uOmnispace.MI is a software solution intended to be used for viewing, processing, evaluating and analyzing of PET, CT, MR, SPECT images. It supports interpretation and evaluation of examinations within healthcare institutions. It has the following additional indications: uOmnispace.MI MM Fusion application is intended to provide tools for viewing, analyzing and reporting PET, CT, MR, SPECT data with its flexible workflow and optimized layout protocols for dedicated reporting purposes on oncology, neurology, cardiology. uOmnispace.MI MM Oncology application is intended to provide tools to display and analyze the follow-up PET, CT, MR data, with which users can do image registration, lesion segmentation, and statistical analysis. uOmnispace.MI Dynamic Analysis application is intended to display PET data and anatomical data such as CT or MR, and supports to do lesion segmentation and output associated time activity curve. uOmnispace.MI NeuroQ application is intended to analyze the brain PET scan, give quantitative results of the relative activity of different brain regions, and make comparison of activity of normal brain regions in AC database or between two studies from the same patient, as well as provide analysis of amyloid uptake levels in the brain. uOmnispace.MI Emory Cardiac Toolbox application is intended to provide cardiac short axis reconstruction, browsing function. And it also performs perfusion analysis, activity analysis and cardiac function analysis of the cardiac short axis.
Device Story
uOmnispace.MI is a post-processing software platform for viewing, manipulating, and analyzing PET, CT, MR, and SPECT images. It functions as a standalone software or with other advanced applications. Inputs include medical images (PET, CT, MR, SPECT); outputs include processed images, quantitative metrics, lesion segmentations, and reports. Used in healthcare institutions by clinicians for oncology, neurology, and cardiology workflows. Features include image fusion, registration (manual/auto/dot), lesion segmentation (threshold-based), and specialized analysis (NeuroQ for brain PET, Emory Cardiac Toolbox for cardiac function). Deep learning algorithms perform automated spine and rib labeling. Clinicians use these tools to assist in interpretation, evaluation, and reporting, potentially improving diagnostic efficiency and workflow.
Clinical Evidence
No clinical studies were required. Performance verification was conducted via bench testing of deep learning-based spine and rib labeling algorithms. Spine labeling validated on 286 CT scans (267 patients); rib labeling on 160 CT scans (156 patients). Metrics used a 5-point scoring system based on gold standard annotations. Average scores were 4.951 for spine labeling and 5.0 for rib labeling, exceeding the acceptance criterion of >4 points. Subgroup analysis demonstrated robustness across age, gender, and ethnicity.
Technological Characteristics
Standalone software platform for medical image processing. Utilizes deep learning-based algorithms for automated anatomical labeling (spine/rib). Connectivity via DICOM (NEMA PS 3.1-3.20). Software developed per IEC 62304. Risk management per ISO 14971. No hardware components; biocompatibility and electrical safety not applicable.
Indications for Use
Indicated for clinicians to view, process, evaluate, and analyze PET, CT, MR, and SPECT images for oncology, neurology, and cardiology applications. Includes tools for image registration, lesion segmentation, statistical analysis, time-activity curve generation, brain PET quantification, and cardiac short-axis analysis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 20, 2023
Shanghai United Imaging Healthcare Co., Ltd. Gao Xin Official Correspondent No.2258 Chengbei Road, Jiading District Shanghai, Shanghai 201807 China
Re: K233176
Trade/Device Name: uOmnispace.MI Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: September 26, 2023 Received: September 28, 2023
Dear Gao Xin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
### 33176 Device Name
uOmnispace.MI
Indications for Use (Describe)
uOmnispace.MI is a software solution intended to be used for viewing, processing, evaluating and analyzing of PET, CT, MR, SPECT images. It supports interpretation and evaluation of examinations within healthcare institutions. It has the following additional indications:
u uOmnispace.MI MM Fusion application is intended to provide tools for viewing, analyzing and reporting PET, CT, MR, SPECT data with its flexible workflow and optimized layout protocols for dedicated reporting purposes on oncology, neurology, cardiology.
u uOmnispace.MI MM Oncology application is intended to provide tools to display and analyze the follow-up PET, CT, MR data, with which users can do image registration, lesion segmentation, and statistical analysis.
