K233502 · Graphy, Inc. · EBF · Nov 1, 2023 · Dental
Device Facts
Record ID
K233502
Device Name
TERA HARZ II
Applicant
Graphy, Inc.
Product Code
EBF · Dental
Decision Date
Nov 1, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3690
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
TERA HARZ II is indicated as an indirect restorative for both anterior and posterior restorations, including occlusal surfaces. The TERA HARZ II material is used for fabricating temporary or permanent restorations such as crowns and bridges, inlays, onlays, veneers and full crown restorations. Fabrication of TERA HARZ II requires a computer-aided and manufacturing (CAD/CAM) system that includes the following : scanner, design software, additive printer, and post-cure unit.
Device Story
TERA HARZ II is a light-cured, methacrylate-based photopolymer resin used by dental professionals for additive manufacturing of dental restorations. Input consists of digital STL files generated from intraoral scans via CAD/CAM software. The device is processed in a UV-light 3D printer (DLP or SLA) at 405-412nm, where photo-initiators trigger polymerization to form the restoration layer-by-layer. Post-printing, the object undergoes UV-light curing for final polymerization and is finished by the clinician (polishing/fitting). Used in dental clinics/labs; output is a permanent or temporary prosthetic. Benefits include customized, high-quality dental restorations. The device is non-sterile and prescription-only.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including flexural strength, depth of cure, water sorption, solubility, color stability, and biocompatibility (ISO 10993-1, ISO 7405). All tests met ISO 10477:2020 requirements.
Technological Characteristics
Methacrylate-based photopolymer resin; UV-curable (405-412nm). Form factor: liquid in 1,000g HDPE bottle. Manufacturing: additive 3D printing (DLP/SLA). Standards: ISO 10477:2020 (performance), ISO 10993-1 (biocompatibility). Non-sterile. No integrated software; requires external CAD/CAM ecosystem (scanner, design software, printer, post-cure unit).
Indications for Use
Indicated for patients requiring indirect anterior or posterior dental restorations, including crowns, bridges, inlays, onlays, veneers, and full crown restorations.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 1, 2023
Graphy Inc. % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059
Re: K233502
Trade/Device Name: Tera Harz II Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth shade resin material Regulatory Class: Class II Product Code: EBF, EBG Dated: October 30, 2023 Received: October 31, 2023
Dear Dave Yungvirt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name TERA HARZ II
Indications for Use (Describe)
TERA HARZ II is intended as an indirect restorative for both anterior restorations, including occlusal surfaces.
The TERA HARZ II material is used for fabricating temporary or permanent restorations such as crowns and bridges, inlays, onlays, veneers and full crown restorations.
Fabrication of TERA HARZ II requires a computer-aided and manufacturing (CAD/CAM) system that includes the following : scanner, design software, additive printer, and post-cure unit.
| | Brand | Type |
|---------------------------------|----------------|----------------------|
| Design : | 3 Shape A/S | TRIOS 3 Basic |
| Instraoral scanner | 3 Shape A/S | 3Shape dental system |
| Design software | | |
| Additive Manufacturing system : | | |
| 3D Printer | UNIZ | NBEE |
| | SprintRay Inc. | Pro 95 |
| Post-Curing : | | |
| Post-cure unit | CureM | U102H |
| | TERA HARZ CURE | THC |
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> </div> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
## For
## TERA HARZ II
[Complying with 21 CFR 807.92]
### l. SUBMISSION SPONSOR
Graphy Inc. #603, #617, Ace Gasan Forhu, 225, Gasan digital 1-ro, Geumcheon-gu, Seoul, Republic of Korea Office Phone: +82-2-864-3056 Fax: 82-2-864-3057 Contact Person: Mr. Moon-Soo Park, Assistant Manager of RA Team
## II. SUBMISSION CORRESPONDENT
## III. DATE PREPARED
04 October 2023
## IV. DEVICE
| Trade or Proprietary Name: | TERA HARZ II |
|----------------------------|------------------------------------------------------------------------------------------------------|
| Common or Usual Name: | Preformed Crown and Bridge |
| Classification Name: | Tooth shade resin material (21 CFR 872.3690)<br>Crown and Bridge, Temporary, Resin (21 CFR 872.3770) |
| Regulatory Class: | II |
| Product Code: | EBF, EBG |
| Classification Panel: | Dental |
## V. PREDICATE DEVICE
Primary Predicate Device: K202846, TERA HARZ / Graphy Inc. (Class II) Referencee Device K222414, TERA HARZ DENTURE / Graphy Inc. (Class II)
## VI. DEVICE DESCRIPTION
The TERA HARZ II is a light-cured, methacylate oligomer based polymerizable resin used by dentist or dental technician for the CAD/CAM manufacturing of indirect restorative for both anterior and posterior restorations, including dcclusal surfaces, such as temporary or permanent crowns and bridges, inlays, onlays, and veneers. Methacrylate based resin is well known materials, commonly used in the dental industry for fixed and removable prosthetic devices due to their physical-chemical and biocompatible properties.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the word "Graphy" in orange font. The font is sans-serif and appears to be bolded. The word is displayed horizontally and is the only element in the image.
