K233115 · E Surgical, LLC · HRX · Feb 6, 2024 · Orthopedic
Device Facts
Record ID
K233115
Device Name
Hawkeye Control Unit with Eyas Endoscope
Applicant
E Surgical, LLC
Product Code
HRX · Orthopedic
Decision Date
Feb 6, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.1100
Device Class
Class 2
Indications for Use
The Hawkeye Control Unit with Eyas Endoscope is intended for use in diagnostic and operative arthroscopic and endoscopic procedures to provide visualization and image capture of an interior body cavity through a surgical incision.
Device Story
The Hawkeye Control Unit with Eyas Endoscope is an arthroscopic system comprising a medical tablet, a console box, and a single-use, hand-held Eyas endoscope. The system is used in operating rooms, treatment rooms, and clinics by healthcare professionals. The endoscope captures real-time images of interior body cavities through a small incision; the console box acts as an interface, powering the tablet and endoscope. The medical tablet provides image display, system control, and data storage. The device uses LED illumination to visualize the surgical site and relays images for processing and display. The system includes audio indicators for operation and warnings. The output is viewed by the clinician on the tablet to assist in diagnostic and operative decision-making, potentially benefiting the patient by providing minimally invasive visualization of joint or body cavity interiors.
Clinical Evidence
Bench testing only. Performance testing included electrical safety (IEC 60601-1, 60601-1-2, 60601-2-18), optical performance (luminance, sensitivity, white balance, SFR, uniformity, color response, field of view), mechanical integrity (activation force, strain relief, leakage, fluid ingress), and package integrity. No clinical data was provided.
Technological Characteristics
System includes a medical tablet, console box, and single-use endoscope. Materials include SUS 304, ABS, polycarbonate, copper clad laminate, polyimide, CaZrO3, epoxy resin, acrylate urethane, and silicone. Illumination via LED (SMD type). Connectivity includes USB, HDMI, and audio jack. Sterilization of endoscope via Ethylene Oxide. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, ISO 10993-1, and ISO 11135.
Indications for Use
Indicated for diagnostic and operative arthroscopic and endoscopic procedures requiring visualization and image capture of interior body cavities via surgical incision.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 6, 2024
E Surgical, LLC % Craig Coombs President Coombs Medical Device Consulting, Inc. 427 14th Ave San Francisco, California 94118
Re: K233115
Trade/Device Name: Hawkeye Control Unit with Eyas Endoscope Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: January 8, 2024 Received: January 9, 2024
Dear Craig Coombs:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Digitally signed by Jesse Muir Jesse Muir -S Date: 2024.02.06 -5 08:00:02 -05'00'
Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233115
Device Name Hawkeye Control Unit with Eyas Endoscope
Indications for Use (Describe)
The Hawkeye Control Unit with Eyas Endoscope is intended for use in diagnostic and operative arthroscopic and endoscopic procedures to provide visualization and image capture of an interior body cavity through a surgical incision.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
### A. Device Information:
| Category | Comments |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | E SURGICAL, LLC<br>150 Isidor Ct,<br>Ste 203, Sparks, NV 89441 |
| Correspondent Contact<br>Information: | Mr. Craig Coombs<br>President<br>Coombs Medical Device Consulting, Inc.<br>427 14th Ave,<br>San Francisco, CA, 94118<br>Tel: 650-380-2474<br>Email: craigJcoombs@gmail.com |
| Device Common Name: | Arthroscope |
| Device Classification Number: | 21 CFR 888.1100 |
| Device Classification &<br>Product Code: | Class II,<br>HRX |
| Device Proprietary Name: | Hawkeye Control Unit with Eyas Endoscope |
#### Predicate Device Information:
| Predicate Device : | Trice medical mi-eye2, mi-eye 2 monitor |
|----------------------------------------------------|-----------------------------------------|
| Predicate Device Manufacturer: | Trice Medical |
| Predicate Device Common Name: | Arthroscope |
| Predicate Device Premarket Notification #: | K162475 |
| Predicate Device Classification: | 21 CFR 888.1100<br>Arthroscope |
| Predicate Device Classification &<br>Product Code: | Class II,<br>HRX |
# B. Date Summary Prepared
30 January 2024
## C. Description of Device
The E Surgical Hawkeye Control Unit with Eyas Endoscope is an arthroscope system, consisting of the Hawkeye Control Unit (Medical Tablet plus Console Box), and the
{5}------------------------------------------------
## E Surgical, LLC
Eyas Endoscope. The Medical Tablet is a touch panel computer for the system function control, endoscope image displays and operation data storage. The Eyas Endoscope is a hand-held, single-use device for accessing the interior of a joint to capture the real-time image of the target site through a small incision. The Console Box is an interface between the Medical Tablet and Eyas Endoscope. There are 10 models of Eyas Endoscope in this system; including 2 different
directions of view (0° and 30°) in combination with 5 different working lengths. All these models must be used in conjunction with Hawkeye Control Unit to perform as intended.
