K233069 · Inari Medical, Inc. · PUR · Mar 26, 2024 · General Hospital
Device Facts
Record ID
K233069
Device Name
Removal System Large Bore 60 cc Syringe
Applicant
Inari Medical, Inc.
Product Code
PUR · General Hospital
Decision Date
Mar 26, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
The Removal System Large Bore 60 cc Syringe is used with the Removal Catheters to inject fluids into or withdraw fluids and/or tissues from the body. It can be used for aspiration removal of fluid and/or tissue in the form of abscess fluid, infected materials, etc.
Device Story
Manual piston syringe; 60 mL volume; used with Removal Catheters for fluid/tissue aspiration or injection. Features large bore Toomey tip adapter for quick-release connection to catheter sideport; twist-to-lock plunger mechanism maintains vacuum/retracted position. Operated by clinician; used in clinical settings. Provides airtight seal for aspiration of abscess fluid or infected materials. Benefits patient by enabling controlled removal of fluids/tissues via catheter system.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1), vacuum/leak testing, tensile testing, and performance testing per ISO 7886-1. Leveraged data from previous submissions (K191368, K231848, K231108, K230494, K191710) for packaging and mechanical performance.
Technological Characteristics
Piston syringe; 60 mL capacity. Materials: polycarbonate barrel, ABS plunger, EPDM rubber seal. Features large bore Toomey tip adapter and twist-to-lock plunger mechanism. Sterilized via EtO (SAL 10^-6) per ISO 11135. Compliant with ISO 7886-1 for graduation and design. Manual operation; no software or electronics.
Indications for Use
Indicated for use with Removal Catheters to inject fluids into or withdraw fluids and/or tissues (e.g., abscess fluid, infected materials) from the body.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
March 26, 2024
Inari Medical, Inc. Kaitlyn Weinkauf Sr. Regulatory Affairs Specialist 6001 Oak Canyon, Suite 100 Irvine, California 92618
Re: K233069
Trade/Device Name: Removal System Large Bore 60 cc Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: PUR Dated: February 22, 2024 Received: February 22, 2024
Dear Kaitlyn Weinkauf:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
# Sincerely, Shruti N. Mistry -S
Shruti Mistry Assistant Director Division of Drug Delivery and General Hospital Devices, and Human Factors Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
Submission Number (if known)
K233069
Device Name
Removal System Large Bore 60 cc Syringe
Indications for Use (Describe)
The Removal System Large Bore 60 cc Syringe is used with the Removal Catheters to inject fluids into or withdraw fluids and/or tissues from the body. It can be used for aspiration removal of fluid and/or tissue in the form of abscess fluid, infected materials, etc.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
PAGE 1 OF 5
## 510(K) SUMMARY
| Date prepared | March 25, 2024 | |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Name | Inari Medical, Inc.<br>6001 Oak Canyon, Suite 100<br>Irvine, CA 92618<br>877.923.4747 | |
| Contact person | Kaitlyn Weinkauf<br>Sr. Regulatory Affairs Specialist | |
| Trade name | Removal System Large Bore 60 cc Syringe | |
| Common name | Vacuum Syringe | |
| Regulation name | Piston Syringe | |
| Classification number | 21 CFR 880.5860 | |
| Product code | PUR | |
| Regulatory class | II | |
| Predicate device(s) | Merit VacLok AT Vacuum Syringe (K163597) | |
| Device description | The Removal System Large Bore 60 cc Syringe is a general piston syringe<br>constructed using a barrel, plunger, plunger seal, and large bore Toomey tip<br>adapter that acts as a quick-release connector. The quick-release connector is<br>compatible with the Removal System Catheters' sideport connector to<br>establish an airtight seal. The twist-to-lock plunger with barrel tabs maintains<br>the retracted position of the plunger. | |
| Indications for Use | The Removal System Large Bore 60 cc Syringe is used with the Removal<br>Catheters to inject fluids into or withdraw fluids and/or tissues from the body.<br>It can be used for aspiration removal of fluid and/or tissue in the form of<br>abscess fluid, infected materials, etc. | |
| Summary of | A tabular comparison of the predicate and subject devices is provided below: | |
substantial equivalence
| | Subject Device<br>Removal System Large<br>Bore 60 cc Syringe | Primary Predicate Device<br>Merit VacLok AT Vacuum<br>Syringe | Comments |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number | K233069 | K163597 | - |
| Manufacturer | Inari Medical, Inc. | Merit Medical Systems, Inc. | - |
| Regulation | 21 CFR 880.5860 Piston<br>Syringe | 21 CFR 880.5860 Piston<br>Syringe | Identical. |
| Product Code | PUR | PUR | Identical. |
| Intended Use | To inject fluids into or<br>withdraw fluids from the | To inject fluids into or<br>withdraw fluids from the | Identical. |
| | body. | body. | |
