K233157 · Innovaquartz, LLC · GEX · May 3, 2024 · General, Plastic Surgery
Device Facts
Record ID
K233157
Device Name
VaporMAX LDD; MOJo LDD; MegaJOULE LDD
Applicant
Innovaquartz, LLC
Product Code
GEX · General, Plastic Surgery
Decision Date
May 3, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VaporMax LDD® family of laser fibers, which includes VaporMAX LDD®, MegaJOULE LDD, and MoJo LDD, are indicated for use in all surgical specialties in which compatible laser systems with operational wavelengths between 500 nm and 2200 nm have received regulatory clearance. The VaporMAX LDD family of laser delivery fibers are intended for use with any cleared surgical laser having a SMA 905 connector, SMA 906 connector or manufacturer specific connectors and adaptors.
Device Story
Fiber optic surgical laser delivery device; transmits laser energy from external surgical laser systems to target tissue. Input: laser energy (500-2200 nm) from compatible laser systems via SMA 905/906 or manufacturer-specific connectors. Transformation: fiber core (low [OH-] fused silica) transmits energy to distal tip; beam conditioning lens and prism redirect output at nearly right angles. Output: redirected laser beam for surgical cutting/ablation. Used in OR/surgical settings by physicians. Ergonomic control knob allows rotational aiming of output without torquing the fiber. Benefits: enables precise tissue interaction in various surgical specialties; improved durability/power handling compared to predicates. Healthcare providers use the device to deliver laser energy to specific anatomical sites during surgery, facilitating tissue removal or coagulation.
Clinical Evidence
Bench testing only. Testing included power delivery in saline and tissue phantoms at 80W (CTH:YAG laser) and 200W (thulium fiber laser). Subject fibers delivered up to 1,000,000 joules without failure. Additional testing included beam profile analysis, divergence measurements, and mechanical integrity (tensile strength of cap/connector retention).
Technological Characteristics
Materials: low [OH-] fused silica core, fluorine-doped fused silica cladding, fluoropolymer secondary cladding, ETFE jacket. Distal tip: fused quartz capsule with beam conditioning lens and prism. Sterilization: Ethylene Oxide (EO) per ISO 11135:2014. Connectivity: SMA 905/906 or proprietary connectors. Biocompatibility: USP Class VI.
Indications for Use
Indicated for use in all surgical specialties requiring laser energy delivery for compatible laser systems (500 nm - 2200 nm). No specific patient population, age, or gender restrictions provided.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of the symbol, there is the FDA logo in blue, followed by the words "U.S. FOOD & DRUG" in a larger font and "ADMINISTRATION" in a smaller font below it, also in blue.
InnovaQuartz LLC Stephen Griffin Chief Innovations Officer 23030 N 15th Avenue Phoenix, Arizona 85027-1315
Re: K233157
Trade/Device Name: VaporMAX LDD; MOJo LDD; MegaJOULE LDD Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 2, 2024 Received: April 3, 2024
Dear Stephen Griffin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510/k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tanisha L. Digitally signed by
Tanisha L. Tanisha L. Hithe -S Hithe -S - Date: 2024.05.03 21:53:12 -04'00'
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233157
Device Name VaporMAX LDD, MegaJOULE LDD, and MoJo LDD
Indications for Use (Describe)
The VaporMax LDD® family of laser fibers, which includes VaporMAX LDD, Mega/OULE LDD and MoJO LDD, are indicated for use in all surgical specialties in which compatible laser systems with operational wavelengths between 500 nm and 2200 nm have received regulatory clearance. The VaporMAX LDD family of laser delivery fibers are intended for use with any cleared surgical laser having a SMA 905 connector or manufacturer specific connectors and adaptors.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for IQ LLC InnovaQuartz. The letters "IQ" are in a large, bold, blue font, with a red swoosh design wrapping around the top and bottom of the letters. To the right of "IQ" is the text "LLC" in a smaller, blue font. Below the letters is the word "InnovaQuartz" in a smaller, gray font.
