← Product Code [FLL](/productcode/FLL) · K233192

# Wireless Smart Thermometer (K233192)

_Famidoc Technology Company Limited · FLL · Apr 5, 2024 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K233192

## Device Facts

- **Applicant:** Famidoc Technology Company Limited
- **Product Code:** [FLL](/productcode/FLL.md)
- **Decision Date:** Apr 5, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2910
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Pediatric

## Indications for Use

Wireless Smart Thermometer is intended to be used at home for the intermittent measurement and monitoring of human body temperature orally, rectally, and under the arm. The devices are reusable for the adult and pediatric patient population (not suitable for neonates).

## Device Story

Wireless Smart Thermometer (multiple models) uses thermistor-based temperature measurement module to collect electrical signals; signals processed via AD conversion and algorithms to determine body temperature. Device features buzzer for audible notifications, display module (LED/LCD), battery voltage management, and button interface. Wireless function (BLE) transfers data to mobile devices. Used at home by patients or caregivers for intermittent monitoring. Output displayed on device and transmitted to mobile app; assists users in tracking temperature trends and identifying high body temperature (fever). Benefits include convenient, reusable home monitoring of body temperature.

## Clinical Evidence

Clinical study conducted per ISO 80601-2-56:2017 with 105 volunteer subjects. Primary endpoint was measurement accuracy compared to standard requirements. Results demonstrated accuracy within acceptable scope defined by ISO 80601-2-56. No adverse events reported.

## Technological Characteristics

Materials: TPE, ABS, stainless steel (biocompatible per ISO 10993-5, -10). Sensing: Thermistor. Energy: CR1632 battery. Connectivity: Bluetooth Low Energy (BLE). Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home use), ISO 80601-2-56 (performance). Software: Moderate level of concern.

## Regulatory Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

## Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; 
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and 
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

## Predicate Devices

- Kinsa QuickCare Thermometer ([K173730](/device/K173730.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services - USA logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 5, 2024

Famidoc Technology Company Limited Amos Zou Management Representative No.212 Yilong Road, Hexi Industrial Zone, Jinxia, Changan Town Dongguan, Guangdong 523853 China

Re: K233192

Trade/Device Name: Wireless Smart Thermometer (Model: FDTH3401, FDTH3402, FDTH3403, FDTH3404, FDTH3405, FDTH3406, FDTH3407, FDTH3408, FDTH3409, FDTH3410, FDTH3411, FDTH-V0-13, FDTH-V0-4) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: March 7, 2024 Received: March 7, 2024

Dear Amos Zou:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Porsche Bennett

Porsche Bennett for David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and

{2}------------------------------------------------

General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K233192

#### Device Name

Wireless Smart Thermometer (Model: FDTH3400, FDTH3402, FDTH3403, FDTH3404, FDTH3405, FDTH3406, FDTH3407, FDTH3408, FDTH3409, FDTH3410, FDTH3411, FDTH-V0-13, FDTH-V0-4)

#### Indications for Use (Describe)

Wireless Smart Thermometer is intended to be used at home for the intermittent measurement and monitoring of human body temperature orally, rectally, and under the arm. The devices are reusable for the adult and pediatic patient population (not suitable for neonates).

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(K) SUMMARY -K233192

This summary of 510(K) substantial equivalence information is being submitted in accordance with the requirement 21 CFR 807.92.

## 1. Submitter of 510(K):

| Company Name:     | Famidoc Technology Company Limited                                                             |
|-------------------|------------------------------------------------------------------------------------------------|
| Address:          | No.212 Yilong Road, Hexi Industrial Zone, Jinxia, Changan Town<br>Dongguan Guangdong 523853 CN |
| Contact person:   | Amos Zou                                                                                       |
| TEL:              | +86 -769-89272488                                                                              |
| E-mail:           | qa@famidoc.com                                                                                 |
| Date of Prepared: | April 4, 2024                                                                                  |

