br>Communication System<br>(Medical Imaging<br>Software) | Same |
| Product Code | LLZ | LLZ | Same |
| Classification | II | II | Same |
| Device description | Autofuse is a software<br>application providing<br>relevant tools for<br>Radiotherapy<br>professionals to use | Velocity is a software<br>application providing<br>relevant tools for<br>Radiotherapy<br>professionals to use | Similar |
#### 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
{5}------------------------------------------------
while creating treatment plans.
The Autofuse device is a Picture Archiving and Communication System (medical imaging software). Autofuse provides medical image processing designed to facilitate the oncology or other clinical specialty workflow by allowing the comparison of medical imaging data from different modalities, points in time, and / or scanning protocols. The product provides users with the means to display, co-register and analyze medical images from multiple modalities including PET. CT. RT Dose and MR; draw Regions of Interest (ROI): and import / export results to/from commercially available radiation treatment planning systems and PACS devices. Coregistration includes deformable registration technology that can be applied to DICOM data including diagnostic and planning image volumes, structures, and dose.
Autofuse is used as a stand-alone application on recommended Off-The-Shelf (OTS) computers supplied by the company or by the end-user.
while creating treatment plans.
The Velocity device is a Picture Archiving and Communication System (medical imaging software). Velocity provides medical image processing designed to facilitate the oncology or other clinical specialty workflow by allowing the comparison of medical imaging data from different modalities, points in time, and / or scanning protocols. The product provides users with the means to display, co-register and analyze medical images from multiple modalities including PET. SPECT. CT, RT Dose and MR; draw Regions of Interest (ROI), calculate and report relative differences in pixel intensities, Standardized Uptake Value (SUV) or other values within those regions; and import / export results to/from commercially available radiation treatment planning systems and PACS devices. Co-registration includes deformable registration technology that can be applied to DICOM data including diagnostic and planning image volumes, structures, dose, and automatic anatomical
{6}------------------------------------------------
| | | | atlas-based<br>segmentation tools. | | |
|------------------------------------|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | | | Velocity is used as a<br>stand-alone application<br>on recommended Off-<br>The-Shelf<br>(OTS)computers<br>supplied by the<br>company or by the end-<br>user. | | |
| | Indications for Use | Autofuse is a software<br>product that provides<br>physicians a means for<br>comparison of medical<br>data including imaging<br>data that is DICOM<br>compliant. It allows the<br>display, annotation,<br>volume operation,<br>volume rendering,<br>registration and fusion of<br>medical images as an aid<br>during use by diagnostic<br>radiology, oncology,<br>radiation therapy<br>planning, and other<br>medical specialties.<br>Autofuse is not intended<br>for mammography. | VelocityAIS (VelocityAI)<br>is a standalone software<br>product that provides<br>the physician a means<br>for comparison of<br>medical imaging data<br>from multiple DICOM<br>conformant imaging<br>modality sources. It<br>allows the display,<br>annotating, volume<br>rendering, registration<br>and fusing of medical<br>images as an aid during<br>use by diagnostic<br>radiology, oncology,<br>radiation therapy<br>planning and other<br>medical specialties.<br>VelocityAIS (VelocityAI)<br>is not intended for<br>mammography<br>diagnosis. | Similar | |
| General Usage/Performance Features | | | | | |
| | Image Study<br>Importation | CT/MR/PET | DICOM,<br>PET/SPECT/CT/MRI;<br>Dose | Similar. The<br>most commonly<br>used imaging<br>modalities used<br>in radiation<br>treatment<br>planning are<br>DICOM-format<br>CT, MR, and<br>PET images. | |
{7}------------------------------------------------
| | | | Autofuse does<br>allow import of<br>DICOM-RT<br>Dose, but it is<br>considered a<br>DICOM-RT<br>object<br>associated with<br>a scan, not an<br>imaging<br>modality itself.<br>This difference<br>does not bring<br>up any new<br>questions of<br>safety or<br>efficacy. | |
|------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Image Structure<br>Import, Save &<br>Export to Treatment<br>Planning Systems | Yes | Yes | Same | |
| Volume Rendering | Yes | Yes | Same | |
| Advanced<br>Visualization and<br>Navigation Features | Yes | Yes | Same | |
| Volume Operations | No | No | Same | |
| Diagnostic Image<br>Registration | Yes | Yes | Same | |
| Image Fusion | Yes; Anatomical and<br>Functional Images | Yes; Anatomical and<br>Functional Images | Same | |
| ROI & Contouring | Yes | Yes | Same | |
| Manual Contouring<br>Tools | Yes | Yes | Same | |
| Image Analysis | Yes | Yes | Same | |
| Plan Review of<br>Imported Plans or<br>Created Dose<br>Composites | No | No | Same | |
| Oncology Workflow<br>Automation | No | No | Same | |
| Image/ROI Export<br>to DICOM RT | Yes | Yes | Same | |
| Secure Login and<br>Data Storage | No | Yes | Velocity uses<br>an application-<br>level login.<br>Autofuse is<br>intended to run<br>on a secure | |
