K233391 · VPIX Medical, Inc. · GWG · Aug 21, 2024 · Neurology
Device Facts
Record ID
K233391
Device Name
cCeLL - In vivo
Applicant
VPIX Medical, Inc.
Product Code
GWG · Neurology
Decision Date
Aug 21, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1480
Device Class
Class 2
Indications for Use
The cCeLL - In vivo is an optic scanner probe placed in direct contact with tissue to create images of the internal microstructure of tissues and is indicated for use with indocyanine green (ICG) for fluorescence imaging as an aid in the visualization of vessels (micro- and macro-vasculature) blood flow in the cerebrovasculature before, during or after cranial diagnostic and therapeutic procedures, such as tumor biopsy and resection.
Device Story
cCeLL - In vivo is an optic scanner probe for real-time NIR fluorescence imaging during adult neurosurgery. System components: 6 mm Pixection ICG/NIR Endoscope (0°), NIR light source, photo-multiplier tube, and sterile probe sheath. Operation: probe contacts tissue/bone; irradiates tissue with laser; measures fluorescence intensity from ICG-injected tissue; generates 2D images of internal microstructure (depth of several hundred µm). Output: real-time HD images displayed on medical-grade monitor (DVI-D/RGB input). Used by neurosurgeons to visualize vessels and blood flow; aids in identifying tumor sites and surrounding tissue during biopsy/resection. Benefits: provides intraoperative visualization of micro- and macro-vasculature to guide surgical decision-making.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including image sensitivity analysis in a small animal model, detection linearity, geometric distortion, dynamic range, illumination/detection uniformity, SNR, sensitivity, and video latency. Biocompatibility, sterility, shelf-life, and electrical safety/EMC testing were performed per recognized standards.
Technological Characteristics
System includes 6 mm Pixection ICG/NIR Endoscope (0°), NIR light source, photo-multiplier tube, and sterile probe sheath. Connectivity: DVI-D or RGB input for HD monitors. Sterilization: Ethylene Oxide (ISO 11135). Biocompatibility: ISO 10993-1. Electrical safety: IEC 60601-1, IEC 60601-1-2. Software: Enhanced level of concern.
Indications for Use
Indicated for adult patients undergoing cranial diagnostic and therapeutic procedures (e.g., tumor biopsy, resection) to visualize cerebrovasculature (micro- and macro-vasculature) and blood flow using indocyanine green (ICG) fluorescence imaging.
Regulatory Classification
Identification
A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.
{0}------------------------------------------------
August 21, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
VPIX Medical, Inc. % Dawn N. Norman, M.S. Partner MRC Global, LLC 9085 E. Mineral Circle, Suite 110 Centennial, CO 80112
Re: K233391
Trade/Device Name: cCeLL - In vivo Regulation Number: 21 CFR 882.1480 Regulation Name: Neurological Endoscope Regulatory Class: Class II Product Code: GWG, OWN Dated: July 18, 2024 Received: July 22, 2024
Dear Dawn Norman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Adam D Pierce -S
Digitally signed by Adam D. Pierce -S Date: 2024.08.21 15:25:30 -04'00'
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices
{2}------------------------------------------------
OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233391
Device Name cCeLL - In vivo
### Indications for Use (Describe)
The cCeLL - In vivo is an optic scanner probe placed in direct contact with tissue to create images of the internal microstructure of tissues and is indicated for use with indocyanine green (ICG) for fluorescence imaging as an aid in the visualization of vessels (micro- and macro-vasculature) blood flow in the cerebrovasculature before, during or after cranial diagnostic and therapeutic procedures, such as tumor biopsy and resection.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> <b> </b> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K233391 510(k) Summary
### i. Submitter Information
ii.
iii.
