The Nest VT Vitrification Device is a cryopreservation storage device intended for use in vitrification procedures to contain and maintain human oocytes (MII), 4-8 cell embryos and blastocyst stage embryos.
Device Story
Nest VT Vitrification Device; sterile, single-use cryopreservation storage device. Composed of stick (124.5 mm) and cap (38.5 mm) made of MBS copolymer; features 6 mm loading well for 1.0 microliter volume. Used in assisted reproduction labs by embryologists; oocytes/embryos loaded onto tip, capped, then plunged into liquid nitrogen for storage. Laser-etched markings indicate tip location, orientation, and cap opening. Tapered design creates seal. Device provides physical containment during ultra-rapid cooling and warming cycles. Benefits include secure storage and maintenance of viability for human oocytes and embryos.
Clinical Evidence
Bench testing only. No clinical data. Performance validated via Mouse Embryo Assay (MEA) with ≥80% blastocyst expansion at 96 hours, sterilization validation (SAL 10^-6), endotoxin testing (<2 EU/device), and mechanical testing (capping/uncapping force, seal integrity, dimensional analysis) after 2-year accelerated aging and simulated transportation.
Indicated for cryopreservation storage of human oocytes (MII), 4-8 cell embryos, and blastocyst stage embryos during vitrification procedures.
Regulatory Classification
Identification
Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 8, 2024
Cook Incorporated Christopher Sontag Regulatory Scientist 1 750 Daniels Way Bloomington, Indiana 47404
Re: K233177 Trade/Device Name: Nest VT Vitrification Device Regulation Number: 21 CFR 884.6160 Regulation Name: Assisted Reproduction Labware Regulatory Class: II Product Code: MQK Received: February 9, 2024
Dear Christopher Sontag:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael T. Bailey -S
For Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233177
Device Name Nest VT Vitrification Device
Indications for Use (Describe)
The Nest VT Vitrification Device is a cryopreservation storage device intended for use in vitrification procedures to contain and maintain human oocytes (MII), 4-8 cell embryos and blastocyst stage embryos.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K233177
#### 1. Submitter Information
| Applicant: | Cook Inc. |
|------------|----------------------------------------------------------|
| Address: | 750 Daniels Way<br>P.O. Box 489<br>Bloomington, IN 47402 |
#### 2. Submission Correspondent
| Contact: | Christopher Sontag |
|----------|--------------------------------|
| Email: | Chris.Sontag@CookMedical.com |
| | RegSubmissions@cookmedical.com |
#### 3. Date prepared: March 7, 2024
#### 4. Device Information
| Device Name: | Nest VT Vitrification Device |
|--------------------|--------------------------------------|
| Common Name: | Cryopreservation Storage Device |
| Regulation Number: | 21 CFR 884.6160 |
| Regulation Name: | Assisted Reproduction Labware |
| Product Code: | MQK (Labware, Assisted Reproduction) |
| Regulatory Class: | Class II |
### 5. Predicate Device Information
| Device Name: | VitriGuard |
|----------------|---------------------|
| 510(k) Number: | K200815 |
| Sponsor: | CooperSurgical Inc. |
The predicate device has not been subject to a design-related recall.
### 6. Device Description
The subject devices, Cook Nest VT Vitrification Devices, are sterile, single-use assisted reproduction storage devices for use in vitrification procedures to contain and maintain oocytes (MII), 4-8 cell embryos, and blastocyst stage embryos.
The Cook Nest VT Vitrification Devices are composed of a stick (124.5 mm) and cap (38.5 mm) design with a combined length of 136.5 mm, a 35 mm space for notations, and a 6 mm long loading well. Visible marks are on the distal tip end, on the body of the stick and at the opening of the cap, which are created through laser etching to indicate the location of the tip, the upright orientation of the device and the opening of the cap, respectively. The stick and cap include a taper design that create a seal when assembled. All the subject devices included in this submission are identical except
{4}------------------------------------------------
for their color. There are 5 different colors: blue, purple, red, yellow, and green. The subject devices are sterilized using gamma radiation to a sterilization assurance level of 10-9, are one-time use, and have a 2 year shelf-life.
During the vitrification procedure, the oocytes or embryos are loaded on the tip of the stick and the stick is capped prior to plunging the device in liquid nitrogen and for subsequent storage.
#### 7. Indications for Use Statement
The Nest VT Vitrification Device is a cryopreservation storage device intended for use in vitrification procedures to contain and maintain human oocytes (MII), 4-8 cell embryos and blastocyst stage embryos.
