External evaluation with fresh and stock clinical isolates and challenge strains
—
Indications for Use
VITEK 2 AST-Yeast Voriconazole is designed for antifungal susceptibility testing of Candida species and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antifungal agents. VITEK 2 AST-Yeast Voriconazole is a quantitative test. Voriconazole has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antifungal. Active in vitro and in clinical infections: Candida krusei Candida parapsilosis Candida tropicalis The VITEK 2 Fungal Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant yeasts to antifungal agents when used as instructed.
Device Story
VITEK 2 AST-Yeast Voriconazole is an automated, quantitative antifungal susceptibility test. It uses a miniaturized 64-well card containing premeasured Voriconazole and nutrient media. Clinical laboratory technicians prepare an organism suspension (saline) via auto-dilution (VITEK 2) or manual dilution (VITEK 2/VITEK 2 Compact). The system fills, seals, and incubates the card, monitoring growth via optical scanner (light attenuation). The system's analysis software determines the Minimum Inhibitory Concentration (MIC) and provides an interpretive category result. Results assist clinicians in selecting appropriate antifungal therapy for Candida infections. The device is intended for professional use in clinical microbiology laboratories.
Clinical Evidence
Performance evaluated using 535 total isolates (459 clinical, 76 challenge). Comparison against CLSI broth microdilution reference method. Overall Essential Agreement (EA) and Category Agreement (CA) were acceptable for C. krusei, C. parapsilosis, and C. tropicalis. C. albicans showed unacceptable performance (high error rates) and was removed from indications. Trending analysis identified MIC biases for specific species, addressed via labeling footnotes. Quality control testing met acceptance criteria.
Technological Characteristics
Miniaturized 64-well test card; automated growth-based detection via visible light transmittance optics; 15-minute measurement intervals. Dimensions/form factor: VITEK 2 AST card. Connectivity: Networked VITEK 2/VITEK 2 Compact systems. Software: VITEK 2 Systems 9.04 or later. Sterilization: Not specified.
Indications for Use
Indicated for antifungal susceptibility testing of Candida krusei, Candida parapsilosis, and Candida tropicalis in clinical laboratories using VITEK 2 or VITEK 2 Compact systems.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K232967
B Applicant
bioMérieux, Inc.
C Proprietary and Established Names
VITEK 2 AST-Yeast Voriconazole (≤ 0.03125 – ≥4 μg/mL)
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| NGZ | Class II | 21 CFR 866.1640 - Antimicrobial Susceptibility Test Powder | MI - Microbiology |
| LON | Class II | 21 CFR 866.1645 - Fully automated short-term incubation cycle antimicrobial susceptibility system | MI - Microbiology |
| LRG | Class II | 21 CFR 866.1640 - Antimicrobial susceptibility test powder | MI - Microbiology |
| LTT | Class II | 21 CFR 866.1640 - Antimicrobial susceptibility test powder | MI - Microbiology |
| LTW | Class II | 21 CFR 866.1640 - Antimicrobial susceptibility test powder | MI - Microbiology |
## II Submission/Device Overview:
### A Purpose for Submission:
To obtain a substantial equivalence determination for Voriconazole testing of Candida species on the VITEK 2 and VITEK 2 Compact Antimicrobial Susceptibility Test Systems.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
B Measurand:
Voriconazole (≤ 0.03125 - ≥ 4 μg/mL)
C Type of Test:
Automated quantitative antifungal susceptibility test
III Intended Use/Indications for Use:
A Intended Use(s):
The VITEK 2 Fungal Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant yeasts to antifungal agents when used as instructed.
B Indication(s) for Use:
VITEK 2 AST-Yeast Voriconazole is designed for antifungal susceptibility testing of Candida species and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antifungal agents. VITEK 2 AST-Yeast Voriconazole is a quantitative test. Voriconazole has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antifungal.
Active in vitro and in clinical infections:
Candida krusei
Candida parapsilosis
Candida tropicalis
The VITEK 2 Fungal Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant yeasts to antifungal agents when used as instructed.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
The ability of the AST card to detect resistance with the following combination(s) is unknown because resistant strains were not available at the time of comparative testing:
Voriconazole (vrc02n): Candida spp.
