K233284 · Hangzhou Qiantang Longyue Biotechnology Co., Ltd. · LHI · May 1, 2024 · General Hospital
Device Facts
Record ID
K233284
Device Name
Vial Adapter with Filter
Applicant
Hangzhou Qiantang Longyue Biotechnology Co., Ltd.
Product Code
LHI · General Hospital
Decision Date
May 1, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The Vial adapter with filter is intended for use by Healthcare Professionals for the transfer and mixing of drugs contained in vials within clinical, hospital, or healthcare environments.
Device Story
Vial adapter with filter facilitates drug reconstitution and transfer from vials to syringes. Device features integrated piercing spike to penetrate vial elastomeric septum; housing seats securely around vial ferrule. Opposite side contains female luer connector for attachment to standard luer syringe. Used by healthcare professionals in clinical/hospital settings. Device enables closed or filtered fluid transfer, reducing contamination risk during drug preparation. Benefits include efficient, sterile drug handling and mixing. Available in 13mm and 20mm sizes.
Clinical Evidence
No clinical data was conducted. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993), sterilization validation (ISO 11137), shelf-life/aging studies (ASTM F1980), and mechanical/functional performance testing (ISO 22413, ISO 80369-7, ISO 8536-4). All tests passed.
Technological Characteristics
Materials: Polycarbonate, polyethylene, acrylic, polyamide. Single-use, sterile, E-beam sterilized (SAL 10-6). Features: Integrated piercing spike (single lumen), filtration membrane, female luer connector. Sizes: 13mm, 20mm. Standards: ISO 10993 (biocompatibility), ISO 80369-7 (connectors), ISO 22413 (fluid transfer), ISO 8536-4 (infusion equipment).
Indications for Use
Indicated for healthcare professionals for the transfer and mixing of drugs contained in vials in clinical, hospital, or healthcare environments. No specific patient population or contraindications stated.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized caduceus. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 1, 2024
Hangzhou Qiantang Longyue Biotechnology Co., Ltd % Esther Zhang Regulatory Affairs Shanghai Ling Fu Technology Co., Ltd. 4F No. 585-2, Wanyuan Rd. Minhang District Shanghai, Shanghai 201102 China
Re: K233284
Trade/Device Name: Vial Adapter with Filter Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: LHI Dated: April 3, 2024 Received: April 3, 2024
Dear Esther Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Porsche Bennett
Porsche Bennett For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn,
{2}------------------------------------------------
General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233284
Device Name Vial Adapter with Filter
Indications for Use (Describe)
The Vial adapter with filter is intended for use by Healthcare Professionals for the transfer and mixing of drugs contained in vials within clinical, hospital, or healthcare environments.
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K233284- 510(k) Summary
#### Submitter l
| Device submitter: | Hangzhou Qiantang Longyue Biotechnology Co., LTD<br>104, 301, 302, building 12, building 1,619 WangMei Roadinping<br>street, Linping District, Hangzhou |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Chunyu Wang<br>Quality Assurance Manager |
| | Phone: 0086-13965638723 |
| | Email: chunyu_wang@nextech-x.com |
| Prepare Date: | April 26, 2024 |
#### ll Correspondent
| Company | Shanghai Ling Fu Technology Co., Ltd. |
|-----------------|-----------------------------------------------------------------------------------------------------|
| | 4F, No.585-2 Wanyuan Road, Minhang District, Shanghai<br>P.R.China |
| Contact person: | Esther ZHANG<br>Regulatory affairs<br>Phone: 0086-13771505757<br>Email: Esther.zhang@llins-tech.com |
#### ��� Subject Device
Trade Name of Device: Vial Adapter with Filter Common Name: Fluid Transfer IV Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product code: LHI Review Panel: General Hospital
#### IV Predicate Device
| Trade name: | Vial Adapter |
|-------------------------|--------------------------------------------------|
| Regulation Name: | Intravascular Administration Set |
| Regulation Number: | 21 CFR 880.5440 |
| Classification: | Class II |
| Product Code: | LHI |
| Premarket Notification: | K222718 |
| Manufacturer: | Hangzhou Qiantang Longyue Biotechnology Co., LTD |
{5}------------------------------------------------
#### > Device description
The sterile device pierces the elastomeric septum of a drug vial with its integrated piercing spike. The device is then pushed fully onto the drug vial and seats securely around the ferrule of the drug vial utilizing the housing of the vial adapter. The connector on opposite side of the vented vial adapter is for the connection of a standard luer syringe for the reconstitution and removal of the contents of the drug vial.
