C&B Ceramic material is used for fabricating permanent restorations such as inlays, onlays, veneers and full crown restorations. C&B Permanent material is used for fabricating permanent restorations such as inlays, onlays, veneers and full crown restorations. C&B Interim material is indicated for the fabrication of temporary dental restorations in conjunction with extra-oral light-curing equipment.
Device Story
Light-curing methacrylate polymer resins with ceramic/glass fillers; used for additive manufacturing (3D printing) of dental prostheses (crowns, bridges, inlays, onlays, veneers). Compatible with DLP and LCD/MSLA 3D printers (385nm-405nm light spectrum). Workflow: import .STL file; generate supports; print; separate parts; clean; post-process with extra-oral light-curing equipment. Used in dental clinics/labs by dental professionals. Provides patient-specific restorative dental prostheses.
Clinical Evidence
Bench testing only. Evaluated appearance, flexural strength, sensitivity to ambient light, water solubility, color stability, and shade stability per ISO 4049 and ISO 10477. Biocompatibility assessment conducted per ISO 10993-1. All results met standard criteria.
Technological Characteristics
Light-curing methacrylate polymer resin with ceramic/glass fillers, photo initiators, inhibitors, UV absorbers, and pigments. Form factor: liquid resin in 500/1000ml HDPE bottles. Energy source: 385nm-405nm light for polymerization. Standards: ISO 4049, ISO 10477, ISO 10993-1. Shelf life: 3 years.
Indications for Use
Indicated for fabrication of permanent (inlays, onlays, veneers, full crowns) or temporary dental restorations.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 27, 2024
Prevest Denpro Limited % Angela Blackwell Senior Consultant Blackwell Device Consulting P.O. Box 718 Gresham, Oregon 97030-0172
Re: K233273
Trade/Device Name: C&B Ceramic; C&B Permanent; C&B Interim Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF, EBG Dated: March 5, 2024 Received: March 6, 2024
Dear Angela Blackwell:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
k233273
Device Name
C& B Ceramic, C&B Permanent, C&B Interim
Indications for Use (Describe)
The C&B Ceramic material is used for fabrications such as inlays, onlays, onlays, veneers and full crown restorations.
The C&B Permanent material is used for fabrications such as inlays, onlays, onlays, veneers and full crown restorations
C&B Interim material is indicated for the fabrication of temporary dental restorations in conjunction with extra-oral lightcuring equipment.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# C & B Ceramic, C&B Interim, C&B Permanent 510(k) Summary - K233273 November 25, 2024
Name and Address: Prevest Denpro Limited Unit II, Export Promotion Industrial Park Bari Brahmana, Jammu 181133 India Contact Person: Atul Modi Email: prevestindia@gmail.com Telephone: (941) 919 4280
Name of device: C&B Ceramic, C&B Interim, C&B Permanent Classification Name: Tooth shade resin material CFR: 21 CFR 872.3690 Primary Product Code: EBF Secondary Product Code: EBG Regulatory Class: II
#### Submission Contact:
Angela Blackwell Blackwell Device Consulting P.O. Box 718 Gresham, OR 97030-0172 (704)450-9934 angela@blackwelldevice.com
## Device Description:
Prevest C&B, Ceramic Resin is a light-curing material with ceramic fillers used for the 3D printing/ fabrication of Ceramic Crown & Bridge dental prosthesis for use in DLP, LCD/MSLA 3D printers. It is reactive to wavelength of light between 385nm and 405nm. Recommended printers and light curing unit are Asiga Max UV, Ackuretta SOL and Acuretta Curie. It is stored in 500 and 1000ml HDPE bottles and is available in several shades: A1 /A2 /A3 /A3.5 /B1 /B2 /B3 /C2 /D3 /Bleach/Extra Bleach based on the shade guide.
Prevest C&B, Interim resin is a light-curing material with glass fillers used for the 3D printing/ fabrication of long term temporary Crown & Bridge dental prosthesis for use in DLP, LCD/MSLA 3D printers. It is reactive to wavelength of light between 385nm and 405nm. Recommended printers and light curing unit are Asiga Max UV, Ackuretta SOL and Acuretta Curie. It is stored in 500 and 1000ml HDPE bottles and is available in several Shades: A1 /A2 /A3 /A3.5 /B1 /B2 /B3/ C2 /D3/Bleach/Extra Bleach based on the shade guide.
