K233123 · Hironic Co., Ltd. · GEI · Jun 24, 2024 · General, Plastic Surgery
Device Facts
Record ID
K233123
Device Name
SILKRO
Applicant
Hironic Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jun 24, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SILKRO with RM. RN. RC. RV. PS handpieces is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis. The SILKRO with RN handpieces is intended for the removal and destruction of skin lesions and coagulation of tissue.
Device Story
Electrosurgical system for tissue coagulation and hemostasis; utilizes high-frequency (RF) current. System includes main unit, five handpieces (RM, RN, RC, RV, PS), foot switch, and LCD touch interface. Operates in monopolar mode (RF energy between active electrode and neutral ground pad) or bipolar mode (RF energy between adjacent needles in electrode tip). Handpieces held at right angles to skin; RF energy generates thermal effect for coagulation. Used in clinical settings by healthcare professionals. Output parameters (intensity, duration, depth) adjusted via touch screen. Benefits include precise tissue destruction and hemostasis for dermatologic/surgical procedures.
Clinical Evidence
No clinical data. Evidence consists of bench testing (electrical safety per IEC 60601-1, 60601-2-2, 60601-1-2) and ex vivo animal studies. Animal testing evaluated histological depth and zone of coagulation/thermal damage at various intensity levels, confirming the thermal effect of the handpieces.
Technological Characteristics
Electrosurgical system; 2MHz operating frequency; 100-240VAC power. Handpieces: RM (bipolar), RN/RC/RV (monopolar). RF intensity 1-10 levels. Sterilization via EO gas. Complies with IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2, IEC 60601-1-6, IEC 62366-1, and ISO 14971.
Indications for Use
Indicated for dermatologic and general surgical procedures requiring electro-coagulation and hemostasis, and for the removal and destruction of skin lesions using specific handpieces.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
June 24, 2024
Hironic Co., Ltd. Gwijin Lee Section Chief 19F, U-TOWER, 767, Sinsu-Ro, Suji-Gu Yongin-Si, Gyeonggi-do 16827 Korea, South
### Re: K233123
Trade/Device Name: SILKRO Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: June 4, 2024 Received: June 4, 2024
### Dear Gwijin Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long H. Chen -S Digitally signed by Long H. Chen -S
Date: 2024.06.24 08:49:08 -04'00'
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K233123
Device Name SILKRO
Indications for Use (Describe)
The SILKRO with RM. RN. RC. RV. PS handpieces is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis.
The SILKRO with RN handpieces is intended for the removal and destruction of skin lesions and coagulation of tissue.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "HIRONIC" in a stylized font. The first letter, "H", is red and has a unique design, while the rest of the letters are black and have rounded edges. The logo appears to be for a company or brand named HIRONIC.
### K233123
### 510(k) Summary
This summary of 510(k) Safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
1. Submitter Information - 807.92(a)(1)
| Applicant | Hironic Co., Ltd. |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address | 19F, U-TOWER, 767, Sinsu-ro, Suji-gu, Yongin-si, Gyeonggi-do,<br>16827, Republic of Korea |
| Phone Number | +82-31-525-7000 |
| Fax Number | +82-31-525-7010 |
| Contact Person | Gwijin Lee |
| Contact Information | 19F, U-TOWER, 767, Sinsu-ro, Suji-gu, Yongin-si, Gyeonggi-do,<br>16827, Republic of Korea<br>t. +82-31-525-7000, m. +82-10-5049-2959, e. ra@hironic.com |
| Preparation Date | Sep 27th, 2023 |
Date 510(k) summary prepared: June 21, 2024
| | 2. Device Name and Code - 807.92(a)(2) | | | | | | | | |
|--|----------------------------------------|--|--|--|--|--|--|--|--|
|--|----------------------------------------|--|--|--|--|--|--|--|--|
| Trade/Device Name | SILKRO |
|---------------------|----------------------------------------------------------------|
| Common Name | Electrosurgical System |
| Classification Name | Electrosurgical cutting and coagulation device and accessories |
| Product Code | GEI |
| Regulation Number | 21 CFR 878.4400 |
| Regulatory Class | II |
| Review Panel | General & Plastic Surgery (ODE) |
3. Legally marketed device(s) to which equivalence is claimed - 807.92(a)(3)
| Predicate Devices 1 | K210084 |
|---------------------|------------------------------------------------------------------------------------|
| | Applicant: Hironic Co., Ltd. |
| | Trade/Device Name: SILKRO |
| | Regulation Number: 21 CFR 878.4400 |
| | Regulation Name: Electrosurgical Cutting and Coagulation Device<br>and Accessories |
| | Regulatory Class: Class II |
| | Product Code: GEI |
| Predicate Devices 2 | K201738 |
| | Applicant: Cartessa Aesthetics |
| | Trade/Device Name: SubNovii Advanced Plasma Technology |
| | Regulation Number: 21 CFR 878.4400 |
| | Regulation Name: Electrosurgical Cutting and Coagulation Device<br>and Accessories |
| | Regulatory Class: Class II |
| | Product Code: GEI |
{4}------------------------------------------------
# анікопіс
### 4. Device Description - 807.92(a)(4)
This device is an instrument used for tissue coagulation by means of high-frequency current. It consists of the main unit, five handpieces, GP Cable, foot switch, LCD touch screen, ground pad and power cable.
