K233099 · Rohrer Aesthetics, LLC · GEI · May 22, 2024 · General, Plastic Surgery
Device Facts
Record ID
K233099
Device Name
Gladiator
Applicant
Rohrer Aesthetics, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
May 22, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PiXel8-RF handpiece is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostatis. The Bodysculp laser lipolysis handpiece intended for non-invasive lipolysis of the flank and abdomen to achieve disruption of adipocyte cells intended for non-invasive a desired a desired aesthetic affect. This treatment is intended for individuals with a Body Mass Index (BMI) of 30 or less. The Body Tone handpiece is intended for Relaxation of muscle spasms, Prevention or disuse atrophy, Increasing local blood circulation. Muscle re-education. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, Maintaining or increasing range of motion. Powered muscle stimulators should only be used under medical supervision for adjunctive therapy for the treatment of medical diseases and conditions.
Device Story
Gladiator is a multi-function console integrating three distinct technologies: bipolar RF microneedling, low-level diode laser, and electronic muscle stimulation (EMS). System operated by clinicians in medical settings. PiXel8-RF handpiece uses disposable microneedle tips to deliver 4MHz bipolar RF energy for electrocoagulation/hemostasis. BodySculp handpiece uses 1060nm diode laser for heat-assisted non-invasive lipolysis of flank/abdomen. BodyTone handpiece provides EMS for muscle rehabilitation and circulation. Clinicians select handpieces and settings via console controls. Output effects include thermal tissue coagulation, adipocyte disruption, and muscle contraction. Benefits include aesthetic body contouring, surgical hemostasis, and physical therapy support. Device relies on facility power.
Clinical Evidence
No clinical studies were conducted. Evidence consists of bench testing, including thermal damage profile comparisons between the 4MHz test device and 2MHz predicate on ex vivo tissue, and verification/validation of electrical safety (IEC 60601-1, 60601-1-2, 60601-2-2) and software life cycle processes (IEC 62304).
Indicated for dermatologic/general surgical electrocoagulation/hemostasis (PiXel8-RF); non-invasive lipolysis of flank/abdomen for aesthetic effect in individuals with BMI ≤ 30 (Bodysculp); and muscle spasm relaxation, prevention/retardation of disuse atrophy, increased local blood circulation, muscle re-education, post-surgical calf muscle stimulation to prevent venous thrombosis, and maintenance/increase of range of motion (BodyTone).
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 22, 2024
Rohrer Aesthetics, LLC Mark Rohrer President 105 Citation Court Homewood, Alabama 35209
Re: K233099
Trade/Device Name: Gladiator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI, OUH, GEX, IMG Dated: April 19, 2024 Received: April 19, 2024
Dear Mark Rohrer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and -CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Long H. Chen - S Digitally signed by Long H.Chen-S
Digitally signed Date: 2024.05.22 13:54:06 -04'00'
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233099
Device Name Gladiator
#### Indications for Use (Describe)
The Gladiator system contains three handpieces, each of which has its own indications for use.
The PiXel8-RF handpiece is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostatis.
The Bodysculp laser lipolysis handpiece intended for non-invasive lipolysis of the flank and abdomen to achieve disruption of adipocyte cells intended for non-invasive a desired a desired aesthetic affect. This treatment is intended for individuals with a Body Mass Index (BMI) of 30 or less.
The Body Tone handpiece is intended for Relaxation of muscle spasms, Prevention or disuse atrophy, Increasing local blood circulation. Muscle re-education. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, Maintaining or increasing range of motion. Powered muscle stimulators should only be used under medical supervision for adjunctive therapy for the treatment of medical diseases and conditions.
