K233442 · Biotex, Inc. · OCX · May 21, 2024 · Gastroenterology, Urology
Device Facts
Record ID
K233442
Device Name
troCarWash™ System
Applicant
Biotex, Inc.
Product Code
OCX · Gastroenterology, Urology
Decision Date
May 21, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
The troCarWash™ system consists of a reusable control unit and a disposable, sterile, single-use trocar intended to remove visual obstructions such as condensation, blood, and other tissue particulates from the distal lens of a laparoscope during surgery and therefore maintain a clear image of the surgical site.
Device Story
System comprises mains-powered reusable control unit and disposable sterile trocar/obturator/tubing set. Used in hospital settings by medical professionals during laparoscopic procedures. Device detects laparoscope position via light sensor when surgeon retracts scope into trocar. Upon detection, system initiates automated <250ms wash/dry cycle using CO2 and saline/surfactant mix to clear distal lens. Output is a clean lens, restoring surgical vision. Benefits include reduced need to remove scope for manual cleaning, maintaining continuous surgical site visualization.
Clinical Evidence
No clinical data. Bench testing only. Performance evaluated using 37 disposable sets (preconditioned with sterilization, accelerated aging, and simulated transport/distribution) to verify wash efficacy across multiple scope brands and angles.
Technological Characteristics
Materials: ABS, polycarbonate, PVC. Sensing: Light sensor for scope position detection. Energy: Mains-powered. Cleaning: CO2 and saline/surfactant mix. Form factor: Reusable control unit with disposable trocar/obturator/tubing. Sterilization: Irradiation (VDmax25). Software: Minor level of concern. Standards: IEC 60601-1, IEC 60601-1-6, IEC 60601-2-18, IEC 60601-1-2, ISO 10993, ISO 11137, ISO 11607.
Indications for Use
Indicated for patients undergoing laparoscopic surgery to maintain clear visualization by removing lens contaminants including condensation, blood, peritoneal fluid, smoke, fat, and tissue smears.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 21, 2024
Biotex, Inc. Wade Munsch Regulatory Affairs Manager 114 Holmes Rd. Houston, Texas 77045
Re: K233442
Trade/Device Name: troCarWash™ System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: OCX, OCT Dated: April 23, 2024 Received: April 23, 2024
Dear Wade Munsch:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Long H. Chen -S Digitally signed by Long H. Chen-S
Digitally signed Date: 2024.05.21 08:22:21 -0400'
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
Submission Number (if known)
K233442
Device Name
troCarWash™ System
#### Indications for Use (Describe)
The troCarWash™ system consists of a reusable control unit and a disposable, sterile, single-use trocar intended to remove visual obstructions such as condensation, blood, and other tissue particulates from the distal lens of a laparoscope during surgery and therefore maintain a clear image of the surgical site.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Contact Details
| Applicant Name | Biotex, Inc. |
|-----------------------------|-----------------------------------------------|
| Applicant Address | 114 Holmes Rd. Houston TX 77045 United States |
| Applicant Contact Telephone | 7137410111 x208 |
| Applicant Contact | Mr. Wade Munsch |
| Applicant Contact Email | wade.munsch@biotexmedical.com |
#### Device Name
| Device Trade Name | troCarWash™ System |
|---------------------|--------------------------------------|
| Common Name | Endoscope and accessories |
| Classification Name | Endoscopic Irrigation/Suction System |
| Regulation Number | 876.1500 |
| Product Code | OCX |
#### Legally Marketed Predicate Devices
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|--------------------|----------------------------------------------------------|--------------|
| K222695 | troCarWash™ System | OCX |
| K102452(Reference) | Defogging and cleaning solution – Flo-X | OCT |
#### Device Description Summary
The troCarWash™ system is a laparoscopic lens cleaning device which is intended to be used during any laparoscopic surgical procedure where there is a potential for contamination of the distal lens of the laparoscope. Consisting of a mains powered reusable control unit, and a single use disposable trocar, obturator, and tubing set, it is intended to maintain surgical vision by removing contaminants such as condensation, blood, peritoneal fluid, smoke, fat, and tissue smears that have contaminated the distal lens of the laparoscope during surgical procedures providing a clear image of the surgical site. The disposable portion of the system is sterilized via irradiation and has limited (<24 hours) patient contact with abdominal tissue. The trocar and obturator are primarily made of ABS and polycarbonate, and the tubing set is made of ABS, Polycarbonate, and PVC tubing. The system is intended for use in typical hospital environments by medical professionals and is suitable for all patients approved for laparoscopic operations. The system incorporates a software that initiates a short (<250 millisecond) wash and dry mechanism to efficiently clean the lens of the laparoscope and may be initiated with a natural single-handed surgical motion of retracting the scope momentarily into the supplied trocar.
#### Intended Use/Indications for Use
The troCarWash™ system consists of a reusable control unit and a disposable, sterile, single-use trocar intended to remove visual obstructions such as condensation, blood, and other tissue particulates from the distal lens of a laparoscope during surgery and therefore maintain a clear image of the surgical site.
#### Indications for Use Comparison
Indications for use between the predicate device and the subject device are the same.