The TERA HARZ II is made by Methacrylate-based resins. It has stored in a black 1,000g of HDPE bottle. It contains materials with shade
A1/A2/A3/B1/B2/B3/C1/C2/C3/D1/D2/D3/OM1/OM2/OM3/M4. This resin is a liquid photopolymer material that is polymerized by UV laser at 405~412nm. the resin can be used to create a customized artificial permanent tooth model with a 3d printer that is cured by ultraviolet light. The liquid UV curing resin is cured at a specific wavelength (395~405nm) by the photo-initiator contained in the resin. Curing in a 3D printer is related to the conditions of the printer equipment, and is typically 100um in layer thickness, and is output at a resolution of 40 to 90μm on the x, y axis. This device should use specific 3D Printer equipment using UV light source, and it is possible to produce three-dimensional printed matter by curing lamination step by step a thickness of 100μm.
However, scanner, design software, 3D printer and post-cure unit are not included with the device.
TERA HARZ II can be used in combination with all lasers and DLP based 3D printers which support dental materials. TERA HARZ II is a resin for the generative production of longterm temporary dental restorations based on image projection systems (405-412 nm). The formulation of TERA HARZ II is optimized for the requirements of a robust production guaranteeing constant high quality. The TERA HARZ II is successfully tested for biocompatibility, certainly meets all mechanical and application demands. The material is used in a 3D printer, which prints the shape determined by a 3D stereo-lithographic drawing.
The material can be used for build processes with layer thicknesses from 25 up to 100 um. After printing, the printed product is recommended to use a UV-light curing for final polymerization.
These fabrications of TERA HARZ II are beginning with the dental clinician prescribing indirect restorative to treat a patient's both anterior restorations, including occlusal surfaces, and decision to use methacrylate-based resins is made by the dental clinician. TERA HARZ II, a permanent or temporary restorations such as crowns and bridges, inlays, onlays, veneers and full crown restorations, is manufactured in a 3D printer that is compatible.
The dental clinician can generate a digital file by scanning the patient's mouth directly using approved Intraoral scanner software. This digital file is a series of CAD files (.stl) for building models that can be used to fabricate permanent or temporary restorations. Commonly used standard dental software is used by dental professionals to virtually design a restoration and generate an industry-standard "STL" 3D dataset which reflects the intended shape and contour. The design software used is 3D Scanner by 3Shape A/S (510(K) Exempt). The specialized prosthetic treatment planning software has a establishment registration for the intended use under FDA Classification Product Code
{5}------------------------------------------------
NOF, regulation 872.3661. This software is used for management of 3D scanned prosthetic models, prosthetic diagnosis by measuring, analyzing, inspecting and visualizing 3D scanned prosthetic models, virtual planning of prosthetic treatments by simulating tooth movements, and design of permanent or temporary restorations based on 3D scanned prosthetic models.
Once dental clinic manufacturing unit receive the data that * .stl CAD files of crown and bridge the 3D printer begins additive manufacturing. The dental clinician (e.g., dentist) generates sequential 3D printed models replicating the approved treatment plan. The permanent or temporary restorations is 3D printed and cured in a post-curing unit. The fabricated permanent or temporary restorations are cut to fit dentition, the cleaned and polished to remove rough edges by the dental clinician. The prescribing physician review and approves the permanent or temporary restorations are provides them to the patient the confirming fit and design.
