← Product Code [EFB](/productcode/EFB) · K233122

# High Speed Air Turbine Handpiece (G100,G200,G300,G400,G450,G500,G600,G700,G700L,G800,G800L) (K233122)

_Litu Tech , Ltd. · EFB · Apr 30, 2024 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K233122

## Device Facts

- **Applicant:** Litu Tech , Ltd.
- **Product Code:** [EFB](/productcode/EFB.md)
- **Decision Date:** Apr 30, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4200
- **Device Class:** Class 1
- **Review Panel:** Dental

## Indications for Use

High Speed Air Turbine Handpiece is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations, root canal preparations and polishing teeth.

## Device Story

High-speed air-powered dental handpiece; utilizes compressed air to drive turbine for high-speed rotation (320,000–430,000 rpm). Models feature varying head angles (45° or 90°) and head sizes (standard or miniature) for specific clinical access, including wisdom teeth. Select models include fiber optic light guides. Cooling system provides air/water or water-only spray. Used by dentists in clinical settings for restorative and endodontic procedures. Connects to dental unit tubing via integrated or user-supplied couplings. Provides mechanical energy for rotary cutting/polishing instruments (burs).

## Clinical Evidence

No clinical data. Bench testing performed per ISO 14457:2017 to evaluate noise levels, air/water flow rates, chuck retention force, torque, rotational speed, dynamic eccentricity, stall torque, and light output.

## Technological Characteristics

Air-powered turbine handpiece; 11 models (G100-G800L). Materials and design comply with ISO 14457:2017. Features include air/water cooling, fiber optic illumination (select models), and various coupling interfaces (4-hole or quick-connect). Rotational speed: 320,000–430,000 rpm.

## Regulatory Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

## Predicate Devices

- High Speed Handpieces and Accessories ([K152146](/device/K152146.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

April 30, 2024

Litu Tech Ltd. % Diana Hong General Manager Mid-Link Consulting Co. Ltd. P.O. Box 120-119 Shanghai, 200120 CHINA

Re: K233122

Trade/Device Name: High Speed Air Turbine Handpiece (G100,G200,G300,G400,G450,G500,G600,G700,G700L,G800,G800L) Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EFB Dated: March 29, 2024 Received: March 29, 2024

Dear Diana Hong:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

# Michael E. Adjodha -S

Michael Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

Submission Number (if known)

Device Name

High Speed Air Turbine Handpiece

(G100,G200,G300,G400,G450,G500,G600,G700,G700L,G800,G800L)

Indications for Use (Describe)

High Speed Air Turbine Handpiece is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations, root canal preparations and polishing teeth.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary

Prepared on: 2023-09-27

### Contact Details

21 CFR 807.92(a)(1)

| Applicant Name                                            | Litu Tech Ltd.                                                                                 |  |  |                     |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------|--|--|---------------------|
| Applicant Address                                         | C1-2 No. 12 Denengxi Road Torch Zone, Zhongshan, Guangdong, China<br>Guangdong 528437 China    |  |  |                     |
| Applicant Contact Telephone                               | +86-760-8697442                                                                                |  |  |                     |
| Applicant Contact                                         | Mr. Guangzhao Liu                                                                              |  |  |                     |
| Applicant Contact Email                                   | info@litutechltd.com                                                                           |  |  |                     |
| Correspondent Name                                        | Mid-Link Consulting Co., Ltd.                                                                  |  |  |                     |
| Correspondent Address                                     | P.O. Box 120-119 Shanghai 200120 China                                                         |  |  |                     |
| Correspondent Contact Telephone                           | 86-21-2281-5850                                                                                |  |  |                     |
| Correspondent Contact                                     | Diana Hong                                                                                     |  |  |                     |
| Correspondent Contact Email                               | info@mid-link.net                                                                              |  |  |                     |
| Device Name<br>21 CFR 807.92(a)(2)                        |                                                                                                |  |  |                     |
| Device Trade Name                                         | High Speed Air Turbine Handpiece<br>(G100,G200,G300,G400,G450,G500,G600,G700,G700L,G800,G800L) |  |  |                     |
| Common Name                                               | Dental handpiece and accessories                                                               |  |  |                     |
| Classification Name                                       | Handpiece, Air-Powered, Dental                                                                 |  |  |                     |
| Regulation Number                                         | 872.4200                                                                                       |  |  |                     |
| Product Code                                              | EFB                                                                                            |  |  |                     |
| Legally Marketed Predicate Devices<br>21 CFR 807.92(a)(3) |                                                                                                |  |  |                     |
| Predicate #                                               | Predicate Trade Name (Primary Predicate is listed first)                                       |  |  | Product Code        |
| K152146                                                   | High Speed Handpieces and Accessories                                                          |  |  | EFB                 |
| Device Description Summary                                |                                                                                                |  |  | 21 CFR 807.92(a)(4) |

The proposed devices, High Speed Air Turbine Handpiece, is a dental abrasive device that rotates at high speed air as the driving force. It has 11 models: G100, G200, G300, G400, G450, G500, G600, G700, G700L, G800 and G800L.

The proposed device has the following features

The cooling system includes air/water spray or only water spray. .

The head types include standard head and miniature head.

The angle of the head and shaft include 45℃ and 90°. In general, dentist usually uses 90° handpiece having teeth treatment. Except for wisdom teeth 45°handpieces is considered more desirable for wisdom teeth operation .It is because the 45° angle is more appropriate for the angle of eruption of wisdom teeth.

{5}------------------------------------------------

High Speed Handpieces are able to run from 320,000rmp to 430,000rmp. G450, G700L and G800L have optic light guide for light function, other models don't have the optic light guide. Coupling is the accessory for the some models of the proposed handpiece to connect with tubes of dental unit. G500 and G600 require additional 4-hole coupling, and the 4-hole coupling is included in the product package. G450,G700, G800 and G800. require additional quick coupling is not included in the product package. The user shall purchase the compatible

## Intended Use/Indications for Use

coupler for G450,G700, G700L, G800 and G800L.

High Speed Air Turbine Handpiece is intended for removing carious material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.

## Indications for Use Comparison

The proposed device and predicate device have the same indications for use.

## Technological Comparison

The proposed device and predicate device have the same principles of operation, type of chuck, head angle. Both they have cooling system, and some models of them have the fiber optics. Regarding the technological characteristics, there are difference on Light intensity, Speed in rpms and Bur retention the performance test has conducted on the proposed device based on ISO 14457:2017, and the test result demonstrated that the proposed device complies with the standard requirements, the differences on technological characteristics do not affect the safety and effectiveness of the proposed device.

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

The proposed device, High Speed Air Turbine Handpiece, is tested per tISO 14457-2017 Dentistry - Handpieces and motors, to evaluate the handpiece's following particular performance. Noise level @ Background noise level 56 dB(A) Flow rate of air supply @ 400kPa Flow rate of spray air @ 200kPa(2.0 bar) Flow rate of water supply @200kPa(2.0 bar) The force required to extract the test mandrels from the chuck system for Type 5 test mandrels Torque @ Type 5 test mandrels Speed of handpiece @0.22MPa Dynamic eccentricity Stall torque The light at the output side of the handpiece

The clinical testing is not applicable.

The conclusions drawn from the nonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicate device K152146.

# 21 CFR 807.92(a)(6)

# 21 CFR 807 92(a)(5)

21 CFR 807 92(a)(5)

---

**Source:** [https://fda-staging.innolitics.com/device/K233122](https://fda-staging.innolitics.com/device/K233122)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).