The iX series Patient Monitors including iX10, iX12, iX15 are intended to be used for monitoring, storing, recording, and reviewing of, and to generate alarms for, multiple physiological parameters of adults and pediatrics (including neonates). The monitors are intended for use by trained healthcare professionals in hospital environments. The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), cardiac output (C.O.), and Anaesthesia gas (AG). The arrhythmia detection and ST Segment analysis are intended for adult patients. The NIBP monitoring supports iCUFS algorithm and iFAST algorithm. The iCUFS algorithm is intended for adult, pediatric and neonatal patients. The iFAST algorithm is intended for adult and pediatric patients (≥3 years of age). Both measurement algorithms are also intended for use with pregnant women, including pre-eclamptic patients. NIBP MAP is not applicable to pregnant women. The Spot Temp with T2A module can only measure temperature of adult and pediatric (>1 year of age) patients. The monitors are not intended for MRI environments. The cardiac output (C.O.) is only intended for adult patients.
Device Story
iX series (iX10, iX12, iX15) are multi-parameter patient monitors for hospital use by clinicians. Inputs: ECG, RESP, TEMP, SpO2, PR, NIBP, IBP, CO2, C.O., and anesthesia gas. Device processes signals via internal algorithms (iCUFS, iFAST, arrhythmia/ST analysis) to display real-time waveforms, numeric values, and generate alarms for abnormalities. Clinicians use output to monitor patient status and guide clinical interventions. Benefits include continuous physiological surveillance and timely detection of adverse events.
Clinical Evidence
Clinical investigation conducted for T2A Quick Temp module per ISO 80601-2-56. Study evaluated 142 subjects (ages >1 to 73 years) comparing Predict mode to direct mode. Results confirmed accuracy met ISO 80601-2-56:2017+AMD1:2018 requirements. No adverse reactions reported. Other parameters supported by extensive bench testing against consensus standards (IEC 60601 series).
Technological Characteristics
Multi-parameter patient monitor; AC power (100-240V) and rechargeable battery. Connectivity: Wi-Fi (802.11a/b/g/n), Bluetooth Low Energy (e-Link). Modules: ECG (3/5/6/10 lead), NIBP (oscillometric), SpO2, IBP, CO2, C.O., AG, and T2A/Infrared temperature. Software-based arrhythmia/ST analysis. Complies with IEC 60601-1, 60601-1-2, 60601-2-25, 60601-2-27, 80601-2-30, 60601-2-34, 60601-2-49, ISO 80601-2-55, ISO 80601-2-56, ISO 80601-2-61.
Indications for Use
Indicated for monitoring, storing, and alarming of physiological parameters in adult, pediatric, and neonatal patients. Arrhythmia/ST analysis and cardiac output indicated for adults only. iFAST NIBP algorithm indicated for adults and pediatrics ≥3 years. Spot Temp module indicated for adults and pediatrics >1 year. All algorithms indicated for use with pregnant women (including pre-eclamptic), excluding NIBP MAP. Not for MRI environments.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Patient Monitor Model iM50, iM60, iM70, iM80 (K202336)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 19, 2024
Edan Instruments, Inc Tracy Yue Regulatory Affairs Engineer 15 Jinhui Road, Jinsha Community, Kengzi Sub-district Pingshan District Shenzhen. Shenzhen 518122 China
Re: K232962
Trade/Device Name: Patient Monitor (iX10, iX12, iX15) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX, DPS, DSI, MLD, DRT, DXN, DSK, FLL, DQA, CCK, CBS, CBR, CCL, CBQ, NHO, NHQ, NHP Dated: March 22, 2024 Received: March 22, 2024
Dear Tracy Yue:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer W. Shih -S
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices
{2}------------------------------------------------
Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232962
Device Name Patient Monitor, Model: iX10, iX12, iX15
#### Indications for Use (Describe)
The iX series Patient Monitors including iX10, iX12, iX15 are intended to be used for monitoring, storing, and reviewing of, and to generate alarms for, multiple physiological parameters of adults and pediatics (including neonates). The monitors are intended for use by trained healthcare professionals in hospital environments.
The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), cardiac output (C.O.), and Anaesthesia gas (AG).
