K233280 · Steadysense GmbH · FLL · Jun 21, 2024 · General Hospital
Device Facts
Record ID
K233280
Device Name
SteadyTemp
Applicant
Steadysense GmbH
Product Code
FLL · General Hospital
Decision Date
Jun 21, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
The SteadyTemp system is a wireless thermometer intended to provide precise body temperature for users of all ages. The SteadyTemp system comprises a single-use temperature measuring sensor, the SteadySense Patch, and the SteadyTemp application. SteadySense Patch measures and stores axillary skin temperature data up to 10 days. The typical application duration when using the system is 7 days.
Device Story
SteadyTemp is a wireless, non-invasive, single-use clinical electronic thermometer for all ages. System consists of a disposable SteadySense patch and a mobile SteadyTemp application. Patch contains a solid-state contact temperature transducer; measures axillary skin temperature; records data locally for up to 77 hours. User applies patch to healthy axillary skin; data transmitted to app via Near-Field Communication (NFC). App displays temperature data to user. Device intended for home healthcare environment. Provides continuous temperature monitoring; aids in tracking body temperature trends. Benefits include non-invasive, continuous, precise temperature measurement without requiring frequent manual checks.
Clinical Evidence
Bench testing only. No clinical data. Performance verified against ISO 80601-2-56:2018 (clinical thermometer accuracy), IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and IEC 60601-1-11 (home healthcare). Biocompatibility testing (irritation, sensitization, cytotoxicity) performed per ISO 10993-1. Software verified per IEC 62304.
Technological Characteristics
Solid-state contact temperature transducer; 3.0V lithium button cell battery; NFC connectivity (2.5 cm range); flexible textile/insulating foam patch with adhesive layer; dimensions 65x60x3 mm; weight 3g. Software follows IEC 62304. Compliant with ISO 10993-1 for patient-contacting materials.
Indications for Use
Indicated for users of all ages requiring precise body temperature measurement via axillary skin contact.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 21, 2024
SteadySense GmbH % Brodie Pedersen Chief Regulatory Officer Borderless MedDev LLC 7118 Teakwood Cir Maple Grove, Minnesota 55369
Re: K233280
Trade/Device Name: SteadyTemp Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: May 23, 2024 Received: May 24, 2024
Dear Brodie Pedersen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Daniel Walloschek
David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
{2}------------------------------------------------
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233280
Device Name SteadyTemp
Indications for Use (Describe)
The SteadyTemp system is a wireless thermometer intended to provide precise body temperature for users of all ages. The SteadyTemp system comprises a single-use temperature measuring sensor, the SteadySense Patch, and the SteadyTemp application. SteadySense Patch measures and stores axillary skin temperature data up to 10 days. The typical application duration when using the system is 7 days.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
(as required by 21CFR 807.92)
## K233280
#### I. SUBMITTER
SteadySense GmbH Kärntner Strasse 518 Seiersberg Pirka 8054 Austria Phone: +43 660 8850100 Phone: 612-272-5541 Contact: Brodie Pedersen Bropdiecp@BorderlessMD.com Phone: 612-272-5541 Date Prepared: May 31, 2024
#### DEVICE II.
Name of Device: SteadyTemp Classification Name: Clinical electronic thermometer. Common or Usual Name: Thermometer, Electronic, Clinical Device Panel: General Hospital Devices, and Human Factors Regulation Number: 880.2910 Regulatory Class: Class 2 Product Code: FLL
#### III. PREDICATE DEVICE
The Steady Temp system is similar in technological characteristics to the predicate device except in the communication technology, NFC vs BLE in the predicate. This yields a difference in the distance of transmission and the reduction in the power consumption of the Steady Temp versus the predicate. These differences do not impact safety or effectiveness.