· uOmnispace.MI Dynamic Analysis application is intended to display PET data and anatomical data such as CT or MR, and supports to do lesion segmentation and output associated time activity curve.
u uOmnispace.MI NeuroQ application is intended to analyze the brain PET scan, give quantitative results of the relative activity of different brain regions, and make comparison of activity of normal brain regions in AC database or between two studies from the same patient, as well as provide analysis of amyloid uptake levels in the brain.
u uOmnispace.MI Emory Cardiac Toolbox application is intended to provide cardiac short axis reconstruction, browsing function. And it also performs perfusion analysis, activity analysis and cardiac function analysis of the cardiac short axis.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" symbol, which is dark blue with a white vertical line running through the center. The logo is simple and modern in design.
# 510 (k) SUMMARY
K233176
- 1. Date of Preparation: September 26, 2023
## 2. Sponsor Identification
Shanghai United Imaging Healthcare Co.,Ltd. No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China
Establishment Registration Number: 3011015597
Contact Person: Xin GAO Position: Regulatory Affairs Specialist Tel: +86-021-67076888-5386 Fax: +86-021-67076889 Email: xin.gao@united-imaging.com
## 3. Identification of Proposed Device
Trade Name: Medical Image Post-processing Software Common Name: Medical image management and processing system Model(s): uOmnispace.MI
Regulatory Information Classification Name: Medical image management and processing system Classification: II Product Code: LLZ Regulation Number: 21 CFR 892.2050 Review Panel: Radiology
## 4. Identification of Predicate Device(s)
Predicate Device 510(k) Number: K192630 Device Name: uWS-MI
Reference Device#1 510(k) Number: K173897
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED IMAGING" is stacked on top of each other and is in a bold, sans-serif font. To the right of the text is a stylized "U" symbol, which is also in a bold font. The logo is simple and modern, and the use of a bold font gives it a strong and confident look.
Device Name: syngo.via MI Workflows
Reference Device#2 510(k) Number: K183170 Device Name: uWS-CT
## 5. Device Description
The uOmnispace.MI is a post-processing software based on the uOmnispace platform (cleared in K230039) for viewing, manipulating, evaluating and analyzing PET, CT, MR, SPECT images, can run alone or with other advanced commercially cleared applications.
This proposed device contains the following applications:
- uOmnispace.MI MM Fusion
- uOmnispace.MI MM Oncology
- . uOmnispace.MI Dynamic Analysis
Additionally, uOmnispace.MI offers the users the options to run the following third-party applications in uOmnispace.MI:
- uOmnispace.MI NeuroQ ●
- uOmnispace.MI Emory Cardiac Toolbox ●
## 6. Indications for use
uOmnispace.MI is a software solution intended to be used for viewing, processing, evaluating and analyzing of PET, CT, MR, SPECT images. It supports interpretation and evaluation of examinations within healthcare institutions. It has the following additional indications:
- uOmnispace.MI MM Fusion is intended to provide tools for viewing, analyzing and reporting PET, CT, MR, SPECT data with its flexible workflow and optimized layout protocols for dedicated reporting purposes on oncology, neurology, and cardiology.
- . uOmnispace.MI MM Oncology application is intended to provide tools to display and analyze the follow-up PET, CT, and MR data, with which users can do image registration, lesion segmentation, and statistical analysis.
- . uOmnispace.MI Dynamic Analysis application is intended to display PET data and anatomical data such as CT or MR, and supports to do lesion segmentation and output associated time activity curve.
- . uOmnispace.MI NeuroQ application is intended to analyze the brain PET scan, give quantitative results of the relative activity of different brain regions, and make comparison of activity of normal brain regions in AC database or between two studies from the same patient, as well as provide analysis of amyloid uptake levels in the brain.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in a bold, sans-serif font, stacked on top of each other. To the right of the words is a stylized "U" shape, which is also in a bold font. The logo is simple and modern, and the use of a bold font gives it a strong and confident look.
- uOmnispace.MI Emory Cardiac Toolbox application is intended to provide cardiac . short axis reconstruction, browsing function. And it also performs perfusion analysis, activity analysis and cardiac function analysis of the cardiac short axis.
