The Midmark Smart M9® and Smart M11® Sterilizers can be used in medical and dental offices, hospitals, clinics, nursing homes, laboratories, and other facilities to sterilize heat and moisture stable reusable items (including dental handpieces) that are compatible with steam sterilization. Refer to Standard Cycle Parameters for detailed information.
Device Story
Compact, portable tabletop steam sterilizers; utilize Steam Flush Pressure Pulse (SFPP) technology to remove residual air via controlled pressure pulses and steam flushes. Input: heat/moisture stable items loaded into stainless-steel chamber. Operation: operator selects cycle via 5-inch LCD touch screen; device automatically executes filling, heating, sterilization, venting, and drying phases. Output: sterilized items; electronic sterilization records stored internally on SD card. Connectivity: Digital Ecosystem Connectivity Module enables transfer of records to Midmark cloud for remote status/history viewing and software updates. Used in medical/dental offices/clinics by staff. Benefits: automates sterilization process; provides record-keeping and remote monitoring capabilities; ensures compliance with sterilization standards.
Clinical Evidence
No clinical data. Bench testing only. Performance evaluated against ANSI/AAMI ST55:2016, ANSI/UL 61010-1, IEC 61010-2-040, IEC 60601-1-2, and ASME Boiler and Pressure Vessel Code.
Indicated for sterilization of heat and moisture stable reusable items, including dental handpieces, compatible with steam sterilization in medical, dental, clinical, laboratory, and nursing home settings.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
Predicate Devices
Midmark and Ritter M9 and M11 UltraClave® Automatic Sterilizers and Ritter M9D AutoClave® Automatic Sterilizers (K163337)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 4, 2024
Midmark Corporation Adam Clutter Quality Systems and Regulatory Affairs Manager 60 Vista Drive Versailles, Ohio 45380
## Re: K233026
Trade/Device Name: Midmark Smart M9® Sterilizer, Midmark Smart M11® Sterilizer Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: Class II Product Code: FLE Dated: March 7, 2024 Received: March 8, 2024
Dear Adam Clutter:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Christopher K. Dugard -S
Christopher K. Dugard, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical
{2}------------------------------------------------
and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233026
Device Name Midmark Smart M9® Sterilizer Midmark Smart M11® Sterilizer
#### Indications for Use (Describe)
The Midmark Smart M9® and Smart M11® Sterilizers can be used in medical and dental offices, hospitals, clinics, nursing homes, laboratories, and other facilities to sterilize heat and moisture stable items (including dental handpieces) that are compatible with steam sterilization. Refer to Standard Cycle Parameters for detailed information.
Type of Use (*Select one or both, as applicable*)
| | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|--|------------------------------------------------------------|
| | <b>X</b> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Standard Cycle Parameters
| Cycle Type | Cycle Parameters<br>Temperature Minimum | Time | Pressure 1 Reference | Drying Time 2 | Items to be Sterilized<br>(Always follow the item manufacturer's recommendations for sterilization.) | M9 Maximum Capacity 4 | M11 Maximum Capacity 4 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|---------|-----------------------|---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Image: Wrapped Cycle Type | 132°C (270°F) | 4 min. | 186 kPa<br>(27.1 psi) | 30 min. | • Pouched or wrapped items manufacturers recommend for exposure at 132°C (270°F) for 4 minutes.<br>• Wrapped cassettes.<br>• Dental Handpieces (wrapped or unwrapped) | 3,629 grams (8.0 lb) or 8 handpieces (2 per tray) with other instruments 3,629 grams (8.0 lb) total | 4,082 grams (9.0 lb) or 8 handpieces (2 per tray) with other instruments 4,082 grams (9.0 lb) total |
| Image: Wrapped Cycle Type | 135°C (275°F) | 3 min. | 214 kPa<br>(31 psi) | 30 min. | • Pouched or wrapped items manufacturers recommend for exposure at 135°C (275°F) for 3 minutes.<br>• Wrapped cassettes.<br>• Dental Handpieces (wrapped or unwrapped) | 3,629 grams (8.0 lb) or 8 handpieces (2 per tray) with other instruments 3,629 grams (8.0 lb) total | 4,082 grams (9.0 lb) or 8 handpieces (2 per tray) with other instruments 4,082 grams (9.0 lb) total |
| Image: Delicate Cycle Type | 121°C (250°F) | 30 min. | 104 kPa<br>(15 psi) | 30 min. | • Textiles and surgical packs wrapped for sterilization.3<br>• Items, except liquids, manufacturers recommend for exposure at 121°C (250°F) for 30 minutes. | 3,629 grams (8.0 lb) or 590 grams (1.3 lb) textile load | 4,082 grams (9.0 lb) or 2 packs each @ 590 grams (1.3 lb) textile load |
| Image: Unwrapped Cycle Type | 132°C (270°F) | 3 min. | 186 kPa<br>(27.1 psi) | 30 min. | • Instruments loose on a tray.<br>• Open glass or metal canisters.<br>• Tubing not used in surgical procedures. (Max. length - 40 in and Min. inside diameter - .187 in)<br>• Loose items manufacturers recommend for exposure at 132°C (270°F) for 3 minutes.<br>Note: The sterility of unwrapped items is compromised on exposure to a non-sterile environment. | 3,629 grams (8.0 lb) | 4,082 grams (9.0 lb) |
| 1 The pressure is absolute. The pressure inside the chamber is the sum of the atmospheric pressure and the pressure shown in this table. The values are included for reference and may not be exact. | | | | | | | |
The pressures shown in this table are at sea level and a resure of saturated steam at the sterilization temperature The pressures on the sterilizer display may be higher.
