The NanoZoomer S360MD Slide scanner system (“NanoZoomer System”) is an automated digital slide creation, viewing, and management system. The NanoZoomer System is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (“FFPE”) tissue. The NanoZoomer System is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. The NanoZoomer System comprises the NanoZoomer S360MD Slide scanner, the NZViewMD Software and a compatible display that has been 510(k) cleared for use with the NanoZoomer system or a 510(k)-cleared display that has been assessed in accordance with the Predetermined Change Control Plan (PCCP) for qualifying additional compatible displays. The NanoZoomer System is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using NanoZoomer System.
Device Story
Automated whole slide imaging system; converts glass slides containing FFPE tissue into high-resolution digital images. System components: NanoZoomer S360MD scanner, NZViewMD viewing software, and compatible FDA-cleared display. Used in clinical pathology settings by pathologists to perform primary diagnosis in lieu of conventional light microscopy. Scanner captures entire slide; software enables viewing, storage, retrieval, and sharing of digital images. Pathologist reviews images on compatible monitor to render diagnostic decisions. PCCP allows for future integration of additional FDA-cleared displays following validated performance testing.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including spatial resolution, luminance, color gamut, and color reproducibility (using △E2000 CIEDE2000 metric) to validate display compatibility.
Technological Characteristics
Automated slide scanner; 360-slide capacity; 20x objective lens; CMOS sensor; LED illumination. Software: NZAcquireMD (acquisition) and NZViewMD (viewing). Connectivity: LAN-based server storage. Standards: IDMS v1.03 for display testing; ISO/CIE standards for colorimetry. PCCP implemented for display interoperability.
Indications for Use
Indicated for use by pathologists as an aid to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. Not indicated for frozen section, cytology, or non-FFPE hematopathology specimens.
Regulatory Classification
Identification
The whole slide imaging system is an automated digital slide creation, viewing, and management system intended as an aid to the pathologist to review and interpret digital images of surgical pathology slides. The system generates digital images that would otherwise be appropriate for manual visualization by conventional light microscopy.
Special Controls
A whole slide imaging system must comply with the following special controls: (1) Premarket notification submissions must include the following information: (i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system. (ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate: (A) Slide feeder; (B) Light source; (C) Imaging optics: (D)Mechanical scanner movement; (E) Digital imaging sensor; (F) Image processing software; (G)Image composition techniques; (H)Image file formats; (I) Image review manipulation software; (J) Computer environment; (K)Display system. (iii)Detailed bench testing and results at the system level, including for the following, as appropriate: (A)Color reproducibility; (B) Spatial resolution; (C) Focusing test; (D) Whole slide tissue coverage; (E) Stitching error: (F) Turnaround time. (iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate: (A)Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included. (B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included. (C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (e.g., main sign-out diagnosis). (D) A detailed human factors engineering process must be used to evaluate the whole slide imaging system user interface(s). (2) Labeling compliant with 21 CFR 809.10(b) must include the following: The intended use statement must include the information described in paragraph (i) (1)(i) of this section, as applicable, and a statement that reads, "It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device." (ii) A description of the technical studies and the summary of results, including those that relate to paragraph (1)(ii) and (1)(iii) of this section, as appropriate. (iii) A description of the performance studies and the summary of results, including those that relate to paragraph (1)(iv) of this section, as appropriate. (iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information:
(i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system.
(ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate:
(A) Slide feeder;
(B) Light source;
(C) Imaging optics;
(D) Mechanical scanner movement;
(E) Digital imaging sensor;
(F) Image processing software;
(G) Image composition techniques;
(H) Image file formats;
(I) Image review manipulation software;
(J) Computer environment; and
(K) Display system.
(iii) Detailed bench testing and results at the system level, including for the following, as appropriate:
(A) Color reproducibility;
(B) Spatial resolution;
(C) Focusing test;
(D) Whole slide tissue coverage;
(E) Stitching error; and
(F) Turnaround time.
(iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate:
(A) Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (
*e.g.,* main sign-out diagnosis).(D) A detailed human factor engineering process must be used to evaluate the whole slide imaging system user interface(s).
(2) Labeling compliant with 21 CFR 809.10(b) must include the following:
(i) The intended use statement must include the information described in paragraph (b)(1)(i) of this section, as applicable, and a statement that reads, “It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device.”
