Molecular Transport Media - MTM

K233324 · Alb Luz · QBD · Jun 17, 2024 · Microbiology

Device Facts

Record IDK233324
Device NameMolecular Transport Media - MTM
ApplicantAlb Luz
Product CodeQBD · Microbiology
Decision DateJun 17, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.2950
Device ClassClass 2

Indications for Use

The Molecular Transport Media - MTM is intended for the stabilization, transportation, and direct lysis of infectious unprocessed nasopharyngeal samples suspected of containing SARS-COV-2 virus RNA. These devices can be used for the collection transport and storage of specimens at 15-35 °C. Specimens collected and stored in a Molecular Transport Media are suitable for use with legally marketed molecular diagnostic devices.

Device Story

MTM consists of pre-filled plastic tubes containing 2 or 3 mL of proprietary liquid medium; includes guanidine-free inactivation buffer, salts, pH buffer, and distilled water. Device collects, inactivates, stabilizes, and preserves SARS-CoV-2 viral RNA from nasopharyngeal swabs. User inserts swab into tube, breaks swab tip, and seals lid. Samples stored at 15-35°C for up to 21 days. Media disrupts/lyses cells to release and stabilize RNA for downstream molecular diagnostic testing. Benefits include viral inactivation for safer handling and extended specimen stability at ambient temperatures.

Clinical Evidence

No clinical data. Bench testing only. Performance validated via Limit of Detection (LoD) studies (1.35E+03 PFU), viral nucleic acid stability studies (21 days at 15-35°C), and viral inactivation studies (complete inactivation at 5 minutes).

Technological Characteristics

Polypropylene tubes; 2 or 3 mL proprietary liquid medium (guanidine-free inactivation buffer, salts, pH buffer, distilled water). Single-use. Sterilized via ethylene oxide (EO). Compatible with standard molecular diagnostic techniques. Storage: 15-35°C.

Indications for Use

Indicated for the collection, transport, and storage of nasopharyngeal swab specimens suspected of containing SARS-CoV-2 virus RNA for laboratory testing. For prescription use only.

Regulatory Classification

Identification

A microbial nucleic acid storage and stabilization device is a device that consists of a container and reagents intended to stabilize microbial nucleic acids in human specimens for subsequent isolation and purification of nucleic acids for further molecular testing. The device is not intended for preserving morphology or viability of microorganisms.

Special Controls

The special controls for this device are: (1) The intended use for the 21 CFR 809.10 labeling must include a detailed description of microorganisms and types of human specimens intended to be preserved. (2) The 21 CFR 809.10(b) labeling must include: (i) A detailed device description, including all device components. (ii) Performance characteristics from applicable analytical studies, including but not limited to, nucleic acid stability and microorganism inactivation. (iii) A limiting statement that erroneous results may occur when the transport device is not compatible with molecular testing. (iv) A limiting statement that the device has only been validated to preserve the representative microorganisms used in the analytical studies. (3) Design verification and validation must include the following: (i) Overall device design including all device components and all control elements incorporated into the analytical validation procedures. (ii) Thorough description of the microorganisms and methodology used in the validation of the device including, but not limited to, extraction platforms and assays used for the detection of preserved nucleic acids. (iii) The limit of detection (LoD) of the molecular test used to establish microorganism nucleic acid stability.

