Tono-Vera® Tonometer and Ocu-Dot® Tonometer Probes (16305, 16306, 16318)
K233516 · Reichert, Inc. · HKY · May 1, 2024 · Ophthalmic
Device Facts
Record ID
K233516
Device Name
Tono-Vera® Tonometer and Ocu-Dot® Tonometer Probes (16305, 16306, 16318)
Applicant
Reichert, Inc.
Product Code
HKY · Ophthalmic
Decision Date
May 1, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1930
Device Class
Class 2
Indications for Use
The Tono-Vera® Tonometer and Ocu-Dot® Tonometer Probes is intended to be used for the measurement of intraocular pressure of the human eye.
Device Story
Hand-held, battery-powered tonometer; uses rebound technology to measure intraocular pressure (IOP). Device propels a sterile, single-use plastic-tipped probe (Ocu-Dot) against the patient's unanesthetized cornea. Features ActiView™ camera-based alignment system to ensure proper positioning and reduce operator-dependent variability; measurement triggers automatically upon alignment. Offers two modes: 3+ (Quick) and 6-measurement. Data can be transferred via Bluetooth to ReichertSync® software for EMR integration. Used by eye care professionals in clinics/hospitals to assist in glaucoma screening and IOP monitoring. Benefits include reduced measurement variability and efficient, non-anesthetic IOP assessment.
Clinical Evidence
Prospective, cross-sectional, single-center clinical study (NCT05345262) conducted May-August 2022. Compared Tono-Vera to Goldmann Applanation Tonometer per ANSI Z80.10-2014 standards. Results confirmed device is safe and effective for IOP measurement without anesthesia. Bench testing established accuracy (±2 SD) of 0.49 (5-<20 mmHg) and 1.82 (>20-60 mmHg), with repeatability (CV%) of 2.30% and 2.00% respectively, outperforming the predicate device.
Technological Characteristics
Handheld, microprocessor-controlled rebound tonometer. Materials: thermoplastic housings and probe tips. Power: 4x 1.5V AA batteries or 3.7V Li-ion rechargeable pack. Connectivity: Bluetooth for data transfer to ReichertSync. Alignment: Camera-based ActiView system. Measurement range: 5-60 mmHg. Sterilization: Sterile, single-use probes.
Indications for Use
Indicated for the measurement of intraocular pressure (IOP) in the human eye. Intended for use by trained eye care professionals (ophthalmologists, opticians, technicians) in clinical settings. No specific age or gender restrictions noted; intended for use on unanesthetized eyes.
Regulatory Classification
Identification
A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.
{0}------------------------------------------------
May 1, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Reichert, Inc. Elizabeth Schultz Manager, Regulatory Affairs 3362 Walden Avenue, Suite 100 Depew. New York 14043
Re: K233516
Trade/Device Name: Tono-Vera® Tonometer and Ocu-Dot® Tonometer Probes (16305, 16306, 16318) Regulation Number: 21 CFR 886.1930 Regulation Name: Tonometer And Accessories Regulatory Class: Class II Product Code: HKY Dated: March 22, 2024 Received: March 25, 2024
Dear Elizabeth Schultz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Elvin Y. Ng -S
Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# K233516 - Elizabeth Schultz
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K233516
Device Name
Tono-Vera® Tonometer and Ocu-Dot® Tonometer Probes (16305, 16306, 16318)
Indications for Use (Describe)
The Tono-Vera® Tonometer and Ocu-Dot® Tonometer Probes is intended to be used for the measurement of intraocular pressure of the human eye.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) #: K233516 | | 510(k) Summary |
|---------------------------------------------------------------|----------------------------------------------------------|-----------------------------------------------------------------------------|
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | | Reichert, Inc. |
| Applicant Address | | 3362 Walden Avenue<br>Suite 100<br>Depew NY 14043<br>United States |
| Applicant Contact Telephone | | 716-686-4568 |
| Applicant Contact | | Ms. Elizabeth Schultz |
| Applicant Contact Email | | Elizabeth.Schultz@ametek.com |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | | Tono-Vera® Tonometer and Ocu-Dot® Tonometer Probes<br>(16305, 16306, 16318) |
| Common Name | | Tonometer and accessories |
| Classification Name | | Tonometer, Manual |
| Regulation Number | | 886.1930 |
| Product Code | | HKY |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K153694 | Icare Ic100 | HKY |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| Device Description [807.92(a)(4)]<br>a. Physical Description: | | |
The Reichert Tono-Vera® Tonometer is a hand-held, battery powered instrument that utilizes rebound methodology to measure the intraocular pressure (IOP) of the eye. The tonometer propels a small, plastic-tipped measurement probe in a controlled manner against a patient's unanesthetized eye in order to calculate IOP. The device has a simple, four-button control system and an LCD that provides information to the operator and displays measurement results.
