K232981 · Synaptive Medical, Inc. · LNH · Oct 11, 2023 · Radiology
Device Facts
Record ID
K232981
Device Name
Synq Software Version 1.3
Applicant
Synaptive Medical, Inc.
Product Code
LNH · Radiology
Decision Date
Oct 11, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Synq Software is indicated for use in conjunction with Synq, a magnetic resonance diagnostic device (MRDD) that produces axial, sagittal, coronal, and oblique cross-sectional images and displays the internal structure and/or function of the head. Depending on the region of interest, contrast agents may be used. These images when interpreted by a trained physician yield information that may assist in diagnosis.
Device Story
Synq Software v1.3 is an upgrade to the Synq (formerly EVRY) MRDD system. It coordinates magnetic field, gradient, and RF transmitter/receiver coil interactions to generate cross-sectional images representing proton spin distribution. Operated by qualified medical professionals in clinical settings, the software enables configuration and initiation of MR scans. New features include additional pulse sequences (e.g., 3D T1 CSE ME-GRE, 2D Spin Echo EPI, DTI, 3D SWI) and protocol updates. Output images are interpreted by radiologists to assist in clinical diagnosis. The device benefits patients by providing diagnostic imaging of head structures and functions.
Clinical Evidence
No clinical trials performed. Substantial equivalence established via bench testing, image performance testing, and safety testing meeting predefined acceptance criteria. A representative subset of clinical images was provided for review, supported by an attestation from a U.S. Board-certified radiologist.
Technological Characteristics
Software-based control system for MRDD. Coordinates magnetic field, gradients, and RF hardware. Implements various pulse sequences (e.g., 3D T1 CSE ME-GRE, 2D Spin Echo EPI, DTI, 3D SWI). Complies with NEMA MS 1-5, 8-9, IEC 62464-1, and IEC 60601-2-33 standards.
Indications for Use
Indicated for use with Synq MRDD to produce cross-sectional images (axial, sagittal, coronal, oblique) of the head for diagnostic assistance by trained physicians. Contrast agents may be used.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
October 11, 2023
Synaptive Medical Inc. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K232981
Trade/Device Name: Syng Software Version 1.3 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH Dated: September 21, 2023 Received: September 21, 2023
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
Submission Number (if known)
K232981
Device Name
Syng Software Version 1.3
#### Indications for Use (Describe)
The Synq Software is indicated for use in conjunction with Syng, a magnetic resonance diagnostic device (MRDD) that produces axial, sagittal, coronal, and oblique cross-sectional images and displays the internal structure and/or function of the head. Depending on the region of interest, contrast agents may be used. These images when interpreted by a trained physician yield information that may assist in diagnosis.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) #: | K232981 | | 510(k) Summary | | Prepared on: 2023-09-28 |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|--------------|----------------|--|-------------------------|
| Contact Details | | | | | 21 CFR 807.92(a)(1) |
| Applicant Name | Synaptive Medical Inc. | | | | |
| Applicant Address | 555 Richmond St West, Suite 800 Toronto Ontario M5V 3B1 Canada | | | | |
| Applicant Contact Telephone | +1-647-243-3334 | | | | |
| Applicant Contact | Mr. Ahmed Hamed | | | | |
| Applicant Contact Email | ahmed.hamed@synaptivemedical.com | | | | |
| Device Name | 21 CFR 807.92(a)(2) | | | | |
| Device Trade Name | Synq Software Version 1.3 | | | | |
| Common Name | Magnetic resonance diagnostic device | | | | |
| Classification Name | System, Nuclear Magnetic Resonance Imaging | | | | |
| Regulation Number | 892.1000 | | | | |
| Product Code | LNH | | | | |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) | | | | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | | |
| K200327 | Evry | LNH | | | |
| Device Description Summary | 21 CFR 807.92(a)(4) | | | | |
| The Synq Software allows a user to configure and initiate a magnetic resonance scan of a subject. In doing so, the software coordinates the interactions of the magnetic field, gradients, radio frequency (RF) transmitter and receiver coil in Synq (Previously known as EVRY, K200327) to produce axial, sagittal, coronal, and oblique cross-sectional images that represent the spatial distribution of protons with spin. | | | | | |
| The Synq Software Version 1.3 upgrades the current software version in the Synq system to include additional imaging applications, functionality, and minor bug fixes. | | | | | |
| The Software should be used only by qualified medical professionals who are trained in magnetic resonance diagnostic devices. | | | | | |
| Intended Use/Indications for Use | 21 CFR 807.92(a)(5) | | | | |
| The Synq Software is indicated for use in conjunction with Synq, a magnetic resonance diagnostic device (MRDD) that produces axial, sagittal, coronal, and oblique cross-sectional images and displays the internal structure and/or function of the head. Depending on the region of interest, contrast agents may be used. These images when interpreted by a trained physician yield information that may assist in diagnosis. | | | | | |
## Indications for Use Comparison
The indications for use in the subject device is the same as the predicate device:
The Synq Software is indicated for use in conjunction with Synq, a magnetic resonance diagnostic device (MRDD) that produces axial,
21 CFR 807.92(a)(5)
{4}------------------------------------------------
sagittal, coronal, and oblique cross-sectional images and displays the internal structure and/or function of the region of interest, contrast agents may be used. These images when interpreted by a trained physician year may assist in diagnosis.
# Technological Comparison
### 21 CFR 807.92(a)(6)
The Subject Device is an updated version of the Predicate Device and any differences are only additive and do not raise concerns on safety and effectiveness.
The substantive changes in the Subject Device are as follows:
· New pulse sequences, including: o 3D T1 CSE ME-GRE (3D Fat Sat T1) o 2D Spin Echo EPI o 2D T2 CSE Spin Echo (2D Fat Sat T2) o 3D T2 Spin Echo o 2D Spin Echo EPI FLAIR o 3D T2 FLAIR o Diffusion Tensor Imaging (DTI) o 3D T2* ME-GRE (3D SWI)
· Minor updates to existing protocols
· Updates to various SOUP/OTS packages
· Minor bug fixes
As such, functionality in the software of the Predicate Device remains substantially unchanged in the Subject Device.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
As per Bench Testing document attached under Bench Testing the image performance testing and safety testing meet all predefined acceptance criteria. Together, with an attestation from a U.S. Board certified radiologist, demonstrate substantial equivalence to the predicate device (EVRY K200327) by conforming to FDA recognized standards and addressing all requirements in FDA MRDD Guidance. Applicable Standards and Guidance include: NEMA MS 1 NEMA MS 2 NEMA MS 3 NEMA MS 4 NEMA MS 5 NEMA MS 8 NEMA MS 9 IEC 62464-1 IEC-60601-2-33 Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices
A small, representative subset of clinical images are provided along with this 510(k) submission, as per quidance document "Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices" issued November 18, 2016. Synq Software Version 1.3 did not require clinical tests since substantial equivalence to the legally marketed predicate device was proven with the verification and validation testing.
The conclusions drawn from the nonctrate that the device is as safe, as effective, and performs as intended, and therefore, substantially equivalent to the legally marketed device (Evry, K200327). Synq Software Version 1.3 did not require clinical tests since substantial equivalence to the legally marketed predicate device was proven with the verification and validation testing.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.