| Submitter: | VPIX Medical, Inc. |
|-------------------------|-------------------------------------|
| | 774, Gyeryong-ro, Jung-gu |
| | Daejeon, 34873 |
| | Republic of Korea |
| Contact Person: | Kyungmin Hwang, President |
| Submitter: | Jin Hee Byon, RA Team Leader |
| Email: | julie.byon@vpixmedical.com |
| Phone: | +82 (0)42 535 0668 |
| Official Correspondent: | Dawn Norman, MS |
| Email: | dawn.norman@askmrcglobal.com |
| Phone: | 01-618-604-3064 |
| Date Prepared: | August 17, 2024 |
| Device Name | |
| Proprietary Name: | cCeLL - In vivo |
| Manufacturer: | VPIX Medical, Inc. |
| Common Name: | Neurological Endoscope |
| Classification Name: | Endoscope, Neurological |
| Regulation Number: | 21 CFR 882.1480; 21 CFR 876.1500 |
| Device Class: | Class II |
| Product Code: | GWG (Primary) |
| | OWN (Secondary) |
| | |
| Predicate Devices | |
| 510(k) Number: | K180146 |
| Proprietary Name: | KARL STORZ ICG Imaging System |
| Manufacturer | KARL STORZ Endoskopie-America, Inc. |
Manufacturer: Common Name: Classification Name: Regulation Number: Device Class: Product Code:
KARL STORZ Endoscopy-America, Inc. Neurological Endoscope Endoscope, Neurological 21 CFR 882.1480; 21 CFR 876.1500 Class II GWG, OWN
{5}------------------------------------------------
#### iv. Device Description
The cCeLL - In vivo is used to provide real-time endoscopic images of near-infrared (NIR) indocyanine green (ICG) dye fluorescence during minimally invasive, neurosurgery in adults.
The overall system includes a 6 mm Pixection ICG/NIR Endoscope (0°) for use in neurosurgery, a light source for emission of NIR illumination, a photo-multiplier tube capable of capturing NIR imaging, and a sterile probe sheath intended for maintaining a sterile barrier between the subject device and the patient. The cCeLL - In vivo can be used with any medical grade high definition (HD) monitor with a DVI-D or RGB input. The patient contacting components contact tissue or bone with a duration of less than 24 hours.
### V. Principle of Operation / Mechanism of Action
This product irradiates the tissue with a laser by contacting the probe to the suspected tumor site of a patient injected with indocyanine green. It is a medical fluorescence imaging device that measures the intensity of the fluorescence signal emitted from the tissue and images the cells of the tissue. It has a function to acquire an image of a part corresponding to an area of several hundred um of tissue at a depth of several to several hundred um by using fluorescence and shows a two-dimensional image through the provided software.
#### vi. Indications for Use
The cCeLL - In vivo is an optic scanner probe placed in direct contact with tissue to create images of the internal microstructure of tissues and is indicated for use with indocyanine green (ICG) for fluorescence imaging as an aid in the visualization of vessels (micro- and macro-vasculature) blood flow in the cerebrovasculature before, during or after cranial diagnostic and therapeutic procedures, such as tumor biopsy and resection.
#### vii. Summary of Substantial Equivalence
The similarities between subject device and predicate device are as follows: Both devices provide images of NIR Indocyanine green (ICG) dye fluorescence during minimally invasive, neurosurgery in adults.
Both can be used with any medical grade HD monitor with a DVI-D input.
The differences between subject device and predicate device are as follows:
The predicate also provides fluorescence during plastic, micro- and reconstructive surgical procedures in adults and pediatric populations. The predicate allows use of a medical grade HD monitor with 3G-SDI input. The predicate device is supplied with an ICG kit while the subject device is not.