*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use for the labeling required under § 809.10 of this chapter must include a detailed description of microorganisms and types of human specimens intended to be preserved. (2) The labeling required under § 809.10(b) of this chapter must include the following: (i) A detailed device description, including all device components; (ii) Performance characteristics from applicable analytical studies, including nucleic acid stability and microorganism inactivation; (iii) A limiting statement that erroneous results may occur when the transport device is not compatible with molecular testing; and (iv) A limiting statement that the device has only been validated to preserve the representative microorganisms used in the analytical studies. (3) Design verification and validation must include the following: (i) Overall device design, including all device components and all control elements incorporated into the analytical validation procedures; (ii) Thorough description of the microorganisms and methodology used in the validation of the device including, extraction platforms and assays used for the detection of preserved nucleic acids; and (iii) The limit of detection (LoD) of the molecular test used to establish microorganism nucleic acid stability.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY ## I Background Information: A 510(k) Number K233324 B Applicant ALB Luz C Proprietary and Established Names Molecular Transport Media - MTM D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | QBD | Class II | 21 CFR 866.2950 - Microbial Nucleic Acid Storage And Stabilization Device | MI - Microbiology | ## II Submission/Device Overview: A Purpose for Submission: To obtain substantial equivalence determination for ALB Luz Molecular Transport Media – MTM for the collection, transport and storage or nasopharyngeal specimens suspected of containing SARS-CoV-2 to the laboratory for downstream testing. B Measurand: Storage and stability of nucleic acids from SARS-CoV-2. C Type of Test: Microbial nucleic acid storage and stabilization device. Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} III Intended Use/Indications for Use: A Intended Use(s): See Indications for Use below. B Indication(s) for Use: The Molecular Transport Media - MTM is intended for the stabilization, transportation, and direct lysis of infectious unprocessed nasopharyngeal samples suspected of containing SARS-COV-2 virus RNA. These devices can be used for the collection transport and storage of specimens at 15-35 °C. Specimens collected and stored in a Molecular Transport Media are suitable for use with legally marketed molecular diagnostic devices. C Special Conditions for Use Statement(s): Rx - For Prescription Use Only D Special Instrument Requirements: None IV Device/System Characteristics: A Device Description: The MTM consists of a pre-filled plastic tube containing either 2 or 3 mL of proprietary liquid medium intended for viral nucleic acid stabilization and transportation and inactivation of nasopharyngeal swab specimens suspected of containing SARS-CoV-2. MTM is intended for use with standard diagnostic/identification techniques that have been adequately validated and found to be compatible with the MTM. The formulation of the MTM includes guanidine-free inactivation buffer, salts, a buffer to maintain a neutral pH, and distilled water. The MTM maintains the integrity of SARS-CoV-2 viral nucleic acid when properly stored. Prior to use, vials should be stored at 15-35°C for up to 18 months. After collection, the transport tube containing the specimen can be stored at 15-35°C for up to 21 days, for transportation to the laboratory and storage. The MTM is available in the following