## VII. INDICATION FOR USE
TERA HARZII is indicated as an indirect restorative for both anterior and posterior restorations, including occlusal surfaces.
The TERA HARZ II material is used for fabricating permanent or temporary restorations such as crowns and bridges, inlays, onlays, veneers and full crown restorations. Fabrication of TERA HARZ II requires a computer-aided and manufacturing (CAD/CAM) system that includes the following: scanner, design software, additive printer, and postcure unit.
| | Brand | Type |
|--------------------------------|-------------------------|----------------------|
| Design: | | |
| Intraoral scanner | 3Shape A/S | TRIOS 3 Basic |
| Design software | 3Shape A/S | 3Shape Dental System |
| Additive Manufacturing System: | | |
| 3D Printer | UNIZ | Nbee |
| | SprintRay Inc. | SprintRay Pro 95 |
| Post-Curing: | | |
| Post-cure unit | CureM<br>TERA HARZ CURE | U102H<br>THC2 |
{6}------------------------------------------------
### VIII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The indications for use and mechanism of action of subject device is identical to the predicate device and supports a determination of substantial equivalence.
However, there two main differences in the predicate device because the material of use and the physical property values are differed. The subject device has more flexible physical properties compared to the predicate device.
So, TERA HARZ DENTURE is selected as a reference device to support difference of raw material. The reference device has exactly the same raw material composition as the subject device. In addition, although there is a difference in physical property values, both products satisfy the acceptance criteria of ISO 10477:2020.
The following table compares the TERA HARZ II to the predicate device with respect to indications for use, principles of operation, technological characteristics, method of manufacturing.
The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device. Any differences in technology characteristics are accompanied by information that demonstrated the device is as safe and as effective as the predicate device and do not raise different questions of safety and effectiveness than the predicate.
It was concluded, therefore, that the technological differences do not raise different questions of safety and effectiveness
{7}------------------------------------------------
| | SUBJECT Device | Primary PREDICATE<br>Device (K202846) | Significant Difference |
|--------------|----------------|---------------------------------------|------------------------|
| Manufacturer | Graphy Inc. | Graphy Inc. | - |
| Trade Name | TERA HARZ II | TERA HARZ | |
| Regulation Description | -Tooth shade resin materials<br>- Crown and Bridge,<br>Temporary, Resin | -Tooth shade resin materials<br>- Crown and Bridge,<br>Temporary, Resin | Same |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| Regulation Number | -21 CFR 872.3690<br>-21 CFR 872.3770 | -21 CFR 872.3690<br>-21 CFR 872.3770 | Same e |
| Product Code | EBF<br>EBG | EBF<br>EBG | Same |
| Class | II | II | Same |
| Indication for Use | TERA HARZ II is indicated<br>as an indirect<br>restorative for both<br>anterior and posterior<br>restorations, including<br>occlusal surfaces.<br>The TERA HARZ<br>material is used for<br>fabricating temporary<br>or permanent<br>restorations such as<br>crowns and bridges,<br>inlays, onlays, veneers<br>and full crown<br>restorations.<br>Fabrication of TERA<br>HARZ requires a<br>computer-aided and<br>manufacturing<br>(CAD/CAM) system<br>that includes the<br>following: scanner,<br>design software,<br>additive printer, and<br>post-cure unit. | TERA HARZ is indicated<br>as an indirect<br>restorative for both<br>anterior and posterior<br>restorations, including<br>occlusal surfaces.<br>The TERA HARZ<br>material is used for<br>fabricating temporary<br>or permanent<br>restorations such as<br>crowns and bridges,<br>inlays, onlays, veneers<br>and full crown<br>restorations.<br>Fabrication of TERA<br>HARZ requires a<br>computer-aided and<br>manufacturing<br>(CAD/CAM) system<br>that includes the<br>following: scanner,<br>design software,<br>additive printer, and<br>post-cure unit. | Similar: The indication for<br>use of the subject device is<br>very similar to the predicate<br>device. |