Prevest C&B, Permanent resin is a light-curing material with glass fillers for the 3D printing/fabrication of Permanent crown and bride dental prosthesis for use in DLP, LCD/MSLA 3D printers. It is reactive to wavelength of light between 385nm and 405nm. Recommended printers and
{5}------------------------------------------------
light curing unit are Asiga Max UV, Ackuretta SOL and Acuretta Curie. It is stored in 500 and 1000ml HDPE bottles and is available in several Shades: A1 / A2 / A3 / A3.5 / B1 / B2 / B3 / C2 / D3/Bleach/Extra Bleach based on the shade guide.
These materials are used for the additive manufacturing of crown and bridge prosthesis following the workflow as depicted below:
Image /page/5/Figure/2 description: The image shows a flowchart of the steps involved in 3D printing. The first step is to import the .STL file, followed by generating supports. The next steps are filling resin in the printer, printing the .STL file, and separating printed parts and supports. The final steps are cleaning and post-processing.
#### Indications for Use:
C&B Ceramic material is used for fabricating permanent restorations such as inlays, onlays, veneers and full crown restorations.
C&B Permanent material is used for fabricating permanent restorations such as inlays, onlays, veneers and full crown restorations.
C&B Interim material is indicated for the fabrication of temporary dental restorations in conjunction with extra-oral light-curing equipment.
## Testing Summary:
C&B Ceramic, C&B Interim, and C&B Permanent were tested for appearance, flexural strength, sensitivity to ambient light, water solubility, color stability, shade stability, according to protocols based on ISO 4049 and ISO 10477.
All test results met the criteria in standard.
Shelf life for C&B Ceramic, C&B Interim, and C&B Permanent is 3 years. All shelf life determinations use the same testing protocols as the characterization testing which are based on ISO 4049 and ISO 10477. The predicate and reference devices use the same ISO standard for their shelf lives are 2 years, 2,5 years or are not given.
A biocompatibility assessment according to ISO 10993-1 was conducted on all three resins.
Mechanism of Action: 3D printing of dental prosthesis
Type of Printer to Use with the materials: DLP, LCD/MSLA 3D printer with a light spectrum of between 385nm and 405nm like Ackuretta SOLand Asiga Max UV.
Predicate Device: Varseo Crown Plus K201668 and Varseo Smile Temp K193553 from Bego
Reference Devices: Freeprint Temp K200273 from Detax GmbH
#### Substantial Equivalence:
{6}------------------------------------------------
The resins have similar ingredients to the predicate and reference devices, the same indications for use, and similar physical parameter testing.
| | C&B Ceramic | C&B<br>Permanent | Varseo Crown Plus<br>K201668<br>Predicate Device | Freeprint Temp<br>K200273<br>Reference Device |
|----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | EBF | EBF | EBF | EBG |
| Indications for<br>Use | C&B Ceramic<br>material is<br>used for<br>fabricating<br>permanent<br>restorations<br>such as inlays,<br>onlays,<br>veneers and<br>full crown<br>restorations. | C&B<br>Permanent<br>material is<br>used for<br>fabricating<br>permanent<br>restorations<br>such as inlays,<br>onlays, veneers<br>and full crown<br>restorations. | Varseo Crown Plus<br>is indicated as an<br>indirect restorative<br>for both anterior<br>and posterior<br>restorations,<br>including occlusal<br>surfaces. The<br>VarseoSmile Crown<br>Plus material is<br>used for fabricating<br>permanent<br>restorations such<br>as inlays, onlays,<br>veneers and full<br>crown restorations. | Freeprint Temp is<br>indicated for the<br>fabrication of<br>temporary dental<br>restorations in<br>conjunction with<br>extra-oral curing<br>light equipment.<br>Duration is less than<br>30 days in oral<br>environment. |