The SILKRO has two operating modes: monopolar mode and bipolar mode. In the monopolar mode. RF energy flows from the main unit and a patient loop is formed by pairing the active electrode tip with the neutral electrode pad. Heat is not generated in the neutral electrode pad due to its low contact resistance, but heat is generated in the active electrode tip which has higher contact resistance. The higher contact resistance heats up the tissue resulting in coagulation. In the bipolar mode, RF energy is delivered between adjacent needles in the electrode tip without use of the neutral electrode pad. The user can select the mode and adjust parameters through the touch screen user interface of electrosurgical device.
RF(HF) enerqy is delivered to the target tissue using a handpiece (RM, RN, RC, RV, PS) and electrode tip, the tip being placed in light contact with the epidermis and the handpiece being held at right angles to the target tissue. As the RF(HF) energy passes through the skin, it generates an electro RF reaction which is capable of coagulating the tissue.
- Electrosurgical Unit Main body
- Five different handpieces
- Neutral electrode pad and neutral electrode pad cable, cleared under K092761
- Handpiece stands
- Foot switch
- Power cord
5. Indication for Use - 807.92(a)(5)
The SILKRO with RM, RN, RC, RV, PS handpieces is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis.
The SILKRO with RN handpiece is intended for the removal and destruction of skin lesions and coaqulation of tissue.
| 6. Summary of the Technological Characteristics of the Device Compared to the Predicate - 807.92(a)(6 |
|-------------------------------------------------------------------------------------------------------|
| (1) Predicate Device 1 |
| | Proposed Device | Predicate Device 1 |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Pending | K210084 |
| Manufacturer | Hironic Co., Ltd. | Hironic Co., Ltd. |
| Trade/Device Name | SILKRO | SILKRO |
| Classification Name | Electrosurgical cutting and<br>coagulation device and accessories | Electrosurgical cutting and<br>coagulation device and accessories |
| Indication for use | The SILKRO with RM, RN, RC, RV,<br>PS handpieces is intended for use in<br>dermatologic and general surgical<br>procedures for electro-coagulation<br>and hemostasis.<br>The SILKRO with RN handpiece<br>is intended for the removal and<br>destruction of skin lesions and<br>coagulation of tissue. | This device is intended for use in<br>dermatologic and general surgical<br>procedures for electro-coagulation. |
| Output energy type | Radio Frequency | Radio Frequency |
| User interface | Color Touch Panel | Color Touch Panel |
| Operating Frequency | 2MHz | 2MHz |
| Rated Input | 100-240VAC, 50/60Hz | 100-240VAC, 50/60Hz |
| Max. Power | RM Handpiece: 20.520 W ± 20 %,<br>500 Ω<br>RN Handpiece: 7.587 W ± 20 %,<br>500 Ω | RM Handpiece: 25W ± 20%, 500Ω<br>RN Handpiece: 7.5W ± 20%, 500Ω<br>RC Handpiece: 34 W ± 20%, 500Ω<br>RV Handpiece: 18 W ± 20%, 500Ω |
{5}------------------------------------------------
# анікопіс
K233123
Page 3 of 4
| | RC Handpiece: 33.754 W ± 20 %,<br>500 Ω | |
|---------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| | RV Handpiece: 18.354 W ± 20 %,<br>500 Ω | |
| RF Intensity | 1 ~ 10 Level | 1 ~ 10 Level |
| RF Duration | RM Handpiece: 50 ms ~ 950 ms<br>RN Handpiece: 50 ms ~ 1000 ms | RM Handpiece: 50 ms ~ 950 ms<br>RN Handpiece: 100 ms ~ 1000 ms |
| Needle Insert Depth | RM Handpiece with RF Micro<br>Needle Tip: 0.5 ~ 3.5 mm | RM Handpiece with RF Micro<br>Needle Tip: 0.5 ~ 3.5 mm |
| | RN Handpiece with A-FIT RF<br>Needle Tip: 1.2/1.5/1.8 mm | RN Handpiece with A-FIT RF<br>Needle Tip: 1.2/1.5/1.8 mm |
| Repetition | RM Handpiece: 0.2/0.5/0.8/1.0/2.0 sec, Single | RM Handpiece: 0.2/0.5/0.8/1.0/2.0 sec, Single |