## 7. Summary of Technological Characteristics
The technology characteristics of the uOmnispace.MI, reflected in this 510(k) submission are substantially equivalent to those of the predicate devices.
The following tables compare the features, principles of operation, fundamental scientific technology and intended use of uOmnispace.MI when compared to the predicate devices.
{6}------------------------------------------------
| Item | Proposed Device<br>uOmnispace.MI | Predicate Device<br>uWS-MI (K192630) | Remark |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General | | | |
| Device Classification<br>Name | Medical image management and processing system | Medical image management and processing system | Same |
| Product Code | LLZ | LLZ | Same |
| Regulation Number | 21 CFR 892.2050 | 21 CFR 892.2050 | Same |
| Device Class | II | II | Same |
| Classification Panel | Radiology | Radiology | Same |
| Advanced<br>Application | Yes | Yes | Same |
| Indications for use | uOmnispace.MI is a software solution intended to be used for viewing, processing, evaluating and analyzing of PET, CT, MR, SPECT images. It supports interpretation and evaluation of examinations within healthcare institutions. It has the following additional indications:<br>uOmnispace.MI MM Fusion application is intended to provide tools for viewing, analyzing and reporting PET, CT, MR, SPECT data with its flexible workflow and optimized layout protocols for dedicated reporting purposes on oncology, neurology, cardiology.<br>uOmnispace.MI The MM Oncology application is intended to provide tools to display and analyze the | uWS-MI is a software solution intended to be used for viewing, manipulation, communication, and storage of medical images. It supports interpretation and evaluation of examinations within healthcare institutions. It has the following additional indications:<br>The Oncology application is intended to provide tools to display and analyze the follow-up PET/CT data, with which users can do image registration, lesion segmentation, and statistical analysis.<br>The Dynamic Analysis application is intended to display PET data and anatomical data such as CT or MR, and supports to do lesion segmentation | The indication for use of the proposed device is expanded and replenished. The proposed device includes more applications and functions which are discussed in the following chapters. The difference will not impact the safety and effectiveness of the device. |
| Item | Proposed Device | Predicate Device | Remark |
| | uOmnispace.MI | uWS-MI (K192630) | |
| | follow-up PET, CT, MR data, with which users can<br>do image registration, lesion segmentation, and<br>statistical analysis. | and output associated time-activity curve. | |
| | uOmnispace.MI Dynamic Analysis application is<br>intended to display PET data and anatomical data<br>such as CT or MR, and supports to do lesion<br>segmentation and output associated time activity<br>curve. | NeuroQ application is intended to analyze the<br>brain PET scan, give quantitative results of the<br>relative activity of 240 different brain regions, and<br>make comparison of activity of normal brain<br>regions in AC database or between two studies<br>from the same patient, as well as provide analysis<br>of amyloid uptake levels in the brain. | |
| | uOmnispace.MI NeuroQ application is intended to<br>analyze the brain PET scan, give quantitative results<br>of the relative activity of different brain regions, and<br>make comparison of activity of normal brain<br>regions in AC database or between two studies from<br>the same patient, as well as provide analysis of<br>amyloid uptake levels in the brain. | Emory Cardiac Toolbox application is intended to<br>provide cardiac short axis reconstruction,<br>browsing function. And it also performs perfusion<br>analysis, activity analysis and cardiac function<br>analysis of the cardiac short axis. | |
| | uOmnispace.MI Emory Cardiac Toolbox<br>application is intended to provide cardiac short axis<br>reconstruction, browsing function. And it also<br>performs perfusion analysis, activity analysis and<br>cardiac function analysis of the cardiac short axis. | | |
Table 1 Substantial equivalent discussion for hasic functions
{7}------------------------------------------------
| Table 2 Substantial equivalent discussion for MM Fusion | | |
|---------------------------------------------------------|--|--|
| Application | Function name | Proposed device | Predicate | Reference | Remark |