Perform an alternative method of testing prior to reporting of results for the following antibiotic/organism combination:
Voriconazole (vrc02n): C. albicans
D Special Instrument Requirements:
For use with the VITEK 2 and VITEK 2 Compact Systems using VITEK 2 Systems 9.04 software (or later)
K232967 - Page 2 of 12
{2}
K232967 - Page 3 of 12
# IV Device/System Characteristics:
## A Device Description:
The VITEK 2 AST card is a miniaturized, abbreviated and automated version of the doubling dilution technique for determining the minimum inhibitory concentration (MIC). Each VITEK 2 AST card contains 64 wells. A control well(s) which contain only nutrient medium is resident on all cards. The remaining wells contain premeasured portions of antimicrobials combined with the nutrient media. The isolate to be tested is diluted to a standardized concentration with 0.45 to 0.5% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System will automatically (or allow operator to manually) dilute the organism suspension to prepare an inoculum for susceptibility cards. Then, the VITEK 2 will fill, seal and place the card into the incubator/reader. The VITEK 2 Compact has a manual filling, sealing, and loading operation. The VITEK 2 Systems monitor the growth of each well in the card over a defined period of time. The analysis program determines when a well demonstrates growth based on attenuation of light measured by an optical scanner. This data is used to determine the minimum inhibitory concentration or "MIC" values for the antimicrobial agent. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.
VITEK 2 AST-YS Voriconazole has the following concentrations in the card: 0.03125, 0.125, 0.25, 1, and 2 (equivalent standard method concentration by efficacy in µg/mL). The Voriconazole MIC result range for the VITEK 2 card is ≤0.03125 – ≥4 µg/mL.
## B Principle of Operation:
The VITEK 2 and VITEK 2 Compact Systems utilize automated growth-based detection using attenuation of light measured by an optical scanner. The optics in the systems use visible light to directly measure organism growth within each of the 64 micro-wells. Transmittance optics is based on an initial light reading of a well before significant growth has begun. Every 15 minutes throughout the incubation cycle (defined period of time based on the VITEK 2 card), light transmittance readings of each well determine organism growth by the amount of light that is prevented from passing through the well. At the completion of the incubation period, the MIC values and their associated interpretive category results for each antimicrobial on the test card are displayed in an automatically generated report.
# V Substantial Equivalence Information:
## A Predicate Device Name(s):
VITEK 2 AST-YS Fluconazole (≤ 0.5 - ≥ 64 µg/mL)
## B Predicate 510(k) Number(s):
K213241
{3}
C Comparison with Predicate(s):
| Device & Predicate Device(s): | Device: K232967 | Predicate: K213241 |
| --- | --- | --- |
| Device Trade Name | VITEK 2 AST-Yeast Voriconazole (≤ 0.03125 - ≥ 4 μg/mL) | VITEK 2 AST-YS Fluconazole (≤ 0.5 - ≥ 64 μg/mL) |
| General Device Characteristic Similarities | | |
| Indications for Use | VITEK 2 AST-Yeast Voriconazole is designed for antifungal susceptibility testing of Candida species and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antifungal agents. VITEK 2 AST-Yeast Voriconazole is a quantitative test. Voriconazole has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antifungal.
The VITEK 2 Fungal Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant yeasts to antifungal agents when used as instructed. | VITEK 2 AST-Yeast Fluconazole is designed for antifungal susceptibility testing of Candida species and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antifungal agents. VITEK 2 AST-Yeast Fluconazole is a quantitative test. Fluconazole has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antifungal.
