The pulse oximeter is a reusable device and intended for spot-checking of oxygen saturation and pulse rate for use with the finger of adult patients in home and healthcare environments. And it is not intended to be used under motion or low perfusion scenarios.
Device Story
Battery-powered pulse oximeter for spot-checking SpO2 and pulse rate; used in home and healthcare environments by patients or clinicians. Device applies sensor to finger; dual light source (660 nm red, 905 nm infrared) transmits through tissue; photodetector measures light absorption ratio between systole and diastole to calculate SpO2. Components include PCB, OLED/LED screen, and plastic shell. Models AOJ-70D and AOJ-70E include Bluetooth connectivity for data transmission to non-device apps. Output displayed on screen or transmitted wirelessly; assists healthcare providers in monitoring oxygenation status. Not for life-supporting or life-sustaining use.
Clinical Evidence
Clinical data referenced from K221039. Study conducted at Sir Run Run Shaw Hospital with 11 healthy adult volunteers (ages 21-47, Fitzpatrick skin tones 2-6). Evaluated SpO2 accuracy against arterial blood CO-Oximetry during stationary conditions (70-100% range). Results showed ARMS < 2%.
Technological Characteristics
Battery-powered; dual-wavelength (660 nm/905 nm) optical sensor; PCB-based electronics; OLED/LED display; plastic housing. Bluetooth connectivity in specific models. Complies with IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, and ISO 80601-2-61. Software documentation level: Basic.
Indications for Use
Indicated for spot-checking oxygen saturation (SpO2) and pulse rate in adult patients in home and healthcare settings. Not for use during motion or low perfusion scenarios.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
AOJ-70A, AOJ-70B, AOJ-70C, AOJ-70D and AOJ-70E Pulse (K202173)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right, the FDA logo is displayed in blue, with the words "U.S. FOOD & DRUG" stacked above the word "ADMINISTRATION". The logo is clean and professional, reflecting the organization's role in regulating food and drugs.
July 1, 2024
Shenzhen AOJ Medical Technology Co., Ltd Jack Wang Deputy Chief Rm 301&4F, Block A, Bldg A,ingfa Intelligent Manufacturing Park,Xiaweiyuan,Gushu Community,Xixiang St, Bao'an District Shenzhen, Guangdong 518126 China
Re: K232989
Trade/Device Name: Pulse Oximeter (AOJ-70A); Pulse Oximeter (AOJ-70B); Pulse Oximeter (AOJ-70C); Pulse Oximeter (AOJ-70D); Pulse Oximeter (AOJ-70E) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: October 11, 2023 Received: October 11, 2023
Dear Jack Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}------------------------------------------------
See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bradley Q. Quinn -S
Bradley Quinn Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
Submission Number (if known)
K232989
Device Name Pulse Oximeter (AOJ-70A); Pulse Oximeter (AOJ-70B); Pulse Oximeter (AOJ-70C); Pulse Oximeter (AOJ-70D); Pulse Oximeter (AOJ-70E)
Indications for Use (Describe)
The pulse oximeter is a reusable device and intended for spot-checking of oxygen saturation and pulse rate for use with the finger of adult patients in home and healthcare environments. And it is not intended to be used under motion or low perfusion scenarios.