|-------------|---------------|-----------------|-----------|-----------|--------|
| | | uOmnispace.MI | device | device#2: | |
{8}------------------------------------------------
| | | uWS-MI<br>(K192630) | uWS-CT<br>(K183170) | |
|------------------------|--------------------------------------|-----------------------|---------------------|-----------------------------------------------|
| MM Fusion | Image Fusion | Yes | / | Same |
| | Study Comparison | Yes | / | Same |
| | 3D Visualization | Yes | / | Same |
| Registration | Manual Registration | Yes | / | Same |
| | Dot Registration | Yes | / | Same |
| | Auto Registration | Yes | / | Same |
| | Default Layout and Customized Layout | Yes | / | Functional Substantial Equivalent<br>Note1 |
| | Position Correction | Yes | / | Same |
| Lesion<br>Segmentation | Fix threshold segmentation | Yes | / | Same |
| | Percentage threshold<br>segmentation | Yes | / | Same |
| | Adaptive threshold<br>segmentation | Yes | / | Same |
| | | CT Liver Segmentation | / | Yes |
| | CT Lung Segmentation | / | Yes | Same |
| | Rib Labeling | / | Yes | Functional Substantially Equivalent<br>Note 2 |
| | Spine Labeling | / | Yes | Functional Substantially Equivalent<br>Note 3 |
| | Save/Film/Report | Yes | / | Same |
{9}------------------------------------------------
| Application | Function name | Proposed device<br>uOmnispace.MI | Predicate<br>device<br>uWS-MI<br>(K192630) | Reference<br>device#2:<br>uWS-CT<br>(K183170) | Reference<br>device#3:<br>syngo.via MI<br>Workflows<br>(K173897) | Remark | |
|----------------|------------------------|--------------------------------------|--------------------------------------------|-----------------------------------------------|------------------------------------------------------------------|--------|-------------------------------------------------|
| MM<br>Oncology | Registration | Manual Registration | Yes | Yes | / | / | Same |
| | Registration | Auto Registration | Yes | Yes | / | / | Same |
| | One-Step Evaluation | | Yes | Yes | / | / | Same |
| | Lesion<br>Segmentation | Fix threshold<br>segmentation | Yes | Yes | / | / | Same |
| | | Percentage threshold<br>segmentation | Yes | Yes | / | / | Same |
| | | Adaptive threshold<br>segmentation | Yes | Yes | / | / | Same |
| | | Lung nodule<br>segmentation | Yes | Yes | / | / | Same |
| | | Liver tumor<br>segmentation | Yes | Yes | / | / | Same |
| | | Lymph node<br>segmentation | Yes | Yes | / | / | Same |
| | | Spread | Yes | Yes | / | / | Same |
| | CT liver Segmentation | | Yes | / | Yes | / | Same |
| | CT lung Segmentation | | Yes | / | Yes | / | Same |
| | Rib Labeling | | Yes | / | Yes | / | Functional Substantial<br>Equivalent.<br>Note 4 |
| | Spine Labeling | | Yes | / | Yes | | Functional Substantial<br>Equivalent |
### Table 3 Substantial equivalent discussion for MM Oncology
{10}------------------------------------------------
| | | | | | Note 5 |
|------------------------|----------------------|-----|-----|-----|--------|
| | Statistical Analysis | Yes | Yes | / | Same |
| | Reference VOI | Yes | Yes | / | Same |
| Response<br>Assessment | PERCIST | Yes | Yes | / | Same |
| | RECIST1.0 | Yes | Yes | / | Same |
| | RECIST1.1 | Yes | / | Yes | Same |
| | Deauville Score | Yes | / | Yes | Same |
| | Save | Yes | Yes | / | Same |
Table 4 Substantial equivalent discussion for Dynamic Analysis
| Application | Function name | Proposed device | Predicate device | Remark |
|------------------------|--------------------------------------|-----------------|------------------|--------|
| | | uOmnispace.MI | uWS-MI (K192630) | |
| Dynamic Analysis | Automatic cine | Yes | Yes | Same |
| | ROI Drawing | Yes | Yes | Same |
| | Data Recombination | Yes | Yes | Same |
| | Curve Analysis | Yes | Yes | Same |
| | Table Quantization Statistics | Yes | Yes | Same |
| | Save | Yes | Yes | Same |
| | Print (Filming) | Yes | Yes | Same |
| Registration | Manual Registration | Yes | Yes | Same |
| | Dot Registration | Yes | Yes | Same |
| | Auto Registration | Yes | Yes | Same |
| Lesion<br>Segmentation | Fix threshold segmentation | Yes | Yes | Same |
| | Percentage threshold<br>segmentation | Yes | Yes | Same |
| | Adaptive threshold<br>segmentation | Yes | Yes | Same |
{11}------------------------------------------------
Note 1: Compared to predicate device optimizes image layout that adds more default layout and adjusts the interaction of customized layout. This difference between the reference device doesn't impact the safety and effectiveness of the subject device.