## 2. Subject Device:

| Device Trade Name: | Wireless Smart Thermometer (Model:<br>FDTH3400, FDTH3401, FDTH3402,<br>FDTH3403, FDTH3404,<br>FDTH3405, FDTH3406, FDTH3407, FDTH3408,<br>FDTH3409, FDTH3410,<br>FDTH3411, FDTH-V0-13, FDTH-V0-4) |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Name    | Clinical electronic thermometer                                                                                                                                                                  |
| Product Code:      | FLL                                                                                                                                                                                              |
| Regulation number  | 21 CRF 880.2910                                                                                                                                                                                  |
| Device Class       | II                                                                                                                                                                                               |

## 3. Predicate Device:

| 510(K)            | K173730                         |
|-------------------|---------------------------------|
| Trade Name        | Kinsa QuickCare Thermometer     |
| Manufacturer      | KINSA, Inc.                     |
| Regulation Name   | Clinical electronic thermometer |
| Product Code:     | FLL                             |
| Regulation number | 21 CRF 880.2910                 |
| Device Class      | II                              |

## 4. Description of Subject Device:

Wireless Smart Thermometer (Model: FDTH3400, FDTH3401, FDTH3402,

FDTH3403, FDTH3404, FDTH3405, FDTH3406, FDTH3407, FDTH3408, FDTH3409,

FDTH3410, FDTH3411, FDTH-V0-13, FDTH-V0-4) utilizes a modular design method and consists of six main modules:

- Buzzer module: Control the product buzzer to produce audible notifications;
- Display module: Display driver content through notifications, and combine to produce various display interfaces;

{5}------------------------------------------------

- Battery voltage management module: Detect the battery power of the product and output the battery power level;
- Temperature measurement module: After collecting the electrical signal of the external temperature sensor through AD, it is restored to the temperature value through various algorithms
- Button module: This module is used to identify whether the product button is active;
- Wireless function module: Transfer the measurement data to the mobile device.

## 5. Indications for Use

Wireless Smart Thermometer is intended to be used at home for the intermittent measurement and monitoring of human body temperature orally, rectally, and under the arm. The devices are reusable for the adult and pediatric patient population (not suitable for neonates).

## 6. Technical Characteristic and Substantial Equivalence

The following table compares the technological characteristics between the subject device to the predicate device.

{6}------------------------------------------------

| 7. Comparison of Technological Characteristics with Predicate Device |  |
|----------------------------------------------------------------------|--|
|----------------------------------------------------------------------|--|