The E Surgical Hawkeye Control Unit with Eyas Endoscope is also capable recording and storing the images in the system for later review.
## D. Indications for Use
The Hawkeye Control Unit with Eyas Endoscope is intended for use in diagnostic and operative arthroscopic and endoscopic procedures to provide visualization and image capture of an interior body cavity through a surgical incision.
# E. Comparison to Predicate Device
The comparison between E Surgical Hawkeye Control Unit with Eyas Endoscope and the predicate devices are:
| Feature | Application Device:<br>E Surgical Hawkeye<br>Control Unit with Eyas<br>Endoscope | Predicate:<br>Trice medical mi-eye2, mi-eye 2<br>monitor (K162475) | Pertinence of Feature to<br>Consideration of<br>Substantial Equivalence |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Hawkeye Control Unit<br>with Eyas Endoscope is<br>intended for use in<br>diagnostic and operative<br>arthroscopic and<br>endoscopic procedures to<br>provide visualization and<br>image capture of an<br>interior body cavity<br>through a surgical incision. | The mi-eye2 is indicated for use<br>in diagnostic and operative<br>arthroscopic and endoscopic<br>procedure to provide<br>illumination and visualization of<br>an interior cavity of the body<br>through either a natural or<br>surgical opening. | Identical |
| Product Code | HRX | HRX | Identical |
| Feature | Application Device:<br>E Surgical Hawkeye<br>Control Unit with Eyas<br>Endoscope | Predicate:<br>Trice medical mi-eye2, mi-eye 2<br>monitor (K162475) | Pertinence of Feature to<br>Consideration of<br>Substantial Equivalence |
| Operating<br>Principle | Transmission of light to<br>illuminate and image an<br>arthroscopic joint, then<br>relaying the image out of<br>the surgical site for<br>processing and display. | Transmission of light to<br>illuminate and image an<br>arthroscopic joint, then<br>relaying the image out of the<br>surgical site for processing and<br>display. | Identical |
| Input Power | 100 - 240 V | 100 - 240 V | Identical |
| Technology | | | |
| Light source | LED(SMD Type) | LED(SMD Type) | Identical |
| Image File<br>Format | PNG | PNG | Identical |
| Video file<br>Format | MPG | AVI(or MP4) | Functionally identical.<br>These two video formats<br>are applied for different<br>media players. |
| Output port | USB,HDMI, Audio jack | USB | Functionally identical.<br>Per cybersecurity report,<br>image and audio output<br>through HDMI and audio<br>jack do not raise safety or<br>effectiveness concerns. |
| Physical Specification (Tablet) | | | |
| Principle<br>Component | Tablet, Console Box,<br>Endoscope | Tablet, Endoscope | Similar.<br>The application device has<br>a console box that powers<br>the tablet and endoscope,<br>and severs as a stand to<br>position the tablet. This<br>design difference does not<br>affect the safety and<br>performance of the<br>product. |
| Feature | Application Device:<br><i>E Surgical Hawkeye<br/>Control Unit with Eyas<br/>Endoscope</i> | Predicate:<br><i>Trice medical mi-eye2, mi-eye 2<br/>monitor (K162475)</i> | Pertinence of Feature to<br>Consideration of<br>Substantial Equivalence |