Type of Use (*Select one or both, as applicable*)
| <div><span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div><span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Applicant<br>Address | Rohrer Aesthetics<br>105 Citation Court<br>Homewood, Alabama 35209 |
|---------------------------------------|----------------------------------------------------------------------------------------------------|
| Contact Person | Mark Rohrer |
| Contact Information | President, Rohrer Aesthetic<br>205-940-2200<br>mrohrer@rohreraesthetics.com |
| Preparation Date | May 22, 2024 |
| Device Trade Name | Gladiator |
| Classification Name | Electrosurgical Cutting and Coagulation Device and<br>Accessories<br>Powered Laser Surgical System |
| Regulation Number | 878.4400, 878.5400, 878,5860 |
| Product Code | GEI, OUH, PKT, IMG |
| Regulatory Class | II |
| Legally Marketed Predicate<br>Devices | Pinxel-RF system (K180654)<br>BodySculpt (K212331)<br>Futura Pro (K102524) |
## Device Description:
The Gladiator is a multi-function system that combines RF Microneedling, Low Level Laser and Electronic Muscle Stimulation (EMS) technologies into a single console. The system is intended for dermatologic and general surgery purpose as well as pain relief. The system is comprised of a single console with three handpieces:
- PiXel8-RF - a bipolar radiofrequency micro needling handpiece with disposable tips
- . BodySculp - a low level laser
- . BodyTone (previously called Futura Pro) – an electronic muscle stimulator
The system and controls are contained in a single console. Electrical power is supplied to the console by the facility's power source.
#### Indications for use:
Each of the handpieces has its own indication for use.
The PiXel8-RF handpiece is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostatis.
{4}------------------------------------------------
The Bodysculp laser lipolysis handpiece intended for non-invasive lipolysis of the flank and abdomen to achieve disruption of adipocyte cells intended for non-invasive aesthetic use to achieve a desired aesthetic affect. This treatment is intended for individuals with a Body Mass Index (BMI) of 30 or less.
The BodyTone handpiece is intended for Relaxation of muscle spasms, Prevention or retardation of disuse atrophy, Increasing local blood circulation, Muscle re-education, Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, Maintaining or increasing range of motion. Powered muscle stimulators should only be used under medical supervision for adjunctive therapy for the treatment of medical diseases and conditions.
| Proposed Device<br>RF Microneedling Handpiece | Pinxel-RF System<br>K180654 | Comparison |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| The RF Microneedling handpiece is<br>intended for use in dermatologic and<br>general surgical procedures for<br>electrocoagulation and hemostasis. | The RF Microneedling handpiece is<br>intended for use in dermatologic and<br>general surgical procedures for<br>electrocoagulation and hemostasis. | Same |
| Proposed Device<br>BodySculp Handpiece | BodySculp<br>K212331 | Comparison |
| The Bodysculp laser lipolysis system is<br>intended for non-invasive lipolysis of<br>the flank and abdomen to achieve<br>disruption of adipocyte cells intended<br>for non-invasive aesthetic use to<br>achieve a desired aesthetic affect. This<br>treatment is intended for individuals<br>with a Body Mass Index (BMI) of 30 or<br>less. | The Bodysculp laser lipolysis system is<br>intended for non-invasive lipolysis of<br>the flank and abdomen to achieve<br>disruption of adipocyte cells intended<br>for non-invasive aesthetic use to<br>achieve a desired aesthetic affect. This<br>treatment is intended for individuals<br>with a Body Mass Index (BMI) of 30 or<br>less. | Same |
| Propose Device<br>BodyTone Handpiece | Futura Pro<br>K101524 | Comparison |
| The BodyTone handpiece is intended<br>for Relaxation of muscle spasms,<br>Prevention or retardation of disuse<br>atrophy, Increasing local blood<br>circulation, Muscle re-education,<br>Immediate post-surgical stimulation of<br>calf muscles to prevent venous<br>thrombosis, Maintaining or increasing<br>range of motion. Powered muscle<br>stimulators should only be used under | The BodyTone handpiece is intended<br>for Relaxation of muscle spasms,<br>Prevention or retardation of disuse<br>atrophy, Increasing local blood<br>circulation, Muscle re-education,<br>Immediate post-surgical stimulation of<br>calf muscles to prevent venous<br>thrombosis, Maintaining or increasing<br>range of motion. Powered muscle<br>stimulators should only be used under | Same |