| Mechanism of Action | The product is a<br>photopolymer liquid resin<br>that is manufactured to<br>produce<br>an output within UV lighting<br>3D printer. The material is for<br>producing<br>polymeric dental prosthesis<br>according to ISO 10477, type<br>2 and class 2. Using<br>a DLP or SLA 3D printer that<br>uses photopolymer materials,<br>it is produced 3D<br>output layer by layer. By<br>reading a STL file which<br>contains 3D information of<br>the dental prosthesis of a<br>patient's oral cavity, the 3D<br>printer can create a | The product is a<br>photopolymer liquid resin<br>that is manufactured to<br>produce<br>an output within UV lighting<br>3D printer. The material is for<br>producing<br>polymeric dental prosthesis<br>according to ISO 10477, type<br>2 and class 2. Using<br>a DLP or SLA 3D printer that<br>uses photopolymer materials,<br>it is produced 3D<br>output layer by layer. By<br>reading a STL file which<br>contains 3D information of<br>the dental prosthesis of a<br>patient's oral cavity, the 3D<br>printer can create a | Same |
{8}------------------------------------------------
| | SUBJECT Device | Primary PREDICATE<br>Device (K202846) | Significant Difference |
|--------------|----------------|---------------------------------------|------------------------|
| Manufacturer | Graphy Inc. | Graphy Inc. | - |
| Trade Name | TERA HARZ II | TERA HARZ | |
| | continuous layer. When the<br>3D printer emits light to<br>the resin with specific pattern<br>of a shape, the photoinitiator<br>in the resin<br>absorbs the pattern of UV<br>wavelength band. It causes a<br>photocuring reaction<br>in a pattern shape and the<br>photopolymer resin is<br>curedlayer by layer and it is<br>continuously carried out to<br>produce the shape that you<br>want. The operation | continuous layer. When the<br>3D printer emits light to<br>the resin with specific pattern<br>of a shape, the photoinitiator<br>in the resin<br>absorbs the pattern of UV<br>wavelength band. It causes a<br>photocuring reaction<br>in a pattern shape and the<br>photopolymer resin is<br>curedlayer by layer and it is<br>continuously carried out to<br>produce the shape that you<br>want. The operation | |
|-------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | is carried out continuously | is carried out continuously | |
| | and finally the output are<br>produced. | and finally the output are<br>produced. | |
| Manufacturing Technology | Additive | Additive | Same |
| Material of Use | Methacrylate-based resins<br>with photo-<br>initiator, inhibitor and<br>pigments | Methacrylate polymer<br>resin (dimethacrylate) | Different : the subject device<br>differs from the predicate<br>device in the material of use.<br>The two products are similar<br>methacrylate-based resins,<br>but they differ slightly in raw<br>material composition. |
| Product state | Pre-mixed resin(liquid) | Pre-mixed resin(liquid) | Same |
| Biocompatibility | Biocompatible<br>according to ISO<br>10993-1 | Biocompatible<br>according to ISO<br>10993-1 | Same |
| Performance Testing | ISO 10477:2020 | ISO 10477:2020 | Same : the subject device<br>and predicate device have<br>same performance testing<br>standard |
| Depth of Cure (%)<br>(Bottom surface shall be not<br>less than 70% of the<br>hardness of top surface) | Avg. 91.3 | Avg. 93.5 | Similar : the specifications<br>are in the same range. This<br>minor variance does not<br>introduce additional safety o<br>efficacy concerns. |
| Flexural Strength<br>(≥50 MPa; ISO 10477) | Avg. 125.5 | Avg. 148.73 | Similar : the specifications<br>are in the same range. This<br>minor variance does not<br>introduce additional safety o<br>efficacy concerns. |
| Water sorption<br>(≤40µg/mm3) | Avg. 22.03 | Avg. 13.03 | Similar : the specifications<br>are in the same range. This<br>minor variance does not<br>introduce additional safety o<br>efficacy concerns. |
| Solubility<br>(≤7.5 µg/mm3) | Avg. 0.12 | Avg. 1.00 | Similar : the specifications<br>are in the same range. This<br>minor variance does not<br>introduce additional safety o<br>efficacy concerns. |
{9}------------------------------------------------
| | SUBJECT Device | Primary PREDICATE<br>Device (K202846) | Significant Difference |
|--------------|----------------|---------------------------------------|------------------------|
| Manufacturer | Graphy Inc. | Graphy Inc. | - |
| Trade Name | TERA HARZ II | TERA HARZ | |
| | | | efficacy concerns. |
|-------------------|-------------------------------------------------------------------|-------------------------------------------------------------------|--------------------|