{6}------------------------------------------------
# Substantial Equivalence Table:
| | Subject Device | Predicate Device | Comparison |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K233391 | K180146 | N/A |
| Device<br>Identification | Trade Name: cCeLL - In vivo<br>Classification Name: Neurological Endoscope | Trade Name: KARL STORZ ICG Imaging System<br>Classification Name: Neurological Endoscope | N/A |
| Product Code | GWG, OWN | GWG, OWN | Identical |
| Regulation | 21 CFR 876.1480<br>(Neurological Endoscope) | 21 CFR 876.1480<br>(Neurological Endoscope) | Identical |
| Device<br>Description | The cCeLL - In vivo is used to provide real-time<br>high-definition (HD) endoscopic images of near-<br>infrared (NIR) indocyanine green (ICG) dye<br>fluorescence during minimally invasive,<br>neurosurgery in adults.<br><br>The overall system includes a 6 mm Pixection<br>ICG/NIR Endoscope (0°) for use in<br>neurosurgery, a light source for emission of NIR<br>illumination, a photo-multiplier tube capable of<br>capturing NIR imaging, and a sterile probe<br>sheath. | The KARL STORZ ICG Imaging System is used to<br>provide real-time high-definition (HD)<br>endoscopic or telescopic images of visible (VIS)<br>and near-infrared (NIR) indocyanine green (ICG)<br>dye fluorescence during minimally invasive,<br>neuro- and endonasal skull base surgery as well<br>as plastic, micro- and reconstructive surgical<br>procedures in general and pediatric populations.<br><br>The overall system includes a 4mm HOPKINS<br>ICG/NIR Endoscope (0°, 30° or 45°) for use in<br>neuro- and endonasal skull base surgery, a 5mm<br>& 10mm HOPKINS ICG/NIR Endoscope (0° or<br>30°) for use in minimally invasive procedures<br>and a VITOM II ICG/NIR Telescope for use in<br>plastic, micro- and reconstructive surgical<br>procedures for VIS and NIR illumination and<br>imaging, a light source with foot switch for<br>emission of VIS and NIR illumination, a color<br>video camera head capable of capturing both<br>VIS and NIR imaging, and a KARL STORZ ICG Kit.<br>Additional accessories used with the KARL<br>STORZ ICG Imaging System include two<br>standards fiber-optic light cables for<br>transmission of VIS and NIR light and the<br>Image1 S Camera Control Unit (CCU). | Similarities<br>Both devices provide HD<br>images of NIR<br>Indocyanine green (ICG)<br>dye fluorescence during<br>minimally invasive,<br>neuro- and endonasal<br>skull base surgery in<br>adults.<br><br>Both can be used with<br>any medical grade HD<br>monitor with a DVI-D<br>input.<br><br>Differences<br>The predicate also<br>provides fluorescence<br>during plastic, micro-<br>and reconstructive<br>surgical procedures in<br>pediatric populations.<br><br>The predicate allows use<br>of a medical grade HD<br>monitor with 3G-SDI<br>input. |
| | Subject Device | Predicate Device | Comparison |
| | The cCeLL - In vivo can be used with any<br>medical grade HD monitor with a DVI-D or RGB<br>input. | The KARL STORZ ICG Imaging System can be<br>used with any medical grade HD monitor with a<br>DVI-D or 3G-SDI input. | The predicate device is<br>supplied with an ICG kit<br>while the subject device<br>is not. |
| Intended Use | The cCeLL - In vivo is intended to provide real-<br>time near-infrared fluorescence imaging. | The KARL STORZ ICG Imaging System is intended<br>to provide real-time visible and near-infrared<br>fluorescence imaging. | Identical except the<br>predicate provides<br>visible fluorescence<br>imaging. |
| Indications for<br>Use | The cCeLL - In vivo is an optic scanner probe<br>placed in direct contact with tissue to create<br>images of the internal microstructure of tissues<br>and is indicated for use with indocyanine green<br>(ICG) for fluorescence imaging as an aid in the<br>visualization of vessels (micro- and macro-<br>vasculature) blood flow in the<br>cerebrovasculature before, during or after<br>cranial diagnostic and therapeutic procedures,<br>such as tumor biopsy and resection. | The KARL STORZ Endoscopic ICG System enables<br>surgeons to perform minimally invasive surgery<br>using standard endoscopic visible light as well as<br>visual assessment of vessels, blood flow and<br>related tissue perfusion, or at least one of the<br>major extra-hepatic bile ducts (cystic duct,<br>common bile duct and common hepatic duct),<br>using near infrared imaging. Fluorescence<br>imaging of biliary ducts with the KARL STORZ<br>Endoscopic ICG System is intended for use with<br>standard of care white light and, when<br>indicated, intraoperative cholangiography. The<br>device is not intended for standalone use for<br>biliary duct visualization. | Similarities<br>Both use HD images of<br>NIR Indocyanine green<br>(ICG) dye fluorescence<br>during minimally<br>invasive, neuro- and<br>endonasal skull base<br>surgery in adults.<br>Both can be used with<br>any medical grade HD<br>monitor with a DVI-D<br>input. |
