The indications for the I.T.S. Intramedullary Nailing System (INS) - Proximal Femur include intramedullary treatment of fractures and stabilization after tumor resection in the area of the proximal femur, as well as combinations that additionally affect the shaft area using the long nails. The indications for use of the I.T.S. INS Proximal Femur include: All Nails: - Stable and unstable pertrochanteric fractures - Intertrochanteric fractures - Combinations of fractures listed above Additionally for Long Nails: - Subtrochanteric fractures - Proximal fractures as listed above associated with shaft fractures - Pathological fractures in regions as listed above - Nonunions and malunions in regions as listed above
Device Story
Intramedullary fixation system for proximal femur fractures; comprises titanium alloy nails, lag screws, set screws, and distal cortical locking screws. Implants inserted into femoral canal to stabilize fractures; lag screw provides proximal fixation; set screw allows translation while preventing rotation; distal screws provide distal fixation. Instrumentation includes targeting devices, drills, guides, k-wires, and screwdrivers. Used in clinical/surgical settings by orthopedic surgeons. Provides mechanical stabilization of bone segments to facilitate healing. Single-use, sterile.
Clinical Evidence
No clinical data was required for this submission. Substantial equivalence established via non-clinical laboratory testing, including fatigue testing of the nail-screw construct and performance testing per ASTM F543 and ASTM F1264.
Technological Characteristics
Materials: Titanium alloy (Ti-6Al-4V ELI) per ASTM F136; stainless steel and carbon fiber reinforced PEEK for instrumentation. Design: Cannulated intramedullary nail with proximal lag screw, set screw, and distal locking screws. Dimensions: Nails 9-14mm diameter, 180-480mm length. Sterilization: Provided sterile for single-use. Standards: ASTM F543, ASTM F1264.
Indications for Use
Indicated for patients requiring intramedullary fixation for stable/unstable pertrochanteric, intertrochanteric, subtrochanteric, and proximal femoral fractures, including those associated with shaft fractures, pathological fractures, nonunions, malunions, and tumor resection stabilization.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{0}------------------------------------------------
February 12, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is the FDA logo in blue, followed by the words "U.S. FOOD & DRUG" in a larger font and "ADMINISTRATION" in a smaller font below it, also in blue.
I.T.S. GmbH Florian Grill Head of Quality & Regulatory Affairs, Head of R&D Autal 28 Lassnitzhoehe, Styria 8301 Austria
Re: K233134
Trade/Device Name: I.T.S. INS Proximal Femur Nail Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB, HWC Dated: January 11, 2024 Received: January 11, 2024
Dear Florian Grill:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joseph P. Russell Digitally signed by Joseph P. Russell -S -2 Date: 2024.02.12 14:12:32 -05'00'
for: Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233134
Device Name I.T.S. INS Proximal Femur Nail
### Indications for Use (Describe)
The indications for the I.T.S. Intramedullary Nailing System (INS) - Proximal Femur include intramedullary treatment of fractures and stabilization after tumor resection in the area of the proximal femur, as well as combinations that additionally affect the shaft area using the long nails.
The indications for use of the I.T.S. INS Proximal Femur include:
All Nails:
- · Stable and unstable pertrochanteric fractures
- · Intertrochanteric fractures
- · Combinations of fractures listed above
Additionally for Long Nails:
- Subtrochanteric fractures
- · Proximal fractures as listed above associated with shaft fractures
- · Pathological fractures in regions as listed above
- · Nonunions and malunions in regions as listed above
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Submission I.T.S. GmbH I.T.S. INS Proximal Femur Nail
# 510(k) Summary
## In Accordance with 21 CFR 807.92 of the Federal Code of Regulations 510(k) Summary
| NAME OF FIRM: | I.T.S. GmbH<br>Autal 28<br>Lassnitzhoehe, 8301<br>Austria<br>www.its-implant.com |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) FIRM CONTACT: | Jennifer Hadfield<br>Qserve Group US, Inc.<br>350 S Main Street, Suite 309<br>Doylestown, Pennsylvania, 18901, United States<br>Tel. No. 215-6003103<br>e-mail: jennifer.hadfield@qservegroup.com |
| TRADE NAME: | I.T.S. INS Proximal Femur Nail |
| DATE: | February 11, 2024 |
| COMMON NAME: | Rod, fixation, intramedullary and accessories<br>Screw, fixation, bone |