The device is intended for use by Healthcare Professionals (HCPs) in a clinical, hospital, or other healthcare environment. The subject device is available by prescription use only and has no known contraindications.
The proposed vented vial adapter is available in 13mm, 20mm diameter to accommodate respective sizes of drug vials.
#### VI Indications for use
The Vial adapter with filter is intended for use by Healthcare Professionals for the transfer and mixing of drugs contained in vials within clinical, hospital, or healthcare environments.
#### VII Comparison of technological characteristics with the predicate devices
The Vial Adapter intended use, technological characteristics, design and performance specifications are either identical or substantially equivalent to the existing legally marketed predicate device. The differences between the subject and predicate devices do not raise different questions of safety and effectiveness.
| Device feature | Subject Device<br>K233284 | Predicate Device<br>K222718 | Comments |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|-----------|
| Indications for use | The Vial adapter with filter<br>is intended for use by<br>Healthcare Professionals<br>for the transfer and mixing<br>of drugs contained in vials<br>within clinical, hospital, or<br>healthcare environments. | The Vial Adapter is<br>indicated for the transfer<br>and mixing of drugs<br>contained in vials. | Identical |
| Product code | LHI | LHI | Identical |
| Regulation<br>number | 21 CFR 880.5440 | 21 CFR 880.5440 | Identical |
{6}------------------------------------------------
| Class | Class II | Class II | Identical |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Principle of operation | Single use | Single use | Identical |
| Size | 13mm, 20mm | 13mm, 20mm, 28mm | Different<br>Comment 1 |
| Material | Polycarbonate,<br>Polyethylene, Acrylic,<br>Polyamide | Polycarbonate | Different<br>Comment 2 |
| Connector | Female Luer fitting | Female Luer fitting;<br>Male Luer fitting | Different<br>Comment 3 |
| Piercing Spike | Plastic - Single Lumen | Plastic - Single Lumen | Identical |
| Filter | Yes | No | Different<br>Comment 4 |
| Performance | Performance test of:<br>Penetration force.<br>Detachment<br>force from Drug<br>Vial.<br>Spike Tip Ductility.<br>Fluid Leakage.<br>Liquid Filtration | Performance test of:<br>Penetration force;<br>Detachment<br>force from Drug<br>Vial;<br>Spike Tip Ductility;<br>Fluid Leakage. | Different<br>Comment 5 |
| Vial Adapter Fit<br>(Vial Side) | Snap Fit to Vial | Snap Fit to Vial | Identical |
| Sterilization<br>Method | Electron beam Irradiation | Electron beam Irradiation | Identical |
| Sterility<br>Assurance Level | SAL 10-6 | SAL 10-6 | Identical |
| Biocompatibility | Conforms to ISO 10993 | Conforms to ISO 10993 | Identical |
| Labeling | Proposed device<br>labeling (IFU) includes<br>transfer and mixing<br>instructions | Proposed device<br>labeling (IFU) includes<br>transfer and mixing<br>instructions | Identical |
| Shelf life | 4 years | 3 years | Different<br>Comment 6 |
{7}------------------------------------------------
### Discussion:
### Comment 1
The subject device consists of Vial Adapters that are 13mm, 20mm for different diameter standard vials. The size of Vial Adapter with filter was demonstrated to meet specification according to internal performance standards. This difference does not raise new questions of safety and effectiveness.
### Comment 2
The material of the subject device is polycarbonate, polyethylene, acrylic/polyamide, and the predicate device material is polycarbonate. Although there is a difference in the material, the biocompatibility test of the subject device complies with the standard requirements. This difference does not raise new questions of safety and effectiveness.
### Comment 3
The connector of the Vial Adapter with filter have only one type, while the predicated device has two types. This difference was addressed through ISO 80369-7 testing and therefore, this difference does not raise new questions of safety and effectiveness.
