← Product Code [NKG](/productcode/NKG) · K233366

# SYMPHONY OCT System (K233366)

_Medos International SARL · NKG · Nov 9, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K233366

## Device Facts

- **Applicant:** Medos International SARL
- **Product Code:** [NKG](/productcode/NKG.md)
- **Decision Date:** Nov 9, 2023
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3075
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The SYMPHONY OCT System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical unction, the cervical spine (C1 to C7) and the upper thoracic spine (T1-T3): - · Traumatic spinal fractures and/or traumatic dislocations; - Instability or deformity; - · Failed previous fusions (e.g. pseudarthrosis); - · Tumors involving the cervical/thoracic spine; · Degenerative disease, including intractable radiculopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and · Degenerative disease of the facets with instability. The SYMPHONY OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The SYMPHONY OCT System is compatible with occipital fusion components (plates, rods and clamps) from the SYNAPSE Occipital-Cervical-Thoracic (OCT) System and the MOUNTAINEER OCT Spinal System. Additionally, the SYMPHONY OCT System is compatible with SYNAPSE OCT System hooks and rods. The SONGER Wire/Cable System may be used with the SYMPHONY OCT System to allow for wire/cable attachment to the posterior cervical spine. The SYMPHONY OCT System may be connected to the EXPEDIUM® Spine System and VIPER® System using connectors and tapered rods. The SYMPHONY OCT System can also be linked to the USS Spinal System and MATRIX Spine System using connectors and tapered rods.

## Device Story

SYMPHONY OCT System is a posterior spinal fixation system; provides immobilization and stabilization of spinal segments. Components include bone anchors (screws), longitudinal rods, and interconnection mechanisms (set screws) with optional transverse connectors. Used by surgeons in clinical settings to build patient-specific constructs for spinal fusion or stabilization. System is compatible with various DePuy Synthes spinal systems (SYNAPSE, MOUNTAINEER, EXPEDIUM, VIPER, USS, MATRIX) and SONGER wire/cable systems. Device provides mechanical support to the spine; aids in fusion or restores structural integrity in tumor patients. This submission specifically addresses a labeling change to allow reprocessing of sterile-packaged implants if they are removed from packaging and rendered non-sterile prior to use.

## Clinical Evidence

No clinical data or additional bench testing was conducted for this submission. Substantial equivalence is based on the identical nature of the device to the predicate and the evaluation of the proposed labeling change.

## Technological Characteristics

Posterior spinal screw system consisting of bone anchors, rods, and interconnection mechanisms. Materials, dimensions, and mechanical design are identical to predicate devices. No changes to sterilization method or biocompatibility. System is purely mechanical; no software or electronic components.

## Regulatory Identification

Posterior cervical screw systems are comprised of multiple, interconnecting components, made from a variety of materials that allow an implant system to be built from the occiput to the upper thoracic spine to fit the patient's anatomical and physiological requirements, as determined by preoperative cross-sectional imaging. Such a spinal assembly consists of a combination of bone anchors via screws (i.e., occipital screws, cervical lateral mass screws, cervical pedicle screws, C2 pars screws, C2 translaminar screws, C2 transarticular screws), longitudinal members (e.g., plates, rods, including dual diameter rods, plate/rod combinations), transverse or cross connectors, interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors), and closure mechanisms (e.g., set screws, nuts). Posterior cervical screw systems are rigidly fixed devices that do not contain dynamic features, including but not limited to: non-uniform longitudinal elements or features that allow more motion or flexibility compared to rigid systems.Posterior cervical screw systems are intended to provide immobilization and stabilization of spinal segments in patients as an adjunct to fusion for acute and chronic instabilities of the cervical spine and/or craniocervical junction and/or cervicothoracic junction such as: (1) Traumatic spinal fractures and/or traumatic dislocations; (2) deformities; (3) instabilities; (4) failed previous fusions (e.g., pseudarthrosis); (5) tumors; (6) inflammatory disorders; (7) spinal degeneration, including neck and/or arm pain of discogenic origin as confirmed by imaging studies (radiographs, CT, MRI); (8) degeneration of the facets with instability; and (9) reconstruction following decompression to treat radiculopathy and/or myelopathy. These systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

