K233359 · Astura Medical · MQP · Mar 8, 2024 · Orthopedic
Device Facts
Record ID
K233359
Device Name
DOMINION Expandable Corpectomy System
Applicant
Astura Medical
Product Code
MQP · Orthopedic
Decision Date
Mar 8, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3060
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DOMINION Expandable Corpectomy System is indicated for vertebral body replacement in the cervical spine (C2-T1) and the thoracolumbar spine (T1-L5). The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft as an adjunct to fusion. When used in the cervical spine (C2-T1), DOMINION spacers are in skeletally mature patients to replace a diseased or damaged vertebral body caused by tumor fracture or osteomyelitis, or for reconstruction following corpectorny performed to achieve decompression of the spinal cord and neural tissues in the cervical degenerative disorders. These spacers are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion. The System is intended to be used with supplemental fixation that has been cleared by the FDA for use in the cervical spine. When used in the thoracolumbar spine (T1-L5), DOMINION spacers are intended for use to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e. fracture). These spacers are designed to provide anterior spinal column support even in the absence of fusion for a prolonged period. The system is intended to be used with supplemental fixation that has been cleared by the FDA for use in the thoracolumbar spine (i.e. posterior screw and rod systems, and anterior screw and rod systems). When used at more than two levels, supplemental fixation should include posterior fixation.
Device Story
Modular spinal implant system; includes expandable columns, modular endplates, and optional screws. Used for anterior column support following vertebral corpectomy. Surgeon inserts and expands column via specialized inserter tools to match patient anatomy; provides mechanical support to spinal column. Used in OR by surgeons. Complements autogenous/allogeneic bone graft to facilitate fusion; provides structural support in tumor cases where fusion may not be achieved. Output is physical stabilization of spinal column.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by non-clinical bench testing, including static/dynamic compression, static/dynamic torsion, expulsion (push-out) testing per ASTM F2077, and subsidence testing per ASTM F2267.
Technological Characteristics
Modular expandable corpectomy spacer. Materials: Titanium Alloy (Ti6-AL4-V ELI) per ASTM F136 and Nitinol (ASTM F2063). Mechanical expansion mechanism. Designed for anterior spinal column support. Non-implantable inserter tools included.
Indications for Use
Indicated for skeletally mature patients requiring vertebral body replacement in cervical (C2-T1) or thoracolumbar (T1-L5) spine. Used for diseased/damaged vertebrae due to tumor, fracture, or osteomyelitis, or reconstruction following corpectomy for decompression. Requires supplemental fixation.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
Predicate Devices
ALTA Anterior Cervical Corpectomy Spacer (K190426)
Reference Devices
DOLOMITE Anterior Cervical Stabilization System (K202065)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out next to it. The full name reads "U.S. Food & Drug Administration".
March 8. 2024
Astura Medical Parker Kelch Quality Manager 4949 West Royal Ln Irving, Texas 75063
Re: K233359
Trade/Device Name: DOMINION Expandable Corpectomy System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: MQP, PLR Dated: February 9, 2024 Received: February 9, 2024
Dear Parker Kelch:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Brent Showalter -S
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233359
#### Device Name DOMINION Expandable Corpectomy System
#### Indications for Use (Describe)
The DOMINION Expandable Corpectomy System is indicated for vertebral body replacement in the cervical spine (C2-Tl) and the thoracolumbar spine (T1-L5). The System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft as an adjunct to fusion.
When used in the cervical spine (C2-T1), DOMINION spacers are in skeletally mature patients to replace a diseased or damaged vertebral body caused by tumor fracture or osteomyelitis, or for reconstruction following corpectorny performed to achieve decompression of the spinal cord and neural tissues in the cervical degenerative disorders. These spacers are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion. The System is intended to be used with supplemental fixation that has been cleared by the FDA for use in the cervical spine. When used in the thoracolumbar spine (T1-L5), DOMINION spacers are intended for use to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e. fracture). These spacers are designed to provide anterior spinal column support even in the absence of fusion for a prolonged period.
