K233460 · Connexicon Medical , Ltd. · OMD · Jul 15, 2024 · General, Plastic Surgery
Device Facts
Record ID
K233460
Device Name
CM00622 LINC Skin Closure System (CM00622 LINC)
Applicant
Connexicon Medical , Ltd.
Product Code
OMD · General, Plastic Surgery
Decision Date
Jul 15, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4011
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Skin Closure System is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including incisions from minimally invasive surgery, and simple thoroughly cleansed, trauma- induced lacerations. The Skin Closure System should be used in conjunction with, but not in place of, deep dermal stitches. Additionally, the adjunct wound closure device component maintains temporary skin edge alignment along the length of the wound during application of the liquid adhesive.
Device Story
CM00622 LINC is a sterile, single-use topical skin closure system; consists of a liquid adhesive (2-octyl cyanoacrylate with D&C Violet No. 2) in a crushable glass ampoule within a pen applicator, and a self-adhering polyester mesh. Used by clinicians to close surgical incisions up to 20 cm and trauma-induced lacerations. Operation: clinician applies mesh to skin edges for temporary alignment; manually compresses applicator body to break ampoule; applies liquid adhesive over mesh. Adhesive polymerizes within minutes to form a physical barrier to microbial penetration for up to 7 days. Acts as an adjunct to deep dermal sutures. Benefits include temporary skin edge alignment and physical wound protection.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing and an animal wound healing study. Bench testing included peel/shear adhesion, tensile properties, lap-shear strength, T-peel adhesion, wound closure strength, adhesive degradation, heat of polymerization, viscosity, dry time, microbial barrier testing, and flow control. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, sensitization, intracutaneous reactivity, pyrogenicity, acute systemic toxicity, and implantation.
Technological Characteristics
Materials: 2-octyl cyanoacrylate adhesive, D&C Violet #2 colorant, quaternary ammonium salt accelerant, polyester mesh with pressure-sensitive adhesive. Form factor: single-use pen applicator with internal glass ampoule. Sterilization: sterile. Standards: ASTM D3330, D3654, D882, F2255, F2256, F2258, F2458. No software or electronic components.
Indications for Use
Indicated for topical approximation of easily approximated skin edges of surgical incisions (including minimally invasive) and simple, thoroughly cleansed, trauma-induced lacerations. Must be used with deep dermal stitches. Not indicated for puncture wounds.
Regulatory Classification
Identification
A tissue adhesive with adjunct wound closure device intended for the topical approximation of skin is a device indicated for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. It may be used in conjunction with, but not in place of, deep dermal stitches. Additionally, the adjunct wound closure device component maintains temporary skin edge alignment along the length of wound during application of the liquid adhesive.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Guidance for Industry and FDA Staff; Class II Special Controls Guidance Document: Tissue Adhesive with Adjunct Wound Closure Device Intended for the Topical Approximation of Skin.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
July 15, 2024
Connexicon Medical Ltd. Martin Brennan Director of Quality and Regulatory Affairs Synergy Centre Institute of Technology Tallaght (ITT) Dublin, Dublin D24 A386 Ireland
Re: K233460
Trade/Device Name: CM00622 LINC Skin Closure System (CM00622 LINC) Regulation Number: 21 CFR 878.4011 Regulation Name: Tissue Adhesive With Adjunct Wound Closure Device For Topical Approximation Of Skin Regulatory Class: Class II Product Code: OMD Dated: June 11, 2024 Received: June 13, 2024
Dear Martin Brennan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Julie A.
Morabito -S
Julie Morabito, Ph.D. Assistant Director
{2}------------------------------------------------
DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K233460
Device Name
CM00622 LINC Skin Closure System (CM00622 LINC)
### Indications for Use (Describe)
The Skin Closure System is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including incisions from minimally invasive surgery, and simple thoroughly cleansed, trauma- induced lacerations. The Skin Closure System should be used in conjunction with, but not in place of, deep dermal stitches. Additionally, the adjunct wound closure device component maintains temporary skin edge alignment along the length of the wound during application of the liquid adhesive.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Connexicon Medical Ltd Synergy Centre,
ITT Tallaght
Dublin 24
Dublin 24
D24 A386
Ireland
### 510(k) Summary
| Submitted by: | Connexicon Medical Limited,<br>Synergy Centre,<br>ITT Dublin,<br>Dublin 24,<br>D24 A386<br>Tel +353 1 5334996 |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Contact Person: | Martin Brennan<br>Director of Quality and Regulatory Affairs |
| Date of Summary: | 15 July 2024 |
| Device Name: | CM00622 LINC Skin Closure System |
| Common Name: | Cutaneous Tissue Adhesive With Mesh |
| Classification Name: | Tissue adhesive with adjunct wound closure device for topical<br>approximation of skin |
| Regulatory Number: | 21 CFR 878.4011 |
| Device Class: | Class II |
| Product Code: | OMD |
| Predicate Device<br>Device Name:<br>510(k) Clearance: | DERMABOND™ PRINEO™ Skin Closure System<br>Primary predicate device (K133864) |
{5}------------------------------------------------
## Device Description
CM00622 Skin Wound Closure System is a sterile, liquid topical skin adhesive containing a monomeric (2-octyl cyanoacrylate) formulation and the colorant D & C Violet No. 2. It is provided in a single-use applicator packaged in a rigid blister. The applicator is composed of a crushable glass ampoule contained within a pen applicator with an attached applicator tip. As applied to skin, the liquid topical skin adhesive is slightly more viscous than water and polymerizes within minutes. In vitro studies have shown that following polymerization, CM00622 LINC acts as a physical barrier (for up to 7 days) to microbial penetration as long as the polymerized liquid topical skin adhesive film remains intact. Clinical studies were not conducted to demonstrate microbial barrier properties and a correlation between microbial barrier properties and a reduction in infection have not been established.