(ii) A description of the technical studies and the summary of results, including those that relate to paragraphs (b)(1)(ii) and (iii) of this section, as appropriate.
(iii) A description of the performance studies and the summary of results, including those that relate to paragraph (b)(1)(iv) of this section, as appropriate.
(iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
INSTRUMENT ONLY
## I Background Information:
A 510(k) Number
K233027
B Applicant
Hamamatsu Photonics K.K.
C Proprietary and Established Names
NanoZoomer S360MD Slide scanner system
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| PSY | Class II | 21 CFR 864.3700 - Whole Slide Imaging System | PA - Pathology |
## II Submission/Device Overview:
A Purpose for Submission:
- Add the BARCO MDPC-8127 display (monitor), cleared independently in K203364, to the NanoZoomer S360 MD Slide scanner system originally cleared in K213883.
- Establish a Pre-Determined Change Control Plan (PCCP) for qualifying and adding additional FDA-cleared displays as interoperable component to the NanoZoomer S360 MD Slide scanner system.
B Measurand:
Not applicable
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
C Type of Test:
Digital pathology whole slide imaging
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The NanoZoomer S360MD Slide scanner system (“NanoZoomer System”) is an automated digital slide creation, viewing, and management system. The NanoZoomer System is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (“FFPE”) tissue. The NanoZoomer System is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.
The NanoZoomer System comprises the NanoZoomer S360MD Slide scanner, the NZViewMD Software and a compatible display that has been 510(k) cleared for use with the NanoZoomer system or a 510(k)-cleared display that has been assessed in accordance with the Predetermined Change Control Plan (PCCP) for qualifying additional compatible displays. The NanoZoomer System is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using NanoZoomer System.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
IV Device/System Characteristics:
A Device Description:
The NanoZoomer S360MD Slide scanner system is comprised of the NanoZoomer S360MD Slide scanner, NZAcquireMD acquisition software, NZViewMD image viewing software and compatible displays. The NanoZoomer S360MD Slide scanner system is an automated system for creating, viewing, and managing digital slides. The NanoZoomer S360MD Slide scanner system creates digital whole slide images (WSIs) of glass slides containing formalin-fixed paraffin-embedded (“FFPE”) tissue for pathology primary diagnosis purposes. The digital WSIs of glass slides may be viewed, stored, retrieved, annotated, and/or shared, permitting the pathologist to make a primary diagnosis based on the review of the WSIs. The NanoZoomer system does not include any automated Image Analysis Applications that would constitute computer aided detection or diagnosis.
K233027 - Page 2 of 11
{2}
The NanoZoomer slide scanner can be loaded with a maximum of 360 glass slides placed in 12 cassettes each with 30 standard-sized slides (length 75.0–76.5 mm, width 24.7–26.5 mm, thickness 0.9–1.2 mm). Each cassette is manually loaded with slides by the operator, with the slide labels facing outward so they can still be read while the cassette is loaded into the scanner. Once the slide’s position is secured, the macro camera located directly above the slide first takes a picture of the slide’s barcode image as illuminated by an overhead LED, and then takes a macro image of the actual tissue trans-illuminated by a flat-panel LED underneath. The slide holder then moves to the micro imaging position for scanning. The system automatically detects areas where tissues are present and scans a single overall rectangular area containing all detected tissues. The scanning area is automatically determined using a macro image of the slide. The NanoZoomer System uses its motorized stage, 20x objective lens, CMOS camera and LED to detect appropriate focusing points and scan the identified areas of interest. A single micro camera acquires one square image at a time and the acquired images (tiles) are stitched together to make a single composite high-resolution image. Images are automatically saved to the hard disk during scanning and may be viewed later by using the included viewing software. The NZAcquireMD software organizes all WSI tiles into a single NDPI file. Each digital image covers an entire slide and typically contains billions of image pixels. The NZAcquireMD software allows the user to operate the scanner and to store and archive digital images.
The NZViewMD software is a software component intended to run on a separate viewing workstation PC with a compatible display. The NZViewMD software opens the WSI images acquired with the NZAcquireMD software and slide scanner from the image storage attached to local network and renders the image data to the compatible display to deliver the image view at various magnification levels. The NZViewMD software has the following functionalities: continuous panning and zooming, comparing multiple slide images simultaneously in multiple windows, creating annotations during review using annotation tools, tracking of visited areas and annotations and digital bookmarks.