| Elements of<br>Comparison            | Subject Devices                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                   |                                             | Predicate Device                                                                                                                                                                                                                     | Comparison<br>Result |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Device Name &<br>510k Number         | Wireless Smart Thermometer (Model: FDTH3400, FDTH3401, FDTH3402, FDTH3403, FDTH3404,<br>FDTH3405, FDTH3406, FDTH3407, FDTH3408, FDTH3409, FDTH3410,<br>FDTH3411, FDTH-V0-13, FDTH-V0-4)                                                                                                                           |                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                   |                                             | Kinsa QuickCare<br>Thermometer- K173730                                                                                                                                                                                              | /                    |
| Device Model                         | FDTH3400~FDTH3405                                                                                                                                                                                                                                                                                                 | FDTH-V0-13                                                                                                                                                                                                                                                                                                        | FDTH3406~FDTH3411                                                                                                                                                                                                                                                                                                       | FDTH-V0-4                                                                                                                                                                                                                                                                                                         |                                             | KSA-110                                                                                                                                                                                                                              | /                    |
| Intended Use and Indications for Use |                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                   |                                             |                                                                                                                                                                                                                                      |                      |
| Intended Use                         | Wireless Smart<br>Thermometer is<br>intended to be used at<br>home for the intermittent<br>measurement and<br>monitoring of human<br>body temperature orally,<br>rectally, and under the<br>arm. The devices are<br>reusable for the adult<br>and pediatric patient<br>population (not suitable<br>for neonates). | Wireless Smart<br>Thermometer is<br>intended to be used at<br>home for the intermittent<br>measurement and<br>monitoring of human<br>body temperature orally,<br>rectally, and under the<br>arm. The devices are<br>reusable for the adult<br>and pediatric patient<br>population (not suitable<br>for neonates). | Wireless Smart<br>Thermometer is<br>intended to be used at<br>home for the<br>intermittent<br>measurement and<br>monitoring of human<br>body temperature<br>orally, rectally, and<br>under the arm. The<br>devices are reusable<br>for the adult and<br>pediatric patient<br>population (not suitable<br>for neonates). | Wireless Smart<br>Thermometer is<br>intended to be used at<br>home for the intermittent<br>measurement and<br>monitoring of human<br>body temperature orally,<br>rectally, and under the<br>arm. The devices are<br>reusable for the adult<br>and pediatric patient<br>population (not suitable<br>for neonates). |                                             | The Kinsa QuickCare<br>Thermometer is used for<br>the intermittent<br>measurement and<br>monitoring of human body<br>temperature, orally,<br>rectally and under the arm.<br>The device is for the adult<br>and pediatric population. | Similar              |
| Performance Specification            |                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                   |                                             |                                                                                                                                                                                                                                      |                      |
| Power supply<br>voltage              | CR1632 battery                                                                                                                                                                                                                                                                                                    | CR1632 battery                                                                                                                                                                                                                                                                                                    | CR1632 battery                                                                                                                                                                                                                                                                                                          | CR1632 battery                                                                                                                                                                                                                                                                                                    |                                             | Battery powered CR2032<br>(3V)                                                                                                                                                                                                       | Similar              |
| Display module                       | LED (No backlight)                                                                                                                                                                                                                                                                                                | LED (No backlight)                                                                                                                                                                                                                                                                                                | LCD (No backlight)                                                                                                                                                                                                                                                                                                      | LCD (No backlight)                                                                                                                                                                                                                                                                                                |                                             | LCD (No backlight)                                                                                                                                                                                                                   | Similar<br>Note 1    |
| Measuring                            | Adjusted mode / Direct                                                                                                                                                                                                                                                                                            | Adjusted mode / Direct                                                                                                                                                                                                                                                                                            | Adjusted mode                                                                                                                                                                                                                                                                                                           | Adjusted mode                                                                                                                                                                                                                                                                                                     |                                             | Adjusted mode                                                                                                                                                                                                                        | Similar              |
| Method                               | mode                                                                                                                                                                                                                                                                                                              | mode                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                   |                                             | Note 2                                                                                                                                                                                                                               |                      |
| Measuring type                       | Axillary / Oral / Rectal                                                                                                                                                                                                                                                                                          | Axillary / Oral / Rectal                                                                                                                                                                                                                                                                                          | Axillary / Oral / Rectal                                                                                                                                                                                                                                                                                                | Axillary / Oral / Rectal                                                                                                                                                                                                                                                                                          | Axillary / Oral / Rectal                    | Same                                                                                                                                                                                                                                 |                      |
