The Ogmend® Implant Enhancement System is intended to augment pedicle screw fixation when a screw has lost purchase due to screw loosening, back-out, or breakage. The Ogmend® Implant Enhancement System is for use with rigid, thoracolumbar pedicle screw systems where a screw is inserted posteriorly, through the pedicle, and into the vertebral body. The Ogmend® Implant Enhancement System is for use in skeletally mature patients.
Device Story
Ogmend® Implant Enhancement System provides supplemental fixation for compromised pedicle screw-to-bone interfaces. System consists of sterile, single-use braided polyethylene terephthalate (PET) sleeve and stainless-steel inserter rod. Used by surgeons in clinical settings to restore stability in thoracolumbar pedicle screw systems. Inserter places sleeve into bone hole; screw is then inserted, creating interference fit, increased surface area contact, and pressure distribution to secure screw to bone. Benefits include restored mechanical stability and support for subsequent healing. Device is disposable and gamma-sterilized.
Clinical Evidence
Bench testing only. Tests included screw axial pullout, sleeve insertion force, screw insertion/extraction torque, sleeve end closure strength, and device breakage/sleeve durability.
Technological Characteristics
Implant: Braided polyethylene terephthalate (PET) monofilament fibers. Inserter: Stainless steel rod. Principle: Interference fit, surface area contact, pressure distribution. Dimensions: Medium (6.5mm diameter, 100mm length) and Large (8.5mm diameter, 150mm length). Sterilization: Gamma irradiation. Single-use.
Indications for Use
Indicated for skeletally mature patients requiring augmentation of pedicle screw fixation due to screw loosening, back-out, or breakage in rigid, thoracolumbar pedicle screw systems.
Regulatory Classification
Identification
A screw sleeve bone fixation device is intended to be implanted in conjunction with a non-resorbable, metallic bone screw where the screw has lost purchase due to loosening, backout, or breakage. The device fits between the screw threads and surrounding bone and provides increased surface area to create an interference fit to restore stability of the implant construct.
Special Controls
In combination with the general controls of the FD&C Act, the screw sleeve bone fixation device is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) In vivo performance testing under anticipated conditions of use must demonstrate:
(i) The device provides sufficient stability to allow for fracture healing; and
(ii) A lack of adverse biologic response to the implant through histopathological and histomorphometric assessment.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must:
(i) Assess the stability of the device in a rescue screw scenario;
(ii) Demonstrate that the device can be inserted and removed without damage to the implant or associated hardware;
(iii) Demonstrate the device can withstand dynamic loading without device failure; and
(iv) Characterize wear particle generation.
(3) The device must be demonstrated to be biocompatible.
(4) The device must be demonstrated to be non-pyrogenic.
(5) Performance data must demonstrate the sterility of the device.
(6) Performance data must support the labeled shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the established shelf life.
(7) Labeling must include:
(i) A detailed summary of the device technical parameters;
(ii) Information describing all materials of the device;
(iii) Instructions for use, including device removal; and
(iv) A shelf life.
{0}------------------------------------------------
October 26, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
Woven Orthopedic Technologies Mr. Brandon Bendes President and CEO 63 Center Street, #3a Manchester, Connecticut 06040
Re: K233223
Trade/Device Name: Ogmend® Implant Enhancement System Regulation Number: 21 CFR 888.3043 Regulation Name: Screw Sleeve Bone Fixation Device Regulatory Class: Class II Product Code: QVI Dated: September 28, 2023 Received: September 28, 2023
Dear Mr. Bendes:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image contains the text "Colin O'neill -S-". The text is arranged vertically, with "Colin" on the top line, "O'neill" on the second line, and "-S-" on the third line. The text is in a simple, sans-serif font and is black in color. The background of the image is white.