There are two Tono-Vera Tonometer models:
· Tono-Vera Tonometer Starter Kit, AA Battery, Model 16305; and
• Tono-Vera Tonometer Starter Kit, Rechargeable, Model 16306.
{5}------------------------------------------------
Both models of the Tono-Vera Tonometer have a required accessory, the Ocu-Dot® Tonometer Probe, model 16318. The Ocu-Dot Tonometer Probe is a sterile, single-use, disposable accessory.
b. Device Features:
The Tono-Vera Tonometer utilizes a camera-based alignment system, called ActiView™ to improve alignment with the apex of the cornea and reduce measurement variability caused by operator technique. The measuring process starts automatically once alignment conditions are fulfilled.
The device has two measurement options: 3+ (Quick) Measurement Option and 6 Measurement Option. Both the 3+ (Quick) Measurement Option and 6 Measurement Option use the same calculation for the IOP result. The 3+ (Quick) Option can calculate a final IOP result with as few as three measurements. The 6 Measurement Option always uses six measurements to calculate the final IOP result.
The Tono-Vera Tonometer can be connected to a computer with ReichertSync® installed on it. Data from ReichertSync can be imported by an EMR system.
The Tono-Vera Tonometer determines IOP by detecting and analyzing the motion of the measurement probe as it briefly contacts the cornea.
c. Patient Contact: There are two direct patient contact points: the adjustable Flexi-Soft Forehead Rest which is used to stabilize the device during the measurement sequence and the Ocu-Dot Tonometer Probe which touches the patient's eye during the measurement sequence.
d. Use Environment: The Tono-Vera Tonometer is a prescription-only device. It is intended to be used by properly trained eve care professionals such as ophthalmologists, opticians, and eye-care technicians. The device will be used in doctors' offices, clinics, and hospitals.
#### Intended Use/Indications for Use
The Tono-Vera® Tonometer and Ocu-Dot® Tonometer Probes is intended to be used for the measurement of intraocular pressure of the human eye.
#### Indications for Use Comparison
The Tono-Vera® Tonometer and Ocu-Dot® Tonometer Probes have the same intended use and Indications for Use as the Icare Ic100 (K153694). Both devices are intended to be used for the measurement of intraocular pressure (IOP) of the human eye. The Indications for Use of the Tono-Vera Tonometer and Ocu-Dot® Tonometer Probes adopts the same language from the Indication for Use of the Icare Ic100 (K153694) in that both devices are intended to be used for the measurement of intraocular pressure of the human eye. Both devices have similar IOP measurement ranges and accuracy profiles. Both devices are similar in size, shape, and display function. The differences between the Tono-Vera® Tonometer and Ocu-Dot® Tonometer Probes and the Icare Ic100 (K153694), do not raise any different questions of safety and effectiveness as confirmed through the Performance Testing.
## Technological Comparison
The Tono-Vera® Tonometer and Ocu-Dot® Tonometer Probes (proposed device) and the Icare Ic100 (K153694) (predicate device) are highly similar in key safety and technological characteristics. Both devices use rebound technology to measure intraocular pressure (IOP). Both devices are active, microprocessor controlled, handheld devices that use a sterile, disposable, single-use probe to contact a patient's unanesthetized eye during the IOP measurement.
The bench test comparison for measurement accuracy and repeatability using proposed device and predicate device was performed to establish basis for substantial equivalence over the entire measurable
21 CFR 807.92(a)(5)
21 CFR 807.92(a)(5)
21 CFR 807.92(a)(6)
{6}------------------------------------------------
IOP range (5 to 60 mmHg) of the proposed device. The results for Pressure Range 5 to <20 mmHg showed the proposed device with Accuracy (±2 SD) of 0.49 and Repeatability (CV%) of 2.30% as compared to the predicate device with Accuracy (±2 SD) of 0.95 and Repeatability (CV%) of 6.30%. Whereas the results for Pressure Range >20 to 60 mmHg showed the proposed device with Accuracy (±2 SD) of 1.82 and Repeatability (CV%) of 2.00% as compared to the predicate device with Accuracy (±2 SD) of 2.57 and Repeatability (CV%) of 3.00%.
