K233196 · JLK, Inc. · QIH · Jun 21, 2024 · Radiology
Device Facts
Record ID
K233196
Device Name
Medihub Prostate
Applicant
JLK, Inc.
Product Code
QIH · Radiology
Decision Date
Jun 21, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Prostate gland segmentation
Deep learning model
Dice coefficient > 0.894
Dice coefficient 0.928
Training: Korea (n=748) and US (n=709); Validation: Korea (n=80) and US (n=136)
—
Stand-alone performance US (n=114)
3 (expert radiologists)
Prostate gland segmentation
Deep learning model
Dice coefficient > 0.894
Improved Dice coefficient for three radiologists by 0.156, 0.011, and 0.008
Training: Korea (n=748) and US (n=709); Validation: Korea (n=80) and US (n=136)
—
Reader performance US (n=73)
3 (expert radiologists)
Indications for Use
MEDIHUB PROSTATE is an image processing software package that performs outlining, processing, viewing, and editing of prostate MR images. The software can support study review and analysis of prostate MR data with computed modules. The analysis result can be presented in various formats, including images overlaid onto source MR images and a structured report. MEDIHUB PROSTATE semi-automatically outlines the prostate based on MR images by contour, and it requires the user to edit with image manipulating tools and confirm the final result. The package provides additional functionalities including registered multiparametric-MRI viewing and combining MR sequences into a single image to support visualization. Edited PI-RADS report and semi annotated prostate region can be viewed in each single image in the final report. MEDIHUB PROSTATE is intended to be used by trained radiologists and urologists. Patient management decisions should not be made solely based on the analysis performed by MEDIHUB PROSTATE.
Device Story
MEDIHUB PROSTATE is a web-based image processing software for multi-parametric prostate MRI analysis. It takes DICOM-compliant T2-weighted, DWI, ADC, and DCE MRI scans as input. The device uses a deep learning-based algorithm to semi-automatically segment the prostate volume; users must review, edit, and approve the output. It provides tools for multiparametric MRI registration, ROI analysis, PSA density calculation, and PI-RADS scoring. The device is used in hospital settings by radiologists and urologists. Output is presented as overlaid images and structured reports. The device assists in clinical interpretation; it does not perform fully automatic delineation or lesion segmentation. The user remains responsible for final clinical decisions. Benefits include improved accuracy and efficiency in prostate region segmentation compared to manual methods.
Clinical Evidence
Clinical testing evaluated the stand-alone performance of the prostate segmentation algorithm on 114 T2 MR images from the University of Missouri Hospital. The primary endpoint was the Dice coefficient compared to ground truth established by three expert radiologists. Results showed a mean Dice coefficient of 0.928 (95% CI: 0.925, 0.931) and a Hausdorff distance of 2.171 (95% CI: 1.097, 3.245). Reader performance testing demonstrated that the algorithm improved the segmentation accuracy of all three radiologists. Performance was consistent across age and race subgroups.
Technological Characteristics
Software-based image processing system; DICOM compatible; web-browser accessible. Employs deep learning for semi-automatic prostate segmentation. Complies with ISO 14971 (risk management), IEC 62304 (software lifecycle), and IEC 62366 (usability). Validated for Siemens 3T Vida/Skyra MRI systems with 3.5-5mm slice thickness.
Indications for Use
Indicated for trained radiologists and urologists to perform outlining, processing, viewing, and editing of prostate MR images for study review and analysis. Validated for use with Siemens 3T Vida and Skyra MRI machines, T2 MRI series (3.5 mm to 5 mm slice thickness), and patients aged 55 years and above.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line.
June 21, 2024
JLK Inc. % John Smith Partner Hogan Lovells US LLP 555 13th Street, NW Washington, District of Columbia 20004
Re: K233196
Trade/Device Name: Medihub Prostate Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: May 10, 2024 Received: May 10, 2024
Dear John Smith:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel M. Krainak, PhD Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233196
Device Name MEDIHUB PROSTATE
#### Indications for Use (Describe)
MEDIHUB PROSTATE is an image processing software package that performs outlining, processing, viewing, and editing of prostate MR images. The software can support study review and analysis of prostate MR data with computed modules. The analysis result can be presented in various formats, including images overlaid onto source MR images and a structured report.
MEDIHUB PROSTATE semi-automatically outlines the prostate based on MR images by contour, and it requires the user to edit with image manipulating tools and confirm the final result. The package provides additional ties including registered multiparametric-MRI viewing and combining MR sequences into a single image to support visualization. Edited PI-RADS report and semi annotated prostate region can be viewed in each single image in the final report.