Name of Device: TempTraq, Model: TT-100, TT-200, TT-300 Predicate submission number: K201977 Regulation Number: 880.2910 Product Code: FLL
{5}------------------------------------------------
#### IV. DEVICE DESCRIPTION
SteadyTemp is a wireless thermometer designed to provide users of all ages with information about their body temperature. The device has two components, the SteadyTemp app, and the battery operated SteadySense patch. Both are required to measure temperature.
The SteadySense patch is a non-invasive, non-sterile, active temperature measurement device capable of detecting sub-degree (< 0.18°F/ 0.1°C) changes in body temperature by continuously measuring armpit temperature. The SteadySense patch is a disposable bandage product that is applied only to healthy skin under the arm. The top is flexible textile over insulating foam and the bottom is a thin adhesive layer to adhere the thermal sensor to the axilla for measurement recording.
SteadySense temperature measurement is a direct mode, solid state contact temperature transducer embedded in the patch. The patch records up to the last 77 hours of temperature data for download to the SteadyTemp app.
The typical application duration is 7 days. The SteadySense patch continuously records body temperature.
The SteadyTemp app serves as an interface for the user and for interacting with the SteadySense patch and displaying the measured temperature data. The data measured by the patch is transmitted via near-field communication (NFC).
#### V. INDICATIONS FOR USE
The SteadyTemp system is a wireless thermometer intended to provide precise body temperature for users of all ages. The SteadyTemp system comprises a single-use temperature measuring sensor, the SteadySense Patch, and the SteadyTemp application. SteadySense Patch measures and stores axillary skin temperature data up to 10 days. The typical application when using the system is 7 days.
### SUMMARY OF TECHNOLIGICAL CHARACTERISTICS WITH VI. THE PREDICATE DEVICE
The Steady Temp system is similar in technological characteristics to the predicate device except in the communication technology, NFC vs BLE in the predicate. This yields a difference in the distance of transmission and the reduction in the power consumption of the Steady Temp versus the predicate. These differences do not impact safety or effectiveness.
{6}------------------------------------------------
| Item | Predicate Device<br>TempTraq, models<br>TT-100, TT-200 and TT-300<br>K201977 | Subject Device<br>SteadyTemp<br>K233280 | Comparison |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Indications for<br>Use | The Wireless thermometers are<br>battery-operated electronic<br>devices with intended use of<br>measuring human body<br>temperature precisely. The<br>devices are single-use and<br>intended for armpit<br>temperature measurement for<br>persons of all ages. | The SteadyTemp system is a<br>wireless thermometer intended<br>to provide precise body<br>temperature for users of all<br>ages. The SteadyTemp system<br>comprises a single-use<br>temperature measuring sensor,<br>the SteadySense Patch, and the<br>SteadyTemp application.<br>SteadySense Patch measures<br>and stores axillary skin<br>temperature data up to 10<br>days. The typical application<br>duration when using the<br>system is 7 days. | Similar<br>See Note 1. |
| Product<br>Code | FLL | FLL | Identical |
| Regulation # | 21CFR880.2910 | 21CFR880.2910 | Identical |
| Display Use<br>Specification | iOS device display, Android<br>device display, Web based | iOS device display, Android<br>device display | Similar<br>See Note 2. |
| Working<br>Voltage | 3.0V DC | 3.0V DC | Identical |
| Battery | Two (2) Blue Spark 1.5 V<br>batteries (103-UT1) zinc<br>manganese dioxide | 3.0 V manganese dioxide<br>lithium button cell battery | Similar<br>See Note 3. |