The VITEK 2 Fungal Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant yeasts to antifungal agents when used as instructed. |
| Test Methodology | Automated quantitative antimicrobial susceptibility test for use with the VITEK 2 and VITEK 2 Compact Systems to determine the in vitro susceptibility of yeast. | Same |
| Inoculum | Saline suspension of organism | Same |
| Test card | VITEK 2 Yeast Susceptibility Test Card | Same |
| Instrument | VITEK 2 and VITEK 2 Compact Systems | Same |
| Analysis Algorithms | Discriminant Analysis | Same |
| General Device Characteristic Differences | | |
| Claimed Species | Active in vitro and in clinical infections: Candida krusei Candida parapsilosis Candida tropicalis | Active in vitro and in clinical infections: Candida albicans Candida parapsilosis Candida tropicalis |
K232967 - Page 4 of 12
{4}
| Device & Predicate Device(s): | Device: K232967 | Predicate: K213241 |
| --- | --- | --- |
| Antifungal Agent | Voriconazole | Fluconazole |
| Antimicrobial Concentrations (μg/mL) | 0.03125, 0.125, 0.25, 1, 2 | 2, 4, 8, 16, 32, 64 |
VI Standards/Guidance Documents Referenced:
CLSI M27M44S, “Performance Standards for Antifungal Susceptibility Testing of Yeasts” Third Edition (August 2022)
CLSI M27, “Reference Method for Broth Dilution Antifungal Susceptibility Testing of Yeast” Fourth Edition (November 2017)
FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009)
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Reproducibility testing for the VITEK 2 AST-Yeast Voriconazole was conducted at three sites (two external and one internal site) using a panel of ten Candida species consistent with the indications for use (i.e., one isolate of Candida albicans, two isolates of Candida parapsilosis, six isolates of Candida krusei, and one isolate of Candida tropicalis). Each isolate was tested in triplicate, using separate inocula, over three days for a total of 270 data points. Inocula were prepared using both the autodilution and manual dilution methods for testing with the VITEK 2 System. In addition, inocula were prepared by the manual dilution method for testing with the VITEK 2 Compact. The mode of MIC values was determined for each isolate and the reproducibility was calculated based on the number of MIC values that fell within $\pm 1$ and $\pm 2$ doubling dilutions of the mode.
The testing resulted in overall best case reproducibility of $>95\%$ for autodilution and manual dilution in the VITEK 2 System as well as the VITEK 2 Compact System (manual dilution only) when evaluating both $\pm 1$ and $\pm 2$ doubling dilutions of the mode. The overall worst case reproducibility was $>89\%$ for autodilution and manual dilution in the VITEK 2 System as well as the VITEK 2 Compact System (manual dilution only) when evaluating both $\pm 1$ and $\pm 2$ doubling dilutions of the mode. The reproducibility data was acceptable.
K232967 - Page 5 of 12
{5}
Table 1. Reproducibility Performance
| | VITEK 2 | | | | VITEK 2 Compact | |
| --- | --- | --- | --- | --- | --- | --- |
| | Manual Dilution | | Auto-Dilution | | Manual Dilution | |
| | +/- 1 dilution of the mode | +/- 2 dilutions of the mode | +/- 1 dilution of the mode | +/- 2 dilutions of the mode | +/- 1 dilution of the mode | +/- 2 dilutions of the mode |
| Best Case | 97.8% | 100% | 97.8% | 100% | 97.8% | 100% |
| Worst Case | 93.7% | 95.9% | 93.7% | 95.9% | 92.9% | 95.2% |
2. Linearity:
Not applicable.
3. Analytical Specificity/Interference:
Not applicable
4. Assay Reportable Range:
Not applicable.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Inoculum Density Check: The DensiCHEK Plus was used to standardize the inoculum to a 1.8-2 McFarland standard. The instrument was standardized daily with all results recorded at each site. Calibration values were within the expected range.
Purity Check: A purity check of all organisms was performed on the dilution tube used to prepare the VITEK 2 card inoculum. Only those cultures that were pure were evaluated in the study.
Growth or Device Failure: The VITEK 2 AST-YS Voriconazole trial had several device failures including isolated mechanical failures that did not repeat (card recognition, mechanical jam, failure to eject, dispenser/pipettor failure, communication errors and an optics failure). These errors did not raise concerns since they were isolated and did not reflect systematic issues. One clinical isolate of C. parapsilosis failed to grow.
Quality Control Testing: The CLSI recommended QC strain Candida krusei (ATCC 6258) was tested a minimum of 20 times at each site by the reference method, the VITEK 2 instrument platform using both auto- and manual dilution methods, and the VITEK 2 Compact instrument platform using the manual dilution method. A summary of the QC performance is provided in Table 2.
The quality control results for the VITEK 2 AST-YS Voriconazole were within the recommended range $\geq 95\%$ which is acceptable.
