Peace N Quiet (0.7.0)

K233435 · Pnq Health · KLW · Feb 27, 2024 · Ear, Nose, Throat

Device Facts

Record IDK233435
Device NamePeace N Quiet (0.7.0)
ApplicantPnq Health
Product CodeKLW · Ear, Nose, Throat
Decision DateFeb 27, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.3400
Device ClassClass 2
AttributesSoftware as a Medical Device, Therapeutic

Indications for Use

The Peace N Quiet device is indicated for use in the temporary relief of tinnitus symptoms. The device plays customized sounds to relieve patients suffering from tinnitus and can be used in a tinnitus management program. The target population is for patients who are 18 years or older. This device should only be used with the advice of a physician, audiologist or other healthcare professional.

Device Story

Peace N Quiet is a mobile application (software as a medical device) for Apple iPhone; facilitates tinnitus management via customized sound therapy. Healthcare professionals (HCPs) use the app to assist patients in self-administering treatments. Input: patient-specific tinnitus characteristics; Output: customized audio stimuli (periodic signals, pseudorandom noise, white/pink noise) played through user-provided earbuds. Operation: HCPs and patients tailor sounds to match perceived tinnitus; device provides frequency/amplitude modulation and individual volume control per ear. Benefit: temporary relief of tinnitus symptoms. Used in home or clinical settings.

Clinical Evidence

No clinical data. Bench testing only; software verification and validation performed per IEC 62304:2015.

Technological Characteristics

Software-only mobile application; compatible with commercial earbuds. Audio output: 85 dBA max; frequency range 500 Hz–16,000 Hz. Stimuli: periodic signals (sine, triangle, sawtooth, square), pseudorandom, white/pink noise. Features: amplitude/frequency modulation, individual ear volume control. Software lifecycle per IEC 62304:2015.

Indications for Use

Indicated for temporary relief of tinnitus symptoms in patients 18 years or older. Used as part of a tinnitus management program under the guidance of a physician, audiologist, or other hearing healthcare professional.

Regulatory Classification

Identification

A tinnitus masker is an electronic device intended to generate noise of sufficient intensity and bandwidth to mask ringing in the ears or internal head noises. Because the device is able to mask internal noises, it is also used as an aid in hearing external noises and speech.