{4}------------------------------------------------
## Technological Comparison
The troCarWash™ System is substantially equivalent to the troCarWash™ System. Both the subject device and the predicate device have the same intended use and are used for the same indications. Specifically, both devices are intended to remove visual obstructions such as condensation, blood, and other tissue particulates from the distal lens of a laparoscope during surgery and therefore maintaining a clear image of the surgical site. In addition, both the subject device and the predicate device use medical grade CO2 and saline to clear the lens of debris, have one mode to clean, have a light sensor to detect the position of the scope as the user pulls the scope back within the trocar to deliver a cleaning cycle, and utilize a trocar-based wash apparatus designed to fit all 10mm scopes regardless of brand or angle. Both the subject device and predicate device require the user to apply a drop of wash solution to the end of the scope. The technical differences between the subject device and the predicate device consist of the following:
- The subject device utilizes a diluted mix of the same FDA-cleared surfactant and saline . as the wash solution to prevent re-fogging.
The reference device supports the scientific justification of the patient contacting type, with indication for use as follows:
"The Defogging and Cleaning Solution is a single-use laparoscopic accessory device intended to facilitate intra-operative defogging and cleaning of the laparoscope lens thereby maintaining visualization of the surgical site", which is substantially equivalent to the subject device wash solution and supports the second half of the indications for use "… to remove visual obstructions such as condensation, blood, and other tissue particulates from the distal lens of a laparoscope during surgery and therefore maintain a clear image of the surgical site" since both devices have similar working mechanisms.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
The following performance data is provided in support of the substantial equivalence determination.
## Packaging Validation
Packaging validation testing was completed per the listed standards:
- ASTM F88/F88M: Standard Test Method for Seal Strength of Flexible Barrier Materials
- ASTM F2096: Standard Test Method for Detecting Gross Leaks in Packaging by Internal . Pressurization (Bubble Test)
- . ASTM F1980: Standard Guide for Accelerate Aging of Sterile Barrier Systems for Medical Devices
- ASTM D4332: Standard Practice for Conditioning Containers, Packages, or Packaging Components for Testing
- ASTM D4169: Standard Practice for Performance Testing of Shipping Containers and Systems
- . ISO 11607-1: Packaging of Terminally Sterilized Medical Devices – Part 1 Requirements for Materials, Sterile Barrier Systems and Packaging Systems
- . ISO 16269-6: Statistical Interpretation of Data - Part 6 Determination of Statistical Tolerance Intervals
The troCarWash™ system passed the seal strength and package integrity testing for environmental conditioning and simulated T&D, and Accelerating Aging.
{5}------------------------------------------------
## Sterilization Validation
Sterilization validation testing was completed per the listed standards:
- ANSI AAMI ISO 11137-1: Sterilization of health care products Radiation part 1: . Requirements for development, validation and routine control of a sterilization process for medical devices
- . ANSI AAMI ISO 11137-2: Sterilization of health care products - Radiation - Part 2: Establishing the Sterilization Dose
- . AAMI/ISO 11737-1: Sterilization of medical devices – Microbiological methods – Part 1: Determination of population of microorganisms on products
- . AAMI/ISO 11737-2: Sterilization of a medical devices – Microbiological methods – Part 2: Tests of sterility performed in the definition, validation, maintenance of a sterilization process
The troCarWash™ system is sterilized via irradiation. To validate the sterilization cvcle, VDmax25 was used, which includes bioburden testing, bioburden recovery, verification dose testing, product sterility testing, and method suitability testing.
#### Biocompatibility Testing
The troCarWash™ system is a limited (<24 hours) patient contacting device. The trocar, obturator, wash solution, and tubing set are the only components that contact the patient, while the reusable controller does not have any contact with the patient. The biocompatibility evaluation for the troCarWash™ system included cytotoxicity, sensitization, acute systemic toxicity, and material-mediated Pyrogenicity. All testing was conducted in accordance with the following standards and quidance documents:
- . ISO 10993-1: Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process
- . ISO 10993-5: Biological Evaluation of Medical Devices - Part 5: Tests for in Vitro Cytotoxicity
- . ISO 10993-10: Biological Evaluation of Medical Devices - Part 10: Tests for Irritation
- and Skin Sensitization
- . ISO 10993-11: Biological Evaluation of Medical Devices – Part 11: Tests for Systemic Toxicitv
- . ISO 10993-12: Biological Evaluation of Medical Devices – Part 12: Sample Preparation and Reference Materials
- . FDA Guidance Document: "Use of international Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing with a risk management process"
Electrical Safety and Electromagnetic Compatibility (EMC and EST)
Electrical safety and EMC testing were conducted on the troCarWash™ system, consisting of the reusable controller and disposable set (trocar, obturator, and tubing set). The system complies with the IEC 60601-1, IEC 60601-1-6, and IEC 60601-2-18 for EST and IEC 60601-1-2 for EMC.
#### Software Verification and Validation Testing
Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device
{6}------------------------------------------------
was considered as a "minor" level of concern, since a failure or latent flaw in the system would not result in injury or damage to health.
#### Bench Performance Testing
Performance testing was conducting in a benchtop model to evaluate system performance. 29 disposable sets (trocar, obturator, and tubing set) preconditioned with sterilization and accelerated aging and 8 disposable sets preconditioned with sterilization and simulated transportation and distribution were tested for wash efficacy. The study demonstrated that with various scope angles and brands, the troCarWash™ system was able to achieve acceptable cleans.
No clinical data was necessary to support a claim of substantial equivalence.
The troCarWash System performance testing demonstrated a sufficient wash efficacy across scope angles tested. The risk controls and bench performance testing ensure the troCarWash system raises no new issues of safety or effectiveness. The descriptive information contained within the predicate 510(k) is sufficiently precise to ensure the substantial equivalence of the troCarWash™ System, and to determine that there are no new issues of safety or effectiveness. There is no need for comparative performance data as the effect of the device in terms of cleaning and defogging the lens is demonstrated through the bench performance testing.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.