| Measuring range                      | $32.00°C ~ 42.90°C$                                                                                                                                                                                                                                                                                               | $32.00°C ~ 42.90°C$                                                                                                                                                                                                                                                                                               | $32.0°C ~ 42.9°C$                                                                                                                                                                                                                                                                                                       | $32.0°C ~ 42.9°C$                                                                                                                                                                                                                                                                                                 | $32°C-42.8°C$                               | Similar                                                                                                                                                                                                                              |                      |
|                                      | $(89.60°F~109.22°F)$                                                                                                                                                                                                                                                                                              | $(89.60°F~109.22°F)$                                                                                                                                                                                                                                                                                              | $(89.6°F~109.2°F)$                                                                                                                                                                                                                                                                                                      | $(89.6°F~109.2°F)$                                                                                                                                                                                                                                                                                                | $(89.6°F-109.2°F)$                          | Note 3                                                                                                                                                                                                                               |                      |
| Measuring<br>accuracy                | $±0.10°C$ :<br>between<br>$35.00°C~42.00$<br>$°C$<br>$(±0.18°F$ between<br>$95.00°F~107.60°F)$<br>$±0.20°C(or±0.36°F)$ :<br>outside this ranges                                                                                                                                                                   | $±0.10°C$ : between<br>$35.00°C~42.00°C$<br>$(±0.18°F$ between<br>$95.00°F~107.60°F)$<br>$±0.20°C(or±0.36°F)$ :<br>outside this ranges                                                                                                                                                                            | $±0.3°C$ : between<br>$34.0°C~42.0°C$<br>$(±0.6°F$ between<br>$93.2°F~107.6°F)$<br>$±0.4°C(or±0.8°F)$ : outside<br>this ranges                                                                                                                                                                                          | $±0.3°C$ : between<br>$34.0°C~42.0°C$<br>$(±0.6°F$ between<br>$93.2°F~107.6°F)$<br>$±0.4°C(or±0.8°F)$ : outside<br>this ranges                                                                                                                                                                                    | $±0.2°F(0.1°C),89.6°F-109.2°F(32°C-42.8°C)$ | Different<br>Note 4                                                                                                                                                                                                                  |                      |
| Display resolution                   | $0.01°C(0.01°F)$                                                                                                                                                                                                                                                                                                  | $0.01°C(0.01°F)$                                                                                                                                                                                                                                                                                                  | $0.1°C(0.1°F)$                                                                                                                                                                                                                                                                                                          | $0.1°C(0.1°F)$                                                                                                                                                                                                                                                                                                    | $0.1°C(0.1°F)$                              | Similar<br>Note 5                                                                                                                                                                                                                    |                      |
| Measurement<br>time                  | Adjusted mode: Less than<br>8s (for water bath)<br>Direct mode: Less than<br>50s (for water bath)                                                                                                                                                                                                                 | Adjusted mode: Less than<br>8s (for water bath)<br>Direct mode: Less than<br>50s (for water bath)                                                                                                                                                                                                                 | Less than 8s (for water<br>bath)                                                                                                                                                                                                                                                                                        | Less than 8s (for water<br>bath)                                                                                                                                                                                                                                                                                  | Adjusted mode: 8 seconds<br>nominally       | Similar<br>Note 6                                                                                                                                                                                                                    |                      |
| High body<br>temperature hint        | $≥37.60°C(99.68°F)$                                                                                                                                                                                                                                                                                               | $≥37.60°C(99.68°F)$                                                                                                                                                                                                                                                                                               | $≥37.6°C(99.7°F)$                                                                                                                                                                                                                                                                                                       | $≥37.6°C(99.7°F)$                                                                                                                                                                                                                                                                                                 | $≥37.8°C(100.4°F)$                          | Different<br>Note 7                                                                                                                                                                                                                  |                      |
| Temperature unit                     | °C/°F                                                                                                                                                                                                                                                                                                             | °C/°F                                                                                                                                                                                                                                                                                                             | °C/°F                                                                                                                                                                                                                                                                                                                   | °C/°F                                                                                                                                                                                                                                                                                                             | °C/°F                                       | Same                                                                                                                                                                                                                                 |                      |