The new display Barco MDPC-8127 was cleared in K203364. Based on the testing data, it has equivalent display characteristics as the predicate display JVC JD-C240BN01. The Barco MDPC-8127 has been validated to be compatible with the NanoZoomer S360MD Slide scanner system by both Hamamatsu and Barco.
The NanoZoomer S360MD Slide scanner system can be connected to a server over a customer-provided local area network (LAN). The supporting networking hardware and software are not part of the system and may be located in a room separately from the workstation, viewing software, and display.
## B Instrument Description Information:
1. Instrument Name: NanoZoomer S360MD Slide scanner system (NanoZoomer System)
2. Specimen Identification:
K233027 - Page 3 of 11
{3}
Glass slides and scanned images are identified based on the previously assigned specimen identifiers such as patient identifiers, barcodes, etc. Digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue.
## 3. Specimen Sampling and Handling:
Specimen sampling and handling are performed upstream and independent of the use of the subject device. Specimen sampling includes surgical pathology specimens such as biopsy or resection specimens which are processed using standard histology techniques. The FFPE tissue sections are H&E stained. Then digital images are obtained from these glass slides using the NanoZoomer S360MD Slide scanner.
## 4. Calibration:
When the system is launched (powered up), automatic system diagnosis is performed by the software using the provided calibration slides. This process evaluates the proper functioning of the light source, imaging sensor, and macro sensor and light source. When necessary, based on the automatic system diagnosis, adjustment of the shading correction and color balance correction features is performed automatically by the system. Monthly cleaning such as dusting of the scan mechanism and optical components is also performed by the user.
System administrators and field service personnel may also complete a system check following the NZAcquireMD Instructions for Use and the NZTuneMD service software and the calibration slides. This checks correction status and acquires correction data for whole slide and scanned images.
Additional calibration and quality control procedures are performed during field service visits following procedures provided in the NanoZoomer S360MD Maintenance Manual. These include annual testing of several critical components in the system and annual calibration of the components and the display.
## 5. Quality Control:
Quality control (QC) activities are performed by the user. Prior to scanning slides using the NanoZoomer S360MD Slide scanner, the user conducts QC of the slides per the laboratory's standards and professional guidelines (e.g., staining, cover-slipping, and barcode placement, etc.). After completing a scan, the user checks the image data and image quality using Quality Check (QC) mode in NZAcquireMD which includes the following steps:
- Display the Scan Result screen by selecting a slide for which scanning has been completed.
- Confirm scan area, focus quality, and barcode information. Confirm that the barcode information displayed above the slide image is consistent with the barcode label.
- Check the slide image to ensure all tissue on the slide are present within the scan area.
Manually verifying tissue block with the scanned images is also performed as needed. Slides are rescanned if necessary.
Before reading the scanned WSIs, the pathologist should ensure that all scanned slide images have been imported for every case and the images are of acceptable quality for diagnostic
K233027 - Page 4 of 11
{4}
purposes. The pathologist reviews scanned images from all the slides associated with a case before rendering a diagnosis.
V Substantial Equivalence Information:
A Predicate Device Name (s):
NanoZoomer S360MD Slide scanner system
B Predicate 510(k) Number(s):
K213883
C Comparison with Predicate(s):
The subject device is similar to the predicate, having similar indications for use, the same intended use, and the same technological characteristics as its predicate device, with the exception of the implementation of a Predetermined Change Control Plan (PCCP) that specifies the protocols and acceptance criteria for validating and adding additional FDA-cleared displays as interoperable component to the NanoZoomer S360 MD Slide scanner system in a controlled manner, such that the device is as safe and as effective as the predicate.
The sponsor's PCCP for validating the use of additional FDA-cleared display is part of the quality system and comprises the following phases:
1. When a new FDA-cleared display is identified as the candidate, the candidate display, and its supplier will be evaluated based on various considerations. A design change request will be generated if the candidate display, and its supplier pass the evaluation.
2. The candidate display will be tested with a set of FDA-approved test methods and acceptance criteria, including a system-level test for the complete WSI system specified in the PCCP. The tests will be conducted by a qualified tester according to the quality management system. If there is a failure detected during the test, the failure(s) will be recorded, and the design change will not be implemented.