Indications for use comparison
{5}------------------------------------------------
| medical supervision for adjunctive<br>therapy for the treatment of medical | medical supervision for adjunctive<br>therapy for the treatment of medical | |
|----------------------------------------------------------------------------|----------------------------------------------------------------------------|--|
| diseases and conditions. | diseases and conditions. | |
Technical Specifications Comparison
## RF Handpiece
| | Proposed Device<br>Gladiator | Predicate 1<br>PINXEL-RF system<br>(K180654) | Comparison |
|-------------------------------------------|---------------------------------------------|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| System Type | Bipolar Radiofrequency | Bipolar Radiofrequency | Same |
| RF Frequency | 4MHz | 2MHz | Different. Thermal testing<br>shows that the performance<br>of the two devices is<br>substantially equivalent. |
| Max Power | Max 25W @ 500ohm | Max 25W @ 500ohm | Same |
| Total Power<br>delivered per<br>treatment | 25W | 25W | Same |
| Power per<br>pin | 25W | 25W | Same |
| RF Duration | 50ms-950ms | 50ms-950ms | Same |
| Tips | 25, 49 and 64 pin<br>microneedle electrodes | 25 and 64 pin microneedle<br>electrodes | Addition of one new<br>electrode with 49 pins. The<br>electrode is mid-way<br>between the sizing of the<br>others and does not impact<br>the needle depth. |
| Needle Insert<br>Depth | 0.5-3.5mm | 0.5-3.5mm | Same |
## BodySculp Handpiece
| | Proposed Device<br>BodySculp Handpiece | BodySculp<br>K212331 | Comparison |
|---------------------|----------------------------------------|----------------------|------------|
| Laser Type | Diode Laser | Diode Laser | Same |
| Wavelength | 1060nm | 1060nm | Same |
| Lipolysis<br>method | Heat-assisted | Heat-assisted | Same |
{6}------------------------------------------------
| | Proposed Device<br>BodySculp Handpiece | BodySculp<br>K212331 | Comparison |
|------------------|-----------------------------------------------------------|-----------------------------------------------------------|------------|
| Spot Size | 4 × 8 cm2 (A single<br>applicator<br>of four applicators) | 4 × 8 cm2 (A single<br>applicator<br>of four applicators) | Same |
| Peak Power | 50W(per applicator) | 50W(per applicator) | Same |
| Power<br>Density | Up to 0.7-1.7W/ cm2 | Up to 0.7-1.7W/ cm2 | Same |
| Pulse width | CW | CW | Same |
| Power<br>supply | AC100-240V/50-<br>60Hz(customizable ) | AC100-240V/50-<br>60Hz(customizable ) | Same |
| Cooling | Contact Cooling | Contact Cooling | Same |
## Bodytone Handpiece
| Item | Description | Futura Pro | Comparison |
|----------------------------------|--------------------------------|--------------------------------|------------|
| | | (K102524) | |
| Number of Output<br>Modes | Two - Biphasic /<br>Monophasic | Two - Biphasic /<br>Monophasic | Same |
| Number of Output<br>Channels | 10 channels | 10 channels | Same |
| Maximum RMS O/P<br>Current | 16.2 mA,m,@500<br>ohm | 16.2 mA,m,@500 ohm | Same |
| Minimum electrode size | 7.0 cm dia. (38.5 cm') | 7.0 cm dia. (38.5 cm') | Same |
| Maximum Current<br>Densitv IRMS\ | 0.421 mA/cm'@ 500<br>ohm | 0.421 mA/cm'@ 500<br>ohm | Same |
| Maximum RMS<br>Power Density | 3.42 mW/cm' @ 500<br>ohm | 3.42 mW/cm' @ 500<br>ohm | Same |
| pk current into 500<br>Ohms | 120mA | 120mA | Same |
| Max charge/Pulse | 27.2 μC | 27.2 μC | Same |
| pk voltage, no load | 80 Volts | 80 Volts | Same |
{7}------------------------------------------------
| Maximum frequency | 100 Hz | 100 Hz | Same |
|-------------------|--------|--------|------|
| Max Pulse width | 300 us | 300 us | Same |
Substantial Equivalence Discussion
The subject device, the Gladiator, is identical in indications for user and nearly identical in technical specifications to its predicate devices. The sole changes to the device functionality, the RF frequency and addition of one new RF microneedling tip do not have an impact on the safety or efficacy of the device. Thermal testing of the device at the new 4Mhz RF frequency shows that the change raises no new concerns of safety and efficacy.