| | | Additionally, the KARL STORZ Endoscopic ICG<br>System enables surgeon to perform minimally<br>invasive cranial neurosurgery in adults and<br>pediatrics and endonasal skull base surgery in<br>adults and pediatrics > 6 years of age using<br>standard endoscopic visible light as well as<br>visual assessment of vessels, blood flow and<br>related tissue perfusion using near infrared<br>imaging.<br>The KARL STORZ VITOM II ICG System is<br>intended for capturing and viewing fluorescent<br>images for the visual assessment of blood flow, | Differences<br>The predicate is<br>indicated for pediatrics<br>cranial neurosurgery and<br>endonasal skull base<br>surgery in adults and<br>pediatrics > 6 years of<br>age.<br>The predicate has<br>multiple light and<br>visualization options.<br>The predicate also |
| | | as an adjunctive method for the evaluation of | The predicate also |
| | Subject Device | Predicate Device | Comparison |
| | | tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures. The VITOM II ICG System is intended to provide a magnified view of the surgical field in standard white light. | provides fluorescence during plastic, micro- and reconstructive surgical procedures in general and pediatric populations.<br>The predicate allows use of a medical grade HD monitor with 3G-SDI input. |
| | The cCeLL - In vivo includes the following components and accessories: | The KARL STORZ ICG Imaging System includes the following components and accessories: | Similarities<br>Both devices use a camera and optical fiber light source to process images. |
| | Main Unit<br>(for NIR illumination and capturing) Pixection (for endoscope) Sterile Probe Sheath | 4mm, 5mm & 10mm HOPKINS ICG/NIR Endoscopes VITOM II ICG Telescope Camera Head (H3Z-FI) Fiber optic Light Cables Light Source Image1 S CCU | |
| Technological<br>Characteristics | The Pixection is intended to be connected to the Main Unit, which connects to the PC for image processing, as well as to the light source via optical fiber as the source of illumination to allow visualization of internal anatomy. Visualization and navigation is performed using NIR imaging for visual assessment and/or confirmation of vessels, blood flow or tissue perfusion is desired. | The endoscopes/telescope are intended to be connected to the optical coupler of the camera head, which connects to the CCU for image processing, as well as to the light source via compatible light cable as the source of illumination to allow visualization of internal anatomy. Visualization and navigation is performed initially using VIS imaging. NIR imaging is selected when visual assessment and/or confirmation of vessels, blood flow or tissue perfusion is desired. | Differences<br>The predicate uses VIS while the subject device uses NIR for Visualization and navigation. |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
#### viiii. Performance Testing
Verification and validation testing demonstrated that the subject device conforms to the recognized safety standards, design input specifications and intended use consistent with the predicate device.
| Test | Test Method Summary | Results |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Image Sensitivity<br>Analysis | Validate the device's ability to visualize cerebral microstructures and<br>vascular systems, including tumor tissue, surrounding normal tissue, blood<br>vessels using clinically relevant ICG concentrations in a small animal model. | The study confirmed that the device effectively<br>visualized cerebral microstructures and vascular<br>systems across a range of clinically relevant ICG<br>concentrations. PASS |
| Image<br>Comparison<br>Analysis | Comparison of the fluorescence signal between the subject and predicate<br>device, following ICG injection in a small animal model, was conducted to<br>determine if the subject device performs equivalently to the predicate<br>device in visualizing vessels of various sizes and changes in blood flow. | The study confirmed that the subject device can<br>visualize vessels of various sizes and changes in<br>blood flow with image quality comparable to the<br>predicate device, as assessed across users with a<br>range of experience. PASS |
| Detection<br>Linearity | To verify the device's accuracy and reliability in capturing fluorescence<br>intensity, detection linearity was assessed by measuring optical power and<br>brightness at various increments for both the predicate and subject<br>devices. | The detection linearity test demonstrated<br>equivalent performance to the predicate device.<br>PASS |
| Geometric<br>Distortion | To verify the device's accuracy and reliability in capturing fluorescence<br>intensity with respect to geometric distortion, the power was measured<br>using an optical power meter and radial distortion was calculated from the<br>acquired image. Performance testing conducted with the predicate and<br>subject devices. | The geometric distortion test demonstrated<br>equivalent performance to the predicate device.<br>PASS |