2. Dry time can be changed from 5 to 60 minne can be changed from 1 to 5 minutes.) Refer to Standard Cycle Operation.
3. Allow a minimum of 6.4 mm (1/4 in) space between each pack and from the chamber wall.
4. The default dry time may need increased due to variation, wrapping materials and the environment to completely dry the chamber contents at these capacities.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "midmark" in a sans-serif font. Above the word is a logo consisting of two overlapping shapes. The shape on the left is blue, and the shape on the right is pink. The two shapes together form an "M" shape.
Designing better care."™
Midmark Corporation 60 Vista Drive Versailles, Ohio 45380 1-800-MIDMARK midmark.com
As required by the Safe Medical Devices Act (SMDA) of 1990 and in accordance with 21 CFR §807.92, a 510(k) summary is provided below with the required information.
#### Table 1. Administrative Information Date Summary March 28, 2024 Prepared 510(k) Sponsor Midmark Corporation Address 60 Vista Drive Versailles, Ohio 45380 Contact Person Adam Clutter Quality Systems and Regulatory Affairs Manager Email: aclutter@midmark.com Phone: 937-526-8474 Fax: 937-526-8482 Midmark Smart M9® Sterilizer, Trade Name Midmark Smart M11® Sterilizer Common Name Table Top Steam Sterilizer Classification Name 21 CFR §880.6880 Steam sterilizer (Product Code FLE) Device Classification Class II Review Panel General Hospital
#### 1. Administrative Information
#### 2. Equivalent Predicate Comparators
| Table 2. Predicate Device | | | |
|---------------------------|------------|------------|---------------|
| Manufacturer Name | Trade Name | 510(k) No. | Decision Date |
{6}------------------------------------------------
| Table 2. Predicate Device | | | |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------|---------|-----------------|
| Midmark Corporation | Midmark and Ritter M9 and M11<br>UltraClave® Automatic Sterilizers and<br>Ritter M9D AutoClave® Automatic<br>Sterilizers | K163337 | August 31, 2017 |
#### 3. Indications for Use
The Midmark Smart M9® and Smart M11® Sterilizers can be used in medical and dental offices, hospitals, clinics, nursing homes, laboratories, and other facilities to sterilize heat and moisture stable reusable items (including dental handpieces) that are compatible with steam sterilization. Refer to Standard Cycle Parameters for detailed information.