#### 8. Comparison of Intended Use and Technological Characteristics with the Predicate Device
The table below includes a comparison of the intended use and technological characteristics of the subject and predicate devices.
| | K233177 | K200815 | Comparison |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Regulation Number | 21 CFR 884.6160 | 21 CFR 884.6160 | Same |
| Product Code | MQK | MQK | Same |
| Classification | Class II | Class II | Same |
| Indications for Use | The Nest VT<br>Vitrification Device is<br>a cryopreservation<br>storage device<br>intended for use in<br>vitrification<br>procedures to contain<br>and maintain human<br>oocytes (MII), 4-8 cell<br>embryos and<br>blastocyst stage<br>embryos. | VitriGuard is<br>intended for use<br>as a<br>cryopreservation<br>storage device in<br>vitrification<br>procedures and<br>indicated to<br>contain and<br>maintain human<br>oocytes (MII), 4-8<br>cell embryos and<br>blastocyst stage<br>embryos. | Same intended use |
| Design | Stick with cap, design<br>feature to allow<br>handling with forceps | Stick with cap,<br>design feature to<br>allow handling<br>with forceps | Same |
| Marks/etching | Marks on distal tip<br>end, on body of the<br>stick and at the<br>opening of the cap | Black markings at<br>the end of the<br>stick and the tip of<br>the device | Different |
| Stick material | MBS Copolymer | Polystyrene | Different |
| Cap material | MBS Copolymer | Polystyrene | Different |
| Media Loading<br>volume | 1.0 microliter | <0.5 microliter | Different |
| Vitrification system | Closed | Closed | Same |
{5}------------------------------------------------
| Cooling rate | -1744 °C/min | -2271°C/min | Different |
|----------------------------------|-------------------------------------------------------------------------|----------------------------------------------|-----------|
| Warming rate | 48818 °C/min | 36377 °C/min | Different |
| Single use | Yes | Yes | Same |
| Sterilization method | Gamma | Gamma | Same |
| Sterilization<br>assurance level | SAL 10-6 | SAL 10-6 | Same |
| Endotoxin tested | <2 EU/device | <2 EU/device | Same |
| MEA testing | 1-Cell MEA was<br>≥80% embryos<br>expanded to blastocyst<br>at 96 hours | Yes, ≥80%<br>expand blast<br>within 96 hours | Same |
As shown in the table above, there are differences in the subject and predicate device indications for use statements; however, they have the same intended use (i.e., vitrification and storage of human oocytes (MII), 4-8 cell embryos, and blastocyst stage embryos).
The subject and predicate devices also have differences in technological characteristics, including cooling/warming rate, stick composition, loading volume, and available markings. These technological differences do not raise different questions of safety and effectiveness.
#### 9. Summary of Non-Clinical Performance Testing
#### Sterilization Testing:
Sterilization information was provided in accordance with the 2016 FDA guidance "Submission and Review of Sterility Information in Premarket Notification (510(k) Submissions for Devices Labeled as Sterile." The subject devices are subjected to a Gamma sterilization process to achieve a Sterility Assurance Level (SAL) of 10 %. The sterilization cycle was validated per VDmax25 (for 25 kGy dose) in accordance with ISO 11137-1:2006 and 11137-2:2013.
#### Simulated Transportation:
Simulated transportation was performed after accelerated aging per ASTM D4169-22.
#### Endotoxin
Evaluation performed on unaged devices using the Gel-Clot Limulus Amoebocyte Lysate (LAL) method per ANSI/AAMI ST72:2011. The acceptance criterion was ≤ 2 EU/device.
#### Cooling/warming rate
A comparative analysis was provided on unaged subject devices against the predicate, K200815, to verify cooling/warming rates for closed vitrification storage after immersion and removal from liquid nitrogen. The subject device performed within the specifications noted in the table above.
#### Stability and Shelf Life:
The following tests were completed to demonstrate that the subject devices maintained their performance in newly manufactured devices and after 2-years of accelerated aging per ASTM F1980-21 and simulated transportation:
- Mouse embryo assay (MEA) - Testing conducted in accordance with the 2021 FDA guidance "Mouse Embryo Assay for Assisted Reproduction Technology Devices." The acceptance criterion for the 1-Cell MEA was ≥80% embryos expanded to blastocyst at 96 hours.
{6}------------------------------------------------
- Package Integrity Testing: ●
- F1886/F1886M-16 o
- o ASTM F88/F88M-15
- ASTM F1929-15 o
- Mechanical Performance Testing:
- Dimensional testing O
- Examination following cryostorage O
- Seal integrity and leakage assessment following cryostorage O
- Capping force and torque testing O
- Uncapping force and torque testing O
#### 10. Conclusion
The results of the performance testing described above demonstrate that the subject devices are as safe and effective as the predicate device and support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.