| | | | workstation and<br>logs OS-level<br>information of<br>the user logged<br>in and running<br>the application.<br>Data storage<br>and user<br>authentication<br>is handled by<br>the underlying<br>workstation OS,<br>not the<br>Autofuse<br>application.<br>This difference<br>does not bring<br>up any new<br>questions of | |
| | | | safety or<br>efficacy. | |
| Operating System<br>Platform | Ubuntu 22.04 | Microsoft Windows XP<br>32 bit only/ Windows 7,<br>Vista, MAC OSX 10.6 | Autofuse is<br>intended for<br>use in a secure<br>Linux-based<br>computing<br>environment.<br>This difference<br>does not bring<br>up any new<br>questions of<br>safety or<br>efficacy. | |
| Multimodality DICOM Import | | | | |
| Import and Display<br>DICOM CT, MR,<br>DS, PET, RTS, RTP | Yes | Yes | Same | |
| Advanced Visualization and Navigation | | | | |
| General Image<br>Viewer with View<br>Layout Selection<br>and Toolbars | Yes | Yes | Same | |
| User Scaling of<br>Image Volumes | No | No | Same | |
| Biological Effective<br>Dose (BED) Scaling | No | No | Same | |
| Y-90 Microsphere<br>Dosimetry | No | No | Same | |
| conversion of<br>SPECT to DS | | | | |
| | Automated Image-Based Registration | | | |
| Manual Registration<br>Editing | No | Yes | Autofuse does<br>not allow<br>manual<br>registration.<br>AAPM TG-132<br>states "direct,<br>unguided<br>manual<br>adjustment of a<br>deformation<br>vector field is<br>not<br>recommended".<br>This difference<br>does not bring<br>up any new<br>questions of<br>safety or<br>efficacy. | |
| Auto Rigid<br>Registration | Yes | Yes | Same | |
| Deformable<br>Registration | Yes | Yes | Same | |
| Inverse Deformable<br>Registration | No | No | Same | |
| Structure Guided<br>Deformable | No | No | Same | |
| | Segmentation (Manual Contouring Tools and Atlas Segmentation) | | | |
| Manual Contouring<br>Tools | Yes | Yes | Same | |
| Atlas Auto-<br>Segmentation | No | Yes | Autofuse does<br>not contain or<br>use atlases for<br>auto-<br>segmentation.<br>This difference<br>does not bring<br>up any new<br>questions of<br>safety or<br>efficacy. | |
| Image Analysis with Volumetric Graphs | | | | |
| Histograms and<br>Voxel Assessment<br>Graphs | Yes | Yes | Same | |
| | DVH Statistics<br>Display | Yes | Yes | Same |
| | Plan Review | | | |
| | Storage and Display<br>of DICOM RT Plans | No | No | Same |
| | Lesion Volume<br>Tracking by<br>Associating<br>Structure with<br>Name Tag | No | No | Same |
| | Navigator Semi-Automated Workflows | | | |
| | Semi-automatic<br>Workflows to Assist<br>with Common<br>Clinical Image<br>Registration &<br>Analysis Tasks | No. Image registration is<br>fully automated with no<br>user-facing workflow. | No | Same |
| | Adaptive Navigators<br>to Assist in Offline<br>Dose Review:<br>Includes workflows<br>to create adaptive<br>CT based on CBCT<br>Registration, copy<br>plan to adaptive CT,<br>and compare dose | No | No | Same |
| | Security | | | |
| | Logging of<br>Database Activity | No | Yes | Database<br>activity is<br>timestamped,<br>tagged by user<br>and logged for<br>all transactions.<br>This difference<br>does not bring<br>up any new<br>questions of<br>safety or<br>efficacy. |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
### 7. PERFORMANCE DATA
#### 7.1 Non-clinical performance Data
Software Verification and Validation Testing Software verification and validation was conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The primary objective of the design validation testing process was to ensure
{11}------------------------------------------------
documented confirmation that the device specifications meet user needs and the intended use through thorough testing in a simulated environment as part of the Design Control Process. Test results demonstrate conformance to applicable requirements and specifications.
The following references provide external standards and quidelines that are used, in part, for the definition of the quidelines provided in this document:
- 21 CFR Part 820, Quality System (QS) Regulation/Medical Device Good Manufacturing . Practices, US Food and Drug Administration
- . ISO 13485:2016. Medical devices - Quality management systems - Requirements for regulatory purposes, International Organization for Standardization
- ISO/DIS 10013 Quality management systems - Guidance for documented information, International Organization for Standardization
- Glossary of Computerized System and Software Development Terminology. Division of . Field Investigations, Office of Regional Operations, Office of Regulatory Affairs, Food and Drug Administration, August 1995.
- . AAPM Report No. 062 - American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy treatment planning (1998) ICRU Report 83 - Prescribing, Recording, and Reporting Intensity-Modulated Photon-Beam Therapy (IMRT)
- AAPM Report No. 132 Use of image registration and fusion algorithms and techniques in ● radiotherapy: Report of the AAPM Radiation Therapy Committee Task Group No. 132 (2017)
Non-clinical validation testing was performed in the Staging Environment (SE), which consists of a network of virtual machines (VMs) within a Kernel Virtual Machine (KVM) hypervisor. The network of VMs is designed to mimic a typical clinical set up including a departmental CT scanner and a radiology PACS communicating with unencrypted DIMSE communications or TLS-encrypted DIMSE communications.
#### 7.2 Biocompatibility
Not applicable. Autofuse is software.
#### 7.3 Clinical performance data
No clinical study is included in this submission.
### 8. CONCLUSION
The conclusions drawn from the nonclinical tests demonstrate that the proposed device is substantially equivalent, and performs as well as or better than the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.