{7}------------------------------------------------

{8}------------------------------------------------

| Memory group |                                    | Last time memory display |                                    | Last time memory display | Full temperature history   | Similar |
|--------------|------------------------------------|--------------------------|------------------------------------|--------------------------|----------------------------|---------|
|              | 8 groups of memories<br>(FDTH3401) |                          | 8 groups of memories<br>(FDTH3406) |                          | available on mobile device | Note 8  |
|              | 7 groups of memories<br>(FDTH3402) |                          | 7 groups of memories<br>(FDTH3407) |                          |                            |         |
|              | 6 groups of memories<br>(FDTH3403) |                          | 6 groups of memories<br>(FDTH3408) |                          |                            |         |
|              | 5 groups of memories<br>(FDTH3404) |                          | 5 groups of memories<br>(FDTH3409) |                          |                            |         |
|              | 3 groups of memories<br>(FDTH3405) |                          | 3 groups of memories<br>(FDTH3410) |                          |                            |         |
|              | Last time memory                   |                          | Last time memory                   |                          |                            |         |

{9}------------------------------------------------

|                                      | display(FDTH3400)                                                                                                                                  |                                                                                                                        | display(FDTH3411)                                                                                                                                  |                                                                                                                        |                                                                                                                                  |                   |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Sensor type                          | Thermistor                                                                                                                                         |                                                                                                                        | Thermistor                                                                                                                                         |                                                                                                                        | Same                                                                                                                             |                   |
| Wireless interface                   | Bluetooth Low Energy (BLE)                                                                                                                         |                                                                                                                        | Bluetooth Low Energy (BLE)                                                                                                                         |                                                                                                                        | Same                                                                                                                             |                   |
| Materials                            | TPE, ABS, Stainless Steel                                                                                                                          |                                                                                                                        | TPE, ABS, Stainless Steel                                                                                                                          |                                                                                                                        | Biocompatible metals and resins                                                                                                  | Similar Note 9    |
| OPERATING & STORAGE CONDITIONS       |                                                                                                                                                    |                                                                                                                        |                                                                                                                                                    |                                                                                                                        |                                                                                                                                  |                   |
| Operation condition                  | Temperature:5 $^{\circ}$ C~40 $^{\circ}$ C<br>(41°F~104°F)<br>Relative humidity:15%~95%RH,<br>No condensing<br>Atmospheric pressure:70 kPa~106 kPa |                                                                                                                        | Temperature:5 $^{\circ}$ C~40 $^{\circ}$ C<br>(41°F~104°F)<br>Relative humidity:15%~95%RH,<br>No condensing<br>Atmospheric pressure:70 kPa~106 kPa |                                                                                                                        | 59°F-95°F(15°C-35°C)15 % to 85%<br>R.H non-condensing                                                                            | Similar Note 10   |
| Transportation and storage condition | Temperature:-25°C~55°C<br>(-13°F~131°F)<br>Relative humidity:15%~95%RH,<br>No condensing<br>Atmospheric pressure:70 kPa~106 kPa                    |                                                                                                                        | Temperature:-25°C~55°C<br>(-13°F~131°F)<br>Relative humidity:15%~95%RH,<br>No condensing<br>Atmospheric pressure:70 kPa~106 kPa                    |                                                                                                                        | -13°Fto +95°F(-25°Cto +35°C)at R.H.up to 90%non-condensingand 95°F to 158°F(35°C to 70°C) at a water vapor pressure up to 50 hPa | Similar Note 11   |
| Shelf life                           | 5 years                                                                                                                                            |                                                                                                                        | 5 years                                                                                                                                            |                                                                                                                        | Unknown                                                                                                                          | Different Note 12 |
| COMPLIANCE STANDARDS                 |                                                                                                                                                    |                                                                                                                        |                                                                                                                                                    |                                                                                                                        |                                                                                                                                  |                   |
| Electrical Safety                    | IEC 60601-1                                                                                                                                        |                                                                                                                        | IEC 60601-1                                                                                                                                        |                                                                                                                        | Same                                                                                                                             |                   |
| EMC                                  | IEC 60601-1-2                                                                                                                                      |                                                                                                                        | IEC 60601-1-2                                                                                                                                      |                                                                                                                        | Same                                                                                                                             |                   |
| Home Use                             | IEC 60601-1-11                                                                                                                                     |                                                                                                                        | IEC 60601-1-11                                                                                                                                     |                                                                                                                        | Same                                                                                                                             |                   |
| Performance                          | ISO 80601-2-56                                                                                                                                     |                                                                                                                        | ISO 80601-2-56                                                                                                                                     |                                                                                                                        | Same                                                                                                                             |                   |
| Biocompatibility                     | All the patient<br>contracting materials<br>are evaluated by the<br>biocompatibility<br>standard ISO<br>10993 -5, -10.                             | All the patient<br>contracting materials<br>are evaluated by the<br>biocompatibility<br>standard ISO<br>10993 -5, -10. | All the patient<br>contracting materials<br>are evaluated by the<br>biocompatibility<br>standard ISO<br>10993 -5, -10.                             | All the patient<br>contracting materials<br>are evaluated by the<br>biocompatibility<br>standard ISO<br>10993 -5, -10. | All the patient<br>contracting<br>materials are<br>evaluated by the<br>biocompatibility<br>standard ISO<br>10993 -5, -10.        | Same              |

{10}------------------------------------------------

{11}------------------------------------------------

#### Note 1

The display module of the subject device is different; however, the subject device performance testing demonstrates that the difference does not raise different questions of safety and effectiveness.