| Dimensions | Tablet:<br>28 cm x 18 cm x 1.83 cm<br>Console Box:<br>16.9 cm x 19.6 cm x 12.2 cm | Tablet:<br>33.5 cm x 21.6 cm x 3.8 cm | Similar.<br>The difference in design<br>mechanisms does not raise<br>any safety or effectiveness<br>concerns. |
| Weight | Tablet:2.2 lb (1 kg)<br>Console: 2.2 lb (1 kg) | 3.72 lb (< 1.69 kg) | |
| Display | 10.1" LCD | 12.3" LCD | |
| Working<br>Condition | Temperature range:<br>10°C to 30°C<br>Humidity range:<br>$30\% \le RH \le 75\%$<br>Atmospheric Pressure:<br>700 hPa to 1013 hPa | Temperature range:<br>10°C to 32°C<br>Humidity range:<br>$30\% \le RH \le 95\%$<br>Atmospheric Pressure:<br>700 hPa to 1060 hPa | Minor Differences.<br>The internal circuit design<br>and components chosen<br>are different, but have<br>specifications for similar<br>working, transport and<br>storage conditions. The<br>temperature, RH and<br>atmospheric pressure<br>specifications are similar<br>and are typical working<br>conditions. These<br>specifications are<br>considered to be<br>substantially equivalent. |
| Transport and<br>Storage<br>Condition | Ambient temperature:<br>-20°C to +60°C<br>Humidity range:<br>25% to 85%, non-<br>condensing.<br>Atmospheric pressure:<br>500 hPa to 1013 hPa | Ambient temperature:<br>-20°C to +60°C<br>Humidity range:<br>10% to 100%, non-condensing.<br>Atmospheric pressure:<br>500hPa to 1060 hPa | |
| Materials | | | |
| Probe & Patient<br>contacting<br>features | SUS 304,<br>ABS, PC (polycarbonate),<br>Copper Clad Laminate,<br>Polyimide, CaZrO3, Expoy<br>Resin, Acrylate Urethane,<br>Silicone | SUS 304 | Biologically identical |
| Feature | Application Device:<br>E Surgical Hawkeye<br>Control Unit with Eyas<br>Endoscope | Predicate:<br>Trice medical mi-eye2, mi-eye 2<br>monitor (K162475) | Pertinence of Feature to<br>Consideration of<br>Substantial Equivalence |
| Audio indicator | Yes, operating & warning<br>tone | No | The application device has<br>operating and warning<br>tone to remind use, while<br>the predicate has visual<br>signal only. |
| Flushing feature | Yes.<br>The Eyas endoscope<br>requires connecting to a<br>cannula to achieve the<br>flushing feature. | Yes.<br>The integrated rigid shaft that<br>extends from the mi-eye2<br>endoscope handle can be<br>utilized to achieve flushing<br>feature. | Functionally identical. |
| Others | | | |
| Endoscope<br>Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Identical |
| Operation<br>Environment | Operating rooms,<br>treatment rooms, and<br>clinics. | Operating rooms, treatment<br>rooms, and clinics. | Identical |
| Standard Met | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-18<br>ISO 10993-1<br>ISO 11135 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-18<br>ISO 10993-1<br>ISO 11135-1 | Almost identical,<br>application device is in<br>conformance with latest<br>version of the listed<br>standards |
{6}------------------------------------------------
### Hawkeye Control Unit with Eyas Endoscope K233115
# E Surgical, LLC
{7}------------------------------------------------
### Hawkeye Control Unit with Eyas Endoscope K233115
# E Surgical, LLC
{8}------------------------------------------------
#### Hawkeye Control Unit with Eyas Endoscope K233115
# E Surgical, LLC
# F. Bench Testing Summary
The Bench Testing for the Hawkeye Control Unit with Eyas Endoscope is summarized below.