The Tono-Vera utilizes a camera-based alignment system, called ActiView, to improve alignment and reduce measurement variability associated with operator technique, whereas the predicate device is a manual device which does not have a comparable alignment system. However, the ActiView Auto Measure Mode can be disabled in the system menu and the operator can choose to take manual measurements instead.
The Tono-Vera has a Data Transfer feature where the IOP measurement results can be transferred via Bluetooth to an EMR system whereas the predicate has no Data Transfer capabilities.
Both the Tono-Vera Tonometer (Model 16305 (AA Battery)) and predicate device have the same power supply, 4 x 1.5V AA batteries. Whereas the second option of Tono-Vera Tonometer (Model 16306 (Rechargeable)) uses a 3.7V Li-ion rechargeable battery pack.
The Tono-Vera device has a measurement range of 5-60 mmHg (display range of 1-99 mmHg) and the predicate device has a measurement range of 7-50 mmHg (display range of 1-99 mmHg.). The predicate device uses six measurements to produce a single IOP result. Tono-Vera 6 Measurement Option also uses six measurements to produce a single IOP result. In addition, Tono-Vera has a second measurement option that can be selected by the operator. The Tono-Vera 3+ (Quick) Measurement option can produce an IOP result in as few as three measurements.
The devices are similar in Display, Weight, and Dimensions. The devices are similar in materials in that both use thermoplastic materials for the device housings and tonometer probe tip. There are minor differences in the Force applied to the eye during the rebound measurement and in the Device Orientation during use. The predicate has a wider range for force applied to the eve during measurement. The predicate device must be oriented roughly horizontally, whereas the Tono-Vera can be oriented horizontally within ± 15 degrees.
Overall, the Tono-Vera Tonometer and Ocu-Dot Tonometer Probes are substantially equivalent to the predicate device, Icare ic 100 Tonometer. They share technological characteristics with regard to physical and measurement attributes. Reichert has performed comprehensive performance testing to demonstrate that the device meets all required specifications. The testing demonstrated that the technological differences between Tono-Vera Tonometer and Ocu-Dot Tonometer Probes and the Icare Ic100 (K153694) do not raise any different questions of safety or effectiveness.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions
#### 21 CFR 807.92(b)
All necessary bench testing were conducted on the Tono-Vera® Tonometer and Ocu-Dot® Tonometer Probes to support a determination of substantial equivalence to the predicate device.
Non-Clinical Testing included:
- Performance Testing Bench
- · Electrical Safety and Electromagnetic Compatibility
- · Biocompatibility Testing
- · Sterilization Validation Testing
- · Shelf-Life and Packaging Validation Testing
{7}------------------------------------------------
The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of the Tono-Vera® Tonometer and Ocu-Dot® Tonometer Probes meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the Tono-Vera® Tonometer and Ocu-Dot® Tonometer Probes does not raise different questions of safety or effectiveness when compared to the predicate Icare Ic100 (K153694) Tonometer.
Reichert conducted a clinical study between May 2022 and August 2022. (National Clinical Trial Number NCT05345262). The primary endpoint of the study was to determine if the Reichert Tono-Vera® Tonometer is equivalent to the Goldmann Applanation Tonometer per the requirements of the ANSI Z80.10-2014 standard. The study was a prospective, cross-sectional, single-center, nonsignificant risk study. The results demonstrate that Reichert Tono-Vera® Tonometer is safe and effective at IOP tonometry measurements without the need for anesthesia per the requirements of the ANSI Z80.10-2014 standard.
The Tono-Vera® Tonometer and Ocu-Dot® Tonometer Probes are substantially equivalent to the predicate Icare Ic100 (K153694) Tonometer. The devices have the same intended use, similar indications for use, and both devices use rebound technology to measure intraocular pressure. The devices have similar IOP measurement ranges and accuracy profiles. They are similar in size, shape. and display function. The Clinical study results demonstrate that Tono-Vera is safe and effective at IOP tonometry measurements without the need for anesthesia per the requirements of the ANSI Z80.10-2014 standard.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.