23030 North Fifteenth Avenue Phoenix, Arizona 85027-1315 623-434-1895 innovaquartz.com
# 510(K) Summary K233157
#### Date Submitted: May 2, 2024
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
#### Sponsor/Owner/Holder:
InnovaQuartz, LLC 23030 N 15th Ave Phoenix, AZ 85027-1315 623-434-1895
#### Company Contact:
Stephen Griffin, CIO 623-434-1895 (main) x101 623-229-5174 (mobile) steveg@innovaquartz.com
#### Registration Number: 3010933841
#### Subject Device Name:
| Trade Name: | VaporMAX LDD®, MegaJOULE LDD, and MoJo LDD |
|-------------------|----------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Laser Fiber |
| Classification: | Laser Instrument, Surgical, Powered Laser surgical instrument for use in general<br>and plastic surgery and in dermatology |
| Product Code: | GEX |
| Regulatory Class: | Class II |
| Panel: | General and Plastic Surgery |
#### Predicate Devices:
| Trade Name: | VaporMAX™ |
|-------------------|----------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Laser Fiber |
| Classification: | Laser Instrument, Surgical, Powered Laser surgical instrument for use in general<br>and plastic surgery and in dermatology |
| Product Code: | GEX |
| Regulatory Class: | Class II |
| Panel: | General and Plastic Surgery |
| 510(k) Number: | K053457 issued to Trimedyne, Inc., 15091 Bake Parkway, Irvine, CA 92618 |
| Trade Name: | Multiwavelength Lateral Emitting Laser Fiber |
| Common Name: | Laser Fiber |
| Classification: | Laser Instrument, Surgical, Powered Laser surgical instrument for use in general<br>and plastic surgery and in dermatology |
| Product Code: | GEX |
| Regulatory Class: | Class II |
| Panel: | General and Plastic Surgery |
| 510(k) Number: | K170366 issued to Laser Peripherals, LLC 13355 10th Ave, Ste 110, Plymouth,<br>MN 55441 |
{5}------------------------------------------------
This statement is based on the similarity of the subject device to the predicate devices in the intended use and optical principles.
# Device Description:
VaporMAX LDD® family of fibers, which includes VaporMAX LDD®, MegaJOULE LDD, and MoJo LDD, are fiber optic surgical laser energy delivery devices consisting of a stainless steel laser connector, e.g. SMA 905 or Trimedyne OmniPulse™ MAX, an anodized aluminum extension sleeve or polymer overnut for accommodating recessed laser ports, fiber/connector strain relief, and a transmitting optical fiber consisting of a low [OH-], fused silica core fiber with fluorine-doped, fused silica cladding, a fluoropolymer secondary cladding and an ethylene tetrafluorethylene copolymer (ETFE, e.g. DuPont Tefzel 210™) protective jacket. The distal tip is laser polished and is disposed within a fused quartz capsule (cap) containing a beam conditioning lens and a prism for redirecting the output at nearly right angles to the fiber longitudinal axis.
The dimensional difference between the optical fiber and the distal cap is equalized with a Udel tube covering the distal length of fiber that may be tasked to pass a cystoscope bridge working channel, sized to match the cap diameter and upon which is dispose a green visual indicator of the fiber rotational orientation. The proximal end of the diameter matching tube is terminated within a patent pending control knob for ergonomic rotational control (aiming). Some models of the VaporMAX LDD, those with subscripted "e", where e" indicates "ergonomic") exploit severing the connection between the redirection prism from the transmitting fiber by providing for free rotation of the cap about the fiber, vial the control knob, thereby aiming the fiber output without torquing the transmitting fiber.
VaporMAX LDD fibers are packaged as a coil upon a coated fiberboard or dense polymer carrier card that is intended to provide for ease of dispensing within the surgical field while immobilizing the product within the sterile pouch. The working end (distal section) of the fiber is maintained in a straight configuration to avoid "fiber jumping" when rotated in surgery (due to the device's "memory" of being stored in a coil). Carrier cards are sealed within non-woven/impermeable pouches, e.g., Tyvek/paper or Tyvek/Mylar, currently validated by IQ for a three-year shelf life. Sealed sterile pouches are protected by an outer, nonsterile fiberboard carton. Both the sterile pouch and carton are labeled with the product information per FDA labeling requirements. All tissue contacting VaporMAX LDD materials of construction, and any materials in fluidic communication with tissue, are USP Class VI biocompatible and are compatible with ethylene oxide (EO) sterilization.
# Intended Use:
The VaporMax LDD® family of laser fibers, which includes VaporMAX LDD®, MegaJOULE LDD, and MoJo LDD, are indicated for use in all surgical specialties in which compatible laser systems with operational wavelengths between 500 nm and 2200 nm have received regulatory clearance. The VaporMAX LDD family of laser delivery fibers are intended for use with any cleared surgical laser having a SMA 905 connector, SMA 906 connector or manufacturer specific connectors and adaptors.