| Measurement<br>Range | 30 - 42 °C (86 °F ~ 108°F) | 30 - 42 °C (86 °F ~ 108°F) | Identical |
| Accuracy | +/- 0.1 °C between<br>30 °C ~ 42.4 °C | Measurement accuracy of +/-<br>0.3°C. 30-42 °C | Similar<br>See Note 4. |
| Temperature<br>Unit | °C or °F | °C or °F | Identical |
| Signal<br>Transmission | Wireless Bluetooth BLE 4.0<br>operating at 2.4Ghz | Near-Field-Communication<br>(NFC): For data transmission<br>between Patch and software<br>application | Difference<br>See Note 5. |
| Receiver | Wireless Bluetooth BLE 4.0<br>enabled smart devices running<br>Apple operating system iOS or<br>Android operating system | Near Field Communication<br>(NFC) enabled smart devices<br>running Apple operating<br>system iOS or Android<br>operating system | Similar<br>See Note 5. |
| Valid<br>Transmission | Up to 40 feet | 2.5 cm | Difference<br>See Note 5. |
| Item | Predicate Device<br>TempTraq, models<br>TT-100, TT-200 and TT-300<br>K201977 | Subject Device<br>SteadyTemp<br>K233280 | Comparison |
| Operating<br>Temperature | 16 C ~ 40 °C (60.8 °F ~ 104 °F) | 15 C ~ 40 °C (59 °F ~ 104 °F) | Similar See<br>Note 12. |
| Operating<br>Humidity | 15%-95% RH | 5%-95% RH | Similar See<br>Note 12. |
| Anatomical<br>Application | Armpit<br>peel-and-stick contact<br>thermometer sensor | Armpit<br>peel-and-stick contact<br>thermometer sensor | Identical |
| Patient<br>Usage | Single-Use | Single-Use | Identical |
| Temperature<br>Measurement<br>Interval | Continuous-transmitter measures<br>body temperature<br>every 10 seconds | Continuous-transmitter measures<br>body temperature<br>Every 300 seconds | Similar See<br>Note 6. |
| Memory<br>Function | TT-100: Can store up to 24<br>hours of readings.<br><br>TT-200: Can store up to 48<br>hours of readings.<br><br>TT-300: Can store up to 72<br>hours of readings | SteadySense Patch: can store up<br>to 77 hours of readings. | Similar See<br>Note 6. |
| Storage | Data back-up- Stored in app<br>and optionally TempTraq<br>Connect | Data back-up- Stored in app. | Similar See<br>Note 7. |
| Run Time | TT-100: 24-Hours<br>TT-200: 48-Hours<br>TT-300: 72-Hours | SteadySense Patch: 10 days | Difference<br>See Note 8. |
| Patient<br>Contacting<br>Materials | ISO 10993-1:<br>Compliant Silicone Gel<br>Adhesive | ISO 10993-1:<br>Compliant adhesive and bandage<br>material | Similar See<br>Note 9. |
| Applications | TempTraq Consumer Mobile<br>Application<br>TempTraq Patient Mobile<br>Application<br>TempTraq Clinician Mobile<br>Application | STEADYTEMP Application | Similar See<br>Note 10. |
| Dimensions | Length: 100.0 mm<br>Height: 50.0 mm<br>Thickness: 2.0 mm<br>Weight: 5.1 g | Length: 65 mm<br>Height: 60 mm<br>Thickness: 3 mm<br>Weight: 3 g | Similar See<br>Note 11. |
| Printed<br>Circuit | Printed circuit | Printed circuit | Similar See<br>Note 12. |
| Storage<br>Temperature | -4 to 122 °F (-20 °C to 50 °C) | +0°C to +40°C | Similar See<br>Note 12 |
| Item | Predicate Device<br>TempTraq, models<br>TT-100, TT-200 and TT-300<br>K201977 | Subject Device<br>SteadyTemp<br>K233280 | Comparison |
| Storage<br>Humidity<br>Range | 15 - 95% RH (non-condensing) | 15 to 95% RH (non-condensing) | Identical |
{7}------------------------------------------------
{8}------------------------------------------------
- Note 1. The Indications for Use of the predicate TempTraq System, models TT-100, TT-200 and TT-300, is similar to the proposed subject device, and is inclusive of both parts of the system. This difference does not affect safety and effectiveness.
- Note 2. The Display Use Specification for the predicate TempTraq System is similar to the proposed device, with an additional web-based display use specification. Software Verification and Validation and Performance testing was conducted to demonstrate this difference does not affect safety and effectiveness.
