The Handheld X-ray System is indicated for use only by a trained and qualified dentist or dental technician for both adult and pediatric subjects as an extraoral diagnostic dental X-ray source to produce X-ray images using intraoral image receptors.
Device Story
Portable, battery-operated X-ray generator for dental imaging; used in dental offices by dentists or technicians. Device emits X-ray beam through patient anatomy to intraoral detectors (film, digital sensor, or phosphor plate). Operator uses control panel and dead-man triggers (local or hard-wired remote) to initiate exposure. Predefined exposure programs available; parameters adjustable by operator. Output is radiographic image for dental assessment, diagnosis, and treatment planning. Benefits include portability for flexible clinical use and diagnostic imaging of teeth and oral structures.
Clinical Evidence
No clinical data. Bench testing only. Performance confirmed via radiation safety characterization (scattered radiation analysis), electrical safety (UL/IEC standards), and image quality evaluation using a cleared digital receptor (K230732) to confirm diagnostic quality.
Technological Characteristics
Handheld X-ray generator; 60/65 kVp, 2.5 mA; DC waveform. Rechargeable Li-ion battery (21.6V, 2.50Ah). Dead-man trigger mechanism. Standards: ANSI/AAMI ES60601-1:2012, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-65, IEC 60601-1-2 Ed. 4.0. Includes integrated safety protection circuit board for battery. Operates with film, digital sensors, or phosphor plates.
Indications for Use
Indicated for use by trained dentists or dental technicians for adult and pediatric patients as an extraoral dental X-ray source to produce images using intraoral receptors.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 3, 2024
Alliage S/A Industrias Médico Odontológica % Daniel Kamm Principal Engineer Kamm & Assosciates 8870 Ravello Ct NAPLES, FL 34114
## Re: K233395
Trade/Device Name: Portable Dental X-Ray (AXR60 S); Portable Dental X-Ray (AXR65 S) Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral Source X-Ray System Regulatory Class: Class II Product Code: EHD Dated: October 2, 2023 Received: October 3, 2023
## Dear Daniel Kamm:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233395
#### Device Name
Portable Dental X-Ray (AXR60 S); Portable Dental X-Ray (AXR65 S).
#### Indications for Use (Describe)
The Handheld X-ray System is indicated for use only by a trained and qualified dentist or dental technician for both adult and pediatric subjects as an extraoral dental X-ray source to produce X-ray images using intraoral image receptors.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "alliage" in a bold, sans-serif font. The letters are a dark blue color, and the word is set against a white background. The letters are closely spaced together, and the overall effect is one of solidity and strength.
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 872.1800.
Date prepared: December 15, 2023
- 1. Company and Correspondent making the submission:
Name: Alliage S/A Indústrias Médica Odontológico. Address: Rod. Abrao Assed, Km 53+450m Recreio Anhanguera, Zipcode 14097-500 Telephone: +55 16 3512-1212. Contact: Daniel Camargo.
- 2. Trade /Proprietary Names:
| Portable Dental X-Ray (AXR60 S); | |
|----------------------------------|-----------------------------------|
| Portable Dental X-Ray (AXR65 S). | |
| Common Name: | Portable X-ray System |
| Regulation Name: | Extra-oral Source X-ray System |
| Regulation Number: | 21 CFR 872.1800 |
| Primary Product Code: | EHD |
| Classification Name: | Unit, X-ray, Extraoral with Timer |
| Regulatory Class: | II |
| 510k Review Panel: | Dental |
- 3. Legally Marketed Predicate Device Information: K173319
| Trade/Device Name: | KaVo NOMAD Pro 2 Handheld X-ray System |
|-----------------------|----------------------------------------|
| Common Name: | Portable X-ray System |
| Regulation Name: | Extra-oral Source X-ray System |
| Regulation Number: | 21 CFR 872.1800 |
| Primary Product Code: | EHD |
| Classification Name: | Unit, X-ray, Extraoral with Timer |
| Regulatory Class: | II |
| 510k Review Panel: | Dental |
- 4. Description: These are hand-held portable battery operated x-ray generators for dental purposes. The Portable Dental X-Ray is an X radiation controlled emissions generator system, i.e., once put into service, is intended to be moved from one location to another by a person, used together with appropriate capture devices to generate intraoral radiological images for dental assessment, diagnosis and treatment. This equipment has exposure programs that can be applied to a variety of patients and has predefined exposure parameters depending on the type of patient. The operator is free to change these parameters depending on the situation. The equipment's human-machine interface consists of a control panel located on the top of the equipment, a local trigger button and a
{4}------------------------------------------------
remote trigger (hard wired). The triggers are "dead-man" triggers, meaning they release and interrupt the exposure. The Portable Dental X-Ray was designed to be used in adult and children patients by trained dentists and dental technicians to produce X-ray images for diagnosis. The Portable Dental X-Ray is indicated for the production of intraoral medical images of teeth, mandible and oral structures; it assists in the diagnosis of diseases, planning of surgical treatment and monitoring. It is exclusively for dental use, and must be used and handled by qualified and trained health professionals according to the User Manual.
Principles of operation: The Portable Dental X-Ray is an autonomous X-ray emitting system, used to produce radiographic images. The X-ray beam passes through the patient's body, where a part of the X-rays is absorbed or spread by internal structures, and the rest of the X-rays are transmitted to a detector (e.g., a film, digital sensor or phosphorous plate) for recording or further processing by a computer. The generator can be used with any of the three detector technologies that require technique factors within the range of capabilities of these models.