CM00622 LINC also incorporates a self-adhering mesh that is applied to the approximated skin edges to provide temporary skin edge alignment of incisions up to 20 cm in length until the liguid topical skin adhesive is applied to achieve skin closure.
### Indications for Use
CM00622 LINC is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, and simple, thoroughly cleansed, trauma-induced lacerations. CM00622 LINC should be used in conjunction with, but not in place of, deep dermal stitches. Additionally, the adjunct wound closure device component maintains temporary skin edge alignment along the length of the wound during application of the liquid adhesive.
## Comparison of Technological Characteristics with the Predicate Device
The technological characteristics of the CM00622 LINC Skin Closure System and the predicate device are similar, in that they both contain the same technological characteristics:
- Adhesive: 2-octyl cyanoacrylate based adhesive with D&C Violet #2 colorant
- . Accelerant: Quaternary ammonium salt
- . Applicator: Sterile single use applicator
- . Mesh: Polyester mesh with pressure sensitive adhesive
### Differences compared to the Predicate Device
| Use | The indications for use of the subject and predicate device are identical except the<br>subject device does not include punctures. The difference is not critical as the<br>indications are lesser for the subject device. |
|-------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technology | The initiator is located in the filter for the subject device as compared to the mesh<br>for the predicate device.<br>The activation method for the subject device involves manually compressing the<br>applicator body to break the glass ampoule whilst a collar is required to be turned<br>to break the glass ampoule for the predicate device.<br>- Performance testing found that the subject device is substantially equivalent to<br>the predicate device and these differences do not affect safety or efficacy. |
| Performance | It was determined through the testing (Bench and Animal) performed in this 510(k)<br>submission, that the subject device is substantially equivalent to the predicate<br>device and any minor differences in performance does not affect safety or efficacy. |
{6}------------------------------------------------
## Performance Data
Testing was performed in accordance with the FDA special controls guidance document for "Tissue Adhesive with Adjunct Wound Closure Device Intended for the Topical Approximation of Skin" Issued on: November 10, 2010
## Performance Testing
The following tests were performed on the CM00622 LINC Skin Closure System to demonstrate substantial equivalence to the predicate device:
- Peel Adhesion of Pressure-Sensitive Tape (ASTM D3330/D3330M-04)
- Shear Adhesion of Pressure-Sensitive Tapes (ASTM D3654/D3654M-06)
- Tensile Properties of Thin Plastic Sheeting (ASTM D882-12) ●
- Lap-shear strength (ASTM F2255-05)
- T-peel adhesion strength(ASTM F2256-05)
- Adhesive strength in tension (ASTM F2258-05)
- Wound closure strength(ASTM F2458-05)
- Adhesive degradation study
- Heat of polymerization
- Viscosity
- Dry time
- Microbial barrier testing
- Device Yield and Quality of Film
- Flow Control
- Animal wound healing study
## Biocompatibility
The biological evaluation of the CM00622 LINC Skin Closure System was performed in accordance with FDA guidance on the use of ISO 10993-1.
The following test reports were provided in this submission:
- . Physical and/or chemical information: Solvent compatibility, Chemical characterization and Toxicological risk assessment.
- . Cytotoxicity.
- Sensitization.
- Intracutaneous reactivity.
- Material mediated pyrogenicity.
- Endotoxin mediated pyrogenicity.
- Acute systemic toxicity.
- Implantation (including sub-acute toxicity).
## Conclusions
Based on the intended use, technological characteristics, safety and performance testing, CM00622 LINC Skin Closure System has been demonstrated to be substantially equivalent to the predicate device, DERMABOND™ PRINEO™ Skin Closure System
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.