#### Standard Cycle Parameters
| Cycle Type | Temperature<br>Minimum | Time | Pressure 1<br>Reference | Drying<br>Time 2 | Items to be Sterilized<br>(Always follow the item manufacturer's<br>recommendations<br>for sterilization.) | M9<br>Maximum<br>Capacity 4 | M11<br>Maximum<br>Capacity 4 |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|---------|-------------------------|------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| Wrapped | 132°C (270°F) | 4 min. | 186 kPa<br>(27.1 psi) | 30 min. | • Pouched or wrapped items manufacturers<br>recommend for exposure at 132°C (270°F)<br>for 4 minutes.<br>• Wrapped cassettes.<br>• Dental Handpieces (wrapped or unwrapped) | 3,629 grams (8.0 lb)<br>or 8 handpieces<br>(2 per tray)<br>with other<br>instruments 3,629<br>grams (8.0 lb) total | 4,082 grams (9.0 lb)<br>or 8 handpieces<br>(2 per tray)<br>with other<br>instruments 4,082<br>grams (9.0 lb) total |
| Wrapped | 135°C (275°F) | 3 min. | 214 kPa<br>(31 psi) | 30 min. | • Pouched or wrapped items manufacturers<br>recommend for exposure at 135°C (275°F)<br>for 3 minutes.<br>• Wrapped cassettes.<br>• Dental Handpieces (wrapped or unwrapped) | 3,629 grams (8.0 lb)<br>or 8 handpieces<br>(2 per tray)<br>with other<br>instruments 3,629<br>grams (8.0 lb) total | 4,082 grams (9.0 lb)<br>or 8 handpieces<br>(2 per tray)<br>with other<br>instruments 4,082<br>grams (9.0 lb) total |
| Delicate | 121°C (250°F) | 30 min. | 104 kPa<br>(15 psi) | 30 min. | • Textiles and surgical packs wrapped for<br>sterilization.3<br>• Items, except liquids, manufacturers<br>recommend for exposure at 121°C (250°F)<br>for 30 minutes. | 3,629 grams (8.0 lb)<br>or<br>590 grams (1.3 lb)<br>textile load | 4,082 grams (9.0 lb)<br>or<br>2 packs each @<br>590 grams (1.3 lb)<br>textile load |
| Unwrapped | 132°C (270°F) | 3 min. | 186 kPa<br>(27.1 psi) | 30 min. | • Instruments loose on a tray.<br>• Open glass or metal canisters.<br>• Tubing not used in surgical procedures.<br>(Max. length - 40 in and Min. inside<br>diameter - .187 in)<br>• Loose items manufacturers recommend for<br>exposure at 132°C (270°F) for 3 minutes.<br>Note: The sterility of unwrapped items is<br>compromised on exposure to a non-sterile<br>environment. | 3,629 grams (8.0 lb) | 4,082 grams (9.0 lb) |
| 1. The pressures shown in this table are at sea level and are for reference only. These are the ideal pressure of saturated steam at the sterilization temperature.<br>The pressures on the sterilizer display may be higher.<br>2. Dry time can be changed from 5 to 60 minutes. (IUSS dry time can be changed from 1 to 5 minutes.) Refer to Standard Cycle Operation.<br>3. Allow a minimum of 6.4 mm (1/4 in) space between each pack and from the chamber wall.<br>4. The default dry time may need increased due to variations in load configuration, wrapping materials and the environment to completely dry the chamber | | | | | | | |
#### 4. Models
#### Table 3. Models
| Model Name | Model Number | Input Voltage |
|-------------------------------|--------------|---------------|
| Midmark Smart M9® Sterilizer | M9-050 | 115V |
| Midmark Smart M9® Sterilizer | M9-052 | 115V |
| Midmark Smart M11® Sterilizer | M11-050 | 115V |
{7}------------------------------------------------
## Midmark® Smart Sterilizers 510(k) Summary
| Table 3. Models | | |
|-----------------|--------------|---------------|
| Model Name | Model Number | Input Voltage |
| | M11-051 | 230V |
| | M11-052 | 115V |
#### 5. Device Description
The Midmark Smart M9® and Smart M11® Sterilizers utilize steam flush pressure pulse (SFPP) technology to achieve sterilization on heat and moisture stable items (including dental handbieces) that are compatible with steam sterilization.
The Midmark Smart M9® and Smart M11® Sterilizers are compact, self-contained portable units. They can be placed on any level support surface where an electrical outlet is available with no other installation required. The M9 sterilizer models are the smaller of the models with a 9 in (22.9 cm) diameter x 15 in (38.1 cm) deep stainless-steel chamber. The M 1 sterilizer models have an 11" (27.9 cm) diameter x 18" (45.7 cm) deep stainless-steel chamber.
The Midmark Smart M9® and Smart M11® Sterilizers use saturated steam at high pressure and temperature as the sterilizing agent to kill infectious bio-organisms on items placed in the chamber for processing. They use the dynamic Steam Flush Pressure Pulse (SFPP) cycle type for the pre-set cycles that include a 4-minute 270°F and a 3-minute 275°F cycle for wrapped or pouched instruments (including dental handpieces) and cassettes, a 30-minute 250°F cvcle for textiles and instrument packs requiring a lower temperature. and a 3-minute 270°F cycle for unwrapped instruments. The 3-minute 275°F is added to align with Table B.2 in AAMI TIR12:2020. In the SFPP type cycle, residual air is removed from the chamber and its contents by a series of controlled pressure pulses and steam flushes.