configurations: - 50 vials with 3 mL of Molecular Transport Media (Catalog Number: MEI00211) - 50 vials with 2 mL of Molecular Transport Media (Catalog Number: MEI00283) B Principle of Operation: The product is intended for use in the collection, inactivation, stabilization, transport, and preservation of SARS-CoV-2 viral nucleic acid from clinical samples collected from a rayon tipped nasopharyngeal swab. After sample collection, the swab is transferred into the tube of the K233324 - Page 2 of 10 {2} MTM. Users then break the swab leaving the swab tip in the transport tube and screw the lid of the tube tightly. The MTM is intended to disrupt/lyse cells and stabilize SARS-CoV-2 RNA. The preserved and stabilized SARS-CoV-2 RNA maintains its integrity for downstream molecular based detection/analysis. Samples should either be tested immediately or stored at 15-35°C for up to 21 days. The media contains the following reagents: - Guanidine-free inactivation buffer - Salts - pH buffer solution - Distilled water V Substantial Equivalence Information: A Predicate Device Name(s): MagXtract Collection Tube B Predicate 510(k) Number(s): K212113 C Comparison with Predicate(s): | Device & Predicate Device(s): | Device: K233324 | Predicate: K212113 | | --- | --- | --- | | Device Trade Name | Molecular Transport Media - MTM | MagXtract Collection Tube | | General Device Characteristic Similarities | | | | Intended Use/Indications For Use | The Molecular Transport Media - MTM is intended for the stabilization, transportation, and direct lysis of infectious unprocessed nasopharyngeal samples suspected of containing SARS-CoV-2 virus RNA. These devices can be used for the collection, transport, and storage of specimens at 15-35°C. Specimens collected | The MagXtract collection tube collection tube is intended for the stabilization, transportation, and direct lysis of infectious unprocessed nasopharyngeal samples suspected of containing SARS-CoV-2 virus RNA. These devices can be used for collection, transport, and storage of specimens at | K233324 - Page 3 of 10 {3} | | and stored in a Molecular Transport Media are suitable for use with legally marketed molecular diagnostic devices. | refrigerated (2-8°C) or ambient temperatures (20-25°C). Specimens collected and stored in MagXtract collection tube are suitable for use with legally marketed molecular diagnostic devices. | | --- | --- | --- | | Sample Source | Same | Upper human respiratory | | Specimen Type | Same | Nasopharyngeal swab suspected of containing SARS-CoV-2 | | Nucleic Acid Preserved | Same | SARS-CoV-2 RNA | | Inactivation Test | Same | Inactivates SARS-CoV-2 virus | | General Device Characteristic Differences | | | | Specimen Stability | The Molecular Transport Media - MTM collection tube preserves SARS-CoV-2 RNA for up to 21 days at 15-35°C. | MagXtract collection tube preserves SARS-CoV-2 RNA for up to 5 days at 2-4°C and 20-25°C. | | Volume in Tube | 2 or 3 mL | 1.2 mL | | Storage Temperature | 15-35°C | Refrigerated: 2-8°C Ambient Temperatures 20-25°C | | Shelf-Life | 18 months | 12 months | VI Standards/Guidance Documents Referenced: Special controls that are applicable to regulation 21 CFR 866.2950. VII Performance Characteristics (if/when applicable): A Analytical Performance: 1. Precision/Reproducibility: Not applicable (N/A). 