K232967 - Page 6 of 12
{6}
Table 2. Quality Control Results for VITEK 2 AST-Yeast Voriconazole
| Organism | VITEK 2 Result Range | BMD Result Range (μg/mL) | VITEK 2 Auto-Dilution | BMD | VITEK 2 Manual Dilution | BMD | VITEK 2 Compact Manual Dilution | BMD |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| C. krusei ATCC 6258
Expected Range: 0.06 – 0.5 μg/mL | | ≤0.004 | | | | | | |
| | | 0.0078 | | | | | | |
| | | 0.0156 | | | | | | |
| | ≤0.0313 | 0.0313 | | 2 | 2 | 2 | | 2 |
| | 0.0625 | 0.0625 | 2 | 83 | 0 | 38 | | 38 |
| | 0.125 | 0.125 | 119 | 45 | 63 | 25 | 66 | 25 |
| | 0.25 | 0.25 | 10 | 1 | 1 | 1 | | 1 |
| | 0.5 | 0.5 | | | | | | |
| | 1 | 1 | | | | | | |
| | 2 | 2 | | | | | | |
| | ≥4 | 4 | | | | | | |
| | | ≥8 | | | | | | |
6. Detection Limit:
Not applicable.
7. Assay Cut-Off:
Not applicable.
B Comparison Studies:
1. Method Comparison with Predicate Device:
Testing of Voriconazole was performed at three external sites and one internal site. Results obtained with VITEK 2 AST-Yeast Voriconazole were compared to results obtained with the CLSI broth microdilution reference panel. The MIC result range for the VITEK 2 AST-Yeast Voriconazole is ≤ 0.03125 - ≥ 4 μg/mL for all species. The testing conditions for the reference method consisted of the following:
- Medium: RPMI 1640
- Inoculum: Direct colony suspension
- Incubation: 35°C; 24 hours (up to 48 hours for isolates that are not growing well)
The VITEK 2 cards were inoculated with test organisms using the auto-dilution method (VITEK 2) and using the manual dilution method (VITEK 2 and VITEK 2 Compact). All test inocula used for the VITEK 2 AST cards and the reference method were standardized using the DensiCHEK instruments. A total of 460 clinical isolates were evaluated using autodilution and VITEK 2. One isolate failed to grow in the VITEK 2 card. The results for the remaining 459 clinical isolates were evaluated. The majority of the isolates were recently isolated from clinical specimens and 201 of the 459 clinical isolates tested were from stock isolates (43.8%). The clinical isolates included: 257 C. albicans isolates, 67 C. krusei isolates, 61 C. parapsilosis isolates, and 74 C. tropicalis isolates.
K232967 - Page 7 of 12
{7}
The challenge set was composed of organisms listed in the Microbiology and Indication and Usage Sections of the FDA approved pharmaceutical package insert. A total of 76 challenge isolates including 41 C. albicans isolates, 9 C. krusei isolates, 13 C. parapsilosis isolates, and 13 C. tropicalis isolates were evaluated at two external sites and one internal site.
## Clinical and Challenge Data – VITEK 2 Autodilution
The results obtained using the autodilution method of the VITEK 2 from the 535 total isolates (459 clinical isolates and 76 challenge isolates) are summarized in Table 3.
Table 3. Performance of Candida spp. for the Auto Dilution and the VITEK 2 AST-YS Voriconazole.