The Midmark Smart M9® and Smart M11® Sterilizers are designed to automate the sterilization process, to the extent possible, and the user interface on the subject models extends this capability to include sterilization record keeping. To use the sterilizer, the operator fills the water reservoir with water (distilled or purified) and loads the included trays with properly cleaned and prepped instruments for sterilization. The loaded trays are then placed inside the chamber, and the chamber door is manually closed by the operator. Based on the cycle parameters that are appropriate for the type of load being processed, the operator then selects the appropriate sterilization cycle on the user interface. On the subject units there are options for the user to enter load type and indicator information. Once the cycle is selected and the operator presses "Start," there is an option to capture the identification of the operator. The sterilizer then automatically performs all the operations necessary to complete the sterilization process without further interaction from the operator. The sterilization cycle is composed of several phases which include Filling, Heating, Sterilization, Venting, and Drying, Audible signals indicate cycle initiation, completion, and/or interruption, and the user interface provides visual communication of device status, operator instructions, and troubleshooting information. The LED light bar also provides an estimate of the cycle progression. At the conclusion of the cycle there is another option to capture the identification of the operator that unloads and approves or rejects the results of the sterilization cycle.
{8}------------------------------------------------
## Midmark® Smart Sterilizers 510(k) Summary
All sterilizer cycle and user maintenance (Routine Care) records are stored internally on an SD card. Midmark has developed a Digital Ecosystem Connectivity Module that is incorporated into the Midmark Smart M9® and Smart M11® Sterilizers which adds the ability to transfer electronic sterilization records to the Midmark cloud and to remotely view the current status of, or cycle history for multiple sterilizers. Optional features and settings that may be distributed to multiple sterilizers from the cloud include distribution of software updates to the connectivity module, compliance settings, and entering the results of biological indicator tests. Connectivity is not needed to perform the sterlizer's intended use, and connecting a sterilizer does not change the intended use. Device specifications are provided in Table 4.
| Table 4. Specifications | | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Specification | Smart M9® Sterilizer | Smart M11® Sterilizer | |
| Physical dimensions: | | | |
| Overall Length with Plug | 20.5 in (52.0 cm) | 23.5 in (59.7 cm) | |
| Overall Width | 15.8 in (40.1 cm) | 17.8 in (45.2 cm) | |
| Overall Height | 16.6 in (42.2 cm) | 18.6 in (47.2 cm) | |
| Required Support Surface | 19.8 in x 23.5 in<br>(50.3 cm x 59.7 cm) | 21.8 in x 26.5 in<br>(55.4 cm x 67.3 cm) | |
| Chamber Size | 9 in diameter x 15 in deep<br>(22.9 cm x 38.1 cm) | 11 in diameter x 18 in deep<br>(27.9 cm x 45.7 cm) | |
| Trays, Large | 7-5/16 in x 12 in x 7/8 in<br>(18.6 cm x 30.5 cm x 2.2 cm) | 9-1/8 in x 15-1/4 in x 7/8 in<br>(23 cm x 38.7 cm x 2.2 cm) | |
| Trays, Small | 5-5/8 in x 12 in x 7/8 in<br>(14.3 cm x 30.5 cm x 2.2 cm) | 6-3/4 in x 15-1/4 in x 7/8 in<br>(17 cm x 38.7 cm x 2.2 cm) | |
| Weights: | | | |
| Weight with Empty Reservoir | 82 lb. (37.2 kg) | 99 lb. (44.9 kg) | |
| Weight with Shipping Carton | 92 lb. (41.7 kg) | 131 lb. (59.4 kg) | |