The system is intended to be used with supplemental fixation that has been cleared by the FDA for use in the thoracolumbar spine (i.e. posterior screw and rod systems, and anterior screw and rod systems). When used at more than two levels, supplemental fixation should include posterior fixation.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------|-----------------------------------------------|
|-----------------------------------------------------------------------------------------|-----------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary: DOMINION Expandable Corpectomy System
| Date Prepared | September 29, 2023 |
|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted By | Astura Medical<br>4949 W Royal Ln<br>Irving, TX 75063 |
| Contact | Parker Kelch<br>4949 W Royal Ln<br>Irving, TX 75063<br>Phone: 469-501-5530 x503<br>Email: parker@asturamedical.com |
| Trade Name | DOMINION Expandable Corpectomy System |
| Common Name | Expandable Corpectomy Spacer |
| Classification Name | Spinal Vertebral Body Replacement Device |
| Class | II |
| Product Code | MQP, PLR |
| CFR Section | 21 CFR section 888.3060 |
| Device Panel | Orthopedic |
| Primary Predicate Device | ALTA Anterior Cervical Corpectomy Spacer (K190426) |
| Reference Predicate<br>Device(s) | DOLOMITE Anterior Cervical Stabilization System (K202065)<br>FORTIFY Corpectomy Spacer (K173982)<br>OLYMPIC Posterior Screw Fixation System (K153446) |
| Device Description | The Dominion Expandable Corpectomy system is a modular system<br>comprised of Expandable Columns, Modular Endplates, and optional<br>screws. The primary purpose of the Expandable Column is to provide<br>mechanical support to the anterior column of the spine after a vertebral<br>corpectomy is performed. The primary function of the Modular Endplates<br>is to provide attachment points for the Expandable Column to the<br>vertebral endplates. The screws are an optional device that can provide<br>enhanced fixation to the vertebral endplates when desired. The system<br>contains Inserters which allow the insertion and expansion of the<br>Expandable Columns. |
| Materials | Titanium Alloy (Ti6-AL4-V ELI) per ASTM F136<br>Nitinol #1 (ASTM F2063) |
| Technological<br>characteristics | Compared to the ALTA corpectomy spacer, the Dominion has the<br>capability to expand to the desired height to suit patient anatomy. Once<br>expanded its primary purpose of mechanical support to the anterior<br>column of the spine is equivalent to that of the static ALTA corpectomy<br>spacers. |
| Substantial Equivalence<br>Claimed to Predicate<br>Devices | The DOMINION Expandable Corpectomy System is substantially<br>equivalent to the predicate devices in terms of intended use, design,<br>materials used, mechanical safety and performances. |
| Indications for Use | The DOMINION Expandable Corpectomy System is indicated for vertebral<br>body replacement in the cervical spine (C2-T1) and the thoracolumbar<br>spine (T1-L5). The System is designed for use with autogenous and/or<br>allogeneic bone graft comprised of cancellous and/or corticocancellous<br>bone graft as an adjunct to fusion. |
| | When used in the cervical spine (C2-T1), DOMINION spacers are intended<br>for use in skeletally mature patients to replace a diseased or damaged<br>vertebral body caused by tumor fracture or osteomyelitis, or for<br>reconstruction following corpectomy performed to achieve<br>decompression of the spinal cord and neural tissues in the cervical<br>degenerative disorders. These spacers are intended to restore the<br>integrity of the spinal column even in the absence of fusion for a limited<br>time period in patients with advanced stage tumors involving the cervical<br>spine in whom life expectancy is of insufficient duration to permit<br>achievement of fusion, with bone graft used at the surgeon's discretion.<br>The System is intended to be used with supplemental fixation that has<br>been cleared by the FDA for use in the cervical spine. |
| | When used in the thoracolumbar spine (T1-L5), DOMINION spacers are<br>intended for use to replace a collapsed, damaged, or unstable vertebral<br>body due to tumor or trauma (i.e. fracture). These spacers are designed<br>to provide anterior spinal column support even in the absence of fusion<br>for a prolonged period. |
| | The system is intended to be used with supplemental fixation that has<br>been cleared by the FDA for use in the thoracolumbar spine (i.e. posterior<br>screw and rod systems, anterior plate systems, and anterior screw and<br>rod systems). When used at more than two levels, supplemental fixation<br>should include posterior fixation. |
| Non-clinical Test<br>Summary | The following analyses were conducted:<br>ASTM F2077 - Static compression ASTM F2077 - Dynamic compression ASTM F2077 - Static Torsion ASTM F2077 - Dynamic Torsion Expulsion (Push-Out) |
| Clinical Test Summary | ASTM F2267 - Subsidence No clinical studies were performed |
| Conclusions: Non-Clinical<br>and Clinical | Astura Medical considers the DOMINION Expandable Corpectomy Spacer<br>to be equivalent to the predicate devices listed above. This conclusion is<br>based upon the devices' similarities in principles of operation, technology,<br>materials and indications for use. |
In accordance with 21 CFR 807.92 of the Federal Code of Regulations
{4}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.