## Performance Testing
Verification and validation activities were successfully completed and establish that the Gladiator control unit performs as intended. Testing included the following:
- IEC 60601-1:2005 + Corr.1:2006 + A1:2012; Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance;
- . IEC 60601-1-2:2014 + A1:2020; Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests;
- . IEC 60601-2-2:2017; Medical Electrical Equipment - Part 2-2: Particular Requirements For The Basic Safety And Essential Performance Of High Frequency Surgical Equipment And High Frequency Surgical Accessories;
- IEC 62304:2006+A1:2015; Medical Device - Software Life Cycle Processes;
- EN ISO 14971:2019+A11:2021; Medical Devices Application Of Risk Management To . Medical Devices
## Thermal Testing
The following summarizes the thermal testing conducted to compare the new frequency of 4Mhz to the predicate's 2MHz frequency. Testing was conducted with 49-pin microneedling cartridges on three tissue types (ex vivo kidney, liver, and skin). Treatment settings included the three power levels: low (2.5 W), medium (7.5 W) and maximum (25 W), at a Pulse Width of 800 milliseconds (ms). Target penetration depth was 3.5 mm (e.g., max depth device delivers). Earlier generation PiXel8, operated on 2 MHz frequency, was used as the Predicate Device. The predicate was also equipped with the same microneedling cartridges as the Test Device and tested on all three types of tissue. Since the Predicate Device was used as a reference, it was tested for the worst cases only; e.g. maximum power level (25 W) and pulse width of 800 ms.
Thermal damage profiles for the Test and Predicate Devices were in close proximity in all test tissues. The Test Device showed somewhat milder thermal damage (moderately smaller CMZs) in treated tissues, which indicates a safer nature of the new generation PiXel8. At the same time, the study demonstrated the ability of the Test Device to achieve consistent thermal damage profiles in line with
{8}------------------------------------------------
the target treatment and were comparable to the profiles produced by the Predicate Device. Thus, it can be concluded that treatment by such a device at the appropriate testings will possess a desirable clinical treatment effect.
Software verification and validation testing was conducted, and documentation provided in accordance with FDA's Guidance or the Content of Premarket Submissions for Software Contained in Medical Devices.
Clinical Evidence – N/A. No clinical studies were conducted as part of this submission.
#### Biocompatibility -
#### PiXel8 RF Microneedling handpiece
The only parts of the device in contact with the patient are the sterile needles of the microneedle tips. Per ISO 10993-1, the tips are classified as a limited duration external communicating device. The tip comes in contact with tissue / bone / dentin. The tips are constructed of stainless steel with gold plating and do not include any coloring. The plastic housing of the handpiece is made of a Polycarbonate resin. No new biocompatibility testing is required for this submission because the materials and the manufacturing process are identical to the original PinXel-8 device (K180654).
## BodySculp
The patient contacting materials of Bodysculp are Sapphire Glass, gold plating, PET and polyester. As defined in Appendix A: ISO 10993-1:2009 these materials would be categorized as surface device contact with intact skin for limited duration. This categorization would require the minimal tests recommended by 10993-1. However, the materials and processing are identical to those used in the predicate device and other previously cleared medical devices, thus do not require additional testing.
No new biocompatibility testing is required for this submission because the materials and the manufacturing process are identical to the original BodySculp device (K212331).
#### Bodytone
The patient contacting materials of Bodytone handpiece is Silicone. As defined in Appendix A: ISO 10993-1:2009 these materials would be categorized as surface device contact with intact skin for limited duration. This categorization would require the minimal tests recommended by 10993-1. However, the materials and manufacturing processes are identical to those used in the predicate device. In addition, the FDA Guidance document "Use of International Standard ISO 10993-1, "Biological evaluation of
{9}------------------------------------------------
medical devices - Part 1: Evaluation and testing within a risk management process" lists silicone as exempt from testing if it is used on intact skin with limited contact.
#### Discussion
The subject device, the Gladiator, is identical in indications and nearly identical in technical specifications to its predicate devices. The sole changes to the device functionality, the RF frequency and addition of one new mid-sized RF microneedling tip do not have an impact on the safety or efficacy of the device.
The testing conducted and above comparison of devices shows that the Gladiator is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.