| Shade consistency | The difference in color from different bataches was not observed. | The difference in color from different bataches was not observed. | same |
| Color stability | Any color change after aging treatments was not detected. | Any color change after aging treatments was not detected. | same |
| OTX or Rx | Rx | Rx | Same |
| Sterile | non-sterile | non-sterile | Same |
| Shelf-life | 2 year | 2 year | Same |
{10}------------------------------------------------
| | SUBJECT Device | REFERENCE Device<br>(K222414) | Significant Difference |
|--------------|----------------|-------------------------------|------------------------|
| Manufacturer | Graphy Inc. | Graphy Inc. | - |
| Trade Name | TERA HARZ II | TERA HARZ DENTURE | |
| Regulation Description | -Tooth shade resin materials<br>- Crown and Bridge,<br>Temporary, Resin | -resin, denture, relining,<br>repairing, rebasing | Different |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Regulation Number | -21 CFR 872.3690<br>-21 CFR 872.3770 | -21 CFR 872.3760 | Different |
| Product Code | EBF<br>EBG | EBI | Different |
| Class | II | II | Same |
| Indication for Use | TERA HARZ II is indicated as an indirect<br>restorative for both anterior and posterior<br>restorations, including occlusal surfaces.<br>The TERA HARZ material is used for<br>fabricating temporary or permanent<br>restorations such as crowns and bridges,<br>inlays, onlays, veneers and full crown<br>restorations.<br>Fabrication of TERA HARZ requires a<br>computer-aided and manufacturing<br>(CAD/CAM) system that includes the<br>following: scanner, design software,<br>additive printer, and post-cure unit. | The TERA HARZ DENTURE is a<br>light-<br>curable resin indicated for fabrication and<br>repair of full and partial removable<br>dentures and baseplates. The material is<br>an alternative to traditional heat-curable<br>and auto polymerizing resins. Fabrication<br>of dental prosthetics with the TERA HARZ<br>DENTURE requires a computer-aided<br>design and manufacturing (CAD/CAM)<br>system that includes the following<br>components: digital denture base files<br>based on digital impression,<br>stereolithographic additive printer, and<br>curing light equipment. | Different: The indication for<br>use of the subject device is<br>different from the reference<br>device. |
| Mechanism of Action | The product is a<br>photopolymer liquid resin<br>that is manufactured to<br>produce<br>an output within UV lighting<br>3D printer. The material is for<br>producing<br>polymeric dental prosthesis<br>according to ISO 10477, type<br>2 and class 2. Using<br>a DLP or SLA 3D printer that<br>uses photopolymer materials,<br>it is produced 3D<br>output layer by layer. By<br>reading a STL file which<br>contains 3D information of<br>the dental prosthesis of a<br>patient's oral cavity, the 3D<br>printer can create a | Denture Base Resin is used to<br>fabricated synthetic<br>substances that form<br>denture base after<br>polymerization. The denture<br>base is the part of a denture<br>that rests on the foundation<br>tissues and to which teeth<br>are attached. | Different |
{11}------------------------------------------------
| | SUBJECT Device | REFERENCE Device<br>(K222414) | Significant Difference |
|--------------|----------------|-------------------------------|------------------------|
| Manufacturer | Graphy Inc. | Graphy Inc. | - |
| Trade Name | TERA HARZ II | TERA HARZ DENTURE | |
| Manufacturing Technology | Additive | Additive | Same |
|--------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| Material of Use | Methacrylate-based resins<br>with photo-<br>initiator, inhibitor and<br>pigments | Methacrylate-based resins<br>with photo-<br>initiator, inhibitor and<br>pigments | Same : The subject device<br>and the reference device<br>have the same raw material. |
| Product state | Pre-mixed resin(liquid) | Pre-mixed resin(liquid) | Same |
| Biocompatibility | Biocompatible<br>according to ISO<br>10993-1 | Biocompatible<br>according to ISO<br>10993-1 | Same |
| Performance Testing | ISO 10477:2020 | ISO 20795-1:2013 | Different : The subject device<br>and the reference device<br>applied different technical<br>standards. |
| OTX or Rx | Rx | Rx | Same |
| Sterile | non-sterile | non-sterile | Same |
| Shelf-life | 2 year | 2 year | Same |
{12}------------------------------------------------
## IX. PERFORMANCE DATA
The following performance data were provided in support of thesubstantial equivalence determination.