## Special Controls

*Classification.* Class II (special controls). The special controls for posterior cervical screw systems are:(1) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(2) Nonclinical performance testing must demonstrate the mechanical function and durability of the implant.
(3) Device components must be demonstrated to be biocompatible.
(4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(5) Labeling must include the following:
(i) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(ii) Intended use and indications for use including levels of fixation;
(iii) Device specific warnings, precautions, and contraindications that include the following statements:
(A) “Precaution: Preoperative planning prior to implantation of posterior cervical screw systems should include review of cross-sectional imaging studies (
*e.g.,* CT and/or MRI) to evaluate the patient's cervical anatomy including the transverse foramen, neurologic structures, and the course of the vertebral arteries. If any findings would compromise the placement of these screws, other surgical methods should be considered. In addition, use of intraoperative imaging should be considered to guide and/or verify device placement, as necessary.”(B) “Precaution: Use of posterior cervical pedicle screw fixation at the C3 through C6 spinal levels requires careful consideration and planning beyond that required for lateral mass screws placed at these spinal levels, given the proximity of the vertebral arteries and neurologic structures in relation to the cervical pedicles at these levels.”
(iv) Identification of magnetic resonance (MR) compatibility status;
(v) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user, and;
(vi) Detailed instructions of each surgical step, including device removal.

## Predicate Devices

- SYMPHONY OCT System ([K192646](/device/K192646.md))
- SYMPHONY OCT System ([K203319](/device/K203319.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

November 9, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Medos International SARL % Sheree Geller Senior Regulatory Affairs Project Lead DePuy Synthes Spine 325 Paramount Drive Raynham, Massachusetts 02767

## Re: K233366

Trade/Device Name: SYMPHONY OCT System Regulation Number: 21 CFR 888.3075 Regulation Name: Posterior Cervical Screw System Regulatory Class: Class II Product Code: NKG, KWP Dated: September 29, 2023 Received: October 2, 2023

Dear Sheree Geller:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Eileen
Digitally signed
by Eileen Cadel -
S
Cadel -S Date: 2023.11.09
14:14:17 -05'00'

for Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K233366

Device Name SYMPHONY OCT System

### Indications for Use (Describe)

The SYMPHONY OCT System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical unction, the cervical spine (C1 to C7) and the upper thoracic spine (T1-T3):

- · Traumatic spinal fractures and/or traumatic dislocations;
- Instability or deformity;
- · Failed previous fusions (e.g. pseudarthrosis);
- · Tumors involving the cervical/thoracic spine;

· Degenerative disease, including intractable radiculopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and

· Degenerative disease of the facets with instability.

The SYMPHONY OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

The SYMPHONY OCT System is compatible with occipital fusion components (plates, rods and clamps) from the SYNAPSE Occipital-Cervical-Thoracic (OCT) System and the MOUNTAINEER OCT Spinal System. Additionally, the SYMPHONY OCT System is compatible with SYNAPSE OCT System hooks and rods.

The SONGER Wire/Cable System may be used with the SYMPHONY OCT System to allow for wire/cable attachment to the posterior cervical spine.