3. If the candidate display meets the acceptance criteria, the design change will be implemented according to the quality management system and the new display may be added as an interoperable component to the NanoZoomer S360 MD Slide scanner system without additional premarket review. The system labeling and company website will be revised to identify the new display according to the quality management system.
The similarities and differences between the subject and predicate devices are summarized in Table 1.
K233027 - Page 5 of 11
{5}
Table 1: Comparison with Predicate
| Device & Predicate Device(s): | K233027 | K213883 |
| --- | --- | --- |
| Device Trade Name | NanoZoomer S360MD Slide scanner system | Same |
| General Device Characteristic: Similarities | | |
| Intended Use/Indications For Use | The NanoZoomer S360MD Slide scanner system (“NanoZoomer System”) is an automated digital slide creation, viewing, and management system. The NanoZoomer System is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (“FFPE”) tissue. The NanoZoomer System is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. The NanoZoomer System comprises the NanoZoomer S360MD Slide scanner, the NZViewMD Software and a compatible display that has been 510(k) cleared for use with the NanoZoomer system or a 510(k)-cleared display that has been assessed in accordance with the | The NanoZoomer S360MD Slide scanner system (“NanoZoomer System”) is an automated digital slide creation, viewing, and management system. The NanoZoomer System is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (“FFPE”) tissue. The NanoZoomer System is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. The NanoZoomer System comprises the NanoZoomer S360MD Slide scanner, the NZViewMD Software and the JVC Kenwood JD- C240BN01A display. The NanoZoomer System is for creation and viewing of digital images of scanned glass slides that would |
K233027 - Page 6 of 11
{6}
| Device & Predicate Device(s): | K233027 | K213883 |
| --- | --- | --- |
| | Predetermined Change Control Plan (PCCP) for qualifying additional compatible displays. The NanoZoomer System is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using NanoZoomer System. | otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using NanoZoomer System. |
| Slide Feeder | 360 Slides | 360 Slides |
| Scanner | NanoZoomer S360MD Slide scanner | NanoZoomer S360MD Slide scanner |
| Viewer | NZViewMD | NZViewMD |
| General Device Characteristic: Differences | | |
| Display | • JVC Kenwood JD-C24BN01A
• BARCO MDPC-8127 Display | • JVC Kenwood JD-C24BN01A |
VI Standards/Guidance Documents Referenced:
1. FDA Guidance document: Technical Performance Assessment of Digital Pathology Whole Slide Imaging Devices. Guidance for Industry and Food and Drug Administration Staff. April 20, 2016.
2. ISO/CIE 11664-1, Colorimetry - Part 1: CIE standard colorimetric observers
K233027 - Page 7 of 11
{7}
3. ISO/CIE 11664-2, Colorimetry - Part 2: CIE standard illuminants
4. ISO/CIE 11664-3, Colorimetry - Part 3: CIE tristimulus values
5. ISO/CIE 11664-4, Colorimetry - Part 4: CIE 1976 L*a*b* colour space
6. ISO/CIE 11664-5, Colorimetry - Part 5: CIE 1976 L*u*v* colour space and u'v' uniform chromaticity scale diagram
7. ISO/CIE 11664-6, Colorimetry - Part 6: CIEDE2000 colour-difference formula
8. ISO/CIE 61966-2-1, Multimedia systems and equipment - Colour measurement and management - Part 2-1: Colour management - Default RGB colour space - sRGB
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
1. Precision/Reproducibility: Not applicable
2. Linearity: Not applicable
3. Analytical Specificity/Interference: Not applicable
4. Assay Reportable Range: Not applicable
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): Not applicable
6. Detection Limit: Not applicable
7. Assay Cut-Off: Not applicable
8. Accuracy (Instrument): Not applicable
9. Carry-Over: Not applicable
### B Other Supportive Instrument Performance Characteristics Data:
#### Display Equivalency Study:
Technical performance testing for the Barco MDPC-8127 display was performed and compared with the JVC JD-C240BN01 display. The test results are summarized in Table 2 below.