Note 2: Compared to predicate device optimizes a new algorithm based on deep learing. This algorithm supports the same function as uWS-CT (K183170 cleared). This difference between the reference device device doesn't impact the safety and effectiveness of the subject device.
Note 3: Compared to predicate device optimizes a new algorithm based on deep learing. This algorithm supports the same function as uWS-CT (K183170 cleared). This difference between the reference device doesn't impact the safety and effectiveness of the subject device.
Note 4: Compared to predicate device optimizes a new algorithm based on deep learning. This algorithm suports the same finction as uWS-CT (K183170 cleared). This difference between the reference device device doesn't impact the safety and effectiveness of the subject device.
Note 5: Compared to predicate device optimizes a new algorithm based on deep learning This algorithm supports the same function as uWS-CT (K183170 cleared). This difference between the reference device and the reference device doesn't impact the safety and effectiveness of the subject device.
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized symbol that resembles the letter "U" with a vertical line running through the center, creating a negative space that suggests the letter "I". The color of the text and symbol is a dark teal.
#### Performance Data 8.
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility
Not Applicable to the proposed device, because the device is stand-alone software.
### Electrical Safety and Electromagnetic Compatibility (EMC)
Not Applicable to the proposed device, because the device is stand-alone software.
### Software Verification and Validation
Software verification and validation testing was provided to demonstrate safety and efficacy of the proposed device. This includes a hazard analysis, and the potential hazards have been classified as a moderate level of concern (LOC). Those documentations include:
- · Software Description
- · Device Hazard Analysis
- · Software Requirements Specification (SRS)
- · Software Architecture Design Chart
- · Software Development Environment Description
- · Software Verification and Validation
- · Cybersecurity Documents
### Animal Study
No animal study was required.
### Clinical Studies
No clinical study was required.
### Performance Verification
#### ● Design of the performance verification test
The identification rate is a commonly used evaluation metric to evaluate the labelling performance [1-3], which is defined as the ratio of correctly identified vertebrae/rib to the total number of vertebrae/rib.
As for spine labeling, the mean identification rate of the five top-performing algorithms are 83.3% [2]. For rib labeling, the mean identification rate (labelaccuracy) of a SOTA algorithm is 91.5% [3]. If the proposed device algorithm can
{13}------------------------------------------------
Image /page/13/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is also in bold. The color of the text and the "U" shape is a dark teal.
achieve comparable performance, it is considered that the proposed device algorithm meets the clinical requirements.
We set a 5-point scoring criteria, and the acceptance criteria was that the average score of the proposed device results is higher than 4 points is equivalent to the mean identification rate of spine labeling is greater than 92% (>83.3%, correctly labeled vertebrae number ≥23, total vertebrae number=25, 23/25=92%), and the mean identification rate of rib labeling is greater than 91.7%(>91.5% , correctly labeled rib number ≥22, total rib number=24, 22/24~91.7%). So, the setting of evaluation criteria in this way can meet clinical requirements.
Moreover, scoring according to the number of incorrect label is a common way to evaluate the labeling algorithm [1]. The number of incorrect label is directly related to the extra workload of the doctor, e.g. manual correction. While the purpose of the labeling algorithm is to improve the work efficiency of doctors, a more intuitive way is to score according to the number of errors in automatic label
To validate the uOmnispace.MI software from a clinical perspective, the deep learning-based spine labeling algorithm and rib labeling algorithm contained in the product underwent a scientific evaluation. The validation results for the proposed device demonstrated the good performance, the robustness and good generalization ability among different subgroups.