| Water Reservoir Capacity | 1.1 gallons (4.1 liters) full,<br>0.6 gallons (2.3 liters) usable | 1.4 gallons (5.3 liters) full,<br>0.8 gallons (3.0 liters) usable | |
| Electrical Ratings: | | | |
| Note: A separate (dedicated) circuit is required for these sterilizers. Sterilizer should not be<br>connected into an electrical circuit with other appliances or equipment unless circuit is rated for<br>the additional load. | | | |
| 115 Volt Models | 115 VAC, 12 Amp, 50/60 Hz | 115 VAC, 12 Amp, 50/60 Hz | |
| 230 Volt Models | Not Applicable | 230 VAC, 6.4 Amp, 50/60 Hz | |
| Table 4. Specifications | | | |
| Specification | | Smart M9® Sterilizer | Smart M11® Sterilizer |
| 115 VAC | F1 | 15 Amp, 250 V, Fast Acting,<br>0.250 in Dia. x 1.250 in L<br>(6.35 mm x 31.75 mm) | 15 Amp, 250 V, Fast Acting,<br>0.250 in Dia. x 1.250 in L<br>(6.35 mm x 31.75 mm) |
| | F2 | 15 Amp, 250 V, Fast Acting,<br>0.250 in Dia. x 1.250 in L<br>(6.35 mm x 31.75 mm) | 15 Amp, 250 V, Fast Acting,<br>0.250 in Dia. x 1.250 in L<br>(6.35 mm x 31.75 mm) |
| | F3 | 2.5 Amp, 250 V, Time<br>Delay/Slow-Blow,<br>5.20 mm Dia. x 20.00 mm L<br>(0.205 in x 0.787 in) | 2.5 Amp, 250 V, Time<br>Delay/Slow-Blow,<br>5.20 mm Dia. x 20.00 mm L<br>(0.205 in x 0.787 in) |
| 230 VAC | F1 | Not Applicable | 8 Amp, 250 V, Fast Acting,<br>5.20 mm Dia. x 20.00 mm L<br>(0.205 in x 0.787 in) |
| | F2 | Not Applicable | 8 Amp, 250 V, Fast Acting,<br>5.20 mm Dia. x 20.00 mm L<br>(0.205 in x 0.787 in) |
| | F3 | Not Applicable | 2.5 Amp, 250 V, Time<br>Delay/Slow-Blow,<br>0.250 in Dia. x 1.250 in L<br>(6.35 mm x 31.75 mm) |
| Safety Valve Setting | | 43 psi (296 kPa) | 43 psi (296 kPa) |
| Heat Emission | | 5000 BTU / hr. during<br>operation | 5000 BTU / hr. during<br>operation |
| <b>Operating Environment:</b> | | | |
| Ambient Temperature Range | | 68°F to 104°F (+20°C to +40°C) | 68°F to 104°F (+20°C to +40°C) |
| Relative Humidity | | < 80% (non-condensing) | |
| Operating Altitude | | < 9842 ft (3000 m) above sea level | |
{9}------------------------------------------------
# 6. Technological Characteristics
Refer to Table 5 for a comparison of Midmark Smart M9® and Smart M11® Sterilizer technological characteristics with those of its predicate. Changes made to the Midmark Smart M9® and Smart M11® Sterilizers do not raise new issues of safety or effectiveness in terms of Indications for Use, design, material, chemical composition, principle of operation, energy source, and performance.
{10}------------------------------------------------
| Characteristic | Subject Device: Midmark<br>Smart M9® and Smart M11®<br>Sterilizers | Predicate Device:<br>Midmark and Ritter M9 and<br>M11 UltraClave® Automatic<br>Sterilizers | Comparison |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| 510(k)<br>Clearance | K233026 | K163337 | Not<br>Applicable |
| Regulation<br>Number | 21 CFR §880.6880 Steam<br>sterilizer | 21 CFR §880.6880 Steam<br>sterilizer | Same |
| Product Code | FLE | FLE | Same |
| Intended Use | The Midmark Smart M9® and<br>Smart M11® Sterilizers can be<br>used in medical and dental<br>offices, hospitals, clinics,<br>nursing homes, laboratories,<br>and other facilities to sterilize<br>heat and moisture stable<br>reusable items (including<br>dental handpieces) that are<br>compatible with steam<br>sterilization. Refer to Standard<br>Cycle Parameters for detailed<br>information. | The Midmark and Ritter M9,<br>M9D and M11 Steam Sterilizers<br>can be used in medical and<br>dental offices, hospitals,<br>clinics, nursing homes,<br>laboratories, and other<br>facilities to sterilize heat and<br>moisture stable reusable items<br>(including dental<br>handpieces) that are<br>compatible with steam<br>sterilization. Refer to<br>Guidelines for loading and<br>Standard Cycle Parameters<br>for detailed information. | Same |