- Note 3. The battery used in the proposed SteadyTemp System differs from the predicate device, with two discrete 1.5V cells vs. a single 3V cell. This difference does not affect safety and effectiveness. Electrical safety testing was conducted in accordance with IEC 60601-1:2021 3rd Edition with amendment 1 and 2.
- Note 4. The temperature accuracy for the proposed system SteadyTemp differs from the predicate Temp Traq device. Performance testing was conducted to demonstrate the subject device conforms with IEC 80601-2-56 for absolute temperature accuracy. The difference does not raise new or different safety and effectiveness questions.
- Note 5. The signal transmission for the proposed system (NFC) differs from the predicate Temp Traq device (BLE). Performance testing was conducted to demonstrate this difference does not affect safety and effectiveness. The NFC communication is passive and much lower power the BLE.
- Note 6. The Memory Function for the proposed SteadyTemp system differs from the predicate device. Performance testing was conducted to demonstrate this difference does not affect safety and effectiveness.
- Note 7. The Storage for the proposed SteadyTemp system differs from the predicate device. Software Verification and Validation and Performance testing was conducted to demonstrate this difference does not affect safety and effectiveness.
- Note 8. The Run Time for two new models (TT-200 and TT-300) of the predicate TempTraq System differs from the proposed SteadyTemp system. Performance testing was conducted to demonstrate this difference does not affect safety and effectiveness.
- Note 9. The patient contacting material for the proposed SteadyTemp system differs from the predicate device. Biocompatibility testing was conducted in accordance with ISO 10993-1 to demonstrate this difference does not affect safety and effectiveness.
- Note 10. The Applications for the proposed SteadyTemp system differ from the predicate device. Software Verification and Validation and Performance testing was conducted to demonstrate this difference does not affect safety and effectiveness.
{9}------------------------------------------------
- Note 11. The dimensions for the proposed SteadyTemp system differ from the predicate device. This difference does not have any impact on safety or effectiveness.
- Note 12. The Printed Circuit for the proposed SteadyTemp system is similar to the predicate device. Storage temperature difference is due to the battery in the device for optimal life and safe storage. Electrical Safety (IEC 60601-1 Edition 3.2) and EMC testing (JEC 60601-1-2 4.1 Edition) was conducted to demonstrate these differences do not impact safety and effectiveness.
#### VII. PERFORMANCE TESTING
The following non-clinical performance tests have been passed successfully:
- ISO 80601-2-56:2018 Medical electrical equipment Part 2-56: . Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.
- IEC 60601-1:2021 3rd Edition with amendment 1 and 2. Medical . electrical equipment - Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2: 2021 4th Edition with amendment 1, Medical . Electrical Equipment - Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility - Requirements and tests.
- IEC 60601-1-11:2021 Edition 1.2, Medical electrical equipment Part ● 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- IEC 62304:2016 Edition 1.1, Medical device software Software life ● cycle processes.
- Biocompatibility testing of patient contacting materials according to ISO 10993-1 including Irritation, Sensitization and Cytotoxicity.
- Bench test results verify that SteadyTemp system can continuously ● record temperature, store temperature data in the device memory, and transmit recordings to the SteadyTemp app via NFC app connection for display by the user. Test results verify that all requirements were met and that the SteadyTemp System performs as designed.
- Software documentation following enhanced documentation . level of FDA guidance and prepared in accordance with IEC 62304 to verify and validate software performance of the SteadyTemp system.
- Cyber Security of the system has been addressed following FDA . guidance assessing threats and addressing the subsequent potential risk to the system through design controls, access limitation and monitoring of software components for new threats.
{10}------------------------------------------------
## VIII. CONCLUSIONS
The analysis of the differences between SteadyTemp and the predicate device does not raise new questions of safety and effectiveness. Based on device technological characteristics comparison, analysis, and performance test results, SteadySense determines that the SteadyTemp system performs within its design specifications and is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.