← Product Code [FLE](/productcode/FLE) · K233504

# Lexa MINI (RIS-303, RIS-305) (K233504)

_W&H Sterilization S.R.L. · FLE · Feb 8, 2024 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K233504

## Device Facts

- **Applicant:** W&H Sterilization S.R.L.
- **Product Code:** [FLE](/productcode/FLE.md)
- **Decision Date:** Feb 8, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6880
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

Lexa MINI is a dynamic-air-removal (pre-vacuum) table-top steam sterilizer intended for use by a healthcare provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of medical and dental instruments that are validated to be sterilized by steam. The Lexa MNI has not been designed to sterlize liquid loads, biomedical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and/or damage to the autoclave. Key program features, including sterilization time, temperature and recommended load type are listed in the following: - Wrapped & Porous (270°F/132°C): 4' sterilization time; 10' drying time (recom.); for all types of load; wrapped items (Lexa MINI RIS-303: 2.2 lbs / Lexa MINI RIS-305: 3.3 lbs); wrapped Porous/Textile load (0.22 lbs) - Wrapped (270°F/132°C): 4' sterilization time; 10' drying time (recom.) for all types of load; except porous/textile load; wrapped items (Lexa MINI RIS-303: 2.2 lbs / Lexa MINI RIS-305: 3.3 1bs) - Low Temperature (250°C/121°C): 30' sterilization time (recom.); for all types of load; wrapped items (Lexa MINI RIS-303: 2.2 lbs / Lexa MINI RIS-305: 3.3 lbs); wrapped Porous/Textile load (0.22 lbs) - Unwrapped (270°F/132°C): 4' sterilization time, 1' drying time (recom.) for all types of load; except porous/textile load; unwrapped items (Lexa MINI RIS-303: 2.2 1bs / Lexa MINI RIS-305: 3.3 Ibs)

## Device Story

Dynamic-air-removal (pre-vacuum) table-top steam sterilizer; utilizes pressurized steam to sterilize medical/dental instruments. Operated by healthcare providers in clinical settings. Device features specific cycles (Wrapped & Porous, Wrapped, Low Temperature, Unwrapped) with defined sterilization/drying times and load weight limits (2.2 lbs for RIS-303; 3.3 lbs for RIS-305). Provides automated sterilization processing; ensures instrument sterility for clinical use.

## Clinical Evidence

No clinical data.

## Technological Characteristics

Dynamic-air-removal (pre-vacuum) steam sterilizer; table-top form factor; steam-based sterilization; specific cycle parameters for temperature (121°C/132°C) and time.

## Regulatory Identification

A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

2/08/2024

W&H Sterilization S.r.l. Paola Zampino Regulatory Affairs Manager Via Bolgara. 2 Brusaporto, IT 24060 Italy

Re: K233504

Trade/Device Name: Lexa MINI (RIS-303, RIS-305) Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: Class II Product Code: FLE Dated: October 23, 2023 Received: October 31, 2023

Dear Paola Zampino:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Christopher K. Dugard -S

Christopher K. Dugard, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical

{2}------------------------------------------------

and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K233504

Device Name Lexa MINI (RIS-303, RIS-305)

#### Indications for Use (Describe)

Lexa MINI is a dynamic-air-removal (pre-vacuum) table-top steam sterilizer intended for use by a healthcare provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of medical and dental instruments that are validated to be sterilized by steam.

The Lexa MNI has not been designed to sterlize liquid loads, biomedical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and/or damage to the autoclave. Key program features, including sterilization time, temperature and recommended load type are listed in the following:

- Wrapped & Porous (270°F/132°C): 4' sterilization time; 10' drying time (recom.); for all types of load; wrapped items (Lexa MINI RIS-303: 2.2 lbs / Lexa MINI RIS-305: 3.3 lbs); wrapped Porous/Textile load (0.22 lbs)

- Wrapped (270°F/132°C): 4' sterilization time; 10' drying time (recom.) for all types of load; except porous/textile load; wrapped items (Lexa MINI RIS-303: 2.2 lbs / Lexa MINI RIS-305: 3.3 1bs)

- Low Temperature (250°C/121°C): 30' sterilization time (recom.); for all types of load; wrapped items (Lexa MINI RIS-303: 2.2 lbs / Lexa MINI RIS-305: 3.3 lbs); wrapped Porous/Textile load (0.22 lbs)

- Unwrapped (270°F/132°C): 4' sterilization time, 1' drying time (recom.) for all types of load; except porous/textile load; unwrapped items (Lexa MINI RIS-303: 2.2 1bs / Lexa MINI RIS-305: 3.3 Ibs)

| Type of Use (Select one or both, as applicable) |                             |
|-------------------------------------------------|-----------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D)    |                             |
| Over-The-Counter Use (21 CFR 801 Subpart C)     | <div> <span> </span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda-staging.innolitics.com/device/K233504](https://fda-staging.innolitics.com/device/K233504)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).