| Dynamic Range | To verify the device's gradation performance in capturing fluorescence<br>intensity across its dynamic range, optical power and brightness were set,<br>fluorescent target removed and the dynamic range was calculated using<br>the acquired image. Performance testing conducted with the predicate and<br>subject devices. | The dynamic range test demonstrated equivalent<br>performance to the predicate device. PASS |
| Illumination &<br>Detection<br>Uniformity | To verify the device's accuracy and reliability in capturing fluorescence<br>intensity, illumination and detection uniformity, the average intensity of<br>each fluorescent dot in the region of interest [diagnostic area] and the | Testing measuring illumination and detection<br>uniformity demonstrated equivalent performance<br>to the predicate device. PASS |
| Test | Test Method Summary | Results |
| | illumination uniformity were calculated using the acquired image.<br>Performance testing conducted with the predicate and subject devices. | |
| SNR &<br>Sensitivity | To verify the device's sensing ability in capturing fluorescence intensity,<br>images were acquired and the signal to noise ratio (SNR) and sensitivity<br>were calculated. Performance testing conducted with the predicate and<br>subject devices. | SNR & sensitivity test demonstrated equivalent<br>performance to the predicate device. PASS |
| Video Latency | To verify the device's dynamic vision capability in capturing fluorescence<br>intensity, video latency was assessed by recoding the initialization and<br>stoppage of motion on the screen and calculating the latency. Performance<br>testing conducted with the predicate and subject devices. | The test results demonstrated equivalent<br>performance to the predicate device.<br>PASS |
| Sterile Probe<br>Sheath Tear<br>Resistance | To verify the robustness of SPS (Sterile Probe Sheath) tear resistance of the<br>SPS was tested for breaking strength at different join interfaces on aged<br>samples. | The test resutls demonstrated that the SPS can<br>withstand forces greater than those expected<br>during clinical use. PASS |
| Electrical Safety<br>/ EMC | IEC 60601-1:2005 + A1:2012 + A2:2021 Medical electrical equipment – Part<br>1: General requirements for basic safety and essential performance<br>IEC 60601-1-2 : 2014 + A1:2020, Medical Electrical Equipment - Part 1-2:<br>General requirements for basic safety and essential performance -<br>Electromagnetic Compatibility | PASS |
| Software /<br>Cybersecurity<br>(Enhanced Level) | FDA's Guidance for Industry and FDA Staff, "Content of Premarket<br>Submissions for Device Software Functions" issued June 14, 2023.<br>FDA guidance's Cybersecurity in Medical Devices: Quality System<br>Considerations and Content of Premarket Submissions, Postmarket<br>Management of Cybersecurity in Medical Devices. | PASS |
| Biocompatibility | Per FDA's Guidance "Use of International Standard ISO 10993-1, "Biological<br>evaluation of medical devices - Part 1: Evaluation and testing within a risk<br>management process", the single use Sterile Probe Sheath (SPS) is<br>categorized as an external communicating device in contact with<br>tissue/bone/dentin for <24 hours. | Biocompatibility testing in accordance with FDA's<br>biocompatibility guidance demonstrated that the<br>patient-contacting components are<br>biocompatible. PASS |
| Test | Test Method Summary | Results |
| | The following biocompatibility endpoints were evaluated: | |
| | Cytotoxicity<br>Sensitization<br>Intracutaneous reactivity<br>Acute systemic toxicity<br>Material medicated pyrogenicity<br>Hemocompatibility (indirect)<br>Neurotoxicity | |
| Sterility / Shelf<br>Life | ISO 11135:2014 + Amd1:2018 Sterilization of health care product –<br>Ethylene Oxide – Requirements for development, validation and routine<br>control of a sterilization process for medical devices | Sterilization and shelf-life testing of the Sterile<br>Probe Sheath (SPS) demonstrated the device is<br>and can remain sterile and functional for the<br>documented shelf life. PASS |
| | ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier<br>Systems and Medical Devices | |
| | ISO 11607–1:2019 Packaging for terminally sterilized medical devices - Part<br>1: Requirements for materials, sterile barrier systems and packaging<br>systems were used to validate the shelf life. | |
{10}------------------------------------------------
{11}------------------------------------------------
#### ix. Conclusion
Based on the similarities of the indications for use, device design, principles of operation, technological characteristics and the results of the non-clinical performance testing, the substantially equivalent to the legally marketed predicate device and does not raise new concerns of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.