The arrhythmia detection and ST Segment analysis are intended for adult patients.
The NIBP monitoring supports iCUFS algorithm and iFAST algorithm is intended for adult, pediatric and neonatal patients. The iFAST algorithm is intended for adult and pediatic patients (≥3 years of age). Both measurement algorithms are also intended for use with pregnant women, including pre-eclamptic patients. NIBP MAP is not applicable to pregnant women.
The Spot Temp with T2A module can only measure temperature of adult and pediatric (> 1 year of age) patients. The monitors are not intended for MRI environments.
The cardiac output (C.O.) is only intended for adult patients.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span> <span style="font-size: 10pt;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-----------------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 10pt;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92
# 1. Submitter:
| Applicant: | Edan Instruments, Inc.<br>#15 Jinhui Road, Jinsha Community,<br>Kengzi Sub-District, Pingshan District,<br>Shenzhen, 518122 P.R.China.<br>Tel: +86(0755) 26858736<br>Fax: +86(0755) 26882223 |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Tracy Yue |
| Contact person: | Tracy Yue |
|-----------------|--------------------|
| Preparing date: | September 21, 2023 |
# 2. Device name and classification:
| Trade name: | Patient Monitor, Model: iX10, iX12, iX15 | | | | |
|--------------------------|-----------------------------------------------------------------------------------------|--|--|--|--|
| Common/Usual Name: | Patient Monitor | | | | |
| Classification Name: | 21 CFR 870.1025<br>Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) | | | | |
| Regulatory class: | Class II | | | | |
| Primary product code: | MHX-Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms) | | | | |
| Subsequent product code: | Regulation number/Device Product Code 21 CFR 868.2375 DPS | | | | |
| Regulation number/Device | Product Code |
|---------------------------------------------------------------------------|--------------|
| 21 CFR 868.2375<br>Electrocardiograph | DPS |
| 21 CFR 870.2340<br>Detector and Alarm, Arrhythmia | DSI |
| 21 CFR 870.1025<br>Monitor, ST Segment with Alarm | MLD |
| 21 CFR 870.2300<br>Monitor, Cardiac (Incl. Cardiotachometer & Rate Alarm) | DRT |
| 21 CFR 870.1130<br>System, Measurement, Blood-Pressure, Non-Invasive | DXN |
| 21 CFR 870.1110<br>Computer, Blood-Pressure | DSK |
| 21 CFR 880.2910<br>Thermometer, Electronic, Clinical | FLL |
| 21 CFR 870.2700<br>Oximeter | DQA |
{5}------------------------------------------------
| 21 CFR 868.2375<br>Monitor, Breathing Frequency | BZQ |
|--------------------------------------------------------------------------------------------|-----|
| 21 CFR 870.1400<br>Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase | CCK |
| 21 CFR 868.1620<br>Analyzer, Gas, Halothane, Gaseous-Phase (Anesthetic<br>Conc.) | CBS |
| 21 CFR 868.1700<br>Analyzer, Gas, Nitrous-Oxide, Gaseous Phase (Anesthetic<br>Conc.) | CBR |
| 21 CFR 868.1720<br>Analyzer, Gas, Oxygen, Gaseous-Phase | CCL |
| 21 CFR 868.1500<br>Analyzer, Gas, Enflurane, Gaseous-Phase (Anesthetic<br>Concentration) | CBQ |
| 21 CFR 868.1500<br>Analyzer, Gas, Desflurane, Gaseous-Phase (Anesthetic<br>Concentration) | NHO |
| 21 CFR 868.1500<br>Analyzer, Gas, Isoflurane, Gaseous-Phase (Anesthetic<br>Concentration) | NHQ |
| 21 CFR 868.1500<br>Analyzer, Gas, Sevoflurane, Gaseous-Phase (Anesthetic<br>Concentration) | NHP |
### 3. Predicate Device(s):
- 1) Edan Instruments, Inc., Patient Monitor Model iM50, iM60, iM70, iM80, K202336 (Primary)
- 2) Shenzhen Mindray Bio-Medical Electronics Co., LTD., VS9 Vital Signs Monitor, K211475 (Reference)