| Standard Cycle<br>Parameters | Wrapped 1 Cycle<br>4 min./ 132°C (270°F) / 30 min. | Pouches Cycle<br>4 min./ 132°C (270°F) / 30 min.<br>Handpiece Cycle<br>4 min./ 132°C (270°F) / 30 min. | Same |
| | Wrapped 2 Cycle<br>3 min./ 135°C (275°F) / 30 min. | Not Available | Different |
| | Delicate Cycle<br>30 min./ 121°C (250°F) / 30<br>min. | Packs Cycle<br>30 min./ 121°C (250°F) / 30<br>min. | Same |
| Unwrapped Cycle<br>3 min./ 132°C (270°F) / 30 min. | Unwrapped Cycle<br>3 min./ 132°C (270°F) / 30 min. | Same | |
| Dry Time | 5 to 60 minutes.<br>(Immediate Use Steam<br>Sterilization dry time can be<br>changed from 1 to 5 minutes.) | 0 to 60 minutes. | Same |
| Table 5. Predicate Comparison | | | |
| Characteristic | Subject Device: Midmark<br>Smart M9® and Smart M11®<br>Sterilizers | Predicate Device:<br>Midmark and Ritter M9 and<br>M11 UltraClave® Automatic<br>Sterilizers | Comparison |
| Wrapped 1<br>Cycle | 3,629 grams (8.0 lb.)<br>or 8 handpieces<br>(2 per tray) with other<br>instruments 3,629 grams (8.0<br>lb.) total | 8 handpieces with other<br>instruments<br>3629 grams (8.0 lb.) | Same |
| Wrapped 2<br>Cycle | 3,629 grams (8.0 lb.)<br>or 8 handpieces (2 per tray)<br>with other instruments 3,629<br>grams (8.0 lb.) total | Not available | Different |
| Delicate Cycle | 3,629 grams (8.0 lb.)<br>or<br>590 grams (1.3 lb.)<br>textile load | 4752 cm3<br>(290 in3) | Similar |
| Unwrapped<br>Cycle | 3,629 grams (8.0 lb.) | 3,629 grams (8.0 lb.) | Same |
| M11 Maximum Capacity | | | |
| Wrapped 1<br>Cycle | 4,082 grams (9.0 lb.)<br>or 8 handpieces<br>(2 per tray)<br>with other<br>instruments 4,082<br>grams (9.0 lb.) total | 8 handpieces with other<br>instruments<br>4,082 grams (9.0 lb.) | Same |
| Wrapped 2<br>Cycle | 4,082 grams (9.0 lb.)<br>or 8 handpieces<br>(2 per tray)<br>with other<br>instruments 4,082<br>grams (9.0 lb.) total | Not Available | Different |
| Delicate Cycle | 4,082 grams (9.0 lb.)<br>or<br>2 packs each @<br>590 grams (1.3 lb.)<br>textile load | 505 in3<br>(8275 cm3) | Similar |
| Table 5. Predicate Comparison | | | |
| Characteristic | Subject Device: Midmark<br>Smart M9® and Smart M11®<br>Sterilizers | Predicate Device:<br>Midmark and Ritter M9 and<br>M11 UltraClave® Automatic<br>Sterilizers | Comparison |
| Unwrapped<br>Cycle | 4,082 grams (9.0 lb.) | 4,082 grams (9.0 lb.) | Same |
| Controls, Features, and Data Records | | | |
| Sterilizer Control | Microcontroller based<br>embedded control | Microcontroller based<br>embedded control | Same |
| User Interface | LCD Touch Screen, 5-inch, 800<br>x 480 color display<br>Glass cover | Membrane switch with LCD<br>display (2 row x 20 character)<br>Polyester overlay | Different |
| | Operator Alerts | Audible signals<br>Visual signals via touch screen<br>display and a full-width LED<br>light bar | Audible signals<br>Visual signals via LCD display |
| Data Records | | Built-in communication<br>module with wireless and<br>ethernet available network<br>connections; optional USB<br>thumb drive accessory | Optional add-on Data Logger<br>with USB thumb drive;<br>optional thermal printer |
| Water Fill System | Gravity fill with Chamber<br>Water Level Sensor | Gravity fill with Chamber<br>Water Level Sensor | Same |
| | Optional autofill solenoid kit | Not available | Different |
| Reservoir Water<br>Level Detection | Visual Indicator | Visual indicator | Same |
| | PCBA Mounted Pressure<br>Transducer | Not included | Different |
| Chamber<br>Pressure Sensor | PCBA Mounted Gauge<br>Pressure Transducer | PCBA Mounted Gauge<br>Pressure Transducer | Same…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.