K233288 · Nakanishi, Inc. · EBW · Jun 24, 2024 · Dental
Device Facts
Record ID
K233288
Device Name
NLZ Built-In Motor System
Applicant
Nakanishi, Inc.
Product Code
EBW · Dental
Decision Date
Jun 24, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The NLZ Built-in Motor System is indicated for use in the field of prophylaxis dentistry, restorative applications including cavity preparation and endodontic therapy, prosthodontics applications such as crown preparations.
Device Story
NLZ Built-In Motor System is a dental treatment system integrated into dental chair units; comprises control unit, Multi-Pad, handpiece, and cord. Functions as a brushless electric micromotor; receives input from dental chair foot pedal to regulate motor speed, rotation direction, and torque. Provides power to compatible handpieces for dental procedures. Used by dental professionals in clinical settings. Software controls brightness, rotation, speed, and torque. Enables both reciprocating and continuous rotation modes for endodontic files. Benefits include enhanced control and versatility during restorative and endodontic treatments.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including motor performance, reprocessing validation, software verification, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and biocompatibility (ISO 10993-1).
Technological Characteristics
Brushless electric micromotor system; integrated with dental chair units. Includes control unit, Multi-Pad, handpiece, and cord. Software-controlled speed, torque, and rotation (reciprocating/continuous). Complies with ISO 14457 (handpieces/motors), IEC 62304 (software), and IEC 60601 series (electrical safety/EMC). Biocompatibility per ISO 10993-1 for limited contact (<24h) with tissue/bone/dentin.
Indications for Use
Indicated for dental patients requiring prophylaxis, restorative procedures (cavity preparation), endodontic therapy, and prosthodontic applications (crown preparations).
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 24, 2024
Nakanishi Inc. % Takahiro Haruyama President Globizz Corporation 1411 W. 190th Street Suite 200 Gardena, California 90248
Re: K233288
Trade/Device Name: NLZ Built-In Motor System Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EBW Dated: May 25, 2024 Received: May 28, 2024
Dear Takahiro Haruyama:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Michael E. Adjodha -S
Michael Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
| Submission Number (if known) | |
|------------------------------|--|
|------------------------------|--|
K233288
Device Name
NLZ Built-In Motor System (NLZ PRO); NLZ Built-In Motor System (NLZ BF SET); NLZ Built-In Motor System (NLZ BFB SET); NLZ Built-In Motor System (NLZ SEL SET); NLZ Built-In Motor System (NLZ SELB SET):
NLZ Built-In Motor System (Multi Pad 2 SET)
Indications for Use (Describe)
The NLZ Built-in Motor System is indicated for use in the field of prophylaxis dentistry, restorative applications including cavity preparation and endodontic therapy, prosthodontics applications such as crown preparations.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) #: K233288
510(k) Summary
Prepared on: 2024-06-23
### Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | Nakanishi Inc |
|---------------------------------|---------------------------------------------------------------|
| Applicant Address | 700 Shimohinata Kanuma Tochigi 323-8666 Japan |
| Applicant Contact Telephone | +81-289-64-7277 |
| Applicant Contact | Mr. Masaaki Kikuchi |
| Applicant Contact Email | nskra@nsk-nakanishi.co.jp |
| Correspondent Name | Globizz Corporation |
| Correspondent Address | 1411 W. 190th Street Suite 200 Gardena CA 90248 United States |
| Correspondent Contact Telephone | 1-310-538-3860 |
| Correspondent Contact | Mr. Takahiro Haruyama |
| Correspondent Contact Email | register@globizz.net |
21 CFR 807.92(a)(2)
| Device Name | |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name | NLZ Built-In Motor System (NLZ PRO);<br>NLZ Built-In Motor System (NLZ BF SET);<br>NLZ Built-In Motor System (NLZ BFB SET);<br>NLZ Built-In Motor System (NLZ SEL SET);<br>NLZ Built-In Motor System (NLZ SELB SET);<br>NLZ Built-In Motor System (Multi Pad 2 SET) |
| Common Name | Dental handpiece and accessories |
| Classification Name | Controller, Foot, Handpiece And Cord |
| Regulation Number | 872.4200 |
| Product Code | EBW |
21 CFR 807.92(a)(3)
Legally Marketed Predicate Devices
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K181858 | Electric Handpiece Motor EM-12L | EBW |
| K161213 | XSmart iQ | EBW |
21 CFR 807.92(a)(4)
Device Description Summary
The NLZ Built-in Motor System is a dental treatment system that is integrated and used in conjunction with dental chair units. It is composed of a motor system, a Multi-Pad, as well as a handpiece (optional) and cord that are compatible with different electric attachments The motor system has a control unit that works as an electric micromotor which can turn on or off and regulates the speedattachments. The motor system has a control unit that works as an electric micromotor which can turn on or off and regulates the speed
{5}------------------------------------------------
of the electric motor by the foot pedal of the dental unit. This product can be connected to dental units currently in use to add on a brushless electric micromotor with an LED light. Depending on the specifications of each chair unit manufacturers, the combination and operation method of the NLZ Built-in Motor System board, harness, and operation method will change.