- 3) Edan Instruments, Inc., Vital signs monitor Model iM3, K180380 (Reference)
- 4) Edan Instruments, Inc., Vital signs monitor Model iM3s, K202892 (Reference)
- 5) Edan Instruments, Inc., Patient Monitor Model RespArray, K220308 (Reference)
- 6) Shenzhen Mindray Bio-Medical Electronics Co., LTD., BeneVision N Series Patient Monitors, K182075 (Reference)
{6}------------------------------------------------
| 4. Device Description: | The iX series Patient Monitors including iX10, iX12, iX15 can perform long-time<br>continuous monitoring of multiple physiological parameters. Also, it is capable of<br>storing, displaying, analyzing and controlling measurements, and it will indicate<br>alarms in case of abnormalities so that doctors and nurses can respond to the patient's<br>situation as appropriate. |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 5. Indication for Use | The iX series Patient Monitors including iX10, iX12, iX15 are intended to be used for<br>monitoring, storing, recording, and reviewing of, and to generate alarms for, multiple<br>physiological parameters of adults and pediatrics (including neonates). The monitors<br>are intended for use by trained healthcare professionals in hospital environments.<br>The monitored physiological parameters include: ECG, respiration (RESP), temperature<br>(TEMP), functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR),<br>non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2),<br>cardiac output (C.O.), and Anaesthesia gas (AG).<br>The arrhythmia detection and ST Segment analysis are intended for adult patients.<br>The NIBP monitoring supports iCUFS algorithm and iFAST algorithm. The iCUFS<br>algorithm is intended for adult, pediatric and neonatal patients. The iFAST algorithm is<br>intended for adult and pediatric patients (≥3 years of age). Both measurement algorithms<br>are also intended for use with pregnant women, including pre-eclamptic patients. NIBP<br>MAP is not applicable to pregnant women.<br>The Spot Temp with T2A module can only measure temperature of adult and pediatric (>1<br>year of age) patients.<br>The monitors are not intended for MRI environments.<br>The cardiac output (C.O.) is only intended for adult patients. |
## 6. Predicate Device Comparison
The table below compares the indication for use and key technological feature of the subject devices to the predicate device (Patient Monitor Model iM50, iM60, iM70, iM80, K202336).
Note: "/" means no such functions in predicate device.
| K# | Subject Device<br>(iX10, iX12, iX15) | Predicate Device<br>(iM50, iM60, iM70, iM80) |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K232962 | K232336 | K202336 |
| Description | The iX series Patient Monitors<br>including iX10, iX12, iX15 are<br>intended to be used for monitoring,<br>storing, recording, and reviewing of,<br>and to generate alarms for, multiple<br>physiological parameters of adults<br>and pediatrics (including neonates)<br>The monitors are intended for use<br>by trained healthcare professionals in<br>hospital environments. The monitored<br>physiological parameters | The monitors are intended to be used for<br>monitoring, storing, recording, and<br>reviewing of, and to generate alarms for,<br>multiple physiological parameters of adults,<br>pediatrics and neonates. The monitors are<br>intended for use by trained healthcare<br>professionals in hospital environments. |
| | include: ECG, respiration (RESP), temperature<br>(TEMP), functional oxygen saturation of<br>arterial hemoglobin (SpO2), pulse rate (PR),<br>non-invasive blood pressure (NIBP), invasive<br>blood pressure (IBP), carbon dioxide (CO2),<br>cardiac output (C.O.), and Anaesthesia gas<br>(AG). | The monitored physiological parameters<br>include: ECG, respiration (RESP),<br>temperature (TEMP), oxygen saturation of<br>arterial blood (SpO2), pulse rate (PR),<br>non-invasive blood pressure (NIBP),<br>invasive blood pressure (IBP), carbon<br>dioxide (CO2), cardiac output (C.O.), and<br>Anaesthesia gas(AG). |