The SYMPHONY OCT System may be connected to the EXPEDIUM® Spine System and VIPER® System using connectors and tapered rods. The SYMPHONY OCT System can also be linked to the USS Spinal System and MATRIX Spine System using connectors and tapered rods.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| 510(k) Summary                     |                                                          |              | Prepared on: 2023-09-30 |
|------------------------------------|----------------------------------------------------------|--------------|-------------------------|
| Contact Details                    |                                                          |              | 21 CFR 807.92(a)(1)     |
| Applicant Name                     | Medos International SARL                                 |              |                         |
| Applicant Address                  | Chemin-Blanc 38 Le Locle 2400 Switzerland                |              |                         |
| Applicant Contact Telephone        | 508-828-3291                                             |              |                         |
| Applicant Contact                  | Ms. Sheree Geller                                        |              |                         |
| Applicant Contact Email            | sgeller1@its.jnj.com                                     |              |                         |
| Correspondent Name                 | DePuy Synthes Spine                                      |              |                         |
| Correspondent Address              | 325 Paramount Drive Raynham MA 02767 United States       |              |                         |
| Correspondent Contact Telephone    | 508-828-3291                                             |              |                         |
| Correspondent Contact              | Ms. Sheree Geller                                        |              |                         |
| Correspondent Contact Email        | sgeller1@its.jnj.com                                     |              |                         |
| Device Name                        |                                                          |              | 21 CFR 807.92(a)(2)     |
| Device Trade Name                  | SYMPHONY OCT System                                      |              |                         |
| Common Name                        | Posterior cervical screw system                          |              |                         |
| Classification Name                | Posterior Cervical Screw System                          |              |                         |
| Regulation Number                  | 888.3075                                                 |              |                         |
| Product Code                       | NKG                                                      |              |                         |
| Legally Marketed Predicate Devices |                                                          |              | 21 CFR 807.92(a)(3)     |
| Predicate #                        | Predicate Trade Name (Primary Predicate is listed first) | Product Code |                         |
| K192646                            | SYMPHONY OCT System                                      | NKG          |                         |
| K203319                            | SYMPHONY OCT System                                      | NKG          |                         |
| Device Description Summary         |                                                          |              | 21 CFR 807.92(a)(4)     |

The SYMPHONY OCT System is a posterior system intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion of the craniocervical junction, the cervical spine (C1 to C7) and the upper thoracic spine (T1-T3). The system is composed of multiple components to allow the surgeon to build an implant s anatomical and physiological requirements. The system consists of bone anchors (such as screws) for connection by longitudinal components (such as rods) via an interconnection mechanism (e.g., set screws) with optional transverse connectors) to link the longitudinal components for additional stability.

Intended Use/Indications for Use

{4}------------------------------------------------

21 CFR 807.92(a)(5)

21 CFR 807.92(a)(6)

The SYMPHONY OCT System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the cranical spine (C1 to C7) and the upper thoracic spine (T1-T3); · Traumatic spinal fractures and/or traumatic dislocations;

· Instability or deformity;

· Failed previous fusions (e.g. pseudarthrosis);

· Tumors involving the cervical/thoracic spine;

• Degenerativedisease, including intractable radior myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and

· Degenerative disease of the facets with instability.

The SYMPHONY OCT System is also intended to restore the integrity of the absence of fusion for a limited time period in patients with advanced stage tumors involving the envectancy is of insufficient duration to permit achievement of fusion.

The SYMPHONY OCT System is compatible with occipital fusion components (plates, rods and clamps) from the SYNAPSE Occipital-Cervical-Thoracic (OCT) System and the MOUNTAINEER OCT Spinal System. Additionally, the SYMPHONY OCT System is compatible with SYNAPSE OCT System hooks and rods.

The SONGER Wire/Cable System may be used with the SYMPHONY OCT System to allow for wire/cable attachment to the posterior cervical spine.

The SYMPHONY OCT System may be connected to the EXPEDIUM® Spine System using connectors and tapered rods. The SYMPHONY OCT System can also be linked to the USS Spinal System using connectors and tapered rods.

# Indications for Use Comparison

The indications for use are identical for the subject and predicate devices.

# Technological Comparison

## The sole change presented is a labeling change to allow for reprocessing of the sterile implants once they are removed from sterile packaging and rendered non-sterile prior to use on a patient. The SYMPHONY OCT System implants identical to the predicate. There is no change to the subject devices including the dimensions, materials, principle of operation, packaging, biocompatibility, or MR Compatibility from the predicate devices. When in sterile packaging, the subject devices are identical to the sterile versions of the predicate. Once removed from sterile packaging but prior to use, the subject devices are identical to the non-sterile versions of the predicate and follow the same cleaning and reprocessing instructions for Use as the predicate non-sterile devices.

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

No additional testing was conducted to support this submission. Evaluation of the subject device intended use, technological characteristics, and performance data demonstrates substantial equivalence with the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K233366](https://fda-staging.innolitics.com/device/K233366)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com