| REGULATORY CLASS: | Classification Name: Intramedullary Fixation Rod / Smooth or<br>threaded metallic bone fixation fastener |
| | Classification: 21 CFR 888.3020 / 888.3040 |
| | Device Class: Class II |
| DEVICE PRODUCT CODE: | HSB HWC |
| SUBSTANTIAL EQUIVALENCE: | |
| PRIMARY PREDICATE<br>REFERENCE DEVICES | Stryker - Gamma3 System (K200869)<br>Stryker - Gamma4 System (K213328)<br>Smith & Nephew - Intertan (K040212)<br>Synthes De Puy – Proximal Femoral Nail (K970097)<br>Orthofix - Chimaera Hip Fracture System (K161466) |
| DEVICE DESCRIPTION: | The I.T.S. INS Proximal Femur Nail consists of the following implants:<br>1) Nails: |
| | The Proximal Femur Nails are designed for the treatment of femoral<br>fractures. The Femur Nails are available in distal diameters ranging<br>from 9 to 14mm and lengths of 180mm (short nail), 240mm<br>(intermediate nail), and 260 to 480mm in 20mm increments (long<br>nail). 180mm and 240mm nails have a single, oblong hole distally, |
while all nails with a length over 260 have a round hole, an oblong hole and another round hole for distal fixation. The oblong hole
{4}------------------------------------------------
## 510(k) Submission I.T.S. GmbH I.T.S. INS Proximal Femur Nail
allows for both static and dynamic locking configurations. All Trochanteric Nails are locked proximally with a Lag Screw. The set screw engages with the lag screw and allows translation while preventing rotation. Endcaps close the top of the nail and are available in 0mm, 5mm, 10mm, 15mm and 20mm lengths.
- 2) Lag Screw
Lag screws are cannulated with a major diameter of 10.5mm and lengths ranging from 70mm to 130mm in 5mm increments. They are available in titanium alloy (Ti-6Al-4V ELI) per ASTM F136.
- 3) Set Screw
There are two types of set screws available. Inferior short and standard with a diameter of 6.6mm. The Length of the inferior short version is 45mm and the standard lengths are 70 to 130mm in 5mm increments. They are available in titanium alloy (Ti-6Al-4V ELI) per ASTM F136.
- 4) Distal Cortical Locking Screw
Distal locking screws are fully threaded, with a diameter of 5.0mm and lengths ranging from 25mm to 120mm in 5mm increments (2.5mm increments available from 25-70mm). Thev are available in titanium alloy (Ti-6Al-4V ELI) per ASTM F136.
- 5) Instrumentation
I.T.S. INS Proximal Femoral Nail Instrumentation consists of implant-specific targeting devices which help facilitate insertion of the nails, lag screws, and distal locking screws. Targeting devices are manufactured from stainless steel and carbon fibre reinforced PEEK. Other instrumentation includes drills, drill guides, k-wires and screwdrivers.
All nails and screws are provided Sterile for single-use.
#### The indications for the I.T.S. Intramedullary Nailing System (INS) -INDICATION FOR USE: Proximal Femur include intramedullary treatment of fractures and stabilization after tumor resection in the area of the proximal femur, as well as combinations that additionally affect the shaft area using the long nails.
The indications for use of the I.T.S. INS Proximal Femur include: All Nails:
- · Stable and unstable pertrochanteric fractures
- Intertrochanteric fractures
- · Combinations of fractures listed above
Additionally for Long Nails:
- · Subtrochanteric fractures
- · Proximal fractures as listed above associated with shaft fractures
- · Pathological fractures in regions as listed above
- · Nonunions and malunions in regions as listed above
{5}------------------------------------------------
#
| CLINICAL TESTING: | Clinical data was not required for this submission. |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| NON-CLINICAL TESTING: | The following non-clinical laboratory testing was performed to<br>establishing equivalence to predicate, including:<br>• Fatigue Testing of the nail screw construct<br>• Performance testing following the consensus standards ASTM F543<br>& ASTM F1264 |
| | Testing demonstrated that the I.T.S. INS Proximal Femoral Nail is<br>substantially equivalent in mechanical performance to the predicate<br>device. |
| SUMMARY OF<br>TECHNOLOGICAL<br>CHARACTERISTICS: | The I.T.S.INS Proximal Femur Nail is substantially equivalent<br>in material, geometry, design, indications and operational principles to<br>the predicate systems legally marketed in the US listed above. |
| CONCLUSIONS: | Based on the similarity in material, geometry, design, indications and<br>operational principles, as well as both the Engineering Analysis and<br>Performance Testing, the I.T.S. INS Proximal Femur Nail has been<br>demonstrated to be substantially equivalent (SE) to the predicate<br>devices. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.