| | The arrhythmia detection and ST Segment<br>analysis are intended for adult patients. | The arrhythmia detection and ST Segment<br>analysis are intended for adult patients. |
| | The NIBP monitoring supports iCUFS<br>algorithm and iFAST algorithm. The iCUFS<br>algorithm is intended for adult, pediatric<br>and neonatal patients. The iFAST algorithm<br>is intended for adult and pediatric patients<br>(≥3 years of age). Both measurement<br>algorithms are also intended for use with<br>pregnant women, including pre-eclamptic<br>patients. NIBP MAP is not applicable to<br>pregnant women. | The monitors are not intended for MRI<br>environments. |
| | The Spot Temp with T2A module can only<br>measure temperature of adult and pediatric<br>(> 1 year of age) patients. | |
| | The monitors are not intended for MRI<br>environments. | |
| | The cardiac output (C.O.) is only intended<br>for adult patients. | |
| ECG module | | |
| Lead Mode | 3 Electrodes; 5 Electrodes; 6 Electrodes; 10<br>Electrodes | 3 Electrodes; 5 Electrodes; 6 Electrodes; 10<br>Electrodes |
| Arrhythmia analyses | ASYSTOLE, VFIB/VTAC, COUPLET, VT ><br>2, BIGEMINY, TRIGEMINY, VENT, R on T,<br>PVC, TACHY, BRADY, MISSED BEATS,<br>IRR, VBRADY, PNC, PNP | ASYSTOLE, VFIB/VTAC, COUPLET,<br>VT > 2, BIGEMINY, TRIGEMINY, VENT,<br>R on T, PVC, TACHY, BRADY, MISSED<br>BEATS, IRR, VBRADY, PNC, PNP |
| | | The arrhythmia detection algorithm is the same as predicate device K202336 |
| ST value | | |
| Measurement Range | -2.0 mV to +2.0 mV | -2.0 mV to +2.0 mV |
| Pace | | |
| Pulse Indicator | Amplitude: ±2 mV to ±700 mV<br>Width: 0.1 ms to 2.0 ms<br>Ascending time: 10 us to 100 us | Amplitude: ±2 mV to ±700 mV<br>Width: 0.1 ms to 2.0 ms<br>Ascending time: 10 us to 100 us |
| PVC | | |
| Range | ADU: (0 to 300) PVCs/ min | ADU: (0 to 300) PVCs/ min |
| | PED/NEO: (0 to 350) PVCs/ min | PED/NEO: (0 to 350) PVCs/ min |
| HR | | |
| Measurement range | ADU: 15 bpm to 300 bpm | ADU: 15 bpm to 300 bpm |
| | PED/NEO: 15 bpm to 350 bpm | PED/NEO: 15 bpm to 350 bpm |
| QT Analysis | | |
| QT/QTc/ΔQTc<br>measurement | QT Range: 200 ms ~ 800 ms<br>QTc Range :200ms ~ 800 ms<br>△ QTc Range: -600 ms ~ 600 ms | QT Range: 200 ms ~ 800 ms<br>QTc Range :200ms ~ 800 ms<br>△ QTc Range: -600 ms ~ 600 ms |
| RESP module | | |
| Principle of Operation | Impedance between RA-LL, RA-LA | Impedance between RA-LL, RA-LA |
| Measurement Range | 0 rpm to 200 rpm | Adult: 0 to 120 rpm<br>Pediatric/neonate: 0 rpm to 150rpm |
| NIBP module (EDAN) | | |
| Technique | Oscillometry | Oscillometry |
| Measurement Mode | iCUFS, iFAST | iCUFS |
| EDAN-NIBP(iCUFS) | | |
| Intended users | Adult, pediatric, neonates | Adult, pediatric, neonates |
| Measuring parameter | SYS, DIA, MAP, PR | |
| Measurement Range | Adult Pediatric Neonate<br>Systolic 25-290 25-240 25-140<br>Diastolic 10-250 10-200 10-115<br>Mean 15-260 15-215 15-125 | Adult Pediatric Neonate<br>Systolic 25-290 25-240 25-140<br>Diastolic 10-250 10-200 10-115<br>Mean 15-260 15-215 15-125 |
| PR from NIBP | | |
| Measurement range | 40 bpm to 240 bpm | 40 to 240 bpm |
| EDAN-NIBP(iFAST) | | |
| Intended users | Adult, pediatric | |
| Measuring parameter | SYS, DIA, MAP | |
| Measurement Range | Adult Mode:<br>SYS: 25 mmHg to 290 mmHg<br>DIA: 10 mmHg to 250 mmHg<br>MAP: 15 mmHg to 260 mmHg | |
| | | |
| | | |
| | | |
| | Pediatric Mode: | |
| | SYS: 25 mmHg to 240 mmHg | |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