### 1. Spine labeling Algorithm
The performance testing for deep learning-based spine labeling algorithm was performed on 286 CT scans (data shown in Table 8-2) during the product development.
#### . Acceptance Criteria
The validation type and acceptance criteria is shown in the Table 8-1 below:
| Validation Type | Acceptance Criteria |
|------------------------------|------------------------------------------------------------------------------|
| Score based on gold standard | The average score of the proposed device<br>results is higher than 4 points. |
Table 8-1 Validation type and acceptance criteria
#### ● Testing Data Information
#### 1) Equipments and Protocols
CT data were acquired with different multidetector (16, 64, 160, 256, 320-slice) CT scanners from five major manufacturers (GE, Phillips, Siemens, Toshiba, UIH), and with tube voltage of 110 - 140 kVp. tube current of 275-1000mA. slice thickness of 0.625-2.0 mm, in-plane spacing of 0.656 - 1.172 mm.
### 2) Sample Size
| Dataset | Patients Number | Samples Number |
|-----------------|-----------------|----------------|
| Testing Dataset | 267 | 286 |
{14}------------------------------------------------
Image /page/14/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold font. The color of the text and the "U" shape is a dark gray.
(Note: One patient may have multiple samples, because of the differences in scanning protocols, such as slice thickness, in-plane spacing)
#### Clinical Subgroup Information 3)
The testing data information is summarized below.
| Table 8-2. Testing data information | |
|-------------------------------------|----------------------------|
| Information of data | 286 spine CTs |
| Gender | Male: 68 |
| | Female: 55 |
| | Unknown: 163 |
| Age | (10, 25]: 4 |
| | (25, 40]: 16 |
| | (40, 60]: 45 |
| | (60, 75]: 47 |
| | (75, 100]: 14 |
| | Unknown: 160 |
| Ethnicity | Asian(Chinese) data: 106 |
| | European data: 160 |
| | The United States data: 20 |
#### . Performance Testing Summary
For testing dataset, the average score of the proposed device results to be validated is 4.951 points and greater than 4 points. Meanwhile, the subgroup analysis shows that (Table 8-3) the proposed device algorithm has good generalization in different subgroups.
| Age | Average score of the proposed device |
|------------------------|--------------------------------------|
| (10, 25] | 5.000 |
| (25, 40] | 5.000 |
| (40, 60] | 5.000 |
| (60, 75] | 5.000 |
| (75, 100] | 5.000 |
| Unknown | 4.914 |
| Gender | Average score of the proposed device |
| Female | 5.000 |
| Male | 5.000 |
| Unknown | 4.913 |
| Ethnicity | Average score of the proposed device |
| Asian(Chinese) data | 5.000 |
| European data | 4.913 |
| The United States data | 5.000 |
Table 8-3. Subgroup performance test
- Standard Annotation Process ●
{15}------------------------------------------------
Image /page/15/Picture/1 description: The image shows the logo for United Imaging. The text "UNITED IMAGING" is displayed in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized letter "U" symbol, which is dark teal in color. The logo is simple and modern in design.
For ground truth annotations, all ground truth are annotated by well-trained annotators. The annotators use an interactive tool to observe the image, and then set annotation points near the center of vertebral body and assign anatomical labels. Finally, a senior clinical specialist will check and modify annotations to make sure the ground truth correct.
#### . Testing & Training Data Independence
The training data used for the training of the spine labeling algorithm is independent of the data used to test the algorithm.
### Confounders
There are no confounders exist for the training methodology.
### 2. Rib labeling Algorithm
The performance testing for deep learning-based rib labeling algorithm was performed on 160 CT scans (data shown in Table 8-6) during the product development.
#### . Acceptance Criteria
The validation type and acceptance criteria is shown in the Table 8-4 below: Table8-4. Validation type and acceptance criteria
| Validation Type | Acceptance Criteria |
|------------------------------|------------------------------------------------------------------------------|
| Score based on gold standard | The average score of the proposed device<br>results is higher than 4 points. |
#### Testing Data Information .
#### 1) Equipments and Protocols
CT data were acquired with different multidetector (16, 64, 160, 256, 320-slice) CT scanners from five major manufacturers (GE, Phillips, Siemens, Toshiba, UIH), and with tube voltage of 100-140 kVp, tube current of 275-1000 mA, slice thickness of 0.625-3.0 mm, in-plane spacing of 0.531-1.955 mm.