### Comment 4
The inclusion of a Filtration Membrane in the subject device (K233284) differentiates it from the predicate device (K222718), which does not include a filter. Despite this difference, the additional filtration functionality in K233284 does not raise new questions of safety and effectiveness. It is supported by adherence to standards detailed below, which set requirements for transfer sets and infusion equipment, including the performance of fluid filters. The subject device has been tested to ensure that the filter performs adequately. Testing includes visual inspection for defects, dimensional measurements of the membrane, and filtration efficiency tests, all of which the device has passed. Therefore, the difference does not raise different questions of safety and effectiveness.
### Comment 5
Performance testing of the Vial adapter with filter inlet and outlet functions demonstrates that the difference does not raise new questions of safety and effectiveness.
### Comment 6
The shelf life of the subject device is 4 years which is longer than the predicated device; however, performance testing after accelerated aging passed all applicable and internal test methods. The difference does not raise new questions of substantially equivalence on safety
{8}------------------------------------------------
and effectiveness.
#### VIII Performance data
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility testing
Biocompatibility of the Vial Adapter was evaluated in accordance with ISO 10993- 1:2018 for the body contact category of "External communication device – Blood path indirect" with a contact duration of "Limited (< 24 hours)". The following tests were performed, as recommended:
| Cytotoxicity | ISO 10993-5: 2009 |
|---------------------------|--------------------|
| Skin sensitization | ISO 10993-10: 2010 |
| Hemolysis | ISO 10993-4: 2017 |
| Intracutaneous reactivity | ISO 10993-23: 2021 |
| Acute systemic toxicity | ISO 10993-11: 2017 |
| Pyrogenicity | ISO 10993-11: 2017 |
## Sterilization and shelf life testing
The sterilization method has been validated to ISO 11137-1 and ISO 11137-2, which has thereby determined the routine control and monitoring parameters. The sterilization process
is validated to a minimum SAL 10-6.
The shelf life of the Vial Adapter is determined based on stability study which includes ageing test. The testing is performed according to the following standards:
> ISO 11607-1: 2019 Packaging for terminally sterilized medical devices - Part 1:
Requirements for materials, sterile barrier systems and packaging systems
- > ISO 11607-2: 2019 Packaging for terminally sterilized medical devices - Part 2:
Validation requirements for forming, sealing and assembly processes
> ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
A ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
ISO 80369-7:2021 Small-bore connectors for liquids and gases in healthcare A
{9}------------------------------------------------
applications Part 7: Connectors for intravascular or hypodermic applications
» ISO 8536-4:2019 Infusion equipment for medical use Part 4: Infusion sets for single use, gravity feed
A ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems
### Performance testing
Performance testing Summary
| Items | Testing standard | Result | |
|------------------------|--------------------------------------|--------------------------------|------|
| Appearance | Internal performance standards | Pass | |
| Particulate | ISO 22413 -2021 | Pass | |
| Tensile strength | ISO 22413 -2021 | Pass | |
| Leakage | ISO 22413 -2021 | Pass | |
| Unobstructed | ISO 22413 -2021 | Pass | |
| Piercing Spike | ISO 22413 -2021 | Pass | |
| Puncture force | ISO 22413 -2021 | Pass | |
| puncture chip | ISO 22413 -2021 | Pass | |
| Dimension | Internal performance standards | Pass | |
| Housing | ISO 22413 -2021 | Pass | |
| Luer Connector | ISO 80369-7 | Pass | |
| Detachment force | Internal performance standards | Pass | |
| Spike tip ductility | Internal performance standards | Pass | |
| Filtration rate | ISO 22413 -2021 | Pass | |
| Chemical property | ISO 8536-4:2019 | Pass | |
| Chemical<br>Properties | Reducing substances<br>(easy oxides) | Internal performance standards | Pass |
| | Metal ions | | |
| | pH | | |
| | Evaporation residues | | |
| | UV absorbance | | |
| Sterile | USP46-NF41 | Pass | |
| Bacterial endotoxin | USP-NF<71> | Pass | |
#### IX Clinical Data
Clinical data was not conducted to support a substantial equivalence determination.
#### × Conclusion
The Vial Adapter with Filter are substantially equivalent to the predicate device (Vial Adapter). The non-clinical testing demonstrates that the differences between the subject and
{10}------------------------------------------------
predicate device do not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.