| DIA: 10 mmHg to 200 mmHg | | | | |
|-------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|-------------------------------------------------------------------------------|-----------------------------------------|-----------------------------|
| MAP: 15 mmHg to 215 mmHg | | | | |
| | | SpO2 module(EDAN) | | |
| Measurement Range | SpO2 | 0% to 100% | SpO2 | 0% to 100% |
| | Pulse Rate | 25 to 300 bpm | Pulse Rate | 25 to 300 bpm |
| | | Nellcor SpO2 module is the same as K202336 | | |
| | | Temperature(TEMP) module | | |
| Number of channels | 2 | | 2 | |
| Measurement Range | 0 °C to 50 °C(32 °F to 122 °F) | | 0 °C to 50 °C(32 °F to 122 °F) | |
| | | Spot Temp module | | |
| | | Measurement site: oral/axillary/rectal | | |
| Quick Temp (T2A<br>module) | Monitor mode: 25 °C ~45 °C | | | |
| | Predict mode: 35.5 °C ~42 °C | | | |
| | Accuracy (Monitor mode): 0.1 °C (25 °C ~<br>45 °C) | | | |
| The Quick Temp (T2A) module is similar to K180380, and the performance of the Predict mode of the module is | | | | |
| | | validated by the clinical trial. | | |
| Infrared Temp<br>(TAT-5000S-RS232<br>Thermometer) | | Measuring Range : 16 °C~ 43 °C (61 °F ~<br>110 °F) | | |
| | | The Infrared Temp module is the same as K202892 | | |
| | | The Quick Temp module and Infrared Temp module is applicable to iX10 and iX12 | | |
| | | IBP module | | |
| | ART, Ao, UAP, BAP, FAP, LV, P1-P4: (-50 to<br>+400) mmHg | | Art: (0 to +300) mmHg | |
| Measurement Range | PA/PAWP: (-6 to +120) mmHg | | PA/PAWP: (-6 to +120) mmHg | |
| | CVP/RAP/LAP/ICP: (-10 to +40) mmHg | | CVP/RAP/LAP/ICP: (-10 to +40) mmHg | P1/P2: (-50 to +300) mmHg |
| | | C.O. Module | | |
| Technique | | Thermodilution Technique | | Thermodilution Technique |
| | C.O.: 0.1 to 20L/min | | C.O.: 0.1 to 20L/min | |
| Measurement range | TB: 23 °C to 43 °C(73.4 °F to 109.4 °F) | | TB: 23° C to 43 °C(73.4 °F to 109.4 °F) | |
| | TI: -1 °C to 27 °C(30.2 °F to 80.6 °F) | | TI: -1 °C to 27 °C(30.2 °F to 80.6 °F) | |
| | | CO2 Module (EDAN G2) | | |
| Intended Patient | Adult, pediatric, neonatal | | Adult, pediatric, neonatal | |
| Measure Parameters | EtCO2, FiCO2, AwRR | | EtCO2, FiCO2, AwRR | |
| Measuring Range | CO2:0 mmHg to 150 mmHg (0 % to 20%) | | CO2:0 mmHg to 150 mmHg (0 % to 20%) | |
| | AwRR: 0 rpm to 150 rpm | | AwRR: 2 rpm to 150 rpm | |
| Respironics and Masimo CO2 module are the same as K202336 | | | | |
| AG module (EDAN G7 module) | | | | |
| Measure Parameters | | CO2、N2O、O2、HAL、ISO、ENF、SEV、 | | CO2、N2O、O2、HAL、ISO、ENF、SEV、 |
{10}------------------------------------------------
| | DES、AwRR、MAC | DES、AwRR、MAC |
|------------------------------------------|---------------------------------|---------------------------------|
| Measuring Range | CO2: 0%~15Vol % | CO2: 0%~15Vol % |
| | N2O: 0%~100 Vol % | N2O: 0%~100 Vol % |
| | HAL/ISO: 0%-8Vol % | HAL/ISO: 0%-8Vol % |
| | ENF: 0%-8 Vol % | ENF: 0%-8 Vol % |
| | SEV: 0%-10 Vol % | SEV: 0%-10 Vol % |
| | DES: 0%-20 Vol % | DES: 0%-20 Vol % |
| | O2: 0%~100 Vol % | O2: 0%~100 Vol % |
| | AwRR: 2rpm-100rpm | AwRR: 2rpm-100rpm |
| Masimo AG module are the same as K202336 | | |
| WI-FI | | |
| IEEE | 802.11a/b/g/n | 802.11a/b/g/n |
| Frequency Band | 2.4 GHz ISM band & 5 G ISM band | 2.4 GHz ISM band & 5 G ISM band |
| | e-Link | |
| e-Link | Bluetooth Low Energy 4.0 | / |
| | Power supply | |
| AC power | | |
| Requirement | 100-240V, 50/60Hz | 100-240V, 50/60 Hz |
| Battery | | |
| Rechargeable<br>Battery | Yes | Yes |
As seen in the comparison tables, the subject and predicate device have similar design features and performance specifications. The technological differences between the subject and predicate device do not raise different questions of safety or effectiveness.