## Manufacturing Validation
A manufacturing validation was performed to demonstrate themanufacturing process for TERA HARZ II
An independent 3rd party software and digital calipers were used to perform point-topoint and critical displacement measurement.All translational measurements were within 0.150 mm of the target input value, the predefined tolerance of the manufacturing process. There were no statistical differences in the difference in the intended and measured values observed from any of the groups. This test has met the pre-established acceptance criteria.
And the test also conducted studies on the effect of manufacturing validation and material reuse on the properties of the final finished device according to FDA's published guidance documents, "Technical Considerations for Additive Manufactured Medical Devices".
The TERA HARZ II is outputted by each different output condition and each flexural strength were measured and the evaluation criteria of the all the specimens were more than 50 MPa. The optimal output condition is that the output angle is 0°, and the output position is centered to confirming that the optimal condition is to be output. In addition, it was confirmed that there was no problem in the number of effective outputs for repeated material output up to 6 times.
## Performance Testing
The predicate devices performed tests for Depth of Cure, Flexural Strength, Water Solubility, Water Sorption, and Biocompatibility content. The predicate/reference devices performed these tests as well and all met the requirements of ISO 10477:2020, Dentistry - Polymerbased crown and veneering materials.
## Comparison Performance Testing
Bench testing was performed on both the subject device and the predicate device (K202846) to evaluate critical properties including Flexural strength, Shaede consistency and Color stability, Water sorption,solubility, and depth of cure. All met the requirements of ISO 10477:2020.
## Biocompatibility
Biocompatibility Tests in accordance with the FDA Guidance Document, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1 : Evaluation and Testing within a risk management process".
The subject device is considered a surface device that is contact with the mucosal
{13}------------------------------------------------
membrane for > 30 days. The ISO 10993-1 standard was followed the following biological safety aspects have been addressed:
- –ISO 7405:2018, Dentistry Evaluation of biocompatibility of medical devices used in dentistry
- –ISO 10993-1:2018, Biological evaluation of medical devices Part 1: Evaluation and testing
- -ISO 10993-3:2014, Biological evaluation of medical devices Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
- –ISO 10993-5:2009, Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
- –ISO 10993-6:2016, Biological evaluation of medical devices Part 6: Tests for local effects after implantation
- –ISO 10993-10:2010, Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization
- –ISO 10993-11:2017, Biological evaluation of medical devices Part 11: Tests for systemic toxicity
## X. CLINICAL DATA
Clinical performance data was not provided for TERA HARZ II.
## XI. CONCLUSIONS
The TERA HARZ II is very similar to the predicate device and demonstrate substantial equivalence to predicate device, K202846.
An analysis for subject device compared to the predicate device show TERA HARZ II and
the TERA HARZ II meet the requirements of Manufacturing Validation, all two devices share the same product code, meet the requirements, and all two are biocompatible.
In addition, an analysis for subject device compared to the predicate device show TERA HARZ II and the TERA HARZ meet the requirements of ISO 10477:2020, Polymer-based crown and veneering materials All two devices meet or exceed the minimum strength requirements, and all two are biocompatible.
Any differences between subject device and the predicate device are material of use and value of physical properties . So reference device, which has same composition of raw
materials, is added to support material of use. Also two products meet the requirements of ISO 10477:2020.
Any differences between subject device and the predicate devices are minimal and present no new risks.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.