2. Linearity: K233324 - Page 4 of 10 {4} N/A 3. Analytical Specificity/Interference: N/A 4. Assay Reportable Range: N/A 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): A. Shelf-Life: The shelf-life for the ALB Luz Molecular Transport Media - MTM was established to be 18 months from the date of manufacture when stored in a clean, dry, ventilated environment at 15-35°C. The shelf-life study was conducted by using a real-time aging performance test for 18 months. In this study, the shelf life was evaluated for physical stability (appearance) and pH stability during the period of storage. i. MTM Lots Tested: The lot numbers utilized for the Shelf-Life Study are summarized in Table 1 below. Table 1: MTM Lot Numbers Utilized for Shelf-Life Study. | Lot Number | Manufacture Date | Expiration Date | | --- | --- | --- | | 10909040422MTM | 4/4/2022 | 10/4/2023 | | 10910040422MTM | 4/4/2022 | 10/4/2023 | | 02102022MTM | 10/2/2022 | 4/2/2024 | ii. Physical Stability: To evaluate physical stability, any changes in appearance, color, and turbidity of the MTM were physically or visually examined at storage time points T = 0, 3, 6, 9, 12, 15 and 18 months. Three lots with three replicates per lot at each timepoint were tested for physical stability at storage conditions of 15°C and 35°C. At each time point, the appearance of the product was inspected visually as colorless and clear liquid, without having precipitate. iii. pH Stability: To evaluate product stability, the pH of the MTM was tested at storage time points T = 0, 3, 6, 9, 12, 15, and 18 months. Three lots with three replicates per lot at each time point were tested for pH stability at storage conditions of 15°C and 35°C. At each time point, the pH was measured by immersing the pH meter electrode into each MTM by adjusting the temperature to 25°C prior to testing. For all the tubes tested, the pH measurement was within the acceptable pH range of 8 +/- 1. The results for physical stability and pH stability collectively support the 18-month stability claim for the MTM for 18 months. B. Sterilization: K233324 - Page 5 of 10 {5} MTM is not claimed to be sterile nor is it intended to be sterilized by the end user. These vials are single use devices. The manufacturing process includes sterilization of the polypropylene tubes prior to being filled by ethylene oxide (EO). Dispensing the MTM liquid into the tubes occurs in three steps: filtration, filling, and closing the tube. ## 6. Detection Limit: The limit of detection (LoD) study, viral nucleic acid stability, and viral inactivation studies were all evaluated using the three different aged lots listed in Table 2 below. Table 2: MTM Lot Numbers: | Lot Number | Date of Manufacture (DOM) | Expiration Date | Age at time of testing | | --- | --- | --- | --- | | 880124MTM | 1/22/2024 | 7/22/2025 | 2 months | | 112311150523MTM | 5/15/2023 | 11/15/2024 | 10 months | | 02102022MTM | 10/2/2022 | 4/2/2023 | 18 months | ## A. Limit of Detection Study: LoD testing was conducted to determine the lowest concentration of SARS-CoV-2 that contains measurable nucleic acids that can be repeatedly recovered from the transport media with a greater than 95% accuracy. The LoD was determined with in both preliminary and confirmatory studies as described below. i. Preliminary LoD Study: A preliminary LoD study was conducted using a virus stock of SARS-CoV-2 (GenBank MT126808.1, Stock: 2.70E+07 PFU/mL, initially diluted to 2.70E+06 PFU/mL, then 10-fold serially diluted) and three pooled negative clinical nasal samples in a 1:1 ratio (75 μL each) resulting in final sample SARS-CoV-2 concentrations of 1.35E+05, 1.35E+04, 1.35E+03, 1.35E+02 and 1.35E+01 PFU/mL. Approximately 100 μL of each concentration was absorbed onto rayon swabs (Copan, CLASSIQSwabs) in triplicate to obtain virus concentrations of 1.35E+04, 1.35E+03, 1.35E+02, 1.35E+01 and 1.35 PFU. The swab was used to transfer each concentration of the virus and matrix into each lot of MTM at 15°C and 35°C, respectively, prior to RNA extraction. SARS-CoV-2 RNA was extracted using MagMAX Viral/Pathogen Nucleic Acid Isolation Kit. Detection of SARS-CoV-2 RNA was conducted using the TaqPath COVID-19 