K233027 - Page 8 of 11
{8}
Table 2: Display Equivalency Test
| Test | Test Method | Results | |
| --- | --- | --- | --- |
| | | Subject Device Barco MDPC-8127 | Predicate Device JVC JD-C240BN01 |
| 1. Spatial resolution | The measurement of the modulation transfer function based on 7.7 Effective Resolution, IDMS v1.03 (Slanted-Edge) | Both horizontal and vertical MTFs are greater than 85% at Nyquist frequency | Both horizontal and vertical MTFs are greater than 58% at Nyquist frequency |
| 2. Pixel defects | Measurements (counts) of pixel defects based on 7.6 Defective Pixels, IDMS v1.03 | Total number of bright and dark pixels ≤ 5 with a distance of 15 mm | Total number of pixel defects ≤ 5 |
| 3. Artifacts | Evaluate for image artifacts such as noise, periodic spatial irregularities (Moiré), or periodic temporal irregularities. 4.6 Artifacts & Irregularities, IDMS v1.03 | < 0.65% | Artifacts not observed |
| 4. Maximum luminance and contrast ratio | Measurements of the maximum luminance and contrast ratio based on 2.4 Vantage-Point Suite of Measurements, IDMS v1.03 | Maximum luminance ≥ 678.6 and Minimum luminance < 0.633 cd/m² Maximum calibrated luminance = 450 cd/m² Contrast ratio ≥ 1000:1 | Maximum luminance ≥ 162 cd/m² Minimum luminance < 0.162 cd/m² Contrast ratio ≥ 1000:1 |
| 5. Luminance uniformity | Measurements of the uniformity of the luminance across the display screen based on 8.1.2 Sampled Vintage- | Non-uniformity on 80% video level < 10% | Uniformity (white point): Non-uniformity < 30% |
K233027 - Page 9 of 11
{9}
K233027 - Page 10 of 11
| Test | Test Method | Results | |
| --- | --- | --- | --- |
| | | Subject Device Barco MDPC-8127 | Predicate Device JVC JD-C240BN01 |
| | Point Uniformity, IDMS v1.03 | | |
| 6. Grayscale | Measurement of the mapping between image values and the luminance based on 6.1 Gray scale, IDMS v1.03 | Maximum luminance deviation ≤ 1.4% | Max luminance deviation ≤ 10% |
| 7. Stability of luminance | Luminance characteristics measured with temperature and time based on 10. Temporal Measurements IDMS v1.03 | Deviation from maximum luminance (450 cd/m2): 0.44%
Maximum luminance deviation < 5% | Maximum luminance deviation ≤ 10% |
| 8. Bidirectional reflection distribution function | Measurements of the reflection coefficients based on 11 Reflection Measurements, IDMS v1.03 | Specular reflection coefficient: 1.90%
Diffuse reflection coefficient: 2.87% | Diffused reflection coefficient (Rd) ≤ 2.00 % |
| 9. Gray tracking | Chromaticity at different luminance levels based on 6.15 Gray-scale Color Changes, IDMS v1.03 | Δu’v’ < 0.01 | Δu’v’ < 0.0028 |
| 10. Color difference | Measurements of the color difference using a spectrometer based on 6.2 Primary Color Scales, IDMS v1.03 | Average color error < 1 ΔE_{00}
Maximum color error < 3 ΔE_{00} | Average color error < 1 ΔE_{00} |
| 11. Color gamut volume | Measurements of the color gamut volume by volume-color reproduction | 2D color gamut area with respect to sRGB: 137.1% | Vcrc with respect to sRGB: 97.3% |
{10}
| Test | Test Method | Results | |
| --- | --- | --- | --- |
| | | Subject Device Barco MDPC-8127 | Predicate Device JVC JD-C240BN01 |
| | capability based on 5.18.1 Relative Gamut Area, IDMS v1.03 | 2D color gamut area within sRGB: 99.6% | 2D color gamut area with respect to sRGB: 98.2% |
| 12. Temporal response | Measurements of the response by blurred edge time based on 10.2.3 Gray-to-Gray Response Time, IDMS v1.03 | Average: 5.01 ms | Average 18.66 ms (Blurred Edge Time) |
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The subject device, NanoZoomer S360MD Slide scanner system, has similar indications for use, the same intended use, and the same principles of operation as its predicate device, K213883. The technological characteristics are identical with the exception of the implementation of a PCCP, and this difference does not raise new questions of safety and effectiveness. Thus, the subject device is substantially equivalent.
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K233027 - Page 11 of 11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.