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health 2
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
= \$98\$\$%
Device Name
Ogmend® Implant Enhancement System
#### Indications for Use (Describe)
The Ogmend® Implant Enhancement System is intended to augment pedicle screw fixation when a screw has lost purchase due to screw loosening, back-out, or breakage. The Ogmend® Implant Enhancement System is for use with rigid, thoracolumbar pedicle screw systems where a screw is inserted posteriorly, through the pedicle, and into the vertebral body. The Ogmend® Implant Enhancement System is for use in skeletally mature patients.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
#### 510K Summary
### MANUFACTURER IDENTIFICATION
Woven Orthopedic Technologies 63 Center Street #3a, Manchester, CT 06040 Phone: 860-259-1260 Establishment Registration #3017505709 Contact Person: Brandon Bendes, President and CEO
# DEVICE IDENTIFICATION
| Device Trade Name: | Ogmend® Implant Enhancement System |
|------------------------|------------------------------------|
| Device Common Name: | Screw Sleeve Bone Fixation Device |
| Classification Name: | Screw Sleeve Bone Fixation Device |
| Classification Number: | 21 CFR 888.3043 |
| Product Code: | QVI |
| Device Class: | Class II |
| Review Panel: | Orthopedic Devices |
| Date Summary Prepared: | September 28, 2023 |
# INDICATIONS FOR USE
The Ogmend® Implant Enhancement System is intended to augment pedicle screw fixation when a screw has lost purchase due to screw loosening, back-out, or breakage. The Ogmend® Implant Enhancement System is for use with rigid, thoracolumbar pedicle screw systems where a screw is inserted posteriorly, through the pedicle, and into the vertebral body. The Ogmend® Implant Enhancement System is for use in skeletally mature patients.
### DEVICE DESCRIPTION
The Ogmend® Implant Enhancement System is a sterile, single-use device intended to provide supplemental fixation to restore stability when the screw-to-bone interface becomes mechanically compromised. The Ogmend® Implant Enhancement System includes a permanently implanted sleeve ("Ogmend® Implant") and an inserter ("Ogmend® Inserter").
The Ogmend® Implant Enhancement System is designed to use the principles of interference fit, surface area contact, and pressure distribution to secure a screw to bone and achieve stability at the screw-to-bone interface to allow for subsequent healing.
The Ogmend® Implant is manufactured from polyethylene terephthalate (PET) monofilament fibers.
The Ogmend® Implant is offered in two sizes:
- Medium: For use with screws ranging in diameter from 3.5mm - 6.5mm.
- Large: For use with screws ranging in diameter from 6.5mm 10.5mm. ●
{4}------------------------------------------------
K233223 - Page 2 of 3
The Ogmend® Inserter is used to insert the Ogmend® Implant into the bone hole and can be used with both size Ogmend® Implants. The Ogmend® Inserter is a stainless-steel rod with beveled ends. It can be used to insert multiple sleeves within one operative procedure. It is provided sterile as a disposable, single-use device. The Inserter is compatible with sterilization by gamma radiation.
# PRIMARY PREDICATE DEVICE
Ogmend® Implant Enhancement System (K223075).
# Comparison of Technological Characteristics to Predicate
The subject Ogmend® Implant is identical to the primary predicate with respect to materials, design, principle of operation, sterilization method, inserter material, and shelf life as detailed below. The subject and predicate devices are also identical with respect to Intended Use and Indications for use.
| Implant Material: | Polyethylene Terephthalate |
|-------------------------|--------------------------------|
| Implant Design: | Braided PET Sleeve Implant |
| Principle or Operation: | Enhance Screw / Bone Interface |
| Sterilization Method: | Gamma Irradiation |
| Inserter Material: | Stainless Steel |
| Shelf Life: | 36 Months |
The distinguishing characteristics between the subject and predicate devices are detailed below:
| Technological<br>Characteristic | Medium Ogmend® Implant<br>(K223075) | Large Ogmend® Implant<br>(proposed) |
|----------------------------------------|-------------------------------------|-------------------------------------|
| Sleeve Diameter | 6.5mm | 8.5mm |
| Sleeve Length | 100mm | 150mm |
| Compatible<br>Pedicle Screw Size Range | 3.5mm - 6.5mm | 6.5mm - 10.5mm |
# PERFORMANCE DATA
The bench tests listed below were performed to support the substantial equivalence of the subject device in addition to the bench tests were previously purposed on the primary predicate in K223075:
| # | Performance Test |
|---|------------------------------------|
| 1 | Screw Axial Pullout |
| 2 | Sleeve Insertion Force |
| 3 | Screw Insertion Torque |
| 4 | Screw Removal/Extraction Torque |
| 5 | Sleeve End Closure Strength |
| 6 | Device Breakage- Sleeve Durability |
Woven Orthopedic Technologies, LLC | 63 E. Center Street, Manchester CT USA 06040 | +1 860.259.1260
{5}------------------------------------------------
K233223 - Page 3 of 3
# SUBSTANTIAL EQUIVALENT
The data presented in this submission demonstrates that the proposed Ogmend® Implant is substantially to intended use, technological characteristics, design, materials, principles of operation, and procedural steps, the pre-determined acceptance criteria, performance with respect to intended use.
Woven Orthopedic Technologies, LLC | 63 E. Center Street, Manchester CT USA 06040 | +1 860.259.1260
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.