| | Total | #EA | %EA | Eval EA Total | Eval EA # | Eval EA % | #CA | %CA | #R | #S | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Candida albicans ≤0.125 (S), 0.25-0.5 (I), ≥1 (R)* | | | | | | | | | | | | | |
| Clinical | 216 | 188 | 87.0 | 32 | 14 | 43.8 | 181 | 83.4 | 8 | 200 | 28 | 3 | 4 |
| Challenge | 41 | 40 | 97.6 | 6 | 5 | 83.3 | 40 | 97.6 | 4 | 37 | 1 | 0 | 0 |
| Combined | 257 | 228 | 88.7 | 38 | 19 | 50.0 | 221 | 86.0 | 12 | 237 | 29 | 3 | 4 |
| Candida krusei ≤0.5 (S), 1 (I), ≥2 (R) | | | | | | | | | | | | | |
| Clinical | 67 | 67 | 100.0 | 67 | 67 | 100.0 | 67 | 100.0 | 0 | 66 | 0 | 0 | 0 |
| Challenge | 9 | 9 | 100.0 | 9 | 9 | 100.0 | 8 | 88.9 | 0 | 8 | 1 | 0 | 0 |
| Combined | 76 | 76 | 100.0 | 76 | 76 | 100.0 | 75 | 98.7 | 0 | 74 | 1 | 0 | 0 |
| Candida parapsilosis ≤0.125 (S), 0.25-0.5 (I), ≥1 (R) | | | | | | | | | | | | | |
| Clinical | 61 | 57 | 93.4 | 7 | 3 | 42.9 | 58 | 95.1 | 0 | 61 | 2 | 1 | 0 |
| Challenge | 13 | 13 | 100.0 | 2 | 2 | 100.0 | 12 | 92.3 | 1 | 11 | 1 | 0 | 0 |
| Combined | 74 | 70 | 94.6 | 9 | 5 | 55.6 | 70 | 94.6 | 1 | 72 | 3 | 1 | 0 |
| Candida tropicalis ≤0.125 (S), 0.25-0.5 (I), ≥1 (R) | | | | | | | | | | | | | |
| Clinical | 74 | 72 | 97.3 | 16 | 14 | 87.5 | 66 | 89.2 | 1 | 67 | 6 | 2 | 0 |
| Challenge | 13 | 13 | 100.0 | 5 | 5 | 100.0 | 13 | 100.0 | 1 | 10 | 0 | 0 | 0 |
| Combined | 87 | 85 | 97.7 | 21 | 19 | 90.5 | 79 | 90.8 | 2 | 77 | 6 | 2 | 0 |
EA – Essential Agreement
CA – Category Agreement
EVAL – Evaluable isolates
R – Resistant
min – minor discrepancies
maj – major discrepancies
vmj – very major discrepancies
S – Susceptible
*Due to unacceptable performance, perform an alternative method of testing prior to reporting of results for voriconazole and C. albicans.
Essential agreement (EA) was calculated for when the VITEK 2 system results were within +/- two doubling dilutions of the reference method results. Evaluable results are those that are on-scale for both the reference method and VITEK 2 system or results in which an off-scale result is clearly more than two doubling dilutions from the on-scale result. Category agreement was calculated for when the VITEK 2 system result interpretations agreed exactly with the reference method result qualitative interpretations.
For C. albicans evaluated using VITEK 2 and autodilution, the overall EA and CA were 88.7% and 86.0%, respectively (Table 3) with 29 minor errors (11.3% of 257 isolates, 3 major errors (1.3% of 237 isolates), and 4 very major errors (33.3% of 12 isolates), which is not acceptable.
K232967 - Page 8 of 12
{8}
Due to the poor performance observed with C. albicans, this organism was removed from the Indications for Use.
For C. krusei evaluated using VITEK 2 and autodilution, the overall EA and CA were acceptable at 100% and 98.7%, respectively (Table 3), with one minor error (1.4% of 74 isolates). For C. parapsilosis evaluated using VITEK 2 and autodilution, the overall EA and CA were both acceptable at 94.6% (Table 3), with three minor errors (4.1% of 74 isolates) and 1 major error (1.4% of 72 susceptible isolates). or C. tropicalis evaluated using VITEK 2 and autodilution, the overall EA and CA were acceptable at 97.7% and 90.8%, respectively (Table 3), with 6 minor errors (6.9% of 87 isolates), and 2 major errors (2.6% of 77 susceptible isolates).
## Challenge Data – VITEK 2 and VITEK 2 Compact Manual Dilution
The 76 challenge isolates were also evaluated with the manual dilution options of the VITEK 2 and VITEK 2 Compact systems (summarized in Table 4).