| Indications for<br>Use | The Removal System Large<br>Bore 60 cc Syringe is used<br>with the Removal Catheters<br>to inject fluids into or<br>withdraw fluids and/or<br>tissues from the body. It<br>can be used for aspiration<br>removal of fluid and/or<br>tissue in the form of<br>abscess fluid, infected<br>materials, etc. | VacLok AT Vacuum<br>Syringe is used to inject<br>fluids into, or withdraw<br>fluids from the body. It can<br>also be used in cases where a<br>vacuum syringe is preferred<br>(e.g. thrombus, abscess<br>fluid, bile, urine, etc.). | Substantially equivalent.<br>The indications for use for<br>the Removal System Large<br>Bore 60 cc Syringe is not<br>identical to the predicate<br>device; however, the<br>differences do not alter the<br>intended therapeutic use of<br>the device, nor do they<br>affect the performance of<br>the device. The subject and<br>predicate device have the<br>same intended use. |
| Design | Standard piston syringe<br>constructed with a clear<br>hollow barrel into which is<br>inserted a closely fitting<br>moveable plunger and<br>plunger seal. Contains<br>twist-to-lock plunger to<br>maintain the retracted<br>position of the plunger.<br>Fitted with a large bore<br>Toomey tip adapter that<br>acts as a quick-release<br>connector. | Standard piston syringe<br>constructed with a clear<br>hollow barrel into which is<br>inserted a closely fitting<br>moveable plunger and<br>tip/seal. Variable camlock<br>technology maintains the<br>retracted position of the<br>plunger. Fitted with male<br>luer lock connector. | Similar in design. The<br>subject device met all<br>design verification and<br>validation acceptance<br>criteria. The minor design<br>differences between the<br>subject and predicate device<br>do not raise new or different<br>questions of safety or<br>effectiveness. |
| Device<br>Description | The Removal System Large<br>Bore 60 cc Syringe is a<br>general piston syringe<br>constructed using a barrel,<br>plunger, plunger seal, and<br>large bore Toomey tip<br>adapter that acts as a quick-<br>release connector. The<br>quick-release connector is<br>compatible with the<br>Removal System Catheters'<br>sideport connector to<br>establish an airtight seal.<br>The twist-to-lock plunger<br>with barrel tabs maintains<br>the retracted position of the<br>plunger. | The VacLok AT Vacuum<br>Syringe is a general piston<br>syringe constructed using a<br>barrel, plunger, piston seal,<br>and locking mechanism. It is<br>designed to lock in any<br>position along the length of<br>the barrel with the capability<br>of holding a vacuum when<br>the cam locking mechanism<br>is engaged. | |
| Barrel<br>Material | Clear polycarbonate | Clear polycarbonate | Identical. |
| Plunger<br>Material | ABS | ABS | Identical. |
| Seal Material | EPDM Rubber | Silicone | The Removal System Large<br>Bore 60 cc Syringe<br>materials are established<br>medical device materials |
| | | | and have successfully<br>passed ISO 10993-1:2018<br>biocompatibility testing as<br>part of the final, finished<br>device intended for<br>marketing. |
| Operational<br>Volume | 60 mL | 20 mL and 30 mL | The subject device met all<br>design verification and<br>validation acceptance<br>criteria. The operatonal<br>volume difference between<br>the subject and predicate<br>device do not raise new or<br>different questions of safety<br>or effectiveness. |
| Graduation | Printed with accurate<br>graduation lines that are<br>compliant with ISO 7886-1 | Printed with accurate<br>graduation lines that are<br>compliant with ISO 7886-1 | Identical. |
| Vacuum Lock<br>Technology | Twist-to-lock plunger with<br>barrel tabs | Variable camlock<br>technology | Similar locking mechanism<br>for vacuum source. The<br>subject device and predicate<br>device locking mechanism<br>differs only in the number<br>of locking positions along<br>the barrel. The subject<br>device contains one locking<br>position at the operational<br>volume of 60 mL, while the<br>predicate device contains<br>ninety locking positions<br>along the barrel length.<br>Both devices are locked and<br>unlocked by twisting the<br>plunger in the barrel. This<br>does not raise new or<br>different questions of safety<br>or effectiveness. |
| Tip | Large bore Toomey tip<br>adapter | Male luer lock | - |
| Principles of<br>Operation | Manually operated by<br>advancing and withdrawing<br>the plunger within the<br>barrel. | Manually operated by<br>advancing and withdrawing<br>the plunger within the barrel. | Identical. |
| Sterilization | EtO | EtO | Identical. |
| Shelf-life | 6 months | Information not available | The subject device met all<br>T=6 month accelerated<br>aging design verification<br>and validation acceptance<br>criteria. |
| Single-use | Yes | Yes | Identical. |
{4}------------------------------------------------
## PAGE 2 OF 5
{5}------------------------------------------------
{6}------------------------------------------------
PAGE 4 OF 5
Summary of substantial equivalence
#### Biocompatibility
The following biocompatibility tests were completed to satisfy ISO 10993-1 requirements for the subject device:
- Cytotoxicity
- Sensitization
Acute Systemic Toxicity
- Intracutaneous Reactivity
- Material-Mediated Pyrogenicity
The passing results demonstrate that the subject device meets the biological safety requirements per ISO 10993-1.