Combo Kit (EUA200010) with the Applied Biosystems 7500 instrument. All procedures for nucleic acid extraction and amplification were performed following the manufacturer's instructions. The TaqPath COVID-19 Combo Kit result interpretation per the manufacturer's instructions are summarized in Table 3 below. Table 3: TaqPath COVID-19 Combo Kit Result Interpretation. | TaqPath COVID-19 Combo Kit Targets | | | Result | | --- | --- | --- | --- | | ORF1ab | N gene | S gene | | | Two or more SARS-CoV-2 targets with Ct≤37 | | | Positive | | Only one SARS-CoV-2 targets with Ct≤37 | | | Inconclusive | | Ct>37 | Ct>37 | Ct>37 | Negative | Results for preliminary LoD are summarized in Table 4. K233324 - Page 6 of 10 {6} Table 4: Summary of Preliminary LoD Results for MTM at ${15}^{ \circ }\mathrm{C}$ and ${35}^{ \circ }\mathrm{C}$ . | Lot | Temp (℃) | SARS-CoV-2 (PFU) | Replicate Number | Positive Number | Mean ± SD (Ct) | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | ORF1ab | N gene | S gene | | 880124MTM | 15 | 1.35E+03 | 3 | 3 | 32.90±1.18 | 34.05±0.80 | 32.22±0.88 | | | | 1.35E+02 | 3 | 2 | 34.13±1.07 | 34.95±0.63 | 34.61±2.60 | | | | 1.35E+01 | 3 | 1 | 36.08±0.52 | 37.41±1.69 | Undetermined | | | | 1.35 | 3 | 3 | 34.85±0.62 | 35.16±0.52 | Undetermined | | | 35 | 1.35E+03 | 3 | 3 | 34.89±1.69 | 35.91±1.21 | Undetermined | | | | 1.35E+02 | 3 | 0 | 36.46±0.31 | Undetermined | Undetermined | | | | 1.35E+01 | 3 | 0 | 37.85±0.23 | Undetermined | Undetermined | | | | 1.35 | 3 | 0 | Undetermined | 34.62±0 | Undetermined | | 112311150523MTM | 15 | 1.35E+03 | 3 | 3 | 31.53±1.02 | 33.10±0.78 | 31.37±1.18 | | | | 1.35E+02 | 3 | 0 | 36.28±0.77 | 39.14±0.32 | Undetermined | | | | 1.35E+01 | 3 | 0 | 37.56±0.60 | Undetermined | Undetermined | | | | 1.35 | 3 | 3 | 34.50±0.48 | 35.69±0.90 | Undetermined | | | 35 | 1.35E+03 | 3 | 2 | 33.98±1.30 | 36.62±3.01 | Undetermined | | | | 1.35E+02 | 3 | 1 | 37.15±1.06 | 35.21±0 | Undetermined | | | | 1.35E+01 | 3 | 0 | 39.96±0.49 | Undetermined | Undetermined | | | | 1.35 | 3 | 0 | Undetermined | Undetermined | Undetermined | | 02102022MTM | 15 | 1.35E+03 | 3 | 3 | 32.37±0.26 | 33.81±0.18 | 32.43±0.80 | | | | 1.35E+02 | 3 | 3 | 34.14±0.63 | 35.46±1.02 | 34.55±0.89 | | | | 1.35E+01 | 3 | 1 | 36.34±0.21 | 36.47±0 | Undetermined | | | | 1.35 | 3 | 1 | 36.69±0.53 | 37.54±0.97 | Undetermined | | | 35 | 1.35E+03 | 3 | 2 | 34.36±0.66 | 35.46±1.45 | Undetermined | | | | 1.35E+02 | 3 | 0 | 36.41±1.29 | 39.89±0 | Undetermined | | | | 1.35E+01 | 3 | 0 | Undetermined | Undetermined | Undetermined | | | | 1.35 | 3 | 0 | Undetermined | Undetermined | Undetermined | The preliminary LoD at SARS-CoV-2 concentration $1.35\mathrm{E} + 04$ PFU the results were $3/3$ for all targets, data not shown. The concentration of $1.35\mathrm{E} + 03$ PFU was determined to be the preliminary LoD for MTM at both $15^{\circ}\mathrm{C}$ and $35^{\circ}\mathrm{C}$ . ii. Confirmatory LoD: The confirmatory study was determined with 20 replicates for SARS-CoV-2 at concentrations of $1.35\mathrm{E} + 04$ , $1.35\mathrm{E} + 03$ and $1.35\mathrm{E} + 02$ PFU. Each replicate was inoculated into each lot of MTM tubes and were tested as previously described in the preliminary LoD study above. At a concentration of $1.35\mathrm{E} + 04$ PFU the results were 20/20 for all targets, data not shown. Results for confirmatory LoD of MTM at $1.35\mathrm{E} + 03$ and $1.35\mathrm{E} + 02$ PFU are summarized in Table 5 below. Table 5: Summary of Confirmatory LoD Results for MTM at ${15}^{ \circ }\mathrm{C}$ and ${35}^{ \circ }\mathrm{C}$ . | Lot | Temp (℃) | SARS-CoV-2 (PFU) | Replicate Number | Positive Number | Mean ± SD (Ct) | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | ORF1ab | N gene | S gene | | 880124MTM | 15 | 