K233348 · Philips Medical System Nederlands B.V. · MOS · Oct 24, 2023 · Radiology
Device Facts
Record ID
K233348
Device Name
16 Breast Coil
Applicant
Philips Medical System Nederlands B.V.
Product Code
MOS · Radiology
Decision Date
Oct 24, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The 16 Breast Coil for Canon 1.5T MRI Scanners are designed to provide magnetic resonance images of breast anatomy when used in conjunction with a magnetic resonance scanner. These images are interpreted by a trained physician. When used with a disposable biopsy grid, the device permits access to breast anatomy for biopsy and localization procedures.
Device Story
16 Breast Coil is a receive-only phased-array coil for 1.5T Canon Orian/Fortian MRI scanners; captures magnetic resonance signals from hydrogen nuclei in breast tissue. Device features patient support with three configurations: 16-channel (diagnostic), 10-channel (bilateral interventional), and 2-channel (unilateral interventional). Includes tabletop compression system for breast immobilization and positioning of imaging elements close to tissue to optimize signal-to-noise ratio. Operated by MRI technicians/radiologists in clinical imaging settings. Output consists of MR images (axial, sagittal, coronal, oblique) interpreted by physicians to inform clinical diagnosis and biopsy guidance. Benefits include improved image quality and reduced motion artifacts via immobilization and optimized coil proximity.
Clinical Evidence
Clinical performance was assessed by a U.S. Board Certified radiologist who evaluated acquired images to confirm sufficient quality for diagnostic use. No comparative clinical trial data or statistical metrics (sensitivity/specificity) were required; substantial equivalence was primarily demonstrated through non-clinical performance testing per FDA guidance 'Magnetic Resonance (MR) Receive-only Coil - Performance Criteria for Safety and Performance Based Pathway'.
Technological Characteristics
Receive-only phased-array coil; 1.5T magnetic field strength; 2/10/16-channel configurations. Rigid housing with variable geometry for element positioning. Energy source: MR scanner. Connectivity: System-specific interface connector for Canon 1.5T scanners. Biocompatibility per ISO 10993-1. Reprocessing per ISO 17664-2. Electrical/mechanical safety per IEC 60601-1, IEC 60601-2-33. Image quality per NEMA MS 1, 3, 9, 14.
Indications for Use
Indicated for patients requiring magnetic resonance imaging of breast anatomy for diagnostic purposes or interventional procedures (biopsy/localization) using Canon 1.5T MRI scanners.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
October 24, 2023
Philips Medical System Nederlands B.V. Shruti Sancheti Regulatory Affairs Manager Veenpluis 6 Best, Noord-Brabant 5684 PC Netherlands
Re: K233348
Trade/Device Name: 16 Breast Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: MOS Dated: September 29, 2023 Received: September 29, 2023
## Dear Shruti Sancheti:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K233348
Device Name
16 BREAST COIL
#### Indications for Use (Describe)
The 16 Breast Coil for Canon 1.5T MRI Scanners are designed to provide magnetic resonance images of breast anatomy when used in conjunction with a magnetic resonance scanner. These images are interpreted by a trained physician. When used with a disposable biopsy grid, the device permits access to breast anatomy for biopsy and localization procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# The 510(k) Summary was prepared in accordance with 21 CFR §807.92.
| Preparation date: | September 29, 2023 |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner: | Philips Medical Systems Nederland B.V.<br>Veenpluis 6<br>5684 PC Best<br>The Netherlands<br>Establishment Registration # 3003768277 |
| Contact person: | Shruti Sancheti<br>Regulatory Affairs Manager<br>Philips Medical Systems Nederland B.V.<br>Phone: +91 9518546814 |
| Subject Device Trade Name: | 16 Breast Coil |
| Classification Name: | Coil, Magnetic Resonance, Specialty |
| Regulation Number: | 21 CFR 892.1000 |
| Review Panel: | Radiology |
| Device Class: | Class II |
| Product Code: | MOS |
| Predicate Device Trade Name: | dS Sentinelle Breast 16ch 1.5T Coil (K213735) |
| Classification Name: | Coil, Magnetic Resonance, Specialty |
| Regulation Number: | 21 CFR 892.1000 |
| Review Panel: | Radiology |
| Device Class: | Class II |
| Product Code: | MOS |
{5}------------------------------------------------
## Device description
The 16 Breast Coil is a receive-only coil to be used with 1.5T Orian and Fortian Canon MR scanners. The coil arrays are designed in a magnetic strength (1.5T) to correspond with the scanner strength.
The 16 Breast Coil is a phased array design consisting of patient support with three different coil combinations (2, 10 or 16 channels):
- 1. 16Ch for diagnostic imaging: performed in 16ch combination that consists of two 4ch Lateral coils and the 8ch Medial array coil.
- 2. 10ch for bilateral interventional procedures requiring lateral access: performed in 10ch configuration that consists of the 8ch Medial array coil and two 1ch Loop coils (right and left).
- 3. 2ch for unilateral interventional procedures allowing both lateral and medial access: performed in 2ch configuration that consists of two 1ch Loop coils (right and left) and medial pluq.
The coils receive magnetic resonance signals generated in hydrogen nuclei (protons) in the Breast while blocking the high-frequency magnetic field applied by the MRI scanner at specified timings.
Images are typically generated as axial, sagittal, coronal and oblique slices and include full coverage of the breast anatomy.