MEDIHUB PROSTATE is intended to be used by trained radiologists. Patient management decisions should not be made solely based on the analysis performed by MEDIHUB PROSTATE.
Limitations:
* MEDIHUB PROSTATE has been validated for use with Siemens 3T Vida and Skyra MRI machines.
* MEDIHUB PROSTATE is also designed for use with the Siemens 3T T2 MRI series, supporting slice thicknesses ranging from 3.5 mm to 5 mm.
* MEDIHUB PROSTATE has been tested on patients aged 55 years and above.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY JLK's MEDIHUB PROSTATE K233196
#### SUBMITTER -
JLK Inc. 5, Teheran-ro 33-gil, Gangnam-gu, Seoul, Republic of Korea Phone: (+82) 02 6925 6189 FAX: (+82) 0505 300 4051 Contact person / E-mail: Dongmin Kim / dmkim@jlkgroup.com Date Prepared: June 14, 2024
# II. DEVICE
Name of Device: MEDIHUB PROSTATE Common or Usual Name: MEDIHUB PROSTATE Classification Name: Medical imaqe management and processing system Regulatory Class: II Product Code: QIH Classification regulation: 21 CFR 892.2050
## III. PREDICATE DEVICE
Device: Quantib Prostate Manufacturer: Quantib B.V. 510(k) No: K202501 Product Code: LLZ Regulatory Class: II
# IV. DEVICE DESCRIPTION
MEDIHUB PROSTATE is an image processing software package for multi-parametric prostate MR image analysis. The analysis may assist trained radiologists in clinical interpretation of prostate MR studies. It can be accessed through a web browser, and provides the following main features:
- . A semi-automatic processing module that outlines the prostate region and performs multiparametric MRI image registration.
- . A user-interaction module in which the user can edit and approve the computed prostate outline and determine PSA density using serum PSA level.
- . A user-interaction module in which the user can view multi-parametric MRI images, and outline and analyze ROIs. This extension will also apply a mathematical operation on the input images to combine information from another MRI sequences into a single combination image.
{4}------------------------------------------------
- A semi-automatic processing module that collects all results for exporting and transferring back to the user.
All measurements are manual except for the prostate volume, which is semi-automatic and requires user review. The method for measuring prostate volume is straightforward and unaffected by patient demographics. Users have the option to outline the prostate either completely manually or with semi-automatic assistance. This device is not intended to be used for fully automatic prostate delineation. It does not involve any segmentation functions by itself. The Al functionality is limited to assessing the total prostate volume, without segmenting lesions.
In semi-automatic mode, our device employs an Al-based algorithm to initially outline the prostate volume, and then it requires the user to edit, review and approve. Additionally, the device calculates the total prostate volume. However, users are responsible for performing all other image annotations and measurements manually. This implies that the final decision should be confirmed by the user, and the user should not rely solely on the device's analysis.
Additional annotations and measurements: all calculated manually
- PI-RADS
- Location of seminal vesicles
- . Prostate zones
- . DWI and DCE graphs
# V. INDICATIONS FOR USE
MEDIHUB PROSTATE is an image processing software package that performs outlining, processing, viewing, and editing of prostate MR images. The software can support study review and analysis of prostate MR data with computed modules. The analysis result can be presented in various formats, including images overlaid onto source MR images and a structured report.
MEDIHUB PROSTATE semi-automatically outlines the prostate based on MR images by contour, and it requires the user to edit with image manipulating tools and confirm the final result. The package provides additional functionalities including registered multiparametric-MRI viewing and combining MR sequences into a single image to support visualization. Edited PI-RADS report and semi annotated prostate region can be viewed in each single image in the final report.
MEDIHUB PROSTATE is intended to be used by trained radiologists and urologists. Patient management decisions should not be made solely based on the analysis performed by MEDIHUB PROSTATE.
Limitations:
- . MEDIHUB PROSTATE has been validated for use with Siemens 3T Vida and Skyra MRI machines.
- . MEDIHUB PROSTATE is also designed for use with the Siemens 3T T2 MRI series, supporting slice thicknesses ranging from 3.5 mm to 5 mm.
- . MEDIHUB PROSTATE has been tested on patients aged 55 years and above.