.
#### Sterilization
.
The Removal System is sterilized using EtO to achieve a sterility assurance level (SAL) of 106 using a validated sterilization process in accordance with the principles of ISO 11135:2014/Amd 1:2018 and AAMI TIR 28:2016.
#### Non-Clinical Testing
In accordance with the design failure modes and effects analysis, verification and validation testing were identified to support the substantial equivalence of the Removal System Large Bore 60 cc Syringe to the predicate device and to support the indications for use. These tests included:
#### Performance Tests
- Packaging Inspection
- Leak Testing Syringe to Quick-Connect Adapter ●
- Vacuum Testing Syringe to Quick-Connect Adapter ●
- Air Leakage ●
- Simulated Pus Analog Removal ●
- Tensile Testing - Large Bore Syringe to Connector
- General, ISO 7886-1, Clause 6.1 ●
- Limits for Acidity or Alkalinity, ISO 7886-1, Clause 6.2 & Annex A
- Limits for Extractable Metals, ISO 7886-1, Clause 6.3 & Annex A ●
- Lubricant, ISO 7886-1, Clause 7 (Paragraph 1) ●
- Tolerance on Graduated Capacity, ISO 7886-1, Clause 8 ●
- Scale, ISO 7886-1, Clause 9.1
- Numbering of Scales, ISO 7886-1, Clause 9.2 ●
- Overall Length of Scale, ISO 7886-1, Clause 9.3 ●
- Position of Scale, ISO 7886-1, Clause 9.4
- Barrel Flanges, ISO 7886-1, Clause 10.2 ●
- Design, ISO 7886-1, Clause 11.1
- Position of Nozzle on End of Barrel, ISO 7886-1, Clause 12.2 ●
- Nozzle Lumen, ISO 7886-1, Clause 12.3
- Freedom from Liquid Leakage Past Plunger, guided by ISO 7886-1, ● Clause 13.2 & Annex D
- Freedom from Air Leakage Past Plunger, ISO 7886-1, Clause 13.2 . & Annex B
{7}------------------------------------------------
#### PAGE 5 OF 5
- Force to Operate the Piston, ISO 7886-1, Clause 13.3 & Annex E
- Fit of Stopper/Plunger in Barrel, ISO 7886-1, Clause 13.4
- Particulate Matter
### Leveraged Performance Tests
- Pouch Seal Visual Inspection (K191368 and K231848)
- Bubble Leak (K231848)
- Dye Penetration (K231848) ●
- Pouch Peel, Seal Strength (K231108)
- Packaging Usability Evaluation for Aseptic Presentation (K230494) ●
- Vacuum Testing Large Bore Syringe (K191710) ●
- Simulated Use, Tensile Large Bore Syringe (K191710) ●
Test results demonstrated that all acceptance criteria were met; therefore, the device conforms to established product specifications.
Animal testing was not required for the determination of substantial equivalence.
Clinical testing was not required for the determination of substantial equivalence.
#### Conclusion
The Removal System Large Bore 60 cc Syringe has the same intended use as the predicate. Non-clinical performance data show that the different technological characteristics between the devices do not raise any new or different questions of safety or effectiveness and support the Removal System Large Bore 60 cc Syringe's substantial equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.