Special Controls

The special control for this device is patient labeling regarding:(1) Hearing health care professional diagnosis, fitting of the device, and followup care, (2) Risks, (3) Benefits, (4) Warnings for safe use, and (5) Specifications.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font. February 27, 2024 PNQ Health % Alexia Haralambous Senior Principal Ram+ 2251 San Diego Avenue Suite B-257 San Diego, California 92110 Re: K233435 Trade/Device Name: Peace N Quiet (0.7.0) Regulation Number: 21 CFR 874.3400 Regulation Name: Tinnitus Masker Regulatory Class: Class II Product Code: KLW Dated: January 29, 2024 Received: January 29, 2024 Dear Alexia Haralambous: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Shuchen Peng -S Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ # Indications for Use Submission Number (if known) K233435 Device Name Peace N Quiet (0.7.0) #### Indications for Use (Describe) The Peace N Quiet device is indicated for use in the temporary relief of tinnitus symptoms. The device plays customized sounds to relieve patients suffering from tinnitus and can be used in a tinnitus management program. The target population is for patients who are 18 years or older. This device should only be used with the advice of a physician, audiologist or other healthcare professional. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY #### DATE PREPARED February 23, 2024 ## MANUFACTURER AND 510(k) OWNER PNQ Health, Inc. 110 Cheshire Lane, Ste 385 Minnetonka, MN 55305 Telephone: +1 (651) 280-7624 Official Contact: Kurtis Goos, Chief Executive Officer ## REPRESENTATIVE/CONSULTANT Alexia Haralambous, MS, RAC Hrishikesh Gadagkar, Ph.D. ROM+ 2251 San Diego Ave, Suite B-257 San Diego, CA 92110, USA Telephone: +1 (267) 778-5114 Email: aharalambous@rqmplus.com hgadagkar@rqmplus.com ### DEVICE INFORMATION Proprietary Name/Trade Name: Peace N Quiet Common Name: Tinnitus masker Regulation Number: 21 CFR 874.3400 Class: II Product Code: KLW Premarket Review: Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices (OHT1), Division of Dental and ENT Devices (DHT1B) Review Panel: Ear, Nose, and Throat ## PREDICATE DEVICE IDENTIFICATION The Peace N Quiet device is substantially equivalent to the following predicate: | 510(k) Number | Predicate Device Name / Manufacturer | Primary<br>Predicate | Reference<br>Device | |---------------|------------------------------------------------------------------------------------|----------------------|---------------------| | K161562 | Sound Options Tinnitus Treatment / Sound Options<br>Tinnitus Treatments, Inc. | ✓ | | | K163094 | TinniLogic Mobile Tinnitus Management Device /<br>Jiangsu BetterLife Medical, Ltd. | | ✓ | {4}------------------------------------------------ #### DEVICE DESCRIPTION The Peace N Quiet tinnitus device is software as a medical device implemented as a mobile application for patients suffering from tinnitus. The Peace N Quiet tinnitus mobile app can be downloaded from the Apple Store onto a personal Apple iPhone device facilitates a qualified healthcare professional (HCP). i.e., physician or audiologist, to provide professional counseling or education to the patient. The HCPs can direct and assist patients to self-administer customized treatments using the Peace N Quiet tinnitus device. ### INDICATIONS FOR USE The Peace N Quiet device is indicated for use in the temporary relief of tinnitus symptoms. The device plays customized sounds to relieve patients suffering from tinnitus and can be used in a tinnitus management program. The target population is for patients who are 18 years or older. This device should only be used with the advice of a physician, audiologist or other hearing healthcare professional. #### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS PNO Health. Inc. believes that the Peace N Ouiet device is substantially equivalent to the predicate device based on the information summarized here: The subject device has the same intended use as the primary predicate, the Sound Options Tinnitus Treatment (K161562), i.e., software intended for the temporary relief of tinnitus symptoms, with a target population of patients who are 18 years or older and suffering from tinnitus. The subject device shares several technological characteristics with the primary predicate (K161562). Both are software-only medical devices that provide audio sound outputs customized to the patient to mask tinnitus as part of a tinnitus management program, and both systems are compatible with pre-specified earbuds (non-medical device). The subject device differs in that its sounds are customized to the patient by a qualified HCP, whereas the primary predicate (K161562) has sounds that are primarily customized by the device manufacturer (Sound Options Tinnitus Treatments, Inc.) in addition to the HCP. Both allow for physical volume control and maintain the same output maximum. 85 dBA, with similar output frequency ranges that fall within normal human hearing frequency ranges. In addition, both utilize primarily amplitude-modulated stimuli. The primary predicate (K161562) relies on broadband frequency sound prescribed and customized to the patient by the HCP, whereas the subject Peace N Quiet device includes frequency modulation (in addition to amplitude modulation) that is customized to the patient within the device by a qualified HCP. The periodic signal types offered by the subject device provide additional options for HCP and patient customization of timitusmatched sounds. To allow the HCP and patient to further tailor the customized sounds to match the patient's perceived tinnitus, the primary predicate (K161562) uses broadband noise adjustments, while the subject device offers the user the option of additive pseudorandom noise adjustments such as white or pink noise. Both sets of options offer the user ways to tailor the customized sounds to come as close as possible to the patient's perceived tinnitus, with the subject device offering a higher number of adjustable parameters as compared to the primary predicate. The TinniLogic Mobile Tinnitus Management Device (K163094) device was used as a reference device to support the technology of temporary relief of tinnitus symptoms through customized {5}------------------------------------------------ masking sounds. The reference device (K163094) utilizes sinusoidal amplitude-modulated sounds with a combination of narrow-band noise, broad-band noise, and other types of noise such as white noise and natural sounds, to allow the user ways to tailor the customized sounds. These options differ slightly from the additive noise options of the subject device, but both the reference device (K163094) and primary predicate noise options offer the user a variety of options intended to allow them to match the patient's perceived tinnitus as closely as possible. These differences do not present new or modified risks, and therefore they do not raise different questions of safety and effectiveness. This information is summarized in the following table. | | Subject Device:<br>Peace N Quiet | Primary Predicate:<br>Sound Options Tinnitus<br>Treatment (K161562) | Reference Device:<br>TinniLogic Mobile<br>Tinnitus Management<br>Device (K163094) | |------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for<br>Use | The Peace N Quiet device<br>is indicated for use in the<br>temporary relief of<br>tinnitus symptoms. The<br>device plays customized<br>sounds to relieve patients<br>suffering from tinnitus<br>and can be used in a<br>tinnitus management<br>program. The target<br>population is for patients<br>who are 18 years or<br>older. This device should<br>only be used with the<br>advice of a physician,<br>audiologist or other<br>hearing healthcare<br>professional. | The device, Sound<br>Options Tinnitus<br>Treatment, version SO<br>2.0, is for the temporary<br>relief of tinnitus<br>symptoms. The device is a<br>software application that<br>embeds sounds and<br>spectral content into<br>music to relieve patients<br>suffering from tinnitus<br>and can be used as part<br>of a tinnitus management<br>program for adults 18<br>years and older. The<br>device is for prescription<br>use by a physician,<br>audiologist or other<br>healthcare professional. | The TinniLogic Mobile<br>Tinnitus Management<br>Device is indicated for<br>use in the temporary<br>relief of tinnitus<br>symptoms. The device is a<br>player to play customized<br>sounds and display<br>professional counseling<br>or education to relieve<br>patients suffering from<br>tinnitus and can be used<br>in a tinnitus management<br>program. The target<br>population is adults (18<br>years or older). This is a<br>medical device and<br>should only be used with<br>the advice of a physician,<br>audiologist or other<br>hearing healthcare<br>professional. | | Target Population | Adult (18 years and<br>older) tinnitus patients | Adult (18 years and<br>older) tinnitus patients | Adult (18 years and<br>older) tinnitus patients | | Use Environment | At-home and healthcare<br>facilities | At-home and healthcare<br>facilities | At-home and healthcare<br>facilities | | Patient Medium | Software only | Software only | Hand-held audio device<br>for use with earphones | | Professional<br>Counseling/<br>Education | Guidance is provided via<br>the HCP and prompts on<br>the software app | Guidance is provided via<br>the HCP | Guidance is provided via<br>the HCP as well as<br>prompts on the provided<br>tablet | | Device<br>Components | Peace N Quiet Software | ● Sound Options<br>software only device,<br>SO 2.0 | ● TinniLogic Mobile<br>Tinnitus Management<br>Device with software<br>pre-installed | | | Subject Device:<br>Peace N Quiet | Primary Predicate:<br>Sound Options Tinnitus<br>Treatment (K161562) | Reference Device:<br>TinniLogic Mobile<br>Tinnitus Management<br>Device (K163094) | | | | | Earphones Accessories (adaptor, power cable, USB cable) | | Sounds | Sounds customized to the<br>patient by a qualified<br>health care professional. | Sounds customized by<br>Sound Options Tinnitus<br>Treatments, Inc.<br>(manufacturer)<br>Customized music tracks<br>are available to patients<br>by download or on CD. | Sounds customized to the<br>patient by a qualified<br>healthcare professional. | | Stimuli Type | Periodic Signal Type (Sine, Triangle, Sawtooth, Square) Amplitude Frequency Phase Duty Cycle (for square type only) Pseudorandom Type (Random, Pseudorandom) Random Noise Type (White, Pink) Amplitude Frequency Limits (Low & High) Modulation Amplitude (A signal played at various amplitudes) Frequency (A signal played at various frequencies) Pulse (A signal that is pulled on and off) | Amplitude-modulated tinnitus matching Broadband frequency sound Broadband noise | Sinusoidal amplitude modulated (SAM) tinnitus pitch matched sounds Narrow-band noise centered at the tinnitus frequency Broad-band noises Pure tone White noise Natural sounds Relax sounds | | Volume Control | Individual volume control<br>per ear of the application<br>software | Patients are to listen to<br>the sound therapy at a<br>comfortable volume level | Individual volume control<br>per ear of the application<br>software | | | Subject Device:<br>Peace N Quiet | Primary Predicate:<br>Sound Options Tinnitus<br>Treatment (K161562) | Reference Device:<br>TinniLogic Mobile<br>Tinnitus Management<br>Device (K163094) | | | | by adjusting the volume<br>control on their personal<br>music-playing device. | Physical volume controls | | | | Safe levels were<br>determined through bench<br>testing for each of the<br>recommended<br>commercial device<br>options described in the<br>user manuals. | | | Maximum Output | 85 dB | 85 dB | 90 dB | | Output Frequency | 500 Hz~16,000Hz | Dependent on patient<br>headphones (commercial) | 50Hz~15,000Hz | | Performance<br>Testing | ● IEC 62304:2015 | ● IEC 62304:2006 | ● IEC 60601-1<br>● IEC 60601-1-2 | {6}------------------------------------------------ {7}------------------------------------------------ ## SUMMARY OF NON-CLINICAL TESTING Software verification and validation testing was performed per IEC 62304:2015 Medical device software – Software life cycle processes to demonstrate safety and performance based on current industry standards. # SUMMARY OF CLINICAL TESTING Clinical testing was not performed as part of this submission. ## CONCLUSION Based on the verification and validation testing performed in accordance with IEC 62304:2015, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. The Peace N Quiet device is demonstrated to be substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...