Table 4. Performance of Candida spp. for the Manual Dilution with the VITEK 2 and the VITEK 2 Compact Systems AST-YS Voriconazole
| VITEK 2 Systems | Total | #EA | %EA | Eval Total | No Eval EA | Eval EA % | #CA | %CA | #R | #S | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Candida albicans ≤0.125 (S), 0.25-0.5 (I), ≥1 (R) | | | | | | | | | | | | | |
| VITEK 2 | 41 | 40 | 97.6 | 6 | 5 | 83.3 | 39 | 95.1 | 4 | 37 | 1 | 1 | 0 |
| VITEK 2 Compact | 41 | 40 | 97.6 | 6 | 5 | 83.3 | 40 | 97.6 | 4 | 37 | 1 | 0 | 0 |
| Candida krusei ≤0.5 (S), 1 (I), ≥2 (R) | | | | | | | | | | | | | |
| VITEK 2 | 9 | 9 | 100.0 | 9 | 9 | 100.0 | 8 | 88.9 | 0 | 8 | 1 | 0 | 0 |
| VITEK 2 Compact | 9 | 9 | 100.0 | 9 | 9 | 100.0 | 8 | 88.9 | 0 | 8 | 1 | 0 | 0 |
| Candida parapsilosis ≤0.125 (S), 0.25-0.5 (I), ≥1 (R) | | | | | | | | | | | | | |
| VITEK 2 | 13 | 13 | 100.0 | 3 | 3 | 100.0 | 13 | 100.0 | 1 | 11 | 0 | 0 | 0 |
| VITEK 2 Compact | 13 | 13 | 100.0 | 3 | 3 | 100.0 | 13 | 100.0 | 1 | 11 | 0 | 0 | 0 |
| Candida tropicalis ≤0.125 (S), 0.25-0.5 (I), ≥1 (R) | | | | | | | | | | | | | |
| VITEK 2 | 13 | 12 | 92.3 | 5 | 4 | 80.0 | 12 | 92.3 | 1 | 10 | 1 | 0 | 0 |
| VITEK 2 Compact | 13 | 13 | 100.0 | 4 | 4 | 100.0 | 13 | 100.0 | 1 | 10 | 0 | 0 | 0 |
EA – Essential Agreement
CA – Category Agreement
EVAL – Evaluable isolates
R – Resistant
min – minor discrepancies
maj – major discrepancies
vmj – very major discrepancies
S – Susceptible
Essential agreement (EA) was calculated for when the VITEK 2 system results were within +/- two doubling dilutions of the reference method results. Evaluable results are those that are on-scale for both the reference method and VITEK 2 system or results in which an off-scale result is clearly more than two doubling dilutions from the on-scale result. Category agreement was calculated for when the VITEK 2 system result interpretations agreed exactly with the reference method result qualitative interpretations.
The overall EA and CA for C. albicans is >95% (Table 4), with the manual dilution option and the VITEK 2 and VITEK 2 COMPACT instruments, however due to the unacceptable overall clinical performance, this organism was removed from the Indications for Use.
The overall EA and CA for C. parapsilosis is >95% (Table 4), with the manual dilution option and the VITEK 2 and VITEK 2 COMPACT instruments, which is acceptable. For C. krusei and
K232967 - Page 9 of 12
{9}
the VITEK 2 manual dilution method, the overall EA was acceptable at 100.0%; however, the CA was 88.9% (Table 4), with one minor error (11.1% of 9 isolates) for both the VITEK 2 and VITEK 2 COMPACT systems. The AST guidance indicates a CA of < 90% may be acceptable if the EA of the evaluable test results is very good and the majority of the discrepancies as minor discrepancies; therefore, the CA performance was considered acceptable. For C. tropicalis, the overall EA and CA were >90%, with the manual dilution option and the VITEK 2 and VITEK 2 COMPACT instruments (Table 4), with 1 minor error (7.7% of 13 isolates), which is acceptable.
As required under 511A(b)(2)(C)(ii)(I) of the Federal Food, Drug and Cosmetic Act, the following statement was added to the package insert to address testing of non-indicated species:
Per the FDA-Recognized Susceptibility Test Interpretive Criteria website, the safety and efficacy of antimicrobial drugs, for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well-controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labeling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved.
## Resistant Strains
There were limited Candida spp. resistant isolates available for evaluation during clinical and challenge testing. The sponsor included the following limitation in the device labeling:
The ability of the AST card to detect resistance with the following combination(s) is unknown because resistant strains were not available at the time of comparative testing:
- Voriconazole (vrc02n): Candida spp.