1.35E+03 | 20 | 20 | 30.70±1.44 | 31.10±1.46 | 29.92±0.78 | | | | 1.35E+02 | 20 | 11 | 35.55±1.34 | 37.01±1.73 | 35.10±0.74 | | | 35 | 1.35E+03 | 20 | 20 | 30.22±0.53 | 30.40±0.65 | 29.55±0.57 | | | | 1.35E+02 | 20 | 1 | 37.61±1.13 | 36.80±1.16 | 37.25±0.66 | | 112311150523MTM | 15 | 1.35E+03 | 20 | 19 | 29.78±0.52 | 30.04±0.62 | 29.54±0.57 | | | | 1.35E+02 | 20 | 15 | 35.48±0.97 | 36.34±1.71 | 35.50±0.91 | | | 35 | 1.35E+03 | 20 | 20 | 30.39±0.71 | 30.15±0.74 | 29.70±0.53 | | | | 1.35E+02 | 20 | 14 | 35.56±1.19 | 35.37±0.81 | 34.73±1.90 | | 02102022MTM | 15 | 1.35E+03 | 20 | 20 | 30.31±0.50 | 26.32±1.18 | 29.50±0.45 | | | | 1.35E+02 | 20 | 19 | 35.27±0.90 | 35.90±1.13 | 35.15±0.70 | | | 35 | 1.35E+03 | 20 | 20 | 29.72±0.67 | 29.88±0.77 | 29.06±0.58 | K233324 - Page 7 of 10 {7} Based on the results, the confirmatory LoD for MTM is $1.35\mathrm{E} + 03$ PFU. # B. Viral Nucleic Acid Stability Study: The SARS-CoV-2 virus stock (diluted to $6.00\mathrm{E} + 04$ PFU/mL) were combined with pooled negative clinical nasal matrix in a 1:1 ratio $(75~\mu \mathrm{L}$ each) resulting in a final concentration of $3.00\mathrm{E} + 04$ PFU. Approximately $100~\mu \mathrm{L}$ of this mixture was absorbed onto rayon swabs in triplicate at a final virus concentration of $3.00\mathrm{E} + 03$ PFU. Each swab was transferred into three MTM tubes of each lot and incubated at $15^{\circ}\mathrm{C}$ or $35^{\circ}\mathrm{C}$ for the following time points: 0, 7, 14 and 21 days post-inoculation. At the end of each time point, SARS-CoV-2 RNA was extracted using MagMAX Viral/Pathogen Nucleic Acid Isolation Kit, analyzed using TaqPath COVID-19 Combo Kit and Applied Biosystems 7500 instrument. All procedures for nucleic acid extraction and amplification were performed following the manufacturer's instructions. The results of the viral nucleic acid stability study summarized in Table 6 below. Table 6: Summary of SARS-CoV-2 Viral Nucleic Acid Stability Results. | Lot | Days | 15°C | | | | | | 35°C | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | ORF1ab | | N gene | | S gene | | ORF1ab | | N gene | | S gene | | | | | Mean Ct | ΔCt | Mean Ct | ΔCt | Mean Ct | ΔCt | Mean Ct | ΔCt | Mean Ct | ΔCt | Mean Ct | ΔCt | | 880124MTM | 0 | 25.89 | 0 | 30.28 | 0 | 26.26 | 0 | 25.94 | 0 | 27.62 | 0 | 26.50 | 0 | | | 7 | 25.41 | 0.48 | 25.87 | 4.41 | 26.00 | 0.26 | 25.50 | 0.44 | 25.33 | 2.29 | 26.22 | 0.28 | | | 14 | 25.45 | 0.44 | 25.57 | 4.71 | 26.35 | -0.09 | 25.78 | 0.16 | 25.62 | 2.00 | 26.45 | 0.04 | | | 21 | 26.75 | -0.86 | 26.62 | 3.66 | 27.94 | -1.68 | 27.02 | -1.08 | 26.42 | 1.20 | 28.47 | -1.98 | | 112311150523MTM | 0 | 25.62 | 0 | 31.40 | 0 | 26.14 | 0 | 26.53 | 0 | 27.64 | 0 | 27.16 | 0 | | | 7 | 24.67 | 0.95 | 25.26 | 6.14 | 25.25 | 0.89 | 25.80 | 0.73 | 25.75 | 1.89 | 26.55 | 0.61 | | | 14 | 24.96 | 0.66 | 25.10 | 6.30 | 25.71 | 0.43 | 27.16 | -0.63 | 26.14 | 1.50 | 27.09 | 0.07 | | | 21 | 26.39 | -0.77 | 26.21 | 5.19 | 27.65 | -1.51 | 27.71 | -1.18 | 27.22 | 0.42 | 28.64 | -1.48 | | 02102022MTM | 0 | 25.34 | 0 | 27.71 | 0 | 25.90 | 0 | 26.46 | 0 | 27.19 | 0 | 27.25 | 0 | | | 7 | 25.58 | -0.24 | 25.83 | 1.88 | 26.36 | -0.46 | 24.50 | 1.96 | 24.48 | 2.71 | 25.25 | 2.00 | | | 14 | 24.75 | 0.59 | 24.75 | 2.96 | 25.54 | 0.36 | 26.46 | 0.00 | 26.17 | 1.02 | 27.07 | 0.18 | | | 21 | 26.03 | -0.69 | 25.88 | 1.83 | 26.94 | -1.04 | 27.00 | -0.54 | 26.33 | 0.86 | 28.30 | -1.05 | | ΔCT calculation (Mean Ct Day 0 minus Mean Ct Day 7, 14 and 21) | | | | | | | | | | | | | | The MTM provided positive detection for all samples tested with Ct variation (ΔCt) less than ±3 Ct values for at least two of the three viral targets (ORF1ab, N gene and S gene) at time points 7, 14 and 21 days post-inoculation comparing to time point 0 for both $15^{\circ}\mathrm{C}$ and $35^{\circ}\mathrm{C}$ . Based on these study results, MTM is able to provide stability to the SARS-CoV-2 viral nucleic acid (i.e., RNA) for 21 days at both 15 and $35^{\circ}\mathrm{C}$ . # C. Viral Inactivation: A viral inactivation study was conducted to test the ability of MTM to inactivate SARS-CoV-2 virus. i. Cytotoxicity: A volume of $30~\mu \mathrm{L}$ of each lot of MTM was transferred to a 24-well microplate containing $270~\mu \mathrm{L}$ of Dulbecco's Modified Eagle Medium with $10\%$ fetal K233324 - Page 8 of 10 {8} bovine serum cell culture media (DMEM with 10% FBS) and serially diluted (10-fold) to obtain dilutions of 1:10E+01 to 1:10E+08 in triplicate (i.e., three tubes per lot of MTM). Next, 300 μL of each dilution was added onto confluent Vero cells (ATCC, Cat: CCL-81) and incubated at 37°C at 5% CO2 for 48 hours. Cytotoxicity was determined visually using an inverted microscope looking for Cytopathic Effect (CPE). Results of the cytotoxicity study are summarized in Table 7 below. Table 7: Summarized of Cytotoxicity Study Results of MTM. | Lots* | Temp (°C) | MTM Dilution | Replicate (MTM tubes) | Cytotoxicity (CPE) | | --- | --- | --- | --- | --- | | 880124MTM 112311150523MTM 02102022MTM | 15 | 1:10E+01 | 3 | Cytotoxic | | | | 1:10E+02 | 3 | | | | | 1:10E+03 | 3 | Non-cytotoxic | | | 35 | 1:10E+01 | 3 | Cytotoxic | | | | 1:10E+02 | 3 | | | | | 1:10E+03 | 3 | Non-cytotoxic | | *Each MTM lot was tested individually and obtained identical results. | | | | | The lowest dilution to indicate normal cell growth (i.e., no cytotoxicity) was determined to be the 1:10E+03 dilution in all lots at 15°C and 35°C of temperature storage. No cytotoxicity was observed for any of the dilutions from 1:10E+04 to 1:10E+08, data not shown. ii. Viral Inactivation: A volume of 75 μL of SARS-CoV-2 virus stock was mixed with 75 μL of pooled clinical negative nasal matrix in a 1:1 ratio to obtain a virus concentration of 1.35E+07 PFU/mL. Approximately 100 μL of this mixture was absorbed onto a rayon swab at a virus concentration of 1.35E+06 PFU and transferred into three tubes of each MTM lot number stored at 15°C and 35°C at time pointes of 0, 5 and 15-minutes post inoculation. At the end of each time point, 30 μL of each sample was added into 270 μL of DMEM with 10% FBS cell culture media and serially diluted (10-fold) to 1:10E+01 to 1:10E+08 in triplicate (i.e., three tubes per lot of MTM). Next, 300 μL of each dilution was added onto confluent Vero cells and incubated at 37°C with 5% CO2 for virus absorption for one hour. Semisolid media containing DMEM with 10% FBS, 1% carboxymethylcellulose, and 1% penicillin/streptomycin antibiotics were added to each well and incubated at 37°C with 5% CO2 for up to 6 days. After the incubation period, the microplate wells containing the Vero cells were fixed with 10% paraformaldehyde, washed and stained with 1% crystal violet. Plaque-forming units (PFUs) were counted visually. No PFUs were obtained after the exposure times of 0, 5 and 15 minutes for all MTM lot numbers at 15°C and 35°C. All the results appear to be acceptable and support SARS-CoV-2 inactivation at 5 minutes exposure with MTM at 15°C and 35°C. 7. Assay Cut-Off: K233324 - Page 9 of 10 {9} N/A ## B Comparison Studies: 1. Method Comparison with Predicate Device: N/A 2. Matrix Comparison: N/A ## C Clinical Studies: 1. Clinical Sensitivity: N/A 2. Clinical Specificity: N/A 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): N/A ## D Clinical Cut-Off: N/A ## E Expected Values/Reference Range: N/A ## VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. ## IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K233324 - Page 10 of 10
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...