#### 2) Sample Size
Table 8-5. Sample size information of testing data
| Dataset | Patients Number | Samples Number |
|-----------------|-----------------|----------------|
| Testing Dataset | 156 | 160 |
(Note: One patient may have multiple samples, because of the differences in scanning protocols, such as slice thickness, in-plane spacing.)
#### 3) Clinical subgroups
The clinical subgroups information of the testing data is summarized below.
Table 8-6. Testing data information
| Information of data | 160 CTs |
|---------------------|-------------------------|
| Gender | Male: 110<br>Female: 39 |
{16}------------------------------------------------
Image /page/16/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in a bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is formed by two vertical bars and a horizontal bar in the middle. The logo is simple and modern, and the use of bold font and the stylized "U" shape makes it easily recognizable.
| | Unknown: 11 |
|-----------|----------------------------|
| Age | (10, 25]: 2 |
| | (25, 40]: 14 |
| | (40, 60]: 51 |
| | (60, 75]: 69 |
| | (75, 100]: 24 |
| Ethnicity | Asian(Chinese) data: 80 |
| | The United States data: 80 |
#### . Performance Testing Summary
For the testing dataset, the average score of the proposed device results to be validated is 5 points and greater than 4 points. Meanwhile, the subgroup analysis shows that (Table 8-12) the proposed device algorithm has good generalization in different subgroups.
| Age | Average score of the proposed device |
|------------------------|--------------------------------------|
| (10, 25] | 5.00 |
| (25, 40] | 5.00 |
| (40, 60] | 5.00 |
| (60, 75] | 5.00 |
| (75, 100] | 5.00 |
| Gender | Average score of the proposed device |
| Female | 5.00 |
| Male | 5.00 |
| Unknown | 5.00 |
| Ethnicity | Average score of the proposed device |
| Asian(Chinese) data | 5.00 |
| The United States data | 5.00 |
Table 8-7. Subgroup performance test
#### . Standard Annotation Process
For ground truth annotations, all ground truth are annotated by well-trained annotators. A threshold based interactive tool is used to generate initial rib mask, then annotators will refine the rib mask and assign anatomical labels. After the first round annotation, they will check each other's annotation. At last, a senior clinical specialist will check and modify annotations to make sure the ground truth correct.
#### . Testing & Training Data Independence
The training data used for the training of the rib labeling algorithm is independent of the data used to test the algorithm.
### Confounders
There are no confounders exist for the training methodology.
### References:
{17}------------------------------------------------
Image /page/17/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in a bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is also in a bold font. The color of the logo is a dark teal.
[1]Major, David , et al. "Automated landmarking and labeling of fully and partially scanned spinal columns in CT images." Medical Image Analysis 17.8(2013):1151-1163.
[2]C, Anjany Sekuboyina A B , et al. "VerSe : A Vertebrae Labelling and Segmentation Benchmark for Multi-detector CT Images." Medical Image Analysis (2021).
[3]Jin L, Gu S, Wei D, et al. "RibSeg v2: A Large-scale Benchmark for Rib Labeling and Anatomical Centerline Extraction". IEEE Transactions on Medical Imaging, 2023. DOI: 10.1109/TMI.2023.3313627.
### Other Standards and Guidance
- . NEMA PS 3.1 - 3.20 Digital Imaging and Communications in Medicine (DICOM) Set (2016).
- ISO 14971 Medical devices - Application of risk management to medical devices (Edition 2.0, corrected version, 2007).
- IEC 62304 Medical device software - Software life cycle processes (Edition 1.1, 2015).
### Summary
The features described in this premarket submission are supported with the results of the testing mentioned above; the uOmnispace.MI was found to have a safety and effectiveness profile that is similar to the predicate device and reference devices.
#### Substantially Equivalent (SE) Conclusion 9.
The proposed device is equivalent to the predicate device with regard to safety and efficacy. This conclusion is based upon a comparison of intended use, technological characteristics, performance specification, device hazards as well as verification and validation results.
In summary, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.