| Mode of Operation | RM Handpiece: Bipolar Type<br>RN, RC, RV Handpiece: Monopolar Type | RM Handpiece: Bipolar Type<br>RN, RC, RV Handpiece: Monopolar Type |
| Electrode(Needle)<br>Type | RF Micro Needle Tip for RM<br>Handpiece: 25 pin, 49 pin<br>A-FIT RF Needle Tip for RN<br>Handpiece: 1 pin | RF Micro Needle Tip for RM<br>Handpiece: 25 pin, 49 pin<br>A-FIT RF Needle Tip for RN<br>Handpiece: 1 pin |
| Single Use | RF Micro Needle Tip for RM<br>Handpiece<br>A-FIT RF Needle Tip for RN<br>Handpiece | RF Micro Needle Tip for RM<br>Handpiece<br>A-FIT RF Needle Tip for RN<br>Handpiece |
| Sterilization | EO Gas | EO Gas |
### (2) Predicate Device 2
| (2) Predicate Device 2 | | |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| | Proposed Device | Predicate Device 2 |
| 510(k) Number | Pending | K201738 |
| Manufacturer | Hironic Co., Ltd. | Cartessa Aesthetics |
| Trade/Device Name | SILKRO | SubNovii Advanced Plasma<br>Technology |
| Classification Name | Electrosurgical cutting and<br>coagulation device and accessories | Electrosurgical cutting and<br>coagulation device and accessories |
| Indications for Use | The SILKRO with PS RN<br>handpieces is intended for the<br>removal and destruction of skin<br>lesions and coagulation of tissue. | The SubNovii is intended for the<br>removal and destruction of skin<br>lesions and coagulation of tissue. |
| Mode of Operation | RN Handpiece<br>Radiofrequency energy ionizes the<br>air creating a Plasma stream | Plasma<br>Radiofrequency energy ionizes the<br>air creating a Plasma stream |
| Output | Monopolar | Monopolar |
| Power Supply | 100-240VAC, 50/60Hz | 110-250 VAC 50/60 Hz |
| Frequency | 2 MHz | 40 kHz |
| Max Power Output | $7.587W @ 500\Omega$ | 5W |
| Electrical Safety<br>Standards | Complies with IEC 60601-1, IEC<br>60601-1-2, IEC60601-2-2 | Complies with IEC 60601-1, IEC<br>60601-1-2 |
{6}------------------------------------------------
## ahironic
### 7. Non-clinical tests submitted - 807.92(b)(1)
- IEC 60601-1:2005 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- IEC 60601-2-2:2017 Medical electrical equipment Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
- IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests is evaluated in accordance with the FDA-recognized consensus standard
- IEC 60601-1-6:2010 Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability is evaluated in accordance with the FDA-recognized consensus standard
- IEC 62304:2006 Medical device software Software life cycle processes is evaluated according to the FDA-recognized consensus standard
- IEC 62366-1:2015 Medical devices Part 1: Application of usability engineering to medical devices is evaluated in accordance with the FDA-recognized consensus standard
- ISO 14971:2019 Medical devices Application of risk management to medical devices
- ex vivo animal testing using models was also conducted to obtain histological data of values for depth and zone of coagulation and thermal damage immediately post treatment; 10 days post treatment. The treatment was performed at the intensity(power) low, mid, high. Based on this animal test, it was confirmed through mechanical and histological evaluation that RM, RN, RV, RC, PS handpieces can affect tissue by thermal effect of each output condition.
### 8. Clinical tests submitted - 807.92(b)(2)
No clinical test was performed.
9. Conclusions drawn from clinical and non-clinical tests submitted
The SILKRO is similar to its predicate devices in regards to indication for use and technological characteristics. The non-clinical test data for proposed device, including biocompatibility, bench testing, hardware and software documentation shows that the device performs substantially equivalent in terms of safety and effectiveness to identified predicate devices for proposed indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.