#### Note 2

The measuring method of models FDTH3400 and FDTH-V0-13 of the subject device are different than the predicate device. The subject device models FDTH3400 and FDTH-V0-13 measure in direct mode measurement time is longer than the adjustment mode measurement time; however, device accuracy was tested in accordance with the ISO 80601-2-56 Standard. Therefore, the difference does not raise different questions of safety and effectiveness.

#### Note 3 and 4

The Measuring Range and Accuracy of the subject device are different; however, the measuring range and accuracy of subject device are both in compliance with IEC 60601-1 and ISO 80601-2-56. Therefore, this difference does not raise different questions of safety and effectiveness.

#### Note 5

The display resolution of the subject device and predicate device are different as the subject device models FDTH3400 and FDTH-V0-13 have a higher display accuracy, However, this difference does not raise any different questions of safety and effectiveness.

#### Note 6

The measurement time of subject device and predicate device is different as the subject device models, FDTH3400 and FDTH-V0-13 measure in direct mode and therefore the measurement time is longer. Performance testing was adequately conducted in accordance with IEC 60601-1 and ISO 80601-2-56 and demonstrate that the different questions of safety and effectiveness.

#### Note 7

The high body temperature alert is different between the subject and predicate device. However, performance testing demonstrates that the difference does not raise new questions of safety and effectiveness.

#### Note 8

The Memory group of the subject device and predicate device are different. The memory measurement storage is indicated in the user manual and outer carton, The difference does not raise different questions of safety and effectiveness.

#### Note 9

The materials between the subject and predicate device are different; however, biocompatibility testing demonstrates the difference does not raise different questions of safety and effectiveness.

#### Note 10 and Note 11

The Temperature, Relative Humidity and Atmospheric pressure of Operation as well as the storage environment of subject devices is different between the subject and predicate device; 

{12}------------------------------------------------

however, they are both in compliance with IEC 60601-1-11 and therefore, the differences do not raise different questions of safety and effectiveness.

## Note 12

The subject device performance was tested over the shelf life to applicable standards as identified below. Therefore, the difference does not raise different questions of safety and effectiveness.Performance Testing: Performance data includes "Non-Clinical Data" and "Clinical Data", brief description of which are shown as below.

## 7.1 Non-Clinical Data:

The following performance data were provided in support of the substantial equivalence determination.

## 7.2 Biocompatibility testing

The biocompatibility evaluation for the FDTH series Upper arm Blood Pressure Monitor and the NIBP CUFF were conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue contacting for duration of less than 24 hours. And the testing included the following tests:

- Cytotoxicity ●
- Skin Sensitization
- Skin Irritation

## 7.3 Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the FDTH series Upper arm Blood Pressure Monitor, consisting of all the modules and accessories in the system complies with the IEC 60601-1: 2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance for safety and the IEC 60601-1-2: 2014 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances - Requirements and tests standard for EMC.

## 7.4 Bench Testing

Bench testing was conducted on the FDTH series Upper arm Blood Pressure Monitor, consisting of all the accessories in the system. The system complies with the IEC 60601-1-11: 2015 MEDICAL ELECTRICAL EQUIPMENT –Part 1-11: General requirements for basic safety and essential performance -Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment, ISO 80601-2-56: 2017 Medical electrical equipment -- Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.

## 7.5 Software Verification and Validation Testing

Software verification and validation testing were conducted, and documentation was provided as recommended by FDA 's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern.

{13}------------------------------------------------

## 7.6 Clinical data:

Clinical testing is conducted per ISO 80601-2-56: 2017 Medical electrical equipment — Part 2-56:Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.

Based on the same product principle, and the clinical validation data on the FDTH3400 can cover all the models included in this submission.

In this clinical study, 105 patients participated in the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any adverse events. The results showed the accuracy of the subject device is within acceptable scope specified in ISO 80601-2-56.

## 8. Conclusions:

The subject device has the same intended use and similar technological characteristics as the predicate device, Kinsa QuickCare Thermometer cleared under K173730. Clinical testing demonstrate that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. The subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K233192](https://fda-staging.innolitics.com/device/K233192)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).