| Applicable<br>Standards | ISO 4049<br>ISO 10477 | ISO 4049<br>ISO 10477 | ISO 4049<br>ISO 10477 | ISO 10477 |
| Mechanism of<br>Action | 3D printing of<br>dental<br>prosthesis | 3D printing of<br>dental<br>prosthesis | 3D printing of<br>dental prosthesis | 3D printing of dental<br>prosthesis |
| Composition | Methacrylate<br>polymer<br>resin with<br>photo<br>initiator,<br>inhibitor, UV<br>absorber and<br>pigments | Methacrylate<br>polymer resin<br>with photo<br>initiator,<br>inhibitor, UV<br>absorber and<br>pigments | Methacrylate<br>polymer resin<br>with photo<br>initiator, inhibitor<br>and pigments | Polymer resin with<br>photo initiator, UV<br>absorber and<br>pigments |
| Flexural<br>Strength<br>ISO 4049 ≥<br>100Mpa<br>ISO 10477 ≥50<br>MPa | >100 MPa | >100 MPa | >100 MPa | >100 MPa |
| Water | <40 µg/mm³ | <40 µg/mm³ | <40 µg/mm³ | <40 µg/mm³ |
| ISO 4049 ≤ 40 µg/mm³<br>ISO 10477 ≤ 40 µg/mm³ | | | | |
| Water Solubility<br>ISO 4049 ≤ 7.5 µg/mm³<br>ISO 10477 ≤ 7.5 µg/mm³ | <7.5 µg/mm³ | <7.5 µg/mm³ | <7.5 µg/mm³ | |
| Shelf Life | 3 years | 3 years | On label only | 2 years |
Crown and Bridge Resins for 3D Printing from Prevest Denpro
{7}------------------------------------------------
| | C&B Interim | Freeprint Temp<br>K200273<br>Reference Device | VarseoSmile Temp<br>K193553<br>Predicate Device |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | EBG | EBG | EBG |
| Indications for Use | C&B Interim<br>material is<br>indicated for the<br>fabrication of<br>temporary<br>dental<br>restorations in<br>conjunction with<br>extra-oral light-<br>curing<br>equipment. | Freeprint Temp is<br>indicated for the<br>fabrication of<br>temporary dental<br>restorations in<br>conjunction with<br>extra-oral curing<br>light equipment.<br>Duration is less than<br>30 days in oral<br>environment. | Varseo Temp resin<br>is indicated for the<br>fabrication of<br>temporary dental<br>restorations in<br>conjunction with<br>extra-oral light-<br>curing equipment. |
| Composition | Polymer resin<br>with photo<br>initiator, UV<br>absorber, filler<br>and pigments | Polymer resin with<br>photo initiator, UV<br>absorber and<br>pigments | Polymer resin,<br>initiator and<br>pigments |
| Applicable<br>Standards | ISO 4049<br>ISO 10477 | ISO 10477 | ISO 4049<br>ISO 10477 |
| Mechanism of<br>Action | 3D printing of<br>dental prosthesis | 3D printing of dental<br>prosthesis | 3D printing of<br>dental prosthesis |
| Flexural<br>Strength | >100 MPa | >100 MPa | >100 MPa |
| ISO 4049 ≥<br>100Mpa<br>ISO 10477 ≥50<br>MPa | | | |
| Water | <40 µg/mm³ | <40 µg/mm³ | <40 µg/mm³ |
{8}------------------------------------------------
| ISO 4049 ≤ 40 µg/mm³<br>ISO 10477 ≤ 40 µg/mm³ | | | |
|---------------------------------------------------------------------|-------------|-------------|---------------|
| Water Solubility<br>ISO 4049 ≤ 7.5 µg/mm³<br>ISO 10477 ≤ 7.5 µg/mm³ | <7.5 µg/mm³ | <7.5 µg/mm³ | <7.5 µg/mm³ |
| Shelf Life | 3 years | 2 years | On label only |
Conclusion: C&B Ceramic, C&B Permanent, and C&B Interim are substantially equivalent to the predicate devices. They have the same indications, similar testing, and very similar ingredients. Both the subject devices and the predicate device have physical parameters which meet requirements of the relevant ISO standards. Shelf life testing is similar to the shelf life testing of predicate or reference device. Reference devices are included to cover any ingredients not covered by the predicate devices. Any differences in ingredients are minor and do not change the substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.