{5}------------------------------------------------
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
MEDIHUB PROSTATE shares similar indications for use, technological characteristics, and principles of operation with its predicate device. At a high level, both the subject device and predicate device are based on the same technological elements as follows:
| Item | MEDIHUB PROSTATE | Quantib Prostate<br>(K202501) | Substantial<br>Equivalence |
|------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Regulatory Data | | | |
| Regulatory<br>Class | Class II | Class II | Equivalent |
| Classification<br>Name | Medical image management<br>and processing system | Picture archiving and<br>communications system | Equivalent1 |
| Regulation<br>Number | 21 CFR 892.2050 | 21 CFR 892.2050 | Equivalent |
| Product Code | QIH | LLZ | Equivalent |
| Use | | | |
| Target users | Trained medical professionals | Trained medical professionals | Equivalent |
| Anatomical site | Prostate | Prostate | Equivalent |
| Where used | Hospital | Hospital | Equivalent |
| Workflow<br>Considerations | User can approve or reject results User can inspect and edit prostate segmentation User can enter PSA value and compute PSA density User can create and update regions of interest (ROIs) User can add and update annotations (text) to analysis User can manually set a PI-RADS score using a PI-RADS interactive worksheet | User can approve or reject results User can inspect and edit prostate segmentation User can enter PSA value and compute PSA density User can create and update regions of interest (ROIs) User can add and update annotations (text) to analysis User can manually set a PI-RADS score using a PI-RADS interactive worksheet | Equivalent |
| Item | MEDIHUB PROSTATE | Quantib Prostate<br>(K202501) | Substantial<br>Equivalence |
| Design | • Semi-automatic prostate<br>segmentation<br>• Multi parametric image<br>review<br>• Biparametric combination<br>image viewer<br>• Basic image manipulation<br>tools, Thresholding tool<br>• User interface to<br>create/update user generated<br>ROIs | • Semi-automatic prostate<br>segmentation<br>• Multi parametric image<br>review<br>• Biparametric combination<br>image viewer<br>• Basic image manipulation<br>tools, Thresholding tool<br>• User interface to<br>create/update user generated<br>ROIs | Equivalent |
| Non-clinical<br>performance | • Bench testing performed to<br>test the functionality of the<br>system<br>• This included<br>characterization of the<br>standalone performance of<br>the prostate segmentation<br>algorithm. | • Bench testing performed to<br>test the functionality of the<br>system<br>• This included<br>characterization of the<br>standalone performance of<br>the prostate segmentation<br>algorithm. | Equivalent |
| Clinical<br>performance | • Prostate segmentation<br>algorithm was tested in a<br>clinical use context<br>• i.e. as a semiautomatic<br>algorithm after correction by<br>trained clinicians. | • Prostate segmentation<br>algorithm was tested in a<br>clinical use context<br>• i.e. as a semiautomatic<br>algorithm after correction by<br>trained clinicians. | Equivalent |
| Standards met | • ISO 14971 – Medical devices<br>-Application of risk<br>management to medical<br>devices<br>• IEC 62304 – Medical device<br>software - Software life cycle<br>processes<br>• IEC 62366 – Medical<br>devices - Application of<br>usability engineering to<br>medical devices | • ISO 14971 – Medical devices<br>-Application of risk<br>management to medical<br>devices<br>• IEC 62304 – Medical device<br>software - Software life cycle<br>processes<br>• IEC 62366 – Medical<br>devices - Application of<br>usability engineering to<br>medical devices | Equivalent |
| SW verification<br>and validation | • Tested in accordance with<br>verification and validation<br>processes and planning<br>• The testing results support<br>that all the system<br>requirements have met their<br>acceptance criteria and are<br>adequate for its intended use | • Tested in accordance with<br>verification and validation<br>processes and planning<br>• The testing results support<br>that all the system<br>requirements have met their<br>acceptance criteria and are<br>adequate for its intended use | Equivalent |
| Item | MEDIHUB PROSTATE | Quantib Prostate<br>(K202501) | Substantial<br>Equivalence |
| Compatibility<br>with the<br>environment<br>and other<br>devices | DICOM compatible | DICOM compatible | Equivalent |
| Reported<br>measures | Prostate gland Volume, PSA<br>density, ROIs within the<br>prostate, with for each ROI:<br>• Volume<br>• Average ADC value<br>• PI-RADS score<br>• Location | Prostate gland volume, PSA<br>density, ROIs within the<br>prostate, with for each ROI:<br>• Volume<br>• Average ADC value<br>• PI-RADS score<br>• Location | Equivalent |
| Required input | DICOM compatible data, type<br>of MRI scans: T2-weighted,<br>DWI, ADC, DCE | DICOM compatible data, type<br>of MRI scans: T2-weighted,<br>DWI, ADC, DCE | Equivalent |
#### Substantial Equivalence Chart
<sup>1</sup> The name of the generic devices regulated under 21 C.F.R. 892.2050 has been changed from Picture archiving and communications system to Medical image management and processing system as part of FDA's implementation of the 21st Century Cures Act.