## MIC Trends:
A trending analysis was conducted using the combined data (clinical and challenge) obtained from the VITEK 2 auto-dilution method for each organism group. This trending calculation takes into account MIC values that are determined to be one or more doubling dilutions lower or higher than the reference method irrespective of whether the device MIC values are on-scale or not. Results that are not clearly at least one dilution lower, at least one dilution higher or in exact agreement with the CLSI reference method are not considered in the trending analysis.
Organism groups for which the difference between the percentage of isolates with higher vs. lower readings was > 30% and for which the confidence interval was determined to be statistically significant were considered to show evidence of trending. Trending that showed higher or lower MIC values compared to the reference is addressed in the labeling.
Table 5. Trending for Candida spp. with Voriconazole and the VITEK 2 Automated Dilution
| Organism | Total Evaluable for Trending | ≥ 1 Dilution lower No. (%) | Exact No. (%) | ≥ 1 Dilution Higher No. (%) | Percent Difference (CI) | Trending Noted |
| --- | --- | --- | --- | --- | --- | --- |
| Candida albicans | 47 | 4 (8.5) | 5 (10.6) | 38 (80.9) | 72.4 (54.7, 82.5) | Yes |
| Candida krusei | 76 | 36 (47.4) | 29 (38.2) | 11 (14.5) | -32.9 (-45.6, -18.4) | Yes |
| Candida parapsilosis | 10 | 0 (0.0) | 2 (20.0) | 8 (80.0) | 60.0 (16.2, 80.3) | Yes |
| Candida tropicalis | 33 | 17 (51.5) | 4 (12.1) | 10 (30.3) | -21.2 (41.7, -2.4) | No |
K232967 - Page 10 of 12
{10}
A trend toward higher MIC values was observed for C. albicans and C. parapsilosis when compared to the CLSI broth microdilution reference method, as summarized in Table 5. The following statement is included as a footnote to the performance table in the device labeling to address the observed trending:
VITEK 2 Voriconazole MIC values for C. parapsilosis tended to be in exact agreement or at least one doubling dilution higher than the reference MIC values.
A trend toward lower MIC values was observed for C. krusei when compared to the CLSI broth microdilution reference method, as summarized in Table 5. The following statement is included as a footnote to the performance table in the device labeling to address the observed trending:
Voriconazole MIC values for C. krusei tended to be in exact agreement or at least one doubling dilution lower than the reference MIC values.
2. Matrix Comparison:
Not applicable.
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
The FDA and CLSI susceptibility interpretive criteria for Voriconazole are as listed in Table 6.
K232967 - Page 11 of 12
{11}
Table 6. FDA Recognized Interpretive Criteria for Voriconazole
| Organism | Minimum Inhibitory Concentrations (μg/mL)1 | | |
| --- | --- | --- | --- |
| | S | I | R |
| Candida krusei | ≤0.5 | 1 | ≥2 |
| Candida parapsilosis | ≤0.125 | 0.25-0.5 | ≥1 |
| Candida tropicalis | ≤0.125 | 0.25-0.5 | ≥1 |
S = Susceptible; I = Intermediate; R = Resistant
1 According to CLSI M27M44S, 3rd Edition and FDA STIC Website
https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is not complete and does not support a substantial equivalence decision.
To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission incorporated by reference a breakpoint change protocol that was reviewed and accepted by FDA in submission K230864 cleared on July 3, 2023. The referenced protocol addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage
(https://www.fda.gov/drugs/development-resources/fda-recognized-antimicrobial-susceptibility-test-interpretive-criteria). The referenced protocol outlined the specific procedures and acceptance criteria that bioMérieux intends to use to evaluate the bioMérieux VITEK 2 AST-Yeast Voriconazole ( $\leq 0.03125 - \geq 4\mu \mathrm{g / mL}$ ) when revised breakpoints for Voriconazole are published on the FDA STIC webpage. The breakpoint change protocol included with the submission indicated that if specific criteria are met, bioMérieux will update the VITEK 2 AST-Yeast Voriconazole device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation.
K232967 - Page 12 of 12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.