{6}------------------------------------------------
# Technological Characteristics Comparison:
The VaporMAX LDD laser fiber is produced as described in US Patents Nos. 9,323,005, 9,488,782 and patents pending. The subject fibers are technologically equivalent to the predicate devices and the intended end use for the subject fibers is the same as the intended use of the predicate devices. All the predicate devices and VaporMAX LDD operate upon the same optical principles, although somewhat refined. The VaporMAX LDD uses similar materials of construction as the predicate devices and has equivalent or better performance than the predicate devices. Table 4-6 illustrated in the comparison between VaporMAX LDD and the two predicate devices in the traditional 510(k) submission.
| Feature | VaporMAX™<br>Predicate | VaporMAX LDD®<br>Subject | LP Side Fire<br>Predicate |
|---------------------|------------------------------|------------------------------|-------------------------------|
| Theory of Operation | Total Internal<br>Reflection | Total Internal<br>Reflection | Total Internal<br>Reflection |
| Fiber total length | 3 meters | 3 meters | 3 meters |
| Fiber core (um) | 430, 550 and 600 | 365, 550 and 660 | 365 and 550 |
| Fiber core material | Low [OH] fused silica | Low [OH] fused silica | Low [OH] fused silica |
| Fiber CCDR | 1.05 and 1.1 | 1.1 | 1.1 |
| Fiber cladding | F-doped fused silica | F-doped fused silica | F-doped fused silica |
| Fiber buffer | Fluoroacrylate | Fluoroacrylate | Fluoroacrylate |
| Fiber jacketing | ETFE | ETFE | ETFE |
| Working length | 35 cm to 41 cm | 36 cm & 38 cm | Variable |
| Control device | Fixed polymer | Fixed polymer | Movable metal |
| Laser connector | SMA 905 + TMED | SMA 905 + TMED | SMA 905, 906 &<br>Proprietary |
| Output angle | 78° to 82° | 82° to 88° | UNKNOWN |
| Divergence | 12.7°max | 12.7° max | UNKNOWN |
| Cap diameter (max) | 2.2 mm | 2.14 mm | UNKNOWN |
| Output area | >0.8 mm² | >1.2 mm² | UNKNOWN |
| OD | 7.5 French | 7.5 French | UNKNOWN |
| Max energy | 4.0 joules | 4.0 joules | UNKNOWN |
| Max power | 100 watts | 200 watts | UNKNOWN |
| Minimum lifetime | 320,000 J | 600,000 J | UNKNOWN |
Figure 1: Table 4-6 Comparison Table
{7}------------------------------------------------
### Performance Testing (Bench and User Evaluation):
Briefly, subject fibers were power tested in simulated surgery like the predicate, under saline and in contact with tissue phantoms, at 80 watts average power (4 joules, 20 Hz), using a CTH:YAG laser operating at 2080nm (often incorrectly identified as "holmium lasers at 2100nm") for 320,000 joules delivered, and similarly tested in contact with tissue phantoms at 200 watts using a thulium fiber laser operating at 1940nm for 600,000 joules and 1,000,000 joules (or one megajoule). The subject fibers maintained the original operating characteristics far further into testing than did predicate devices and none of the subject fibers failed even under the most abusive test conditions*, leading to the new product name of "MegaJOULE" (which may or may not be retained as a product identifier).
Beam profiles were obtained for the subject VaporMAX LDD and predicate VaporMAX prior to the accumulation of multiple incremental changes and with each step that has potential for altering the output of the laser fiber and studies were performed to access the reproducibility of key performance variables (as measured by beam profiles and divergence as well as in simulated surgery with tissue phantoms). The subject fibers' construction integrity was then tested with tensile strength measurements of cap retention on the fiber, orientation tube retention within the control knob, and fiber retention within the laser connector, all fibers surpassed design specifications. The subject devices are determined to be as safe and as effective as the lower power capable predicate devices even when used at the maximum power output of modern surgical lasers.
* To test caps retention and erosion under extreme conditions, abusive tests were performed, e.g., cutting a tissue phantom in half, e.g., a London broil steak, at 100 watts with less than 10,000 joules).
#### Manufacturing, Packaging and Sterilization Facility:
Subject devices are designed, manufactured, packaging and sterilized within the InnovaQuartz manufacturing facility located in Phoenix, Arizona. All manufacturing equipment has been validated as appropriate for the intended functions and the product sterilization is compliant with ISO 11135:2014 standards. Shelf life is three years when stored as directed.
# Conclusion:
A direct comparison of key characteristics demonstrates that subject devices are substantially equivalent to the predicate device in terms of materials of construction, intended uses, technological considerations, & performance. The subject devices are as safe, as effective, and perform as well as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.