## Intended Use/Indications for Use
The NLZ Built-in Motor System is indicated for use in the field of prophylaxis dentistry, restorative applications including cavity preparation and endodontic therapy, prosthodontics applications such as crown preparations.
## Indications for Use Comparison
The indications for use are the same. Both devices are indicated for use in preventive applications such as cavity preparation and endodontic therapy, and prosthodontic applications including crown preparations.
## Technological Comparison
The subject device is equipped with functionality for driving files in both reciprocating and continuous rotation mode; this feature is absent in the predicate device but present in the reference device.
Similarities: The NLZ Built-in Motor System is a dental treatment system that is integrated and used together with the dental chair unit. It transmits power to compatible handpieces and control functions. Software can control functions such as brightness, direction of rotation, speed, and torque. The NLZ Built-in Motor System shares these technical characteristics with the predicate device.
Differences: The differences between the NLZ Built-in Motor System and the predicate device are the ability to drive files in both reciprocating and continuous rotation modes, and the size. However, this ability to drive in both reciprocating and continuous rotation modes is present in the reference device. The size difference is more a matter of the user's personal preference, which does not affect substantial equivalence.
Therefore, these differences between the NLZ Built-in Motor System and the predicate device do not affect substantial equivalence.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The following non-clinical tests were performed to demonstrate substantial equivalence to the predicate device.
Performance Testing: The NLZ Built-in Motor System was subjected to verification testing for motor performance, reprocessing, software, electrical safety, and EMC to support substantial equivalence. The results of these tests demonstrate compliance with the requirements of the following standards and guidance.
· ISO 14457:2017 "Dentistry - Handpieces and motors"
· ISO 17665-1:2006 "Sterilization of health care products - Moist heat - Part 1: Requirements for the development validation and routine control of a sterilization process for medical devices"
· ISO 17664-1:2021 "Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices - Part 1: Critical and semi-critical medical devices"
• ISO 17664-2:2021 "Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices - Part 2: Non-critical medical devices."
· IEC 62304:2006+AMD1:2015 "Medical device software - Software life cycle processes"
· IEC 60601-1:2005+AMD1:2012+AMD2:2020 "Medical electrical equirements for basic safety and essential performance"
· IEC 60601-1-2:2014 "Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances - Requirements and tests"
· FDA quidance document "Dental Handpieces - Premarket Notification [510(k)] Submissions"
- · FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling"
- · FDA guidance document "Content of Premarket Submissions for Device Software Functions"
Biocompatibility: Based on the intended use, the NLZ Built-in Motor System is classified as an External Communicating Device and a device with limited contact (less than 24 hours) with Tissue/Bone/Dentin. Biocompatibility testing was performed to support substantial equivalence. The test results indicate compliance with the requirements of the following standards and quidance:
· ISO 10993-1:2018 "Biological evaluation of medical devices - Part1: Evaluation and testing within a risk management process"
· FDA guidance document "Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk process"
Nonclinical tests concluded that the NLZ Built-in Motor System is substantially equivalent to its predicates.
21 CER 807 92(a)(5)
## 21 CFR 807,92(a)(6)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.