The 16 Breast Coil is tuned to receive RF frequency corresponding to the proton precession in a 1.5 tesla magnetic field, which is governed by the Larmor equation.
The Variable Coil Geometry design of the 16 Breast Coil allows each imaging element to be independently positioned and configured for each patients can then be positioned quickly and effectively as the imaging elements can be positioned as close to the breast as possible optimizing the signal-to-noise ratio for each individual patient. For clinical imaging, coil housings are placed next to the tissue to help minimize motion artifacts due to patient motion during scanning.
The subject 16 Breast Coil also includes a tabletop compression system which facilitates immobilization of the breast for imaging and interventional procedures and serves to hold the individual imaging coils in proximity to the breast(s). The intent of this is to reduce motion artifacts and ensure the imaging elements are positioned as close to the breast(s) as possible to optimize signal-to-noise ratio and image quality.
## Indications for use
The 16 Breast Coil for Canon 1.5T MRI Scanners are designed to provide magnetic resonance images of breast anatomy when used in conjunction with a magnetic resonance scanner. These images are interpreted by a trained physician. When used with a disposable biopsy grid, the device permits access to breast anatomy for biopsy and localization procedures.
{6}------------------------------------------------
Both the subject and predicate devices are 16 channel coils and intended to be used in conjunction with a 1.5T Magnetic Resonance Scanner(s) to produce diagnostic images of the breast anatomy that can be interpreted by a trained physician. When used with a disposable biopsy grid, the device permits access to breast anatomy for biopsy and localization procedures. The indications for use of the new device falls within the intended use of the predicate device and, therefore, the two devices have the same intended use.
The minor modifications in the indications for use statement for the subject 16 Breast Coil in comparison to dS Sentinelle Breast 16ch 1.5T Coil are as below:
a) New device trade name, b) Compatibility with 1.5T Orian and Fortian Canon MR scanners.
## Fundamental Scientific Technology:
Based on the information provided above, the subject 16 Breast Coil is considered substantially equivalent to the primary predicate device dS Sentinelle Breast 16ch 1.5T Coil (K213735) in terms of fundamental design, materials and scientific technology. At a high level, the 16 Breast Coil and the predicate coil are based on the following equivalent elements:
- Principle of operation, ●
- Prescription Use Only, .
- Anatomy of interest is the breast, .
- Same magnetic field strength (1.5T), .
- 2/10/16-Channel combinations, receive only phased-array coil with decoupling . methodology.
- . Rigid housing design that allows each imaging element to be independently positioned and configured for each patient,
- Compression plates supported by the device are used to immobilize the breast ● tissue.
- Patient support to allow for three different imaging combinations, .
- Energy source from the MR scanner. ●
- Designed to support access to the breast anatomy for both diagnostic breast . imaging and interventional exams.
The following technological differences exist between the subject and predicate devices:
- The subject device is compatible with 1.5T Orian and Fortian Canon MR scanners, . whereas the predicate device is compatible with the 70 cm bore Philips Ingenia 1.5T MR system; different system interface connector and geometry for Canon MR scanners' tabletop;
- Different fabrics used for support pads in subject and predicate devices. .
Clinical and non-clinical testing demonstrates that the safety and effectiveness requirements as outlined in FDA quidance Magnetic Resonance (MR) Receive-only Coil - Performance Criteria for Safety and Performance Based Pathway, issued December 11, 2020 were met. No new safety or efficacy concerns are raised as a result of these differences.
## Summary of Non- Clinical and Clinical Performance Data:
The subject 16 Breast Coil has undergone the following testing in accordance with FDArecognized consensus standards and as recommended in FDA quidance documents Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices, issued November 18, 2016 and Magnetic Resonance (MR) Receive-only Coil – Performance Criteria
{7}------------------------------------------------
## for Safety and Performance Based Pathwav, issued December 11, 2020:
Performance Testing - Non-Clinical:
- IEC 60601-1 General electrical/mechanical safety .
- . IEC 60601-1-6 General requirements for basic safety and essential performance -Collateral standard: Usability
- IEC 60601-1-2 EMC - Immunity, electrostatic discharge
- . IEC 60601-2-33 Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diaqnosis
- . NEMA MS 1, 3, 9 Image signal-to-noise (SNR) and Image uniformity
- NEMA MS 14 Surface heating
- ISO 10993-1 Biological safety evaluation ●
- ISO 17664-2 Cleaning and disinfection validations to support reprocessing instructions ●
#### Performance Testing - Clinical:
Acquired Image quality was assessed by the U.S. Board Certified radiologist to confirm images produced on the subject coil are of sufficient quality for diagnostic use.
#### Substantial Equivalence Conclusion:
Substantial equivalence of the 16 Breast Coil is demonstrated through the Safety and Performance Based Pathway for magnetic resonance (MR) receive-only coils.
The subject device has substantially equivalent indications for use and technological characteristics as the predicate device. Substantially equivalent performance is demonstrated by meeting all criterion in the guidance "Magnetic Resonance (MR) Receiveonly Coil -Performance Criteria for Safety and Performance Based Pathway" issued on December 11. 2020.
The summary includes the conclusions drawn from the nonclinical and clinical tests (discussed above) that demonstrate that the device is as safe, as effective, and performs as well as or better than the predicate device according to 807.92(b)(3).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.