## 7. Performance Data:
#### Non-clinical data:
Electrical safety and electromagnetic compatibility (EMC)
iX Series Patient Monitors were assessed for conformity with the relevant requirements of the following standards and found to comply:
- ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
- IEC 60601-1-2:2014+A1:2020 Medical electrical equipment Part 1-2: General requirements for basic ● safety and essential performance – Collateral Standard: electromagnetic disturbances – Requirements and tests.
{11}------------------------------------------------
# Performance testing-Bench
Edan has conducted functional and system level testing to validate the performance of the results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards.
- IEC 60601-1-8:2020 Medical electrical equipment - part 1-8: general requirements for basic safety and essential performance - collateral standard: general requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- IEC 60601-2-25:2011 Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs
- IEC 60601-2-27:2011 Medical electrical equipment--Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment
- IEC 80601-2-30:2018 Medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
- IEC 60601-2-34:2011 Medical electrical equipment - part 2-34: particular requirements for the basic safety, including essential performance, of invasive blood pressure monitoring equipment
- IEC 60601-2-49:2018 Medical electrical equipment –Part 2–49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment
- ISO 80601-2-55:2018 Medical electrical equipment - part 2-55: particular requirements for the basic safety and essential performance of respiratory gas monitors
- ISO 80601-2-56:2017+A1:2018 Medical electrical equipment - part 2-56: particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
- ISO 80601-2-61:2017 Medical electrical equipment - part 2-61: particular requirements for basic safety and essential performance of pulse oximeter equipment
- IEEE ANSI C63.27:2017 American National Standard for Evaluation of Wireless Coexistence
- ANSI AAMI EC57:2012 Testing and reporting performance results of cardiac rhythm and ST segment measurement algorithms
## Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and Food and Drug Administration Staff, "Content of Premarket Submissions for Device Software Functions".
## Clinical data:
To meet the requirements for the validity and accuracy of the TEMP measurement of the Quick Temp Module (T2A), EDAN conducted clinical investigation according to the requirements of the standard ISO 80601-2-56
{12}------------------------------------------------
Second edition 2017-03 + AMD1:2018: Medical electrical equipment- Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement. A clinical accuracy study evaluated 142 valid cases of sublingual and axillary temperature, which were performed on the following two age groups: >1 and <5 years old and ≥5 years old in accordance with ISO 80601-2-56:2017/Amd.1:2018(E) to compare the direct mode of F3000 Temp Module of M3A Vital signs monitor. The age of subjects is from >1 year old to 73 years old. The total number of hyperthermia subjects shall be no less than 30% and no more than 50% of all subjects in the selected age group. Statistical results show that, the temperature measurement function of the T2A module in Predict mode meets the requirements of standard ISO 80601-2-56:2017+AMD1:2018 for temperature measurement and meets the acceptance criteria in clinical protocol. During the entire clinical trial, all subjects are in good condition, and none of the subjects have discomfort symptoms or adverse reactions.
#### Summarv
Both the clinical and non-clinical performance testing showed that the subject devices are as safe and as effective as the predicate device.
#### 8. Conclusion
The performance data demonstrates that iX series Patient Monitor are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.