| Hawkeye Control Unit | |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | |
| Console Box | Visual Inspection/ Power Supply/ Plug Insertion & Extraction Force |
| Medical Tablet | Visual Inspection/ Luminance Adjustment/ Capturing and Recording / Battery Capacity<br>Check/ Tablet Charging/ Software Version/ Firmware Version/ HDMI Signal Output/ Image Function/<br>USB Storage/Tone (button operation)/ Tone (warning alert) |
| Package | |
{9}------------------------------------------------
E Surgical, LLC
# Hawkeye Control Unit
| Package Inspection | Package and Device Integrity, Label Visibility |
|--------------------|------------------------------------------------|
|--------------------|------------------------------------------------|
| Eyas Endoscope | |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | |
| Electrical | EEPROM Function/ Continuity Resistance |
| Optical | Luminance Intensity/ Sensitivity/ White Balance/ SFR/ Uniformity/ Color Response/ Dust<br>Spot/ Dark/ Noise/ Stray Light/ Color Performance/Direction of View/ Field of View |
| Mechanical | Activation Force/ Plug Insertion & Extraction Force/ Dynamic Strain Relief/ Static Strain<br>Relief/ Weld Integrity / Camera Tube Wobble/ Activation Over Time/ Leakage/Fluid Ingress<br>Test |
| Package | |
| Package Integrity | Visual Inspection/Peel Open / Seal Width/ Seal Strength / Dye Leak |
## G. Summary of Supporting Data
The application E Surgical Hawkeye Control Unit with Eyas Endoscope were tested and found to be in compliance with the pertinent portions of the following standards,
| Standards<br>Body & # | Standard Name | Standard Version |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| IEC 60601-1 | Medical electrical equipment - Part 1: General requirements for<br>basic safety and essential performance | 2005 + AM1:2012 |
| IEC 60601-1-2 | Medical electrical equipment - Part 1-2: General requirements for<br>basic safety and essential performance - Collateral Standard:<br>Electromagnetic disturbances - Requirements and tests | 2017 + AM1:2020 |
| IEC 60601-2-18 | Medical electrical equipment - Part 2-18: Particular requirements for<br>the basic safety and essential performance of endoscopic equipment | 2009 |
| IEC 62304 | Medical device software — Software life cycle processes | 2006+AM1:2015 |
| ISO 10993-1 | Biological Evaluation of Medical Devices – Part 1: Evaluation and<br>testing within a risk management process. | 2018 |
| ISO 11135 | Sterilization of health-care products — Ethylene oxide —<br>Requirements for the development, validation and routine control of<br>a sterilization process for medical devices | 2014 + AM1: 2018 |
{10}------------------------------------------------
## E Surgical, LLC
| Standards<br>Body & # | Standard Name | Standard Version |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| ISO 11607-1 | Packaging for terminally sterilized medical devices – Part 1:<br>Requirements for materials, sterile barrier systems and packaging<br>systems; | 2019 |
| ISO 11607-2 | Packaging for terminally sterilized medical devices – Part 2: Validation<br>requirements for forming, sealing and assembly processes | 2019 |
## G. Conclusion
After comparing the indications for use, technology and design of the E Surgical Hawkeye Control Unit with Eyas Endoscope, along with all electrical safety (including IEC 60601-1; IEC 60601-1-2; IEC 60601-2-18) and relevant performance, in accordance with FDA-recognized consensus standards for electrical safety, E Surgical concludes that the Hawkeye Control Unit with Eyas Endoscope is substantially equivalent to the predicate Trice medical mi-eye2, mi-eye 2 monitor (K162475).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.