Type of Use (Select one or both. as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) #: | K232989 | 510(k) Summary | Prepared on: 2024-05-29 |
|-----------|---------|----------------|-------------------------|
|-----------|---------|----------------|-------------------------|
| Contact Details | | 21 CFR 807.92(a)(1) |
|-----------------|--|---------------------|
|-----------------|--|---------------------|
| Applicant Name | Shenzhen AOJ Medical Technology Co., Ltd |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant Address | Room 301&4F, Block A, Building A,ingfa Intelligent Manufacturing Park,<br>Xiaweiyuan, Gushu Community, Xixiang Street, Bao'an District, Shenzhen<br>Guangdong 518126 China |
| Applicant Contact Telephone | 86 755-27786026 |
| Applicant Contact | Mr. Jack Wang |
| Applicant Contact Email | regulation@aojmedical.com |
| Device Name | | 21 CFR 807.92(a)(2) |
|-------------|--|---------------------|
|-------------|--|---------------------|
| Device Trade Name | Pulse Oximeter (AOJ-70A);<br>Pulse Oximeter (AOJ-70B);<br>Pulse Oximeter (AOJ-70C);<br>Pulse Oximeter (AOJ-70D);<br>Pulse Oximeter (AOJ-70E) |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Oximeter |
| Classification Name | Oximeter |
| Regulation Number | 870.2700 |
| Product Code | DQA |
| Legally Marketed Predicate Devices | | |
|------------------------------------|----------------------------------------------------------|--------------|
| 21 CFR 807.92(a)(3) | | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K202173 | AOJ-70A, AOJ-70B, AOJ-70C, AOJ-70D and AOJ-70E Pulse | DQA |
| K200414 | Pulse Oximeter, A310B, A330B | DQA |
| Device Description Summary | 21 CFR 807.92(a)(4) |
|----------------------------|---------------------|
|----------------------------|---------------------|
## Device Description Summary
The Pulse Oximeter is a battery powered device in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR).
21 CFR 807.92(a)(4)
The Pulse Oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660 m, which is red light; the other is 905 nm, which is Infrared light. Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photodetector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.
The device mainly composed of PCB board, On/Off button, mode button, OLED&LED screen, battery compartment, Bluetooth® module and plastic shell. There are five models AOJ-70A, AOJ-70C, AOJ-70E. Only AOJ-70E. Only AOJ-70D have wireless connection function via Bluetooth®.
The device is a spot-check pulse oximeter and does not intended for life-supporting or life-sustaining.
{5}------------------------------------------------
# Intended Use/Indications for Use
The pulse oximeter is a reusable device and intended for spot-checking of oxygen saturation and pulse rate for use with the finger of adult patients in home and healthcare environments. And it is not intended to be used under motion or low perfusion scenarios.
# Indications for Use Comparison
The indication for use of the subject device is extended for home environment and is the secondary predicate device.
# Technological Comparison
The subject device is a modification device cleared by K221039.The modifications are add Bluetooth communication to a non-device APP and enclosure and display screen changes.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The following performance data were referenced in support of the substantial equivalence determination. Biocompatibility testing
The biocompatibility testing is referenced to K221039.
Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the oximeters. The device complies with the IEC 60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance for safety, IEC 60601-1-11 MEDICAL ELECTRICAL EQUIPMENT – Part 1-11: General requirements for basic safety and essential performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment, and the lectrical equipment -Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests standard for EMC.
Bench Testing
Bench testing was conducted on the oximeters which complies with the ISO 80601-2-61 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.
Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software for this device was considered as a "Basic" Documentation level.
The clinical data is referenced to K221039.The Clinical Study was conducted at Sir Run Run Shaw Hospital, Zhejiang University School of Medicine. The study was conducted in accordance to ISO 14155-1,ISO 80601-2-61:2017, and the FDA Guidance Document for Pulse Oximeters. The subject device of this study was to evaluate the SpO2 accuracy performance of the subject device during stationary (nonmotion) conditions over a wide range of arterial blood oxygen saturation levels as compared to arterial blood CO-Oximetry. After Institutional Review Board (IRB) approval, 11 healthy adult volunteer subjects (ages 21-47yr, with skin tones varying from Fitzpatrick 2-6) were included in the study conducted November 2021 to evaluate the SpO2 accuracy performance. The SpO2 accuracy performance results showed that the subject device have an ARMS<2 during steady state conditions over the range of 70-100%.
Based on the non-clinical performance data as documented in the device development, the subject devices were found to be as safe, as effective, and perform as well as the legally marketed devices identified above.
## 21 CFR 807.92(a)(5)
### 21 CFR 807.92(a)(6)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.