Image receptor compatibility: These devices will work with film, image receptor plates, and digital image receptors within the technique capability of the generator, which is 60 or 65 kVp and 2.5 mA with up to a 1 second exposure time. For example, a typical digital image receptor can work effectively at 60 kVp, 2.5 mA, 0.2 seconds, well within the capability of these generators.
# 5. Indications for use:
The Handheld X-ray System is indicated for use only by a trained and qualified dentist or dental technician for both adult and pediatric subjects as an extraoral diagnostic dental X-ray source to produce X-ray images using intraoral image receptors.
| PRODUCT | K173319 KaVo NOMAD Pro 2<br>Handheld X-ray System | Portable Dental X-Ray (AXR60 S);<br>Portable Dental X-Ray (AXR65 S). |
|------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for use | The KaVo NOMAD Pro 2 Handheld X-ray<br>System is indicated for use only by a<br>trained and qualified dentist or dental<br>technician for both adult and pediatric<br>subjects as an extraoral diagnostic dental<br>X-ray source to produce X-ray images<br>using intraoral image receptors. | The Handheld X-ray System is indicated for<br>use only by a trained and qualified dentist or<br>dental technician for both adult and pediatric<br>subjects as an extraoral diagnostic dental X-<br>ray source to produce X-ray images using<br>intraoral image receptors. SAME |
| Models | Only one, 60 kV (DC) 2.5 mA | Two, one with 60 kV 2.5 mA and one with 65<br>kV 2.5 mA ALMOST IDENTICAL |
| Timer Range | 0.02-1.00 sec. (in 0.01 second steps.) | 0.01 to 1 second (With 0.01 second steps)<br>Sligthly greater range of steps. |
| Duty Cycle | 1:60 | 1:60 SAME |
| Waveform | DC | DC SAME |
| mA | 2.5 mA fixed | 2.5 mA fixed SAME |
| Where Used | Dental offices | SAME |
| Operating<br>Temp. Range | -5°C to +40°C | +10°C to +35°C (Dental office environment) |
| PRODUCT | K173319 KaVo NOMAD Pro 2<br>Handheld X-ray System | Portable Dental X-Ray (AXR60 S);<br>Portable Dental X-Ray (AXR65 S). |
| Supply<br>Voltage<br>(For battery<br>charging) | AC Line 110/120V or 220/240V | AC Line: 100-240 V~, 50-60 Hz<br>SAME (Charger is UL listed) |
| Battery | Rechargeable Lithium-ion,<br>22.2V; 1.7 A-hr | Rechargeable Lithium-ion,<br>21,6V - 2.50Ah |
| Electrical<br>Safety<br>Standards | AAMI ES60601-1:2005/(R)2012 And<br>A1:2012 | ANSI/AAMI ES60601-1: A1:2012,<br>C1:2009/(R)2012 and A2:2010/(R)2012<br>SAME |
| EMC | IEC60601-1-2 Ed. 4 | IEC60601-1-2 Ed. 4 SAME |
| X-Ray<br>Performance | 21 CFR 1020.30, 1020.31; IEC60601-<br>1-3; IEC60601-2-65 | 21 CFR 1020.30, 1020.31; IEC 60601-1-<br>3:2008 (Second Edition) + A1:2013, IEC<br>60601-1-6:2010 (Third Edition) + A1:2013,<br>IEC 60601-2-65:2012 (First Edition) SAME |
| Photos | Image: KaVo NOMAD Pro 2 | Image: AXR60 S, AXR65 S |
- Comparison with predicate device: 6. A comparison table follows:
{5}------------------------------------------------
- 7. Non-clinical Testing Performed: Safety, EMC, and Performance Data:
Electrical Safety: The device has a certificate of compliance issued by UL, Report Reference E503536-D1002-0/A0/C0-UL, Standard(s) for Safety: ANSI/AAMI ES60601-1: A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012, CSA CAN/CSA-C22.2 NO. 60601-1:14. Additional Standards: IEC 60601-1-3:2008 (Second Edition) + A1:2013, IEC 60601-1-6:2010 (Third Edition) + A1:2013, IEC 60601-2-65:2012 (First Edition). The AC Power supply used for charging the battery is a medical grade UL listed, file UL62368-1. The battery pack consists of six cells connected in series connected to an integrated safety protection circuit board.
Electromagnetic Compatibility (EMC): IEC 60601-1-2 Ed. 4.0 (2014) – Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard:
{6}------------------------------------------------
Electromagnetic disturbance – Requirements and tests. The device was evaluated according to IEC TR 60601-4-2, Medical electrical equipment – Part 4-2: Guidance and interpretation – Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems.
FDA Guidance Document Utilized: Radiation Safety Considerations for X-ray Equipment Designed for Hand-Held Use. The scattered radiation from the device (in operation) was characterized in the entire volume surrounding the device and was found to be zero nGy in the volume space that would be occupied by the operator hand holding the device.
Image Evaluation: Using a digital image receptor (cleared in K230732) images were acquired and were found to be of diagnostic quality for dental applications.
Software (Firmware) Validation: Documentation was provided in compliance with FDA Guidance Document: Content of Premarket Submissions for Device Software Functions (for a Basic Documentation Level.) Risk analysis documentation was provided.
Bench Testing was performed to confirm compliance with the FDA Radiation Safety requirements of the Code of Federal Regulations including: accuracy and reproducibility specifications. (kV, ma, time) and aluminum equivalence.
- 8. Clinical Testing Performed: Clinical testing is not required for a finding of substantial equivalence.
- 9. Conclusions: Based on a comparison of intended use, indications, technological characteristics, principle of operation, features and performance data, the Portable Dental X-Ray (AXR60 S); and Portable Dental X-Ray (AXR65 S) devices are deemed to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.