{6}------------------------------------------------
{7}------------------------------------------------
#### VII. PERFORMANCE DATA
#### Data Description
Image /page/7/Figure/3 description: The image shows a flow chart of the study population dataset and the clinical test dataset. The study population dataset includes data from Korea (n=908), US (n=981), and ProstateX (n=19). The training dataset includes data from Korea (n=828) and US (n=845), which is further divided into training dataset Korea (n=748) and US (n=709) and validation dataset Korea (n=80) and US (n=136). The clinical test dataset includes stand-alone performance US (n=114) and reader performance US (n=73 from stand-alone performance).
Study Population Dataset: The Study Population Dataset is the same as algorithm development dataset. The dataset consists of three sets, Korea, US (the University of Missouri Health Care), and ProstateX (open dataset is split into Training/Validation/Testing dataset. In Phase I, a Stand-alone Performance test was conducted, and in Phase II, a Reader Performance test was conducted. In the Study Population Dataset, the Training/Validation datasets and the Testing dataset were stratified at a ratio of 9:1. Additionally, within the Training/Validation split, the Training dataset and the Validation dataset were further stratified at a ratio of 9:1. The data used
{8}------------------------------------------------
to demonstrate performance of the device included Siemens 3T Vida and Skyra systems with slice thickness ranging from 3.5 to 5 mm.
## Algorithm Performance Testing
A performance check test was conducted on the prostate region segmentation algorithm of MEDIHUB PROSTATE. The performance was obtained by comparing the ground truth produced by three expert radiologists and the results of the prostate region segmentation algorithm. The performance was measured by the mean value of Dice coefficient and Hausdorff Distance of the test group. The results were obtained and calculated by computer coding, and the score for each case was recorded. Utilizing the test datasets of Korea patients, and ProstateX (open dataset), the performance of the prostate region segmentation function was evaluated. Calculating the Dice coefficient between ground-truth and algorithm prediction vields the performance. The passffail criterion for each dataset is whether the mean Dice coefficient exceeds 0.894, which is determined by analyzing the results of the most recent state-of-the-art prostate region segmentation algorithm published in the past 5 years; this value is the average of performance in articles whose performance is greater than the average of the entire references. The value is calculated as the lower limit of 99%, confidence interval based on the performances of whole reference. As a consequence, the criteria were met by all three test datasets.
#### Clinical Testing (Stand-alone Performance)
The purpose of this study is to verify the stand-alone performance of the prostate region segmentation algorithm. MEDIHUB PROSTATE was developed by applying a deep learning technique on T2 prostate MRI. The clinical test investigator collected data that met the inclusion criteria and no subjects met the exclusion criteria. Based on a total of 114 T2 MR images (University of Missouri Hospital) the final validity evaluation was performed on prostate region seqmentation ground truth produced by three expert radiologists. In this clinical test, it was confirmed that mean Dice coefficient was higher than the predefined primary efficacy evaluation criterion. The clinical testing results demonstrated that the overall Dice coefficient and Hausdoff distance were 0.928 and 2.171, respectively, with the 95% confidence intervals for these measurements being [0.925, 0.931] for the Dice coefficient and [1.097, 3.245] for the Hausdoff distance.
#### Ground Truthing Process
The ground truthing was conducted by expert-level radiologists. They independently annotated the prostate images, and these annotations were then consolidated into a definitive ground truth through a majority rule approach. The rationale for employing consensus among three radiologists, resolved through discussion and mutual agreement in cases of ties, ensures a reliable and unbiased representation of the prostate, crucial for the accurate clinical performance evaluation of our device.
| Gender, n(%) | Male, 114 (100%) |
|--------------|----------------------------------|
| Age | Mean ± SD<br>66.38 ± 5.33 |
| Age | Median (Min, Max)<br>66 (57, 78) |
#### Clinical Testing Dataset Distribution
#### Demographic information on subjects (N=114)
{9}------------------------------------------------
| Age group, n (%) | ട് -------------------------------------------------------------------------------------------------------------------------------------------------------------------------- | 14 (12.3%) |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| | 60~69 | 63 (55.3%) |
| | 70~79 | 37 (32.5%) |
| | White | 106 (92.9%) |
| Races | African American | 7 (6.1%) |
| | Asian | 1 (0.9%) |
* As indicated in a paper by Mavropoulos (2007)², there is no significant association between race and prostate size. Given that race or ethnicity does not impact measuring prostate area or volume.
# 95% confidence interval of the dice coefficient and Hausdorff distance calculated by comparing ground truth and the result of algorithms by race
| Race | Dice Coefficient | Hausdorff Distance |
|---------------------|--------------------------|--------------------------|
| White | 0.929<br>*[0.925, 0.933] | 2.173<br>*[2.066, 2.280] |
| African<br>American | 0.925<br>*[0.902, 0.949] | 2.085<br>*[1.711, 2.459] |
| Asian | 0.917<br>*[nan, nan] | 2.929<br>*[nan, nan] |
* 95% Confidence Interval
#### 95% confidence interval of the dice coefficient and Hausdorff distance calculated by comparing ground truth and the result of algorithms by age
| Age | Dice Coefficient | Hausdorff Distance |
|------|--------------------------|--------------------------|
| < 60 | 0.938<br>*[0.925, 0.950] | 1.871<br>*[1.638, 2.103] |
| ≥ 60 | 0.927<br>*[0.923, 0.931] | 2.224<br>*[2.114, 2.335] |
* 95% Confidence Interval
#### Clinical Testing (Reader Performance)
The purpose of this study is to compare the reader performance with and without the prostate region segmentation algorithm of MEDIHUB PROSTATE. Dice coefficients of the investigators' performances are calculated by comparing the ground truth of the prostate region segmentation with each investigator's segmentation results. The results were better when all three expert radiologists were with aid of the prostate region segmentation algorithm of MEDIHUB PROSTATE. And regardless of by race and age, the results were better in all cases with aid of the prostate region segmentation algorithm of MEDIHUB PROSTATE. The improvements in the Dice coefficient
<sup>6</sup> Mavropoulos, J. C., Partin, A. W., Amling, C. L., Terris, M. K., Kane, C. J., Aronson, W. J., .. & Freedland, S. J. (2007). Do racial differences in prostate size explain higher serum prostate-specific antigen concentrations among black men'?. Urology, 69(0), 1138-1142.
{10}------------------------------------------------
for prostate outlining performance for the three radiologists when using the prostate region segmentation algorithm of MEDIHUB-PROSTATE were 0.156, 0.011, and 0.008, respectively.
| Properties | Value |
|-------------------------|----------------------------------------|
| Sites | University of Missouri Health Care |
| Field Strengths | 3T |
| MR Scanner Manufacturer | SIEMENS, Siemens HealthCare GmbH |
| Thickness | 3.5mm, 5mm |
| Purpose of Al | Prostate region semi auto-segmentation |
Properties of Performance Testing Cases
#### Safety Implications
MEDIHUB PROSTATE is intended to support trained radiologists. The software does not make diagnoses; it provides quantification results that must be interpreted by a trained radiologists and urologists before using the device. MEDIHUB PROSTATE is both software applications running on off-the-shelve hardware. In conclusion, MEDIHUB PROSTATE does not introduce any new safety issues compared to the predicate device Quantib Prostate.
#### VIII. DISCUSSION & CONCLUSIONS
#### Discussion
MEDIHUB PROSTATE employs a deep learning model to semi-automatically outline the prostate region. Its performance is evaluated using the Dice coefficient against ground-truth data from Korea patients, US patients, and an open dataset (ProstateX). The pass/fail criterion, set at a mean Dice coefficient of 0.894, is determined by recent state-of-the-art algorithms. Results show that all three test datasets meet this criterion, with Dice coefficients surpassing 0.866 in every case, demonstrating substantial equivalence.
#### Conclusions
MEDIHUB PROSTATE is semi-automatic image processing software. Clinical test results indicate its safety as it does not affect patients according to the software's results. By utilizing the semiautomatic drawing function of the software, it demonstrates efficiency in improving the accuracy of user's prostate region segmentation, making it as safe and effective as the Quantib Prostate. MEDIHUB PROSTATE has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. In addition, the minor technological differences between the MEDIHUB PROSTATE and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the MEDIHUB PROSTATE is as safe and